HP (Hewlett-Packard)メーカー3325Bの使用説明書/サービス説明書
ページ先へ移動 of 23
__________ *This bulletin supersedes TB 9 - 6625 - 2049 - 35, 15 January 1996. *TB 9 - 6625 - 2049 - 35 SUPERSEDED COPY DATED 15 JANUARY 1996 DEPARTMENT OF THE ARMY TECHNICAL BULLETIN CALIBRATION PROC.
TB 9 - 6625 - 2049 - 35 2 Paragraph Page SECTION II I. CALIBRATION PROCESS - Continued Triangle and ramp amplitude ............................ 16 17 Dc offset ............................................................... 17 18 Dc of fset with ac ..
TB 9 - 6625 - 2049 - 35 3 Table 1 . Calibration Description Test instrument parameters Performance specifications Harmonic distortion 1 (relative to fundam ental at full output) Fundamental frequency: 0.
TB 9 - 6625 - 2049 - 35 4 Table 1. Calibration Description - Continued Test instrum ent parameters Performance specifications Amplitude accuracy with dc offset and no attenuation (range 1) into 50 Ω load Tolerance relative to programmed amplitude Sine wave: .
TB 9 - 6625 - 2049 - 35 5 5. Accessories Required. The accessories required for this calibration are common usage accessories, issued as indicated in paragraph 4 above, and are not listed in this calibration procedure.
TB 9 - 6625 - 2049 - 35 6 SECTION III CALIBRATION PROCESS 6. Preliminary Instructions a . The instructions outlined in paragraphs 6 and 7 are preparatory to the calibration process. Personnel should become familiar with the entire bulletin b efore beginning the calibration.
TB 9 - 6625 - 2049 - 35 7 8. Harmonic Distortion a. Performance Check (1) Connect TI EXT REF IN 1, 10 MHz to spectrum analyzer 10 MHz REF IN/OUT . (2) Connect TI SIGNAL to spectrum analyzer INPUT 50 Ω Ω . (3) Press keys and enter values using DATA keys as listed in (a) through (h) below: (a) SIGNAL off (option 002 not lit).
TB 9 - 6625 - 2049 - 35 8 (a) 50 Hz . (b) ENTRY AMPTD . (c) 10 VOLT . (13) Set audio analyzer to measure distortion in dB. Audio analyzer indication will be ≤ - 65 dB. (14) Perform (15) through (19) below for option 002 only . (15) Connect equipm ent as shown in figure 1 and set resistance standard to 470 Ω .
TB 9 - 6625 - 2049 - 35 9 9. Amplitude Modulation Distortion a. Performance Check (1) Press keys and enter values using DATA keys as listed in (a) throug h (i) below: (a) SIGNAL off (option 002 not lit). (b) FUNCTION sine wave. (c) ENTRY FREQ . (d) 1 MHz .
TB 9 - 6625 - 2049 - 35 10 (a) SIGNAL off (option 002 not lit). (b) FUNCTION square wave. (c) ENTRY FREQ . (d) 1 MHz . (e) ENTRY AMPTD . (f) 1 V RMS . (2) Connect TI SIGNAL to oscilloscope CH 1 . (3) Set oscilloscope CH 1 COUPLING 50 Ω Ω to ON . (4) Set osc illoscope controls for duty cycle measurement.
TB 9 - 6625 - 2049 - 35 11 (4) Set oscilloscope controls to measure ramp retrace time from the 90 to 10 percent points. Ramp retrace time will be ≤ 3 µ s. ( 5) Press FUNCTION negat ive ramp key and repeat (4) above. b. Adjustments . No adjustments can be made.
TB 9 - 6625 - 2049 - 35 12 (15) Set frequency counter controls to measure frequency. If frequency counter does not indic ate between 59.99970 and 60.00030 MHz, perform b below. b. Adjustments (1) Disconnect TI rear panel 10 MHz OVEN OUTPUT from EXT REF IN option 001 only.
TB 9 - 6625 - 2049 - 35 13 (12) Adjust A9R7 Fine Adj. (fig. 3) and A9 Coarse Adj. (fig. 3) for a minimum indication on frequency difference meter. (13) Disconnect TI from frequency difference meter. (14) Reconnect TI rear panel 10 MHz OVEN OUTPUT to EXT REF IN .
