Agilent PN 8510-16 Controlling T est Port Output Power Flatness Product Note UNCORRECTED POWER CORRECTED POWER OUTPUT POWER Agilent 8510C Network Analyzer.
2 Introduction Designers and manufacturers of active devices oft en need to control the pow er lev el at the t est port 1 of t heir pow er -sensitive devices, but f ind dif- ficulty in o vercoming inser tion losses. The insertion losses occur as a result of connecting components in the measurement pat h between the source and the DUT .
3 System Configuration The basic 8510C measurement system conf igura- tion needed for test por t f latness correction is sho wn in Figure 1. A sing le channel pow er meter such as the A gilent 437B, E4418A, or E4418B is required, along with a compatible sensor , for per- forming t he po wer f latness calibration.
4 Flatness-Correction Operation The basic setup procedure t o obtain f lat t est port output po wer is illustrated in Table 1. To simplify the execution of t he procedure, t he front panel hardk eys are enclosed in [brackets]. The sof tkeys are enclosed in {braces}.
5 T able 1. Flatness-correction calibration procedure 8510C Keystrokes Description Set up the power meter (see T able 2) For proper operation, the E4418B must be set up before initiating the flatness-correction routine. See T able 2 for specific instructions.
6 T able 2. Agilent E4418B power meter setup E4418B Keystrokes Description Preset and zero the power meter [PRESET/LOCAL] Return the power meter to a known state. {CONFIRM} [ZERO/CAL] {ZERO} Set up the 437B emulation on the power meter [SYSTEM/INPUTS] The 437B command set must be activated in order to perform the power flatness calibration.
7 The follo wing er ror messages may occur while attempting t o enable the f latness cor rection and/or reducing the test por t pow er level: • If f latness cor rection is enabled before reducing the test por t pow er level, IF Overload or Source 1 W arning—RF Unleveled ma y be display ed on the analyzer as the source becomes unlev eled.
8 V erifying a flatness-correction calibration T o verify t he f latness-correction calibration, t he po wer sensor should be reconnected t o the t est port t o measure the t est por t po wer at individual frequencies. Since the measurement system is calibrated in 50W , inaccuracies will occur when a device is not well mat ched.
9 Using test port 1 calibrations on test port 2 In most cases, port 1 will be t he input por t of t he DUT . When por t 2 must be used as the input por t to t he device, t he user may choose to use a por t 1 f latness-cor rection calibration on por t 2 since the port 1 and por t 2 signal pat hs are symmetrical.
10 Practical Application Examples Absolute output power measurements with flatness correction One benefit of t he f latness-correction capability is the ability t o measure the absolute output pow er of active devices.
11 Amplifier measurement example Step-by-step instr uctions for setting up and apply- ing f latness cor rections for t he measurement of an amplifier are sho wn in Table 7 . Gain, absolute output po wer , and gain compression measurements procedures are cov ered.
Agilent T echnologies’ T est and Measurement Support, Services, and Assistance Agilent Technologies aims to maximize the value you receive, while minimizing your risk and problems. W e strive to ensure that you get the test and measurement capabilities you paid for and obt ain the suppor t you need.
デバイスAgilent Technologies PN 8510-16の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Agilent Technologies PN 8510-16をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはAgilent Technologies PN 8510-16の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Agilent Technologies PN 8510-16の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Agilent Technologies PN 8510-16で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Agilent Technologies PN 8510-16を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はAgilent Technologies PN 8510-16の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Agilent Technologies PN 8510-16に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちAgilent Technologies PN 8510-16デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。