Emersonメーカー5081FGの使用説明書/サービス説明書
ページ先へ移動 of 88
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 http://www.raihome.com Model 5081FG Two-Wire In Situ Oxygen Analyzer (550° to 1600°C).
Emerson Process Management Rosemount Analytical Inc. Process Analytic Division 6565P Davis Industrial Parkway Solon, OH 44139 T (440) 914 1261 F (440) 914 1271 E gas.
HIGHLIGHTS OF CHANGES Effective May, 2005 Rev . 1.0 Page Summary Throughout Changed IB revision number and release date. Cover Changed E-mail addres s. Inside Cover Changed mailing and E-mail addresse s. Page 7-1 Changed repair facility address. Page 8-1 Changed Figure 5-1, item 10 part number in Tab le 8-1.
.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management i Model 5081FG TABLE OF CONTENTS PREFACE..............................................................................................
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 ii Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG LIST OF ILLUSTRATIONS Figure 1-1. Typical System Package .............................................................
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management iii Model 5081FG Figure 4-3. Menu Tree for HART/AMS on the Two-Wire In Situ Oxygen Analyzer (Sh eet 1 of 3)...........................
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 iv Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Di vision of Emerson Process Management P-1 Model 5081FG PREFACE The purpose of this manual is to provide information con cerning the components, func- tions, installation and maintenance of the Model 5081FG T wo-Wire In Situ Oxygen Ana- lyzer (550° to 1600° C).
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 P-2 Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG IMPORTANT SAFETY INSTRUCTIONS FOR THE WIRING AND INSTALLATION OF THIS APPARATUS The following safety instructions apply specifically to all EU member states.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Di vision of Emerson Process Management P-3 Model 5081FG BELANGRIJK Veiligheidsvoorschriften voor de aansluiting en installatie van dit toestel. De hierna volgende veiligheidsvoorschriften zijn voor al bedoeld voor de EU lidstaten.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 P-4 Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG VIGTIGT Sikkerhedsinstruktion for tilslutning og installering af dette udstyr. Følgende sikkerhedsinstruktioner gælder specifikt i alle EU-medlemslande.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Di vision of Emerson Process Management P-5 Model 5081FG BELANGRIJK Veiligheidsinstructies voor de bedradin g en installatie van dit apparaat. Voor alle EU lidstaten zijn de volgende veiligheidsinstructies van toepassing.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 P-6 Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG TÄRKEÄÄ Turvallisuusohje, jota on noudatettava tämän laitteen asentamisessa ja kaapeloinnissa. Seuraavat ohjeet pätevät erityisesti EU:n jäsenvalti oissa.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Di vision of Emerson Process Management P-7 Model 5081FG IMPORTANT Consignes de sécurité concernant le raccor dement et l’installation de cet appareil.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 P-8 Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG Wichtig Sicherheitshinweise für den Anschluß und die Installation dieser Geräte. Die folgenden Sicherheitshinweise sind in allen Mitgliederstaaten der europäischen Gemeinschaft gültig.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Di vision of Emerson Process Management P-9 Model 5081FG IMPORTANTE Norme di sicurezza per il cablaggio e l’installazione dello strumento.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 P-10 Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG VIKTIG Sikkerhetsinstruks for tilkobling og installasjon av dette utstyret. Følgende sikkerhetsinstruksjoner gjelder spesifikt alle EU medlemsland og land med i EØS-avtalen.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Di vision of Emerson Process Management P-11 Model 5081FG IMPORTANTE Instruções de segurança para ligação e instalação deste aparelho. As seguintes instruções de segurança aplicam-se especificament e a todos os estados membros da UE.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 P-12 Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG IMPORTANTE Instrucciones de seguridad para el montaje y cableado de este aparato.