Emersonメーカー1500の使用説明書/サービス説明書
ページ先へ移動 of 264
Configuration and Use Manual MMI-20019023, Rev AA March 2012 Micro Motion ® Model 1500 Transmitters with Analog Outputs Configuration and Use Manual.
Safety messages Safety messages are provided throughout this manual to protect personnel and equipment. Read each safety message carefully before proceeding to the next step. Micro Motion customer service Location Telephone number Email U.S.A. 800-522-MASS (800-522-6277) (toll free) flow.
Contents Part I Getting Started Chapter 1 Before you begin ............................................................................................................3 1.1 About this manual ............................................................
4.5.1 Configure Density Measurement Unit ......................................................................... 45 4.5.2 Configure slug flow parameters .................................................................................. 46 4.5.3 Configure Density Damping .
Chapter 7 Completing the configuration .................................................................................... 101 7.1 Test or tune the system using sensor simulation ...................................................................... 101 7.
10.3 Flow measurement problems .................................................................................................. 156 10.4 Density measurement problems ...................................................................................
B.2.3 Make a HART/Bell 202 connection ............................................................................ 202 B.2.4 Make a Modbus/RS-485 connection ..........................................................................206 B.3 Menu maps for ProLink III .
Contents vi Micro Motion ® Model 1500 Transmitters with Analog Outputs.
Part I Getting Started Chapters covered in this part: • Before you begin • Quick start Getting Started Configuration and Use Manual 1.
Getting Started 2 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
1 Before you begin Topics covered in this chapter: • About this manual • Transmitter model code • Communications tools and protocols • Additional documentation and resources 1.1 About this manual This manual provides information to help you configure, commission, use, maintain, and troubleshoot the Micro Motion Model 1500 transmitter.
Communications tools, protocols, and related information Table 1-1: Communica- tions tool Supported protocols Scope In this manual For more information ProLink II • HART/Bell 202 • Modbus/RS-485 • Service port Complete configuration and commissioning Basic user information.
Additional documentation and resources (continued) Table 1-2: Topic Document Hazardous area installa- tion See the approval documentation shipped with the transmitter, or download the appropriate documentation from the Micro Motion web site at www.micromotion.
Before you begin 6 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
2 Quick start Topics covered in this chapter: • Power up the transmitter • Check flowmeter status • Make a startup connection to the transmitter • Characterize the flowmeter (if required) • Verify mass flow measurement • Verify the zero 2.
Immediately after power-up, the transmitter runs through diagnostic routines and checks for error conditions. During the power-up sequence, Alarm A009 is active. This alarm should clear automatically when the power-up sequence is complete. 2. Check the status LED on the transmitter.
• To change the protocol, baud rate, parity, or stop bits, choose ProLink > Configuration > RS-485 . • To change the address, choose ProLink > Configuration > Device . To change the communications parameters using ProLink III, choose Device Tools > Configuration > Communications .
• For curved-tube sensors, set Flow Cal ( Flow Calibration Factor ). 3. Set the density characterization parameters. • For straight-tube sensors, set D1 , D2 , DT , DTG , K1 , K2 , FD , DFQ1 , and DFQ2 . • For curved-tube sensors, set D1 , D2 , TC , K1 , K2 , and FD .
Tag on newer curved-tube sensors (all sensors except T-Series) Figure 2-2: Tag on older straight-tube sensor (T-Series) Figure 2-3: Tag on newer straight-tube sensor (T-Series) Figure 2-4: Quick start.
Density calibration parameters (D1, D2, K1, K2, FD, DT, TC) If your sensor tag does not show a D1 or D2 value: • For D1 , enter the Dens A or D1 value from the calibration certificate. This value is the line-condition density of the low-density calibration fluid.
2.5 Verify mass flow measurement Check to see that the mass flow rate reported by the transmitter is accurate. You can use any available method. • Connect to the transmitter with ProLink II and read the value for Mass Flow Rate in the Process Variables window ( ProLink > Process Variables ).
Important Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem, then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a low- severity alarm is active. Procedure 1. Prepare the flowmeter: a.
Important Do not verify the zero or zero the flowmeter if a high-severity alarm is active. Correct the problem, then verify the zero or zero the flowmeter. You may verify the zero or zero the flowmeter if a low- severity alarm is active. Procedure 1. Prepare the flowmeter: a.
Terminology used with zero verification and zero calibration (continued) Table 2-2: Term Definition Live Zero The real-time bidirectional mass flow rate with no flow damping or mass flow cutoff ap- plied. An adaptive damping value is applied only when the mass flow rate changes dra- matically over a very short interval.
Part II Configuration and commissioning Chapters covered in this part: • Introduction to configuration and commissioning • Configure process measurement • Configure device options and preference.
Configuration and commissioning 18 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
3 Introduction to configuration and commissioning Topics covered in this chapter: • Configuration flowchart • Default values and ranges • Disable write‐protection on the transmitter configuration • Restore the factory configuration 3.
Configuration flowchart Figure 3-1: Integrate device with control system Configure device options and preferences Configure process measurement Configure mass flow measurement Configure volume flow me.
3.2 Default values and ranges See Section D.1 to view the default values and ranges for the most commonly used parameters. 3.3 Disable write-protection on the transmitter configuration ProLink II ProL.
Introduction to configuration and commissioning 22 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
4 Configure process measurement Topics covered in this chapter: • Configure mass flow measurement • Configure volume flow measurement for liquid applications • Configure gas standard volume (GSV.
Options for Mass Flow Measurement Unit The transmitter provides a standard set of measurement units for Mass Flow Measurement Unit , plus one user-defined special measurement unit. Different communications tools may use different labels for the units.
Overview A special measurement unit is a user-defined unit of measure that allows you to report process data, totalizer data, and inventory data in a unit that is not available in the transmitter. A special measurement unit is calculated from an existing measurement unit using a conversion factor.
4.1.2 Configure Flow Damping ProLink II ProLink > Configuration > Flow > Flow Damp ProLink III Device Tools > Configuration > Process Measurement > Flow Field Communicator Configure .
Valid values for Flow Damping Table 4-2: Core processor type Update Rate setting Valid damping values Standard Normal 0 , 0.2 , 0.4 , 0.8 , ... 51.2 Special 0 , 0.
Effect of Mass Flow Cutoff on volume measurement Mass Flow Cutoff does not affect volume measurement. Volume data is calculated from the actual mass data rather than the reported value. Interaction between Mass Flow Cutoff and AO Cutoff Mass Flow Cutoff defines the lowest mass flow value that the transmitter will report as measured.
4.2 Configure volume flow measurement for liquid applications The volume flow measurement parameters control how liquid volume flow is measured and reported.
Prerequisites Before you configure Volume Flow Measurement Unit , be sure that Volume Flow Type is set to Liquid . Procedure Set Volume Flow Measurement Unit to the unit you want to use. The default setting for Volume Flow Measurement Unit is l/sec (liters per second).