TB 9 - 6625 - 2049 - 35 14 (6) Adj ust frequency counter sample rate control to HOLD and press RESET pushbutton. Record frequency counter indication. (7) Press ENTRY PHASE key and enter - 1 deg using DATA keys. (8) Press frequency counter RESET pushbutton and record frequency counter indi cation.
TB 9 - 6625 - 2049 - 35 15 Table 3. Amplitude Test instrument settings Multimeter indications ENTRY AMPTD ENTRY FREQ (V ac) 3.536 V RMS 1 kHz 3.495 3.577 3.536 V RMS 100 kHz 3.495 3.577 1.061 V RMS 100 kHz 1.048 1.073 1.061 V RMS 1 kHz 1.048 1.073 1.061 V RMS 100 Hz 1.
TB 9 - 6625 - 2049 - 35 16 Table 4. Flatness Test instrument and measuring receiver setting s (MHz) 1 2 4 6 8 10 12 14 16 18 20 b. Adjustments . No adjustments can be made. 15. Square Wave Amplitude and Flatness a. Performance Check (1) Connect TI SIGNAL to multimeter INPUT HI and LO using 50 Ω feedthrough termination.
TB 9 - 6625 - 2049 - 35 17 (a) 1 kHz . (b) ENTRY AMPTD . (c) 1 V RMS . (d) AMPTD CAL . (7) Multimeter will indicate between 0.99 and 1.01 V ac. (8) Disconnect TI from multimeter. (9) Connect TI SIGNAL to oscilloscope CH 1 using 50 Ω feedthrough termination.
TB 9 - 6625 - 2049 - 35 18 (6) Enter 10 kHz using DATA keys. Multimeter will indicate between 2.742 and 3.032 V ac. (7) Press keys and enter values using DATA keys as listed in (a) through (c) below: (a) FUNCTION positive ramp. (b) 100 Hz . (c) AMPTD CAL .
TB 9 - 6625 - 2049 - 35 19 Table 5. Dc Offset - Continued Test instrument DC OFFSET Multimeter indications (V dc) se ttings Min Max 14.99 mV 0.01493 0.01505 - 14.99 mV - 0.01505 - 0.01493 4.999 mV 0.004979 0.005019 - 4.999 mV - 0.005019 - 0.004979 1.499 mV 0.
TB 9 - 6625 - 2049 - 35 20 (4) Enter - 4.5 VOLT using DATA keys. Multimeter will indicate between - 4.35 and - 4.65 V dc. (5) Press ENTRY FREQ key and enter 999.9 kHz using DATA keys. Multimeter will indicate between - 4.44 and - 4.56 V dc. (6) Press ENT RY DC OFFSET key and enter 4.
TB 9 - 6625 - 2049 - 35 21 (a) ENTRY FREQ . (b) 100 kHz . (c) ENTRY AMPTD . (d) 10 V RMS . (e) AMPTD CAL . (5) Multimete r will indicate between 9 and 11 V ac. (6) Press keys and enter values using DATA keys as listed in (a) through (f) below: (a) FUNCTION square wave.
TB 9 - 6625 - 2049 - 35 22 Figure 5. Power supply assembly A2 – test instrument top view. (5) Press presently active FUNCTION key to remove ac output and activate ENTRY DC OFFSET key (lit). (6) Enter 0 VOLT using DATA keys and press AMPTD CAL key. (7) Connect multimeter INPUT HI to TP AMP OUT (fig.
TB 9-6625-2049-35 By Order of the Secretary of the Army: Official: ERIC K. SHINSEKI General, United States Army Chief of Staff JOEL B. HUDSON Administrative Assistant to the Sec retary of the Army 0029703 DISTRIBUTION: To be distributed in accordance with IDN 342200, requirements for calibration procedure TB 9-6625-2049-35.
デバイスHP (Hewlett-Packard) 3325Bの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
HP (Hewlett-Packard) 3325Bをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはHP (Hewlett-Packard) 3325Bの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。HP (Hewlett-Packard) 3325Bの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。HP (Hewlett-Packard) 3325Bで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
HP (Hewlett-Packard) 3325Bを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はHP (Hewlett-Packard) 3325Bの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、HP (Hewlett-Packard) 3325Bに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちHP (Hewlett-Packard) 3325Bデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。