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Di vision of Emerson Process Management P-13 Model 5081FG VIKTIGT Säkerhetsföreskrifter för kablage och installation av denna apparat. Följande säkerhetsföreskrifter är tillämpliga för samtliga EU-medlemsländer.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 P-14 Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-1 Model 5081FG SECTION 1 DESCRIPTION AND SPECIFICA.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 1-2 Description and Specifications Rosemount Analytic al Inc. A Division of Emerson Process Management Model 5081FG Table 1-1. Product Matrix 5081FG High T emperature Oxyg en Flue Gas Analyzer High T emperature Analyz er - Instruction Book Code Sensing Probe Length 1 20 in.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-3 Model 5081FG 1-2 SYSTEM OVERVIEW a.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 1-4 Description and Specifications Rosemount Analytic al Inc. A Division of Emerson Process Management Model 5081FG The Model 5081 Transmitter is a two -wire transmitter providing an isolated output, 4- 20 mA, that is proportional to the measured oxygen concentration.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-5 Model 5081FG TWO-WIRE IN SITU OXYGEN ANAL YZER T.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 1-6 Description and Specifications Rosemount Analytic al Inc. A Division of Emerson Process Management Model 5081FG DUCT ST ACK GASES PRESSURE REGULA TOR FLOWMETER OPTIONAL ADAPTER PLA TE 4-20 mA SIGNAL 37240003 MODEL 5081 TRANSMITTER OXYGEN PROBE INSTRUMENT AIR SUPPL Y (REFERENCE AIR) Figure 1-3.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Description and Specifications 1-7 Model 5081FG 1-3 SPECIFICATIONS Net O 2 Range......................................................
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 1-8 Description and Specifications Rosemount Analytic al Inc. A Division of Emerson Process Management Model 5081FG ELECTRONICS Enclosure ...............................................................
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-1 Model 5081FG SECTION 2 INSTALLATION 2-1 PRE-INSTALLATION a. Inspect Carefully inspect the shipping container for any evidence of damage.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 2-2 Installation Rosemount Analytical In c. A Division of Emerson Process Management Model 5081FG O O 1.25 NPT PROCESS CONNECTION SIDE VIEW BOTTOM VIEW INST ALL WITH PORT A T THE BOTTOM 3/4 NPT CONDUIT PORT DIM “A” 7.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-3 Model 5081FG 29750002 PLA TE DIMENSIONS DIMENSION ANSI 4512C34G01 DIN 4512C36G01 JIS 4512C35G01 “C” DIA. 4.75 (121) 5.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 2-4 Installation Rosemount Analytical In c. A Division of Emerson Process Management Model 5081FG REFRACT OR Y ST A CK OR DUCT MET AL W ALL WELD PIPE T O MET AL W ALL 1.25 NPT 2 IN. NPT SCHEDULE 40 PIPE CUST OMER SUPPLIED AD APTER 29750004 2 IN.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-5 Model 5081FG 29750005 REFRACT OR Y ST A CK OR DUCT MET AL WA L L A 1.5 + A 2 IN., 1.2 5 NPT PIPE COUPLING DIMENSION A -- 1-5/8, 2-1/ 2, 3, OR 4 IN.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 2-6 Installation Rosemount Analytical In c. A Division of Emerson Process Management Model 5081FG 5. Use high temperature material (alu- mina wool) to seal around the probe during insertion. This prevents hot gases from escaping or cold air fro m entering the stack or duct.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-7 Model 5081FG d. Installing Model 5081 Transmitter 1. Ensure all components are avail able to install the Model 5081 Transmitter.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 2-8 Installation Rosemount Analytical In c. A Division of Emerson Process Management Model 5081FG COVER LOCK TERMINAL END 3/4 -14 NPT 2 PLACES CIRCUIT END 1.32 (33.5) 1.00 (25.4) TERMINAL BLOCK (TB) TERMINAL END CAP OMITTED FOR CLARITY IN THIS VIEW .
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-9 Model 5081FG 3. For correct viewing orientation, the dis- play may be changed 90 degrees, us- ing the following procedure: (a) Refer to Figure 2-8.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 2-10 Installation Rosemount Analytical In c. A Division of Emerson Process Management Model 5081FG pairs are recommended. Ground the trans- mitter housing to an earth ground to prevent unwanted electromagnetic interference (EMI) or radio frequency interference (RFI).
