760000 AL TERNA TING PRESSURE / L OW AIR L OSS MA TTRESS SYSTEM USER MANUAL Impor tant: Do not operate the Mattress System without rst reading and understanding this manual! Save this manual for future use.
2 760000-INS-LAB-RevA08 Contents INTRODUCTION ..................................................................................... 3 INTENDED USE OF THIS DEVICE ........................................................ 3 IMPORT ANT SAFETY PRECAUTIONS .
760000-INS-LAB-RevA08 3 Intr oduCtIon Use this manual for set-up and operation of the Alternating Pr essure / Low Air Loss Mattr ess System. Read all instruc- tions befor e using the mattress system.
4 760000-INS-LAB-RevA08 d anger T o reduce the risk of electr ocution: DANGER: Always unplug this product immediately after using. DANGER: Do not use this product while bathing. DANGER: Do not place or store this product where it can fall or be pulled into a tub or sink.
760000-INS-LAB-RevA08 5 W ARNING: Never block the air openings of this product or place it on a soft surface, such as a bed or couch, where the openings may be blocked. Keep the air opening free of lint, hair , and other similar particles. W ARNING: Never drop or insert any object into any opening or hose.
6 760000-INS-LAB-RevA08 For U.S. and Canada only E228589 53DG Medical Equipment: Air Pump with respect to electrical shock, re and mechanical hazards only in accordance with UL60601-1 and CAN/CSA C22.
760000-INS-LAB-RevA08 7 pump: rear main power switch power cord receptacle fuse hanger filter pump: front p anel cycle time adjust button; cycle time indicators at right of button decrease pr essure b.
8 760000-INS-LAB-RevA08 incr ease pressur e under the buttocks). Press the therapy adjust button at the left end of the r ow to cycle between the settings. The selected therapy mode setting's indicator will illuminate. F r ont panel four th row Panel lock button: The panel lock pr otects the panel set- tings fr om accidental change.
760000-INS-LAB-RevA08 9 Ser vice indicator: Mechanical failure, service requir ed; contact your Graham-eld equipment pr ovider . Inst alla tIon unp aCkIng 1. Befor e unpacking, check for obvious damage to the car- ton or its contents. If damage is evident, please notify the carrier and your Graham-Field equipment pr ovider .
10 760000-INS-LAB-RevA08 s Caution: Place the device in a position wher e the car e- giver and/or doctor can access it easily . 5. Connect the mattr ess-pump air hose quick connector; ensur e that the connector has clicked secur ely into place befor e continuing.
760000-INS-LAB-RevA08 11 Note: The mattress system will automatically enter Autofirm mode for the quickest inflation at set-up; initial inflation takes approximately twenty minutes. 3. The pump will begin delivering air into the mattr ess. 4. When the selected pr essure level is r eached, the pump will enter alternating mode.
12 760000-INS-LAB-RevA08 transport Cap In case of power failur e or trans- port: Disconnect the quick connec- tor and cap the mattr ess with the transport cap to slow deation. Cpr funCtIon If CPR must be performed on the patient while the mattr ess is in use, to immediately deate the mat- tr ess, either: 1.
760000-INS-LAB-RevA08 13 general maIntenanCe 1. Check main power cor d for abrasion or excessive wear . 2. Check mattr ess cover for signs of wear or damage. En- sur e mattress cover and tubes ar e connected correctly . 3. Check airow fr om the air hose connector .
14 760000-INS-LAB-RevA08 st orage 1. Lay the mattr ess out at and upside down. 2. Roll fr om the head end towar d the foot end. 3. Str etch the foot-end strap around the r olled mattress to pr event unrolling. 4. Stor e mattress and pump in a dust-fr ee environment with no exposur e to direct sunlight.
760000-INS-LAB-RevA08 15 Mattress Specication Model 8" mattress Dimensions (L x W x H) 78.7" x 35.4" x 8" (200 x 90 x 20.3 cm) Weight 16.
USA, Corporate Headquar ters: Graham-Field Health Products 2935 Nor theast Parkway Atlanta, Georgia 30360 telephone: 800-347-5678, 770-447-1609 fax: 800-726-0601, 678-291-3232 www .grahamfield.com 760000-INS-LAB-RevA08 © 2008 GF Health Products, Inc.
デバイスGraham Field 760000-INS-LAB-RevA08の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Graham Field 760000-INS-LAB-RevA08をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはGraham Field 760000-INS-LAB-RevA08の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Graham Field 760000-INS-LAB-RevA08の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Graham Field 760000-INS-LAB-RevA08で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Graham Field 760000-INS-LAB-RevA08を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はGraham Field 760000-INS-LAB-RevA08の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Graham Field 760000-INS-LAB-RevA08に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちGraham Field 760000-INS-LAB-RevA08デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。