Options for Volume Flow Measurement Unit for liquid applications (continued) Table 4-3: Unit description Label ProLink II ProLink III Field Communicator Imperial gallons per minute Imp gal/min Imp gal.
a. x base units = y special units b. Volume Flow Conversion Factor = x/y 4. Enter Volume Flow Conversion Factor . 5. Set Volume Flow Label to the name you want to use for the volume flow unit. 6. Set Volume Total Label to the name you want to use for the volume total and volume inventory unit.
Interaction between Volume Flow Cutoff and AO Cutoff Volume Flow Cutoff defines the lowest liquid volume flow value that the transmitter will report as measured.
The GSV flow measurement parameters include: • Volume Flow Type • Standard Gas Density • Gas Standard Volume Flow Measurement Unit • Gas Standard Volume Flow Cutoff Restriction You cannot implement both liquid volume flow and gas standard volume flow at the same time.
Note ProLink II and ProLink III provide a guided method that you can use to calculate the standard density of your gas, if you do not know it. 4.3.3 Configure Gas Standard Volume Flow Measurement Unit.
Options for Gas Standard Volume Measurement Unit (continued) Table 4-4: Unit description Label ProLink II ProLink III Field Communicator Normal liter per second NLPS NLPS NLPS Normal liter per minute .
Base Gas Standard Volume Unit is the existing gas standard volume unit that the special unit will be based on. 2. Specify Base Time Unit . Base Time Unit is the existing time unit that the special unit will be based on. 3. Calculate Gas Standard Volume Flow Conversion Factor as follows: a.
Procedure Set Gas Standard Volume Flow Cutoff to the value you want to use. The default value for Gas Standard Volume Flow Cutoff is 0.0. The lower limit is 0.
- The frequency output will report the actual flow rate, and the actual flow rate will be used in all internal processing. • If the gas standard volume flow rate drops below 10 SLPM, both outputs will report zero flow, and 0 will be used in all internal processing.
Options for Flow Direction (continued) Table 4-5: Flow Direction setting Relationship to Flow Direction ar- row on sensor ProLink II ProLink III Field Communicator Bidirectional Bidirectional Bi direc.
Effect of Flow Direction on the mA output: Lower Range Value = 0 Figure 4-1: Flow Direction = Forward mA output -x 0 x Reverse flow Forward flow 20 12 4 Flow Direction = Reverse, Negate Forward mA out.
• Under conditions of reverse flow or zero flow, the mA output is 4 mA. • Under conditions of forward flow, up to a flow rate of 100 g/sec, the mA output varies between 4 mA and 20 mA in proportion to the flow rate.
Effect of Flow Direction on frequency outputs Flow Direction affects how the transmitter reports flow values via the frequency outputs. The frequency outputs are affected by Flow Direction only if Frequency Output Process Variable is set to a flow variable.
Effect of the Flow Direction parameter and actual flow direction on flow values reported via digital communications Table 4-8: Flow Direction setting Actual flow direction Forward Zero flow Reverse Fo.
4.5.1 Configure Density Measurement Unit ProLink II ProLink > Configuration > Density > Density Units ProLink III Device Tools > Configuration > Process Measurement > Density Field C.
4.5.2 Configure slug flow parameters ProLink II • ProLink > Configuration > Density > Slug High Limit • ProLink > Configuration > Density > Slug Low Limit • ProLink > Config.
The default value for Slug High Limit is 5.0 g/cm 3 . The range is 0.0 to 10.0 g/cm 3 . 3. Set Slug Duration to the number of seconds that the transmitter will wait for a slug flow condition to clear before performing the configured slug flow action. The default value for Slug Duration is 0.
Procedure Set Density Damping to the value you want to use. The default value is 1.6 seconds. The range depends on the core processor type and the setting of Update Rate , as shown in the following table: Core processor type Update Rate setting Density Damping range Standard Normal 0 to 51.
Density Damping controls the rate of change in the density process variable. Added Damping controls the rate of change reported via the mA output. If mA Output Process Variable is set to Density , and.
4.6.1 Configure Temperature Measurement Unit ProLink II ProLink > Configuration > Temperature > Temp Units ProLink III Device Tools > Configuration > Process Measurement > Temperatur.
Overview Damping is used to smooth out small, rapid fluctuations in process measurement. Damping Value specifies the time period (in seconds) over which the transmitter will spread changes in the reported process variable. At the end of the interval, the reported process variable will reflect 63% of the change in the actual measured value.
Prerequisites You will need the flow factor, density factor, and calibration pressure values for your sensor. • For the flow factor and density factor, see the product data sheet for your sensor. • For the calibration pressure, see the calibration sheet for your sensor.
Option Setup Polling for pressure a. Ensure that the primary mA output has been wired to support HART polling. b. Choose ProLink > Configuration > Polled Variables . c. Choose an unused polling slot. d. Set Polling Control to Poll As Primary or Poll as Secondary , and click Apply .
The flow factor is the percent change in the flow rate per PSI. When entering the value, reverse the sign. Example: If the flow factor is 0.000004 % per PSI, enter − 0.000004 % per PSI. 5. Enter Density Factor for your sensor. The density factor is the change in fluid density, in g/cm 3 /PSI.
9. If you want to use digital communications, click Apply , then perform the necessary host programming and communications setup to write temperature data to the transmitter at appropriate intervals.
Option Setup A user-configured static pressure val- ue a. Set Pressure Unit to the desired unit. b. Set Compensation Pressure to the desired value. Polling for pressure a.
Options for Pressure Measurement Unit (continued) Table 4-13: Unit description Label ProLink II ProLink III Field Communicator Millimeters water @ 68 °F mm Water @ 68°F mm Water @ 68°F mmH2O Millim.
Configure process measurement 58 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
5 Configure device options and preferences Topics covered in this chapter: • Configure response time parameters • Configure alarm handling • Configure informational parameters 5.1 Configure response time parameters You can configure the rate at which process data is polled and process variables are calculated.
Procedure 1. Set Update Rate as desired. Option Description Normal All process data is polled at the rate of 20 times per second (20 Hz). All process variables are calculated at 20 Hz. This option is appropriate for most applications. Special A single, user-specified process variable is polled at the rate of 100 times per sec- ond (100 Hz).
Special mode and process variable updates Table 5-1: Always polled and updated Updated only when the petroleum measurement application is disa- bled Never updated • Mass flow • Volume flow • Gas.
Option Description Normal Transmitter calculates process variables at the standard speed. Special Transmitter calculates process variables at a faster speed. 5.2 Configure alarm handling The alarm handling parameters control the transmitter’s response to process and device conditions.
Tip ProLink II allows you to set Fault Timeout in two locations. However, there is only one parameter, and the same setting is applied to all outputs. 5.
