DRA-RTM-8 ANALOG MUL TIPLEXER User’s Guide http://www .omega.com e-mail: info@omega.com.
USA and Canada: Sales Service: 1-800-826-6342 / 1-800-TC-OMEGASM Customer Service: 1-800-622-2378 / 1-800-622-BESTSM Engineering Service: 1-800-872-9436 / 1-800-USA-WHENS M TELEX: 996404 EASYLINK: 62968934 CABLE: OMEGA Mexico and Latin America: T el: (95) 800-TC-OMEGASM F AX: (95) 203-359-7807 En Espanol: (95) 203 359-7803 e-mail: espanol@omega.
C o n t e n t s 1. GENERAL DESCRIPTION 2. MOUNTING INSTRUCTIONS 3. REPLACING FUSES 4. ASSEMBL Y 5. CURRENT INPUTS 6. CONNECTING TRANSMITTERS TO THE MUL TIPLEXER 7. CONNECTING Pt-100 TO THE MUL TIPLEXER 8. CONNECTING THE MUL TIPLEXER TO A PLC 9. CONTROL 9.
1. GENERAL DESCRIPTION 2 W ARNING: Never install a fuse rated more than 800mA The DRA-RTM-8 is a multiplexer for 16 analog inputs - eight of which, marked 1-8, are direct inputs for Pt-100 sensors, while the remaining (9-16), are for 4- 20mA current loops.
3 Insert the two printed cards into their slots. Connect the flat cable between them. Connect the front panel flat cables. The panel must be inserted into the grooves on both sides of the case while pressing down until a distinct "click" is heard.
4 Figure 4. 7. CONNECTING Pt-100 TO THE MUL TIPLEXER The Pt-100 probe should be connected according to fig 4. The three wires connecting the probe should be identical. The distance of the probe can be up to 200 meters. A shielded cable is recommended.
5 9.2 ADDRESS The required channel is selected by four address lines. The operating voltages are: Logical "1" 5V < Vi < 60V Logical "0" 0V < Vi< 0.
6 ADDRESS BUS E/T OUTPUT CHANNEL A0 A1 A2 A3 1 1 1 1 1 0 0 0 0 1 0 0 1 1 0 0 1 0 1 0 1 1 1 1 1 9 10 11 12 13 1 1 1 x 1 1 1 x 0 1 1 x 1 0 1 x 1 1 1 0 14 15 16 TEST MODE ADDRESS BUS E/T OUTPUT CHANNEL A0 A1 A2 A3 0 0 0 0 0 0 0 0 0 1 0 0 1 1 0 0 1 0 1 0 1 1 1 1 1 1 2 3 4 5 0 0 0 x 1 1 1 x 0 1 1 x 1 0 1 x 1 1 1 0 6 7 8 TEST MODE 9.
7 * The calibrator is set according to DIN 43760 Pt-100 table ( a = 0.00385) 10.2 CALIBRA TION T ABLES Note: Logic state of "0" is when the DIP switch lever is down. 10.2.1 "ZERO" - COARSE CALIBRA TION T ABLES ZERO TEMP O C CHANNELS 1-4 SW5 SW4 SW6 -50.
8 12. SPECIFICA TIONS 11. MUL TIDROP CONFIGURA TION In the disabled state (E=0), the multiplexer outputs no current and exhibits a high Z state. This mode allows the connection of several DRA-RTM-8 un.
RETURN REQUESTS / INQUIRIES Direct all warranty and repair requests/inquiries to the OMEGA Customer Service Department. BEFORE RETURNING ANY PRODUCT(S) TO OMEGA, PURCHASER MUST OBT AIN AN AUTHORIZED RETURN (AR) NUMBER FROM OMEGA’S CUSTOMER SERVICE DEP AR TMENT (IN ORDER TO A VOID PROCESSING DELA YS).
Where Do I Find Everything I Need for Process Measurement and Contr ol? OMEGA…Of Course! TEMPERA TURE ⻬ Thermocouple, R TD & Thermistor Probes, Connectors, Panels & Assemblies ⻬ W ire: T.
デバイスOmega Vehicle Security DRA-RTM-8の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Omega Vehicle Security DRA-RTM-8をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはOmega Vehicle Security DRA-RTM-8の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Omega Vehicle Security DRA-RTM-8の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Omega Vehicle Security DRA-RTM-8で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Omega Vehicle Security DRA-RTM-8を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はOmega Vehicle Security DRA-RTM-8の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Omega Vehicle Security DRA-RTM-8に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちOmega Vehicle Security DRA-RTM-8デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。