® 271 Programmable 10 MHz DDS Function Generator Getting Started Manual PN 2423297 January 2005 © 2005 Fluke Corporation , All rights reserv ed. Printed in USA All product names are trademarks of their respec tive companies.
LIMITED WARRA NTY AND LIMITATION OF LIA BILITY Each Fluke product is w arranted to be free from defects i n materi al and workmanship under normal use and serv ice. The warranty period is one y ear and begins on the date of ship ment. Parts, product repair s, and services ar e w arranted for 90 days.
Safety This function generator is a Safety Class I instrument accor d ing to IEC classification and has been designed to meet th e re- quirements of EN61010-1 (Safet y Requirements for Electrical Equipment for Measurem e nt, Control and Laboratory Use).
271 Getting Started Manual • The apparatus shall be disconnected fro m all voltage sources before it is opened for any adjustment, replacement, maintenance or repair.
Safety (continued) The following symbols are used on the instru ment and in this manual: Caution - refer to the accompanying docume ntation, incorre ct operatio n may damage the instrument. Terminal connected to chassis ground. Mains supply OFF. Mains supply ON.
.
EMC Compliance This instrument meets the requirements of the EMC Directive 89/336/EEC. Compliance was demonstrated by meeting t he test limits of the following standards: Emissions EN61326 (1998) EMC produc t standard for Electrical Eq uipment for Measurement, Con t rol and L aborator y Use.
271 Getting Started Manual e) EN61000-4-5 (1995) Surge: 0 .5 kV (line to line), 1 kV (line to ground) Performance A. f) EN61000-4-6 (1996) Con ducted RF: 3 V, 80 % AM at 1kHz (ac line only; signal conn ections <3 m not tested) Performance A.
Table of Contents Title Page Introduc tion ................................................................................................................... . 1 Before You Start .......................................................................
271 Getting Started Manual AUX OUT ............................................................................................................. 8 EXT TRIG ...............................................................................................
Getting Started Introduction This Getting Started Manual for the model 271 Programmable 10 MHz DDS Function Generator is designed to provide an initia l und erstanding of the way the instrument is operated.
271 Getting Started Manual Externally Applied Voltages Caution - Front Panel Sockets To avoid ris k of damage to the instrume nt: Do not apply external voltages to the M AIN OUT or AUX OUT sockets. Do not apply external voltages exc eeding ±10 V to the TRIG IN socket.
Getting Started Controls and Connections Controls and Connections shx0001f.gif Figure 1. Model 271 Fron t Pan el 3.
271 Getting Started Manual The Liquid Crystal Display The 4-line x 20-character LC D panel is used to displ ay and edit all the parameter values. T he paradigm used for selecting fields and editing v alues is described belo w. If necessary (because of temperature variati ons etc.
Getting Started Controls and Connections FIELD and DIGIT Keys and the Rotary Control The FIELD keys move the display cursor bet ween fields. The fields in the display shown in figure 1 above are " FREQ= ", " 10.00000 ", " kHz ", " VhiZ =" and so on.
271 Getting Started Manual Numeric, Units and SET Keys You can use these keys to enter parameter values directly. For example, to set the period to 125 µ s, select the first field ( FREQ ), change it to period ( PER ) by pressing a DIGIT key, then enter 1, 2, 5, us on the keypad.
Getting Started Controls and Connections MODE Keys Alternate presses of a MODE key will turn the mode ( TRIG , GA T E , AM , etc.) on or off, and when the function is on the associated lamp is lit. If you press the EDIT key then press a MODE key you will see the edit menu for that mode.
271 Getting Started Manual Front Panel Inputs and Outputs MAIN OUT MAIN OUT is the 50 or 600 output from the main generator. It will provide up to 20 V p-p into a high- impedance load or 10 V p-p into a matched 50 or 600 load. It can tolerate a short circuit for 60 seconds.
Getting Started Controls and Connections Rear Panel Connectors CLOCK IN/OUT This socket operates both as an input and as an output, depending on the mode in which the instrument is being used.