Option Description Informa- tional Actions when fault is detected: • The alarm is posted to the Alert List. • The status LED (if available) changes to red or yellow (depending on alarm se- verity). Actions when alarm clears: • The status LED (if available) returns to green and may or may not flash.
Status alarms and Status Alarm Severity (continued) Table 5-2: Alarm code Status message Default severity Notes Configurable? A023 Internal Totals Corrupt (Core Processor) Fault Applies only to flowmeters with the standard core processor. No A024 Program Corrupt (Core Processor) Fault Applies only to flowmeters with the standard core processor.
Status alarms and Status Alarm Severity (continued) Table 5-2: Alarm code Status message Default severity Notes Configurable? A109 Basic Event 2 On Informational Applies only to basic events. Yes A110 Frequency Output Satura- ted Informational Can be set to either Informational or Ignore , but cannot be set to Fault .
- Descriptor - Message - Date • Sensor parameters - Sensor Serial Number - Sensor Material - Sensor Liner Material - Sensor Flange Type 5.3.1 Configure Descriptor ProLink II ProLink > Configurati.
5.3.3 Configure Date ProLink II ProLink > Configuration > Device > Message ProLink III Device Tools > Configuration > Informational Parameters > Transmitter Field Communicator Config.
Overview Sensor Material lets you store the type of material used for your sensor’s wetted parts in transmitter memory. This parameter is not used in processing and is not required. Procedure 1. Obtain the material used for your sensor’s wetted parts from the documents shipped with your sensor, or from a code in the sensor model number.
To interpret the model number, refer to the product data sheet for your sensor. 2. Set Sensor Flange Type to the appropriate option. Configure device options and preferences 70 Micro Motion ® Model 1.
6 Integrate the meter with the control system Topics covered in this chapter: • Configure the transmitter channels • Configure the mA output • Configure the frequency output • Configure the discrete output • Configure events • Configure digital communications 6.
Postrequisites For each channel that you configured, perform or verify the corresponding input or output configuration. When the configuration of a channel is changed, the channel’s behavior will be.
Procedure Set mA Output Process Variable as desired. The default setting is Mass Flow Rate . Options for mA Output Process Variable The transmitter provides a basic set of options for mA Output Process Variable , plus several application-specific options.
• URV is the value of mA Output Process Variable represented by an output of 20 mA. The default value for URV depends on the setting of mA Output Process Variable . Enter URV in the measurement units that are configured for mA Output Process Variable .
Overview AO Cutoff (Analog Output Cutoff) specifies the lowest mass flow rate, volume flow rate, or gas standard volume flow rate that will be reported through the mA output.
Result: • If the mass flow rate drops below 15 g/sec but not below 10 g/sec: - The mA output will report zero flow. - The frequency output will report the actual flow rate. • If the mass flow rate drops below 10 g/sec, both outputs will report zero flow.
Valid values for Added Damping (continued) Table 6-3: Setting of Update Rate Process variable Update rate in effect Valid values for Added Damping Special 100 Hz variable (if assigned to the mA output) 100 Hz 0.0 , 0.04 , 0.12 , 0.30 , 0.64 , 1.32 , 2.
Note For some faults only: If Last Measured Value Timeout is set to a non-zero value, the transmitter will not implement the fault action until the timeout has elapsed. Procedure 1. Set mA Output Fault Action to the desired value. The default setting is Downscale .
The frequency output parameters include: • Frequency Output Polarity • Frequency Output Scaling Method • Frequency Output Maximum Pulse Width • Frequency Output Fault Action and Frequency Output Fault Value Restriction The process variable assigned to the primary mA output is automatically assigned to the frequency output.
Options for Frequency Output Polarity (continued) Table 6-5: Polarity Reference voltage (OFF) Pulse voltage (ON) Active Low As determined by power sup- ply, pull-up resistor, and load (see the installation manual for your transmitter) 0 6.
Rate Factor The maximum flow rate that you want the frequency output to report. Above this rate, the transmitter will report A110: Frequency Output Saturated .
6.3.3 Configure Frequency Output Maximum Pulse Width ProLink II ProLink > Configuration > Frequency/Discrete Output > Frequency > Freq Pulse Width ProLink III Device Tools > Configurati.
6.3.4 Configure Frequency Output Fault Action and Frequency Output Fault Level ProLink II • ProLink > Configuration > Frequency/Discrete Output > Frequency > Freq Fault Action • ProLin.
CAUTION! If you set mA Output Fault Action or Frequency Output Fault Action to None , be sure to set Digital Communications Fault Action to None . If you do not, the output will not report actual process data, and this may result in measurement errors or unintended consequences for your process.
The default setting for Discrete Output Source is Flow Direction . Options for Discrete Output Source Options for Discrete Output Source Table 6-8: Option Label Condition Discrete output voltage ProLi.
Configure Flow Switch parameters ProLink II • ProLink > Configuration > Flow > Flow Switch Setpoint • ProLink > Configuration > Flow > Flow Switch Variable • ProLink > Confi.
Overview Discrete outputs have two states: ON (active) and OFF (inactive). Two different voltage levels are used to represent these states. Discrete Output Polarity controls which voltage level represents which state. Procedure Set Discrete Output Polarity as desired.
Typical discrete output circuit Figure 6-1: A. 15 V (Nom) B. 3.2 K Ω C. Out+ D. Out − 6.4.3 Configure Discrete Output Fault Action ProLink II ProLink > Configuration > Frequency/Discrete Out.
Procedure Set Discrete Output Fault Action as desired. The default setting is None . Options for Discrete Output Fault Action Options for Discrete Output Fault Action Table 6-10: Label Discrete output.
6.5.1 Configure a basic event ProLink II ProLink > Configuration > Events ProLink III Device Tools > Configuration > Events > Basic Events Field Communicator Not available Overview A basic event is used to provide notification of process changes.
defined setpoints. You can define up to five enhanced events. For each enhanced event, you can assign one or more actions that the transmitter will perform if the enhanced event occurs. Procedure 1. Select the event that you want to configure. 2. Specify Event Type .
Options for Enhanced Event Action Options for Enhanced Event Action Table 6-11: Action Label ProLink II ProLink III Field Communicator Standard None (default) None None None Start sensor zero Start Se.
6.6.1 Configure HART/Bell 202 communications ProLink II ProLink > Configuration > Device > Digital Comm Settings ProLink III Device Tools > Configuration > Communications > Communica.
Tip In typical installations, burst mode is disabled. Enable burst mode only if another device on the network requires burst mode communication. 4. (Optional) Configure HART Variables .
Label Description ProLink II ProLink III Field Communi- cator Dynamic vars & PV current Process Variables/ Current Process variables/ current The transmitter sends PV, SV, TV, and QV values in measurement units and the PV’s actual milliamp reading in each burst (e.
