CHANNELMAG FLOWMETERS FOR OPEN CHAN NELS AND NON- FULL PIPES RTM- Expol ink s.r.o., Prikop 8, 6020 0 Brno, Czech Republic Instructio n Manual.
2 FOREWORD This in struction manual is designed to help you gain a thorough understanding o f the operatio n of th e eq uipment. ARKON recommends that y ou read this manual completely before placing the equip ment in service. Although ARKON de signs reliab ility into all equipment, there is alway s the possibility of a malfunction occurring.
3 IMPO RTANT – PLEASE READ The installation and ope ration of this product may put y ou at risk of serious injury or even death. Ta ke wh atever precautio ns are necessary to ensure your safety b efore making a n installation or working on one. Never work alo ne or unsupervised.
4 Chann elMag Table of Co ntent s Section 1 Introduction 1.1 Scope: Chann elMag CM2 6 1.2 Scope: Chann elMag PM2 7 1.3 Safety 8 Section 2 Installation 2.1 Media Consideration s 9 2.2 Location and Mou nting 9 2.3 Straight Leng ths of Chan nel / Pipe 10 2.
5 Table of Co ntent s 4.8 Erratic Signal 24 4.9 Signal or Indi cation “ Sw ing ” 24 Section 5 Recommended Spares 5.1 Recomm ended Spares 25 24 24 25.
6 Chann elMag Section 1 Introd uction The ChannelMag C M2 is suitable f or open channels having diver se c ross sectional shapes from 6” to 200 feet ( 150 mm to 6 0 m). They meet or exceed the r equirements of I SO 9 213 and are calibrate d in a long open tank, b elieved to be the largest NIST traceable facility available.
7 Chan el Mag Sectio n I Intro ducc ion The ChannelMag P M2 Series is suitable for fu ll or partia lly full pipes from 2 0” to 120 ” ( 500 - 3000 mm). They meet or exceed the requirements of I SO 92 13 and ar e calibrated in a long open tank, belie ved to b e the largest NIST tr aceable facility availab le.
8 ChannelMag Sectio n 1 Introdu ction Before installation, make sure the oper ating e nvironments of the flow meter and ancillary eq uipment are co nsistent w ith appro priate national approvals such as USA UL or NEC, Canadian CSA, European CE. The installation m ust b e carried out b y qualified personnel only.
9 Chann elMag Section 2 Installa tion The ChannelMag is suitable for such media as water, raw sewage, slurries, algae laden r iver water, and similar. The minim um media conductivity is 0.5 microS/cm for op eration with pulsed AC transmitters. Most water based media has a conductivity of 200 – 800 microS/cm.
10 Channel / Pipe ChannelMag Sectio n 2 Installa tion the electro de cables and the ener gizer coil cab le within the junction box. Do not use silicone rubbe r (RTV) for po tting, since it contains acetic acid which can corr ode the j unction box terminals.
11 ChannelMag Sectio n 2 Installa tion 2.3.3 Important Note 2.4.1 CM2 Sensors Nominal CM2 Width Actual CM2 Width L S R H M Weight Each in. mm . in. mm. in. mm. in. mm . in. mm. in. mm. in. mm . lb kg. 8.0 200 10.0 254 40.0 1016 8.0 203 14.5 370 2.50 64 6.
12 ChannelMag Sectio n 2 Installa tion The height “E” o f the magnetic enhancement plates is custom made and normally extends 2” (50mm) below the top of the channel. T he wi dth W is a djustable to fi t the channel width. W includes the mounting brackets.
13 ChannelMag Sectio n 2 Installa tion 2.4.2 Installation of 2 x CM2 ChannelMag Sensors Adjust width before installation into channel 2.0 0” 50 m Lifting brackets. Do not lift by the Actual CM2 Wid th enhancement plates. For CM2 sensors installed in a rec ess in the bed o f the channel, the installation configuration remains the sam e.
14 ChannelMag Sectio n 2 Installa tion The CM2/D is normally used with a FM X167 hydrostatic level transducer and installed as shown below. A 2-par t re- enterable gel type 444 2, manuf actured b y 3M Company, St. Paul, Minnesota, is normally supplied to retain the CM2/D in the hole made for it.
15 ChannelMag Sectio n 2 Installa tion step s hould be cemented wit h an approach angle of not more than 15 degrees. The r ecess must extend over the ful l width of the bed of t he channel to acc omm odat e the Magnetic Enhancement Plates. Note: CM2/8” ChannelM ag sensors ar e normally supplied w ith ram ps.
16 ChannelMag Sectio n 2 Installa tion 2.4.5 PM2 Sensors i n Non-Full Pipes The ChannelMag PM Serie s sensors a re used in pa rtially filled pip es from 20 ” - 1 20” (50 0 - 3000 mm ) diam e ter. For pipe s 20” - 36” (500 - 915 mm ) a single Channe lMag sensor is installed in a retention band at the bottom o f a horizo ntal p ipe.
17 ChannelMag Sectio n 2 Installa tion 2.4.6 Installation of PM2 Sensors With M agnetic Enhancement Ba n d Center spacer used for co mpound sensor in stallation o nly Outer mountin g points used witho.
18 Enhancement Plate Expander Mechanism The CM2 ChannelMag is equipp ed with a gro unding plate, normally manufactured from AISI 316 stainless steel. This makes sure there are no unequal po tentials between the media and the shielding of the ChannelMag c onnection cab les.
19 ChannelMag Sectio n 2 Installa tion 2.6 Cable Ty pes Exciter Coils Electrodes Refer ence Coils Black to 4 Black to 24 Black to 31 White to 5 White to 25 White to 32 Shield to 6 S hield to 26 Shield.
20 ChannelMag Sectio n 2 Installa tion 2.6.1 ChannelMag Junction Box Wiring Diagram ChannelMag Sectio n 2 Installa tion CAUTION SHIELD BLACK RED OR CLEA R.
21 The Junction Box The cable ends and t he junction box supplied f or mult i-Channel Mag sy stems are t o NEMA 6 and IP68 f or pe rmanent subm ersion t o 30 f eet (10m) w ater colum n. As such, t he cabl e ends are supplie d w ith s hrink-fit sleeves .
22 ChannelMag Sectio n 2 Installa tion For m ultiple ChannelMag system s, the electrode and reference cable s may all run in the same conduit to their respective transmitters, but sep arated from the exciter c oil cables b y a minim um of 2 feet (0.5m).
23 Chann elMag Section 4 Maint enance 4.1 Cleaning 4.2 Zero Span Checking 4.3 Effect of Moisture 4.4 Effect of a 4.5 Effect of One Sensor Of Two I noperative Under typical oper ating conditio ns the ChannelMag sensor r equires no maintenance.
24 ChannelMag Sectio n 4 Maint enanc e 4.6 Transmitter Fault 4.7 All Checks Good, But No Signal 4.8 Erratic Signal 4.9 Signal Or I ndication “Swing” Should the resistances referenced above be correc t, b ut the z ero and sp an cannot be achieved, then the fault lies w ith the transmitter.
25 Chann elMag Section 5 Recommend ed Sp ares 5.1 Recommended Spares Since the ChannelMag sensor is a solid state sensor, completely encapsulated, there a re no recommended spares. Make sure when ord ering a new ChannelMag sensor or transmitter that the serial number found on the transmitter label is advised.
デバイスArkon Channelmagの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Arkon Channelmagをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはArkon Channelmagの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Arkon Channelmagの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Arkon Channelmagで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Arkon Channelmagを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はArkon Channelmagの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Arkon Channelmagに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちArkon Channelmagデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。