271 Getting Started Manual TRIG/SWEEP OUT The TRIG/SWEEP output socket provides both trigger and marker signals, depending on the instrument's mode of operation. shx0016f.gif Its primary use is for triggering an oscilloscope or other recording device, but it can also provide z-a xis bright-up markers during sweep and frequenc y-hopping modes.
Getting Started Using the Instrument Using the Instrument While familiarizing yourself with the instrument you will find it useful to observe the outputs on an oscillo scope.
271 Getting Started Manual Generating Continuous Signals Select SINE on the function keys, if it is not already selected. If there is no signal at the oscilloscope input i t will almost certainly be because the genera tor's output is switched off.
Getting Started Using the Instrument Set the amplitude to +10 dBm You can do this by keying AMP L , 1 , 0 , dBm . The display will now look like this: PER =80.00000us O UT =+10.0dBm 50 Ω DC=+0.00mV (+0.00mV) SYM=50.0% (50.0%) If you switch the amplitude display back to VhiZ you should see: PER =80.
271 Getting Started Manual Effect of the output attenuator The instrument has an output attenuator with fixed steps of 20 dB. The attenuator foll ows the output stages so any dc offset added to the waveform is subject to the attenuator.
Getting Started Using the Instrument Asymmetric waveforms Symmetry adjustments work in much the same way as dc offset adjustments. However, because of the lengthy calculations which the instrument makes foll owing a change to the symmetry setting, you may find it easier to use direct numeric entry rath er than the rotary control.
271 Getting Started Manual instrument's AUX OUT socket in order to see the phase effects more clearly. There are some limitations on using the phas e control at frequencies above 30 kHz - chapter 5 of the Users Manual explains these in detail.
Getting Started Using the Instrument At this point the lamp next to the SWEEP key should be flashing and the display sho uld look like this: MODE=BEG-END LAW=L OG RAMP TIME=0.05 s TRIG SRC=CONTINUOUS MORE->>> All these settings are fine except for the sweep time, which is too short if you are listening to the signal.
271 Getting Started Manual Generating a Triggered Burst In this exercise you will use the internal trigger gen erator to initiate a burst of 500 cycles of a 2 kHz tone t wice every second. The period of the inter nal trigger generator is thus 500 ms. Press RECALL , 0 , ENTER to reset to the factory defaults.
Getting Started Using the Instrument FSK Mode The FSK (frequency shift ke ying) mode can also be driven from the internal trigger generator. Setting up is ver y straightforward; the follo wing settings give a continuous waveform which switches between 800 Hz and 1.
271 Getting Started Manual set the levels to produce five monochrome s tripes from left to right: grey, white, grey, black and grey again. Our line signal thus looks like the dra wing below: shx0019f.
Getting Started Using the Instrument step name start (µs) amplitude start scaled to 0-1024 step length amplitude scaled to 0-511 0 0 30% 0 29 153 1 sync pulse 1.8 0% 29 75 0 2 back porch 6.5 30% 104 90 153 3 50% grey 12.1 65% 194 165 332 4 white 22.4 100% 358 165 511 5 50% grey 32.
271 Getting Started Manual The second screen appears: VALS=A BS AUTO=YES STEP=01 ACTIVE LENGTH=0256 LEVEL=+000 Continue entering the pairs of lengt h and level values in the table until the last step (08).
Getting Started Using the Instrument Saving and Recalling Settings Having done all this work it would be wise to save the settings for future use. You can do this simply by pressing the STORE key, followed by a number in the range 1 to 9 and the ENTER key.
271 Getting Started Manual arbitrary waveforms 01 to 05. These first five store locations are available for waveforms downloaded from a PC. When you pres s the ENTER key the output switches to the sel.
デバイスFluke 271の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Fluke 271をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはFluke 271の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Fluke 271の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Fluke 271で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Fluke 271を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はFluke 271の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Fluke 271に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちFluke 271デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。