Options for HART variables (continued) Table 6-12: Process variable Primary Varia- ble (PV) Secondary Variable (SV) Third Variable (TV) Fourth Varia- ble (QV ) Mass total ✓ Line (Gross) Volume total.
• Protocol • Modbus Address ( Slave Address ) • Parity , Stop Bits , and Baud Rate • Floating-Point Byte Order • Additional Communications Response Delay Restriction To configure Floating-Point Byte Order or Additional Communications Response Delay , you must use ProLink II.
Bit structure of floating-point bytes Table 6-14: Byte Bits Definition 1 SEEEEEEE S=Sign E=Exponent 2 EMMMMMMM E=Exponent M=Mantissa 3 MMMMMMMM M=Mantissa 4 MMMMMMMM M=Mantissa 6. (Optional) Set Additional Communications Response Delay in “delay units.
Options for Digital Communications Fault Action Options for Digital Communications Fault Action Table 6-15: Label Description ProLink II ProLink III Field Communicator Upscale Upscale Upscale • Process variable values indicate that the value is greater than the upper sensor lim- it.
Integrate the meter with the control system 100 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
7 Completing the configuration Topics covered in this chapter: • Test or tune the system using sensor simulation • Back up transmitter configuration • Enable write‐protection on the transmitter configuration 7.
Option Required values Sine Period Minimum Maximum 4. For density, set Wave Form as desired and enter the required values. Option Required values Fixed Fixed Value Sawtooth Period Minimum Maximum Sine Period Minimum Maximum 5. For temperature, set Wave Form as desired and enter the required values.
• All mass flow rate, temperature, and density values shown on the display or reported via outputs or digital communications • The mass total and mass inventory values • All volume calculations .
7.3 Enable write-protection on the transmitter configuration ProLink II ProLink > Configuration > Device > Enable Write Protection ProLink III Device Tools > Configuration > Write-Prote.
Part III Operations, maintenance, and troubleshooting Chapters covered in this part: • Transmitter operation • Measurement support • Troubleshooting Operations, maintenance, and troubleshooting .
Operations, maintenance, and troubleshooting 106 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
8 Transmitter operation Topics covered in this chapter: • Record the process variables • View process variables • View transmitter status using the status LED • View and acknowledge status alarms • Read totalizer and inventory values • Start and stop totalizers and inventories • Reset totalizers • Reset inventories 8.
8.2 View process variables ProLink II ProLink > Process Variables ProLink III View the desired variable on the main screen under Process Variables .
Status LED states (continued) Table 8-1: LED behavior Alarm condition Description Solid red Active high-severity alarm Alarm condition that will cause measurement error (outputs in fault) 8.
8.4.2 View and acknowledge alerts using ProLink III You can view a list containing all alerts that are active, or inactive and have been unacknowleged. From this list, you can acknowlege individual alerts or choose to acknowledge all alerts at once. 1.
• To refresh the list of active or unacknowledged alarms, press Service Tools > Alerts > Refresh Alerts . 8.4.4 Alarm data in transmitter memory The transmitter maintains three sets of data for every alarm that is posted.
Tip You can use the inventories to keep a running total of mass or volume across multiple totalizer resets. 8.6 Start and stop totalizers and inventories ProLink II ProLink > Totalizer Control >.
8.7 Reset totalizers ProLink II ProLink > Totalizer Control > Reset Mass Total ProLink > Totalizer Control > Reset Volume Total ProLink > Totalizer Control > Reset Gas Volume Total P.
Overview When you reset an inventory, the transmitter sets its value to 0. It does not matter whether the inventory is started or stopped. If the inventory is started, it continues to track process measurement. Tip When you reset a single inventory, the values of other inventories are not reset.
9 Measurement support Topics covered in this chapter: • Options for measurement support • Use Smart Meter Verification • Zero the flowmeter • Validate the meter • Perform a (standard) D1 and D2 density calibration • Perform a D3 and D4 density calibration (T‐Series sensors only) • Perform temperature calibration 9.
9.2.1 Smart Meter Verification requirements To use Smart Meter Verification, the transmitter must be paired with an enhanced core processor, and the Smart Meter Verification option must be ordered for the transmitter.
Smart Meter Verification has an output mode called Continuous Measurement that allows the transmitter to keep measuring while the test is in progress. If you choose to run the test in Last Measured Value or Fault modes instead, the transmitter outputs will be held constant for the two minute duration of the test.
Postrequisites View the test results and take any appropriate actions. Run a Smart Meter Verification test using ProLink III 1. Choose Device Tools > Diagnostics > Meter Verification > Run Test . You may need to wait a few seconds while ProLink II synchronizes its database with the transmitter data.
Option Description Outputs Held at Last Value During the test, all outputs will report the last measured value of their assigned process variable. The test will run for approximately 140 seconds. Outputs Held at Fault During the test, all outputs will go to their configured fault action.
2. (Optional) Click Next to view and print a test report. 3. (Optional) Click Export Data to CSV File to save the data to a file on your PC. View test result data using ProLink III 1. Choose Device Tools > Diagnostics > Meter Verification and click Previous Test Results .
• If the meter fails the second test, the flow tubes may be damaged. Use your process knowledge to determine the possibilities for damage and the appropriate actions for each. These actions might include removing the meter from service and physically inspecting the tubes.
2. To schedule a single test or the first test in recurring execution, specify a value for Hours Until Next Run . 3. To schedule recurring execution, specify a value for Hours Between Recurring Runs . 4. To disable scheduled execution: • To disable execution of a single scheduled test, set Hours Until Next Run to 0.
Important In most cases, the factory zero is more accurate than the field zero. Do not zero the flowmeter unless one of the following is true: • The zero is required by site procedures. • The stored zero value fails the Zero Verification procedure.
• Set Zero Time to a lower value, then retry. • If the zero continues to fail, contact Micro Motion. Tip You can restore the factory zero using a communications tool such as ProLink II. Restore the factory zero only if your flowmeter was purchased as a unit, it was zeroed at the factory, and you are using the original components.
Postrequisites Restore normal flow through the sensor by opening the valves. Need help? If the zero fails: • Ensure that there is no flow through the sensor, then retry. • Remove or reduce sources of electromechanical noise, then retry. • Set Zero Time to a lower value, then retry.
4. Modify Zero Time , if desired. Zero Time controls the amount of time the transmitter takes to determine its zero- flow reference point. The default Zero Time is 20 seconds. For most applications, the default Zero Time is appropriate. 5. Click Calibrate Zero .
e. Observe the drive gain, temperature, and density readings. If they are stable, check the Live Zero or Field Verification Zero value. If the average value is close to 0, you should not need to zero the flowmeter. 2. Press Service Tools > Maintenance > Zero Calibration > Perform Auto Zero .
Overview Meter validation compares flowmeter measurements reported by the transmitter to an external measurement standard. If the transmitter value for mass flow, volume flow, or density measurement is significantly different from the external measurement standard, you may want to adjust the corresponding meter factor.