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Installation 2-11 Model 5081FG 4-20 mA (+) 4-20 mA (-) CELL AND THERMOCOUPLE 4-20 m.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 2-12 Installation Rosemount Analytical In c. A Division of Emerson Process Management Model 5081FG 1 FLOWMETER 0.2-2.0 SCFH 771B635H08 2 2" PRESSURE GAGE 0-15 PSIG 275431-006 3 COMBINA TION FIL TER-REG.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Startup and Operation 3-1 Model 5081FG SECTION 3 STARTUP AND OPERATION Install all protective equipment cov ers and safety ground leads before equipment startup.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 3-2 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG 3-3 REESTABLISHING PROPER CALIBRATION CHECK .
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Startup and Operation 3-3 Model 5081FG 3-4 OPERATION a. Overvie w This section explains the operato r controls and displays of the Two-Wire In Situ Oxy- gen Analyzer.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 3-4 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG ENTER EXIT EXIT PROG ENTER ENTER EXIT EXIT T.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Startup and Operation 3-5 Model 5081FG d. Navigation The IRC in Figure 3-6 is used to interact with the Model 5081 Transmitter and navi- gate through the screens on the LCD.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 3-6 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG 3-5 PROGRAM MENU The PROGRAM MENU branch of the menu tree allows you to program and edit some proc- ess parameters, faults, outputs, and security codes.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Startup and Operation 3-7 Model 5081FG % mA O CONCENTRA TION 2 INDICA TOR F AUL T V ALUE 26020010 Figure 3-9. FAULT VAL c. FAULT VAL Refer to Figure 3-9.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 3-8 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG f. RESET MAX CELL T Refer to Figure 3-12. The transmitter tracks the maximum cell temperature obtained.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Startup and Operation 3-9 Model 5081FG % O CONCENTRA TION 2 26020015 Figure 3-15. TRIM 20 mA? i. TRIM 20 mA? Refer to Figure 3-15. Use this screen to trim the 20 mA value of the 4-20 mA output.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 3-10 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG % O CONCENTRA TION 2 TRACK O DURING CALCHECK 2 26020051 Figure 3-18. SET O 2 TRACKING l. SET O 2 TRACKING Refer to Figure 3-18.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Startup and Operation 3-11 Model 5081FG % 26020017 Figure 3-20. SHOW FAULT 3-6 DIAGNOSTICS MENU The DIAGNOSTICS MENU branch of the menu tree allows you to examine outputs, cu rrent faults, and unit information.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 3-12 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG O CELL IMPEDANCE ST A TUS 2 26020020 % O CONCENTRA TION 2 Figure 3-23. CELL IMPEDANCE d. CELL IMPEDANCE Refer to Figure 3-23.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Startup and Operation 3-13 Model 5081FG % CELL TEMPERA TURE MAXIMUM 26020021 Figure 3-26. MAX CELL T g. MAX CELL T Refer to Figure 3-26.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 3-14 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG a. IN MANUAL? Refer to Figure 3-27. Failure to remove the analy zer from automatic control loops prior to per- forming this procedure may result in a dangerous operating condition.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Startup and Operation 3-15 Model 5081FG % % O CONCENTRA TION 2 LOW CALIBRA TION CHECK GAS O V ALUE 2 26020056 Figure 3-29. ACCEPT LOW O 2 d.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 3-16 Startup and Operation Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 4-1 Model 5081FG SECTION 4 HART/AMS 4-1 OVERVIEW The HART Communicato r is a handheld com- munications interface device.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 4-2 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG LOOP CONNECTORS USE INTERFACE 00275 0013 ONLY SERIAL PORT.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 4-3 Model 5081FG LOOP CONNECTORS USE INTERFACE 00275 0013 ONLY SERIAL PORT.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 4-4 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG O2 value Cell T emp Imp VIEW FLD DEV V ARS PV is O2 value.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 4-5 Model 5081FG URV SELECT O2 RANGE Alarm V alue Ta g BASIC SETUP DET AIL.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 4-6 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG Manufacturer Model Dev id Ta g Descriptor Message Date Fi.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management HART/AMS 4-7 Model 5081FG 4-6 HART COMMUNICATOR STA RT CAL- CHECK METHOD To perform a calibration check on the Two-Wire In Situ Oxygen Analyzer with the HART Com- municator, use the following procedure.