OMEGA HH508/HH509/HH509R Digital Thermometer.
OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com Servicing Nor th America: USA: ISO 9001 Certified Canada: One Omega Drive, Box 4047 976 Bergar Stamford, CT 06907-0047 Laval (Quebec) H7L5A1 Tel: (203) 359-1660 Tel: (514) 856-6928 FAX: (203)359-7700 FAX: (514) 856-6886 e-mail: info@omega.
1 The symbol on the instrument indicates that the operator must refer to an explanation in this manual. WARNING To avoid electrical shock, do not use this instrument when working voltages at the measurement surface over 24V AC or DC. WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens.
2 SPECIFICATIONS ENVIRONMENTAL Ambient Operating Ranges: 0°C to 50°C (32°F to 122°F) <80% R.H. Storage Temperature: -20°C to 60°C (-4°F to 140°F) <70% R.H. GENERAL Display: 3½ digit liquid crystal display (LCD) with a maximum reading of 1999.
3 ELECTRICAL Temperature Scale: Celsius or Fahrenheit user-selectable. Measurement Range: Thermocouple Range K-TYPE(0.1°C) -200°C to 1372°C, -328°F to 1999°F J-TYPE(0.1°C) -210°C to 1200°C, -346°F to 1999°F T-TYPE(0.1°C) -200°C to 400°C, -328°F to 752°F E-TYPE(0.
4 OPERATING INSTRUCTIONS Power Switch The key turns the thermometer on or off. In the MIN MAX record mode can not power off, must leave MIN MAX record mode then power off. °C °F Selecting the Temperature Scale Readings is displayed in either degrees Celsius(°C) or degrees Fahrenheit(°F).
5 MIN MAX record Mode Press MIN MAX key to enter the MIN MAX Recording mode, (displays the Maximum reading, Minimum reading, "MAX-MIN" reading and Average reading stored in record mode). In the this mode the automatic power-off feature is disabled and key and all function key are disabled.
6 TIMER STOP-WATCH Mode (for HH501AR / HH501AS) Press TIMER STOP-W key to enter stop-watch mode stop-watch is displayed. Press START/STOP key to toggle stop-watch starts and stops (Time goes up to 19 minutes & 59.9 seconds). Press CLEAR key to display 0.
7 SET mode (Relative value set and Hi/Lo Limits value set) 1. Press SET key to enter Relative valuse SET mode (Press ENTER key can escape relative valuse set mode), REL set mode. _ = = =.= is displayed (the ".=" is displayed when you set up the fourth umber).
8 Hi/Lo LIMITS mode Press Hi/Lo LIMITS key to enter the Hi/Lo Limits comparative mode, "LIMIT" is displayed. In the this mode the automatic power-off feature is disabled and other funtion key are disabled, when input temperature value exceed Hi.
9 OPERATOR MAINTENANCE WARNING To avoid possible electrical shock, disconnect the thermocouple connectors from the thermometer before removing the cover. Battery Replacement Power is supplied by a 9 volt "transistor" battery. (NEDA 1604, IEC 6F22).
10 RS-232 Operation Using Quick BASIC The following example shows how to send command instruction and receive data responses over the RS-232 interface using Quick BASIC.
11 IF C$ = "-" THEN DEC = DEC * -1 END IF B$ = MID$(A$, 10, 6) DEC1 = 0 N = 6 FOR I = 0 TO 5 C$ = MID$(B$, N, 1) N = N - 1 D = INSTR(F$, C$) - 1 DEC1 = DEC1 + (16 ^ I) * D NEXT DEC1 = DEC1 /.
12 RS-232 Transmit and Receive Comammand Summary 1. Control letter should use Capital Letter. Baud rate: 1200 Parity check: EVEN Data bits: 7 Stop bits: 1 "A" Activate data transmission &quo.
13 08 — Thermocouple Type "K" ® K type "J" ® J type "T" ® T type "E" ® E type "S" ® S type OUTPUT DATA FORMATS The data format consists of 32 bytes.
14 16 — none — " - " 17 — MSD 18 — LSD 19 — MSD 20 — LSD 21 — MSD 22 — LSD (H)ours (M)inutes 23 — Record mode "R" ® Record mode "M" ® MAX mode " .
15 COMMUNICATION 508BR RS-232C "COM1" PORT 4612-IR232 RS232C CABLE 1. Install the RSR232C cable between computer "COM1" and Thermometer RS232C port.
1 HH501(A,B) K / HH501(A,B) J / HH501(A, B)T / HH501( A,B)E HH508BR/HH508/HH509/ HH509R CA LIBRA TION PROCEDURE Note: The following calibration procedure s hould perform only by qualified technicians who have access to the items as fol lowing. Equipment: The class of calibrator ha d better 10 times g reater than the measur ed meter.
W ARRANTY / DISCLAIMER OMEGA ENGINEERING, INC. warr ants this unit to be free of defects in materials and workmanship for a period of 13 months from date of purchase. OMEGA Warr anty adds an additional one (1) month grace period to the normal one (1) year product warranty to cover handling and shipping time.
M-3203/0799 Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! TEMPERATURE þ Thermocouple, RTD & Thermistor Probes, Connectors, Panels & Assemblies þ W.
デバイスOmega Vehicle Security HH508の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Omega Vehicle Security HH508をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはOmega Vehicle Security HH508の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Omega Vehicle Security HH508の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Omega Vehicle Security HH508で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Omega Vehicle Security HH508を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はOmega Vehicle Security HH508の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Omega Vehicle Security HH508に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちOmega Vehicle Security HH508デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。