3. Configure the meter factor in the transmitter. Example: Calculating the meter factor for mass flow The flowmeter is installed and validated for the first time. The mass flow measurement from the transmitter is 250.27 lb. The mass flow measurement from the reference device is 250 lb.
3. Ensure that the calculated meter factor is between 0.8 and 1.2, inclusive. If the meter factor is outside these limits, contact Micro Motion customer service.
Procedure See Figure 9‐1 . D1 and D2 density calibration using ProLink II Figure 9-1: Enter density of D1 fluid Calibration in Progress light turns green Calibration in Progress light turns red D1 c.
• If LD Optimization is enabled on your meter, disable it. To do this, choose Device Tools > Configuration > LD Optimization . LD Optimization is used only with large sensors in hydrocarbon applications. In some installations, only Micro Motion customer service has access to this parameter.
9.5.3 Perform a D1 and D2 density calibration using the Field Communicator Prerequisites • During density calibration, the sensor must be completely filled with the calibration fluid, and flow through the sensor must be at the lowest rate allowed by your application.
D1 and D2 density calibration using the Field Communicator Figure 9-3: Enter density of D1 fluid Density Calibration Complete message Calibration in Progress message D1 calibration Close shutoff valve.
• Perform the D3 calibration if you have one calibrated fluid. • Perform both the D3 and D4 calibrations if you have two calibrated fluids (other than air and water). The calibrations must be performed without interruption, in the order shown. Make sure that you are prepared to complete the process without interruption.
D3 or D3 and D4 density calibration using ProLink II Figure 9-4: Enter density of D3 fluid Calibration in Progress light turns green Calibration in Progress light turns red D3 calibration Close shutof.
- Minimum difference of 0.1 g/cm 3 between the density of the D4 fluid and the density of the D3 fluid. The density of the D4 fluid must be greater than the density of the D3 fluid. - Minimum difference of 0.1 g/cm 3 between the density of the D4 fluid and the density of water.
• For D3 density calibration, the D3 fluid must meet the following requirements: - Minimum density of 0.6 g/cm 3 - Minimum difference of 0.1 g/cm 3 between the density of the D3 fluid and the density of water. The density of the D3 fluid may be either greater or less than the density of water.
D3 or D3 and D4 density calibration using the Field Communicator Figure 9-6: Enter density of D3 fluid Density Calibration Complete message Calibration in Progress message D3 calibration Close shutoff.
Important Consult Micro Motion before performing a temperature calibration. Under normal circumstances, the temperature circuit is stable and should not need an adjustment.
Temperature calibration using ProLink III Figure 9-8: Enter temperature of low- temperature fluid T emperature Offset calibration W ait until sensor achieves thermal equilibrium Fill sensor with low- .
Measurement support 142 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
10 Troubleshooting Topics covered in this chapter: • Status LED states • Status alarms • Flow measurement problems • Density measurement problems • Temperature measurement problems • Milli.
10.1 Status LED states The status LED on the transmitter indicates whether or not alarms are active. If alarms are active, view the alarm list to identify the alarms, then take appropriate action to correct the alarm condition.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A003 No Sensor Response The transmitter is not receiving one or more basic electrical sig- nals from the sensor.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A005 Mass Flow Rate Overrange The sensor is signaling a flow rate that is out of range for the sen- sor. 1. If other alarms are present, resolve those alarm conditions first.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A008 Density Overrange The sensor is signaling a density reading below 0 g/cm 3 or above 10 g/cm 3 .
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A010 Calibration Failure This alarm is typically caused by flow through the sensor during the zero, or by a zero offset result that is out of range.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A016 Sensor RTD Failure The sensor RTD is signaling a resistance that is out of range for the sensor. 1. Check the wiring between the sensor and the transmitter.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A019 RAM Error (Transmitter) Power to the transmitter must be cycled to clear this alarm. 1. Check that all wiring compartment covers are installed prop- erly.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A026 Sensor/Transmitter Communi- cations Failure The transmitter has lost communication with the core processor on the sensor.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A031 Low Power The core processor on the sensor is not receiving sufficient pow- er. Check the wiring between the transmitter and the sensor. Power to the transmitter must be cycled to clear this alarm.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A100 mA Output 1 Saturated The calculated mA output value is outside of the meter's config- ured range. 1. Check the Upper Range Value and Lower Range Value parame- ters.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A110 Frequency Output Saturated The calculated frequency output is outside the configured range. 1. Check the Frequency Output Scaling Method parameter.
Status alarms and recommended actions (continued) Table 10-2: Alarm code Description Recommended actions A117 Density Overrange (Petrole- um) 1. Check your process conditions against the values reported by the flowmeter. 2. Verify the configuration of the petroleum measurement ta- ble type and density.
10.3 Flow measurement problems Flow measurement problems and recommended actions Table 10-3: Problem Possible causes Recommended actions Flow indication at no flow conditions or zero offset • Misali.
Flow measurement problems and recommended actions (continued) Table 10-3: Problem Possible causes Recommended actions Erratic non-zero flow rate when flow is steady • Slug flow • Damping value too.
10.4 Density measurement problems Density measurement problems and recommended actions Table 10-4: Problem Possible causes Recommended actions Inaccurate density reading • Problem with process fluid.
10.5 Temperature measurement problems Temperature measurement problems and recommended actions Table 10-5: Problem Possible causes Recommended actions Temperature reading significantly different from process temper- ature • RTD failure • Wiring problem • Check junction box for moisture or verdi- gris.
10.6 Milliamp output problems Milliamp output problems and recommended actions Table 10-6: Problem Possible causes Recommended actions No mA output • Wiring problem • Circuit failure • Channel not configured for desired output • Check the power supply and power supply wiring.
Milliamp output problems and recommended actions (continued) Table 10-6: Problem Possible causes Recommended actions mA output consis- tently out of range • Incorrect process variable or units assig.
Frequency output problems and recommended actions Table 10-7: Problem Possible causes Recommended actions No frequency output • Stopped totalizer • Process condition below cutoff • Fault conditi.
For more information on using sensor simulation using ProLink II, see Section 7.1 . 10.9 Check power supply wiring If the power supply wiring is damaged or improperly connected, the transmitter may not receive enough power to operate properly. Prerequisites You will need the installation manual for your transmitter.
Procedure 1. Before opening the wiring compartments, disconnect the power source. CAUTION! If the transmitter is in a hazardous area, wait five minutes after disconnecting the power. 2. Verify that the transmitter is connected to the sensor according to the information provided in your transmitter installation manual.
Procedure 1. Test the mA output(s). a. Choose ProLink > Test > Fix Milliamp 1 or ProLink > Test > Fix Milliamp 2 . b. Enter 4 mA in Set Output To . c. Click Fix mA . d. Read the mA current at the receiving device and compare it to the transmitter output.