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 4-8 HART/AMS Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Maintenance and Service 5-1 Model 5081FG SECTION 5 MAINTENANCE AND SERVICE Install all protective equipment cov ers and safety ground leads after e quip- ment repair or service.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 5-2 Maintenance and Service Ros emount Analytical Inc. A Di vision of Emerson Process Management Model 5081FG 14 13 12 ANALOG BOARD CPU BOARD RIBBON CABLE 37240008 16 15 9 1 2 3 6 8 7 5 4 11 10 1. Housing 2.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Maintenance and Service 5-3 Model 5081FG 8. Connect the ribbon cable co nnector between display board (12) and the CPU board (11). Ensure the cabl e connector is fully seated.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 5-4 Maintenance and Service Ros emount Analytical Inc. A Di vision of Emerson Process Management Model 5081FG a.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-1 Model 5081FG SECTION 6 TROUBLESHOOTING Install all protective equipment cov ers and safety ground leads after trouble- shooting.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 6-2 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG IMPEDANCE, SPEED OF RESPONSE (t ), sec 90 0 26020060 100 200 300 400 500 AIR TO 0.4% O 2 0.4% O TO AIR 2 Ω Figure 6-2.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-3 Model 5081FG 6-4 IDENTIFYING AND CORRECTING FAULT INDICATIONS A fault in the operation of the Two-Wire In Situ Oxygen Analyzer is indicated by the faulted op- eration display.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 6-4 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG % 26020024 Figure 6-5. Fault 1, Open Thermocouple a. Fault 1, Open Thermocouple The thermocouple connection i s open.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-5 Model 5081FG % 26020026 Figure 6-7. Fault 3, Shorted Thermocouple c. Fault 3, Shorted Thermocouple The thermocouple connection s are shorted.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 6-6 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG % 26020028 Figure 6-9. Fault 5, O 2 Cell Open e. Fault 5, O 2 Cell Open The O 2 cell connection is open.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Troubleshooting 6-7 Model 5081FG % 26020030 Figure 6-11. Fault 7, Reversed O 2 Cell g. Fault 7, Reversed O 2 Cell The O 2 cell connection s are reversed.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 6-8 Troubleshooting Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emers on Process Management Return of Material 7-1 Model 5081FG SECTION 7 RETURN OF MATERIAL 7-1 If factory repair of defective equipment is re- quired, proceed as follows: a.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 7-2 Return of Material Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emer son Process Management Replacement Parts 8-1 Model 5081FG SECTION 8 REPLACEMENT PARTS Table 8-1.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 8-2 Replacement Parts Rosem ount Analytical Inc. A Division of Emerson Process Management Model 5081FG.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Index 9-1 Model 5081FG SECTION 9 INDEX This index is an alphabetized listing of part s, terms, and procedures having to do with the Haz- ardous Area Oxygen/Combustibles T ransmitter.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 9-2 Index Rosemount Analytical Inc. A Division of Emerson Process Management Model 5081FG M Materials of Construction, 1-7 Materials of Construction, .
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 Rosemount Analytical Inc. A Division of Emerson Process Management Index 9-3 Model 5081FG U UNIT SER # Screen, 3-13 UPPER RANGE VAL Screen, 3-7 Z Zirc.
.
3724 3791/05-05 WARRANTY Goods and part(s) (ex cluding consumables) ma nufactured by Seller are warranted to be free from defects in workmanship and material un der normal use and service for a period of twelve (1 2) months from the date of shipment by Seller.
Instruction Manual IB-106-5081, Rev. 1.0 May 2005 © Rosemount Analytical Inc. 2005 Model 5081FG WORLD HEADQUARTERS Emerson Process Management Rosemount Analytical Inc. Process Analytic Division 6565P Davis Industrial Parkway Solon, OH 44139 T 440 914 1261 T 800 433 6076 gas.
デバイスEmerson 5081FGの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Emerson 5081FGをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはEmerson 5081FGの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Emerson 5081FGの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Emerson 5081FGで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Emerson 5081FGを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はEmerson 5081FGの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Emerson 5081FGに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちEmerson 5081FGデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。