OMEGA OS643W Infrared Thermometer.
OMEGAnet SM On-Line Service http://www.omega.com Internet e-mail info@omega.com For immediate technical or application assistance: USA and Canada: Mexico and Latin America: Sales Service: 1-800-826-63.
1 INTRODUCTION This instrument is a portable easy use 3½ digit, compact-sized digital thermometer designed to use infrared or K-type thermocouple as input. Meter with Backlit LCD, Auto-hold function and auto power off (15 seconds approx.) feature after releasing MEAS button to extend battery life and with analog signal output function.
2 WARNING To avoid damage or burns, do not make temperature measurement in microwave ovens. The symbol on the instrument indicates that the operator must refer to an explanation in this manual.
3 CAUTION • Do not use the unit near any device which generates strong electromag- netic radiation or near a static electrical charge, as these may cause errors. • Do not point the lens at the sun or at any other source of strong light. If you do, the sensor may be damaged.
4 SPECIFICATIONS GENERAL Display: 3½ digit liquid crystal display (LCD) with maximum reading of 1999. Overrange : (OL) or (-OL) is displayed. Low battery indication: the " " is displayed when the battery voltage drops below the operating level.
5 Infrared Radiation ELECTRICAL Temperature Range: 0°F to 500°F, -20°C to 260°C. Display Resolution: 1°F / 1°C. Accuracy: ±3% of reading or ±6°F/3°C, whichever is greater @ 64.4 to 82.4°F (@ 18 to 28°C) ambient operating temperature. Temperature Coefficient: ±0.
6 Spot size increases with distance from the probe tip as shown.
7 K-type thermocouple ELECTRICAL Temperature Range: -58°F to 1999°F, -50°C to 1300°C. Display Resolution: 1°F / 1°C. Accuracy: Accuracy is specified for operating temperatures over the range of 64°F to 83°F (18°C to 28°C), for 1 year, not including thermocouple error.
8 OPERATING INSTRUCTIONS MEAS (MEASURE) Button Depress MEAS button to turn on the meter for measuring temperature. Releasing MEAS button to stop measuring temperature and automatically hold the display reading, the meter turns off automatically after 15 seconds.
OPERATION Infrared Radiation Measurements 1. When the power is off, pressing the MEAS button turns on the power. 2. Use "IR / K" switch to select IR Temperature function. 3. Use " " switch to turn on or turn off the display Back-Light.
10 K-type thermocouple Measurements 1. When the power is off, pressing the MEAS button turns on the power. 2. Use "IR / K" switch to select K Temperature function. 3. Use " " switch to turn on or turn off the display Back-Light. 4.
11 MEASUREMENT CONSIDERATIONS 1. Theory of Measurement Every object emits infrared energy in accordance with its temperature. By measuring the amount of this radiant energy, it is possible to determine the temperature of the emitting object.
12 4. Emissivity All objects emit invisible infrared energy. The amount of energy emitted is proportional to the object's temperature and its ability to emit IR energy. This ability, called emissivity, is based upon the material that the object is made of and its surface finish.
13 MAINTENANCE Battery Replacement Power is supplied by four 1.5V (AAA size) batteries. The " " appears on the LCD display when replacement is needed. To replace the batteries, remove the screw from the back of the meter and lift off the battery cover case.
Substance Thermal Substance Thermal emissivity emissivity Asphalt 0.90 to 0.98 Cloth (black) 0.98 Concrete 0.94 Human skin 0.98 Cement 0.96 Lather 0.75 to 0.80 Sand 0.90 Charcoal (powder) 0.96 Earth 0.92 to 0.96 Lacquer 0.80 to 0.95 W ater 0.92 to 0.96 Lacquer (matt) 0.
WARRANTY OMEGA warrants this unit to be free of def ects in materials and workmanship and to give satisfactory service for a period of 13 months from date of purchase. OMEGA W arranty adds an additional one (1) month g race period to the normal one (1) year product warranty to cov er handling and shipping time .
It is the policy of OMEGA to comply with all worldwide safety and EMC/EMI regulations that apply . OMEGA is constantly pursuing certification of its products to the European New Approach Directives. OMEGA will add the CE mark to every appropriate device upon certification.
Where Do I Find Everything I Need for Process Measurement and Control? OMEGA...Of Course! HEATERS þ Heating Cable þ Cartridge & Strip Heaters þ Immersion & Band Heaters þ Flexible Heaters .
デバイスOmega Vehicle Security OS643Wの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Omega Vehicle Security OS643Wをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはOmega Vehicle Security OS643Wの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Omega Vehicle Security OS643Wの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Omega Vehicle Security OS643Wで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Omega Vehicle Security OS643Wを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はOmega Vehicle Security OS643Wの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Omega Vehicle Security OS643Wに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちOmega Vehicle Security OS643Wデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。