• If the mA output reading was significantly off (±200 microamps), or if at any step the reading was faulty, verify the wiring between the transmitter and the remote device, and try again. • If the discrete output reading is reversed, check the setting of Discrete Output Polarity .
e. Click UnFix FO . 3. Test the discrete output(s). a. Choose Device Tools > Diagnostics > Testing > Discrete Output Test . b. Set Fix To: to ON . c. Verify the signal at the receiving device. d. Set Fix To: to OFF . e. Verify the signal at the receiving device.
2. Test the frequency output(s). Note If the Weights & Measures application is enabled on the transmitter, it is not possible to perform a loop test of the frequency output, even when the transmitter is unsecured. a. Press Service Tools > Simulate > Simulate Outputs > Frequency Output Test , and choose the frequency output level.
- Terminate the shielding at the output device. If this is impossible, terminate the shielding at the cable gland or conduit fitting. - Do not terminate the shielding inside the wiring compartment. - 360-degree termination of shielding is unnecessary.
When HART Address is changed, some configuration tools will automatically change Loop Current Mode . Tip Always verify Loop Current Mode after setting or changing HART Address . Procedure 1. Set HART Address as appropriate for your HART network. The default address is 0.
10.19 Check mA Output Fault Action mA Output Fault Action controls the behavior of the mA output if the transmitter encounters an internal fault condition. If the mA output is reporting a constant value below 4 mA or above 20 mA, the transmitter may be in a fault condition.
• For the relevant status alarms, change the setting of Alarm Severity to Ignore . 3. If there are no active fault conditions, continue troubleshooting. 10.23 Check Flow Direction If Flow Direction is set inappropriately for your process, the transmitter may report unexpected flow values or totals.
3. Monitor the density of your process fluid output under normal process conditions. 4. Check the settings of Slug Low Limit , Slug High Limit , and Slug Duration . Tip You can reduce the occurrence of slug flow alarms by setting Slug Low Limit to a lower value, Slug High Limit to a higher value, or Slug Duration to a higher value.
Possible causes and recommended actions for excessive (saturated) drive gain (continued) Table 10-8: Possible cause Recommended actions Incorrect sensor characteriza- tion Verify the characterization parameters.
Possible causes and recommended actions for low pickoff voltage Table 10-10: Possible cause Recommended actions Air entrainment • Increase the inlet or back pressure at the sensor. • If a pump is located upstream from the sensor, increase the distance between the pump and sensor.
Possible causes and recommended actions for electrical shorts (continued) Table 10-11: Possible cause Recommended action Liquid or moisture inside the sensor case Contact Micro Motion. Internally shorted feedthrough Contact Micro Motion. Faulty cable Replace the cable.
Note The CMF400 fixed resistor applies only to certain specific CMF400 releases. Contact Micro Motion for more information. There should be no open circuits, that is, no infinite resistance readings. The left pickoff and right pickoff readings should be the same or very close ( ±5 Ω ).
10.29 Check the core processor LED The core processor has an LED that indicates different meter conditions. 1. Maintain power to the transmitter. 2. Remove the core processor lid. The core processor is intrinsically safe and can be opened in all environments.
Standard core processor LED states (continued) Table 10-13: LED state Description Recommended actions Core processor internal failure The meter requires factory service. Enhanced core processor LED states Table 10-14: LED state Description Recommended action Solid green Normal operation No action required.
Terminal pair Function Expected resistance 3–4 RS-485/A and RS-485/B 40 k Ω to 50 k Ω 2–3 VDC– and RS-485/A 20 k Ω to 25 k Ω 2–4 VDC– and RS-485/B 20 k Ω to 25 k Ω 5. If any resistance measurements are lower than specified, the core processor may not be able to communicate with a transmitter or a remote host.
Appendix A Using ProLink II with the transmitter Topics covered in this appendix: • Basic information about ProLink II • Connect with ProLink II • Menu maps for ProLink II A.1 Basic information about ProLink II ProLink II is a software tool available from Micro Motion.
ProLink II messages As you use ProLink II with a Micro Motion transmitter, you will see a number of messages and notes. This manual does not document all of these messages and notes. Important The user is responsible for responding to messages and notes and complying with all safety messages.
• You cannot make concurrent connections if the connections use the same terminals. You can make concurrent connections if the connections use different terminals.
4. Start ProLink II. 5. Choose Connection > Connect to Device . 6. Set Protocol to Service Port . Tip Service port connections use standard connection parameters and a standard address. You do not need to configure them here. 7. Set the COM Port value to the PC COM port that you are using for this connection.
Procedure 1. Attach the signal converter to the serial port or USB port on your PC. 2. To connect directly to the transmitter terminals: a. Connect the leads from the signal converter to terminals 21 and 22. Tip HART connections are not polarity-sensitive.
Connection over local loop Figure A-3: A C D E R1 R3 R2 B A. PC B. Signal converter C. Any combination of resistors R1, R2, and R3 as necessary to meet HART communication resistance requirements D. DCS or PLC E. Transmitter Note This figure shows a serial port connection.
Connection over multidrop network Figure A-4: B A C D A. Signal converter B. 250–60 0 Ω resistance C. Devices on the network D. Master device 5. Start ProLink II. 6. Choose Connection > Connect to Device . 7. Set Protocol to HART Bell 202 . Tip HART/Bell 202 connections use standard connection parameters.
Option Description Primary Use this setting if no other host is on the network. The Field Communicator is not a host. 12. Click Connect . Need help? If an error message appears: • Verify the HART address of the transmitter. • Ensure that you have specified the correct COM port.
Connection to transmitter terminals Figure A-5: A B C A. PC B. Signal converter C. Transmitter Note This figure shows a serial port connection. USB connections are also supported. 3. To connect over the RS-485 network: a. Attach the leads from the signal converter to any point on the network.
6. Set the connection parameters to the values configured in the transmitter. If your transmitter has not been configured, use the default values shown here.
A.3 Menu maps for ProLink II Main menu Figure A-7: File Preferences • Use External T emperature • Enable Inventory T otals Reset • Enable External Pressure Compensation • Copper RTD Installed .
Main menu (continued) Figure A-8: Data Logging* Enable/Disable Custody T ransfer T ools Plug-ins ProLink Configuration Output Levels Process V ariables Status Alarm Log Diagnostic Information Calibrat.
Configuration menu Figure A-9: Density • Dens Units • Dens Damping • Slug High Limit • Slug Low Limit • Slug Duration • Lo w Density Cutoff • K1 • K2 • FD • D1 • D2 • T emp Coe.
Configuration menu (continued) Figure A-10: T Series • FTG • FFQ • DTG • DFQ1 • DFQ2 • K3 • D3 • D4 • K4 Frequency/Discrete output • Frequency • T ertiary V ariable • Freq Fact.
