OmegaメーカーTX82Bの使用説明書/サービス説明書
ページ先へ移動 of 20
www .omega.com e-mail: info@omega.com omega.com ® OMEGA ® U ser ’ s Guide TX82B Panel Mounted Current Loop Indicator Shop on line at.
Ser vicing Nor th America: USA: One Omega Drive, P .O. Box 4047 ISO 9001 Certified Stamford CT 06907-0047 TEL: (203) 359-1660 F AX: (203) 359-7700 e-mail: info@omega.com Canada: 976 Bergar Laval (Quebec) H7L 5A1 TEL: (514) 856-6928 F AX: (514) 856-6886 e-mail: info@omega.
T ABLE OF CONTENTS 1.0 GENERAL INFORMA TION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1 2.0 SPECIFICA TIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
UNP ACKING INSTRUCTIONS Remove the Packing List and verify that you have received all equipment, including the following: QTY DESCRIPTION 1 Current Loop Meter Indicator with all applicable connnectors attached.
1 1.0 GENERAL INFORMA TION Model TX82B two-wire current-loop indicator accepts 1-5 mA, 4-20 mA, or 10-50 mA process signals and digitally displays the process variable in percentage or engineering units of such measurements as pressure, flow , temperature and level.
2 2.0 SPECIFICA TIONS 2.1 INPUT Current 1-5 mA, 4-20 mA or 10-50 mA Protection 200 mA max forward and 1000 mA max reverse V oltage Drop 2.5 V max forward and 1.
3 2.4 DISPLA Y T ype 7-segment LCD Color Black digits with white background Symbols -1.8.8.8.0.0, 12.7 mm (0.5 in) height Polarity Minus sign Overrange Three least-significant digits blank Extra digit.
4 3.0 MECHANICAL ASSEMBL Y AND INST ALLATION 3.1 SAFETY CONSIDERA TIONS T o ensure safe operation, follow the guidelines below: VISUAL INSPECTION: Do not attempt to operate the instrument if damage is found. SIGNAL WIRING: Insert the proper plus and minus signal wires into the plug-in screw-clamp connector terminals marked plus and minus.
3.3 INST ALLA TION/REMOVAL The TX82B is housed in a 1/8 DIN case. The electronic circuitry can be installed or removed from the front and is attached to the case with two M4 screws through the rear panel. Panel Mounting 1. Remove the two thumbnuts on the rear of the case.
6 Figure 3-2. DIN Case Dimensions 4.0 SIGNAL INPUT CONNECTIONS (P3) The signal input connections for all meters are made at connector P3 as follows (see Figure 3-1): P3 Connection Signal 1 Signal Hi 2 Signal Lo 3 No Connection 4.15 [105.4] SLEEVE 3.50 [88.
7 5.0 CONFIGURA TION The standard TX82B meter is factory-configured for an input of 4-20 mA to display 00.0 to 100.0. Field configuration for input current range, decimal point location, dummy right-h.
8 Figure 5-1 Main Board Jumper Locations A AH J F C D E B G BJ E H G R4 U3 U2 P3 VR1 S1 S2 S2 SP AN ADJ. ZERO ADJ. VR2 R30 S1 F C D.
9 6.0 CUSTOMER CONFIGURA TION AND CALIBRA TION Use this procedure to determine the configuration of the TX82B customized setup. The procedure is general; customers can specify any two current inputs and their corresponding digital readings. Pin-groups are shown in Figure 5-1.
10 6.2 CONFIGURA TION PROCEDURES 1. Remove all push-on jumpers. 2. For an input current range of 1-5 mA, no jumper is required. For 4-20 mA input, install a push-on jumper at S1-H. For 10-50 mA input, install a push-on jumper at S1-J. 3. If N2 (Section 6.
11 Figure 7-1 Wiring Diagram for TX82B Usage for FM (Factory Mutual) SAFE AREA APPARATUS 2 EARTHED BUS BAR + - 3 1 4 INTER- CONNECTING CABLES SHUNT ZENER DIODE SAFETY BARRIER SAFE AREA HAZARDOUS AREA .
.
.
.
WARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warrants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warranty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
M1060/0303 10579ML-99 J Where Do I Find Everything I Need for Process Measurement and Contr ol? OM EGA…Of Course! Shop on line at www .omega.com TEMPERA TURE Thermocouple, RTD & Thermist.
デバイスOmega TX82Bの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Omega TX82Bをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはOmega TX82Bの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Omega TX82Bの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Omega TX82Bで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Omega TX82Bを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はOmega TX82Bの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Omega TX82Bに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちOmega TX82Bデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。