Configuration menu (continued) Figure A-11: T emperature • T emp Units • T emp Cal Factor • T emp Damping • External T emperature • External RTD ProLink > Configuration Additional configu.
Configuration menu (continued) Figure A-12: ProLink > Configuration Device • Model • Manufacturer • Hardware Rev • Distributor • Software Rev • ET O • CP Software Rev • CP ET O • .
Configuration menu (continued) Figure A-13: ProLink > Configuration Additional configuration options Discrete Events • Event Name • Event T ype • Process V ariable • Low Setpoint (A) • Hi.
Configuration menu (continued) Figure A-15: ProLink > Configuration V ariable mapping • PV is • SV is • TV is • QV is System • Weights and Measures Approval • Software Rev • T otalize.
Appendix B Using ProLink III with the transmitter Topics covered in this appendix: • Basic information about ProLink III • Connect with ProLink III • Menu maps for ProLink III B.1 Basic information about ProLink III ProLink III is a configuration and service tool available from Micro Motion.
ProLink III messages As you use ProLink III with a Micro Motion transmitter, you will see a number of messages and notes. This manual does not document all of these messages and notes. Important The user is responsible for responding to messages and notes and complying with all safety messages.
• You cannot make concurrent connections if the connections use the same terminals. You can make concurrent connections if the connections use different terminals.
4. Start ProLink III. 5. Choose Connect to Physical Device . 6. Set Protocol to Service Port . Tip Service port connections use standard connection parameters and a standard address. You do not need to configure them here. 7. Set the PC Port value to the PC COM port that you are using for this connection.
Procedure 1. Attach the signal converter to the serial port or USB port on your PC. 2. To connect directly to the transmitter terminals: a. Connect the leads from the signal converter to terminals 21 and 22. Tip HART connections are not polarity-sensitive.
Connection over local loop Figure B-3: A C D E R1 R3 R2 B A. PC B. Signal converter C. Any combination of resistors R1, R2, and R3 as necessary to meet HART communication resistance requirements D. DCS or PLC E. Transmitter Note This figure shows a serial port connection.
Connection over multidrop network Figure B-4: B A C D A. Signal converter B. 250–60 0 Ω resistance C. Devices on the network D. Master device 5. Start ProLink III. 6. Choose Connect to Physical Device . 7. Set Protocol to HART Bell 202 . Tip HART/Bell 202 connections use standard connection parameters.
Option Description Primary Use this setting if no other host is on the network. The Field Communicator is not a host. 12. Click Connect . Need help? If an error message appears: • Verify the HART address of the transmitter. • Ensure that you have specified the correct COM port.
Connection to transmitter terminals Figure B-5: A B C A. PC B. Signal converter C. Transmitter Note This figure shows a serial port connection. USB connections are also supported. 3. To connect over the RS-485 network: a. Attach the leads from the signal converter to any point on the network.
6. Set the connection parameters to the values configured in the transmitter. If your transmitter has not been configured, use the default values shown here.
B.3 Menu maps for ProLink III Device Tools: Main Figure B-7: Configuration: Process Measurement Figure B-8: Using ProLink III with the transmitter Configuration and Use Manual 209.
Configuration: I/O Figure B-9: Configuration: Events Figure B-10: Using ProLink III with the transmitter 210 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
Configuration: Communications Figure B-11: Configuration: Informational Parameters Figure B-12: Using ProLink III with the transmitter Configuration and Use Manual 211.
Device Tools: Calibration Figure B-13: Calibration: Density Calibration Figure B-14: Using ProLink III with the transmitter 212 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
Calibration: Temperature Calibration Figure B-15: Device Tools: Configuration Transfer Figure B-16: Using ProLink III with the transmitter Configuration and Use Manual 213.
Diagnostics: Testing Figure B-17: Diagnostics: Meter Verification Figure B-18: Using ProLink III with the transmitter 214 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
Device Tools: Trending Figure B-19: Using ProLink III with the transmitter Configuration and Use Manual 215.
Using ProLink III with the transmitter 216 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
Appendix C Using the Field Communicator with the transmitter Topics covered in this appendix: • Basic information about the Field Communicator • Connect with the Field Communicator • Menu maps for the Field Communicator C.
If Micro Motion is not listed, or you do not see the required device description, use the Field Communicator Easy Upgrade Utility to install the device description, or contact Micro Motion. Field Communicator menus and messages Many of the menus in this manual start with the On-Line menu.
Field Communicator connection to transmitter terminals Figure C-1: A B C A. Field Communicator B. 250–60 0 Ω resistance C. Transmitter terminals 2. To connect to a point in the local HART loop, attach the leads from the Field Communicator to any point in the loop and add resistance as necessary.
Field Communicator connection to multidrop network Figure C-3: A B C D A. Field Communicator B. 250–60 0 Ω resistance C. Devices on the network D. Master device 4. Turn on the Field Communicator and wait until the main menu is displayed. 5. If you are connecting across a multidrop network: a.
On-Line menu Figure C-4: Configure 1 Manual Setup 2 Alert Setup Service T ools 1 Alerts 2 V ariables 3 T rends 4 Maintenance 5 Simulate On-Line Menu 2 3 Overview 1 Check Status 2 Primary Purpose V ari.
Overview menu Figure C-5: Identification 1 T ag 2 Model 3 Xmtr Serial Num 4 Sensor Serial Num 5 Date 6 Descriptor 7 Message 1 Revisions 1 Universal 2 Field Device 3 DD Revision 4 T ransmitter Software 5 CP Software 6 ETO Number Mat. of Construction 1 T ube Wetted Mat.
Configure menu Figure C-6: Manual Setup 1 Characterize 2 Measurements 3 Inputs/Outputs 4 Info Parameters 5 Communications 2 1 Alert Setup 1 Configure Alerts 2 Discrete Output 3 Discrete Events On-Line.
Manual Setup menu Figure C-7: Characterize 1 Sensor T ype 2 Sensor T ag Parameters 1 2 Inputs/Outputs 1 Set Up Channels 2 Set Up mA Output 3 Set Up Frequency Output 4 Set Up Discrete Output 5 Set Up R.
Manual Setup menu: Characterize Figure C-8: On-Line Menu > 2 Configure > 1 Manual Setup > 1 Characterize 2 1 2 1 Sensor T ype Straight T ube Curved T ube Density Parameters 1 D1 2 D2 3 DT 4 D.
Manual Setup menu: Measurements Figure C-9: * Displayed only if V olume Flow T ype = Liquid. Menu numbers are adjusted as required. On-Line Menu > 2 Configure > 1 Manual Setup > 2 Measurement.
Manual Setup menu: I/O Figure C-10: On-Line Menu > 2 Configure > 1 Manual Setup > 4 Inputs/Outputs Set Up mA Output 1 Primary V ariable 2 mA Output Settings 3 mA Fault Settings Set Up Frequen.
Manual Setup menu: I/O (continued) Figure C-11: Discrete Output 1 DO Assignment 2 DO Polarity 3 DO Fault Action 4 Flow Switch Source 5 Flow Switch Setpoint 6 Hysteresis (0.
Alert Setup menu Figure C-12: On-Line Menu > 2 Configure > 2 Alert Setup Configure Alerts 1 Fault T imeout 2 MAO Fault Action 3 MAO Fault Level 4 FO Fault Action 5 FO Fault Level 6 Comm Fault Ac.
Service Tools menu Figure C-13: 2 Process V ariables 1 Mass Flow Rate 2 V olume Flow Rate 3 Gas Standard V olume 4 Density 5 T emperature Diagnostic V ariables 1 Drive Gain 2 LPO Amplitude 3 RPO Ampli.
Service Tools menu: Variables Figure C-14: * If V olume Flow T ype = GSV , GSV variables are displayed. On-Line Menu > 3 Service T ools > 2 V ariables Process V ariables 1 Mass Flow Rate 2 V olu.
Service Tools menu: Maintenance Figure C-15: On-Line Menu > 3 Service T ools > 4 Maintenance Routine Maintenance 1 T rim mA Output 2 Meter V erification * 1 Zero Calibration 1 Mass Flow Rate 2 V.
Service Tools menu: Simulate Figure C-16: Simulate Outputs 1 mA Output Loop T est 2 Frequency Output T est 3 Discrete Output T est On-Line Menu > 3 Service T ools > 5 Simulate 1 Simulate Sensor .
Using the Field Communicator with the transmitter 234 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
Appendix D Default values and ranges D.1 Default values and ranges The default values and ranges represent the typical factory transmitter configuration. Depending on how the transmitter was ordered, certain values may have been configured at the factory and are not represented in the default values and ranges.
Transmitter default values and ranges (continued) Table D-1: Type Parameter Default Range Comments Density units g/cm 3 Density cutoff 0.2 g/cm 3 0.0 – 0.5 g/cm 3 D1 0 g/cm 3 D2 1 g/cm 3 K1 1000 µsec 1000 – 50,000 µsec K2 50,000 µsec 1000 – 50,000 µsec FD 0 Temp Coefficient 4.
Transmitter default values and ranges (continued) Table D-1: Type Parameter Default Range Comments Base volume time sec Volume flow conversion factor 1 Variable map- ping Primary variable Mass flow Secondary variable Density Tertiary variable Mass flow Quaternary variable Volume flow mA output 1 Primary variable Mass flow LRV –200.
Transmitter default values and ranges (continued) Table D-1: Type Parameter Default Range Comments Fault action Downscale AO fault level – downscale 2.0 mA 1.0 – 3.6 mA AO fault level – upscale 22 mA 21.0 – 24.0 mA Last measured value timeout 0.
Transmitter default values and ranges (continued) Table D-1: Type Parameter Default Range Comments Polarity Active low Backlight in- tensity 63 0 – 63 Refresh rate 200 milliseconds 100 – 10,000 mi.
Default values and ranges 240 Micro Motion ® Model 1500 Transmitters with Analog Outputs.
Appendix E Transmitter components and installation wiring Topics covered in this appendix: • Installation types • Power supply terminals • Input/output (I/O) wiring terminals E.1 Installation types Model 1500 and Model 2500 transmitters can be installed two different ways, only one of which applies to your specific installation.
• Remote core processor with remote sensor – A remote core process with remote sensor installation separates all three components – transmitter, core processor, and sensor – all of which are installed separately. A 4-wire cable connects the transmitter to the core processor, and a 9-wire cable connects the core processor to the sensor.
E.2 Power supply terminals Power supply wiring terminals Figure E-3: A B A. Primary power supply (DC) B. Power supply jumper to other Model 1500 or Model 2500 transmitters (optional) Transmitter compo.
E.3 Input/output (I/O) wiring terminals I/O wiring terminals Figure E-4: A B C D A. mA/HART B. Unused C. Frequency output or discrete output D. Service port or Modbus/RS‐485 Transmitter components a.
Appendix F NE 53 history F.1 NE 53 history Date Version Type Change Operating in- struction 08/2000 1.x Expansion Added writing of the device tag using Modbus 3600204 A Adjustment Improved communication handling with the HART Tri-Loop product Feature Indication of outputs option board type appears on display at power-up 05/2001 2.
Date Version Type Change Operating in- struction The display start/stop totalizers function can be enabled or disabled Petroleum measurement application improve- ments Live zero available as display v.
Date Version Type Change Operating in- struction 09/2006 5.x Expansion Discrete output assignable as a flow switch 20001715 B Discrete output fault indication configurability Discrete input support fo.
Date Version Type Change Operating in- struction Adjustment The following combinations are not allowed: • mA Output Fault Action = None and Digital Communications Fault Action = NAN • Frequency Ou.
Index A Added Damping 76 Additional Communications Response Delay 96 address HART address 93 Modbus address 96 air calibration , See calibration, density alarms alarm codes 144 configuring alarm handl.
troubleshooting 172 volume flow 32 D damping Added Damping 76 density damping 47 flow damping 26 interaction between Added Damping and process variable damping 77 on mA outputs 76 temperature damping .
Fault Timeout configuring 62 effect on Fault Action 62 Field Communicator connecting to the transmitter 218 device description (DD) 217 menu maps 220 overview 217, 218 startup connection 8 Floating-Po.
M mA outputs Added Damping configuring 76 interaction with density damping 48 interaction with flow damping 27 AO cutoff configuring 74 interaction with volume flow cutoff 33 configuring 72 Fault Acti.
primary variable (PV) 95 process measurement effect of Calculation Speed 61 effect of Update Rate 59, 60 process variables See also density measurement See also gas standard volume flow measurement Se.
status See also alarms status alarms , See alarms status LED 108, 144 T temperature calibration , See calibration, temperature temperature measurement configuring 49 damping configuring 50 effect on p.
Z zero procedure using ProLink II 124 using ProLink III 125 using the Field Communicator 126 using the zero button 123 restore factory zero using ProLink II 124 using ProLink III 125 using the Field C.
*MMI-20019023* MMI-20019023 Rev AA 2012 Micro Motion Inc. USA Worldwide Headquarters 7070 Winchester Circle Boulder, Colorado 80301 T +1 303-527-5200 T +1 800-522-6277 F +1 303-530-8459 www.micromotion.com Micro Motion Europe Emerson Process Management Neonstraat 1 6718 WX Ede The Netherlands T +31 (0) 318 495 555 F +31 (0) 318 495 556 www.
デバイスEmerson 1500の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Emerson 1500をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはEmerson 1500の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Emerson 1500の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Emerson 1500で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Emerson 1500を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はEmerson 1500の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Emerson 1500に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちEmerson 1500デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。