15 Discovery Way, Acton, MA 01720 Phone: (978)263-3 584, Fax: (978)263-50 86 Web Site: www.princetoninstruments.com Operating Instructions A cton Se ries SP - 2300i 0.300 Meter Focal Length Triple Grating Imaging Monochromator / Spectrograph Rev 4 .0 9.
1 A cton SP-2300i Operating Inst ructions CONTENTS: Page I. Brief Introduction and SP- 2300i Description 2 II. Specifications 3 III. SP - 2300i Setup 4 A. Unpacking Note 4 B. Connecting the SP -2300i to a Computer 4 C. Cables and Connecti ons for SP-2300i 5 D.
2 I : Brief Introductio n and SP-2300i Descr iption Brief Introduction: This instruction manual is i ntended to assist you in s et - up and operation of your new SP -2300i monochromator/spectrograp h.
3 II : SP -2300i Specifications Every SP-2 300i monochromator or spec trograph includes a triple grating turret for 1, 2, or 3 gratings, 32 -bit microprocessor controll ed scanning, buil t-in RS - 232 and USB interfaces, and micrometer controlled e ntrance slit.
4 III : SP -2300i Setup Section III-A: Unpacking and Inspection Carefully unpack and examine the SP-2300i and any ac cessories purc hased. Note: Report any damage i mmediately t o the carrier and t o Princeton Instrum ents, and save all pac ki ng materia l.
5 Section III-C: Cables and Connections for SP-2300i The following computer cables are supplied wi th the SP-2300i: CC - 499 - 2 IBM AT or compatible 9 pin female ( DB9S) connector to 9 pin male connector (DB9P). 3650 -USB- 06 USB Cable Type A to Type B.
6 Section III-D: Mounting Accessories to the SP -2300i Slit A sse mblies All Acton accessories come with thei r own set of instructions for proper mounting and operation. The instructions below are only general i nformation. Pleas e refer to the individual instructions for detailed information.
7 Section III-E: Detector Mounting Instructions: Mounting Focal Plane Detectors to the SP-2300i Spectrograph The standard mounting f lange for CCDs and diode arrays acc ommodates detectors with two diff erent bolt circles. There are three (3) equal ly spaced #10 - 32 tapped holes on a 3.
8 Section III-F: Slit Width A djustments in the Model 716 Bilateral Slits The slit width of eac h bilateral slit assembly (716 type) is adjustable f rom 0.010 millimeters to 3 millimeters (10 to 3,000 µm) by a micrometer knob loc ated on the top of the sl it housing.
9 IV : SP-2300 i Operation Section IV-A: Initialization W hen power is turned ON to the SP -2300i, it ini tializes to a wavelength of 0.0 nm for grating number 1. If the power is switched OFF and then ON again to the SP-230 0i, it will re-ini tialize.
10 NM/MIN Sets the scan rate i n nm/min to 0.01 nm/min resolution with units nm/min ?NM/MIN Returns present scan rate i n nm/min to 0.01 nm/min resoluti on with units nm/min Grating Contro l Commands : GRAT ING Places specified grating in position to the wavelength of the wavelength on the present grating.
11 Slit Width Cont rol Command s (for Motorized Slits only ): FRONT-EX IT-SLIT Designates front ex it slit to receive slit control commands. Note : The designation remains in effec t until changed by another slit des ignator. This command does not have to be repeated unti l the designate d sl it is changed.
12 Section IV-C : Focusing and A lignment of A rra y Detectors W ith the array detector properly mounted to the SP-2300 i, use the followi ng procedure to align and foc us the array detector to the SP-230 0i optical system. It is as sumed that the array detection s ystem is running with SpectraSense or W inSpec.
13 Section IV-D : Conv erting the SP-2300i from Monochromator to Spectrograph, or Spectrograph to Monochromator Monochromator to Sp ectrograph: This procedure requires that you ha ve the monochromator version, Model SP- 2355 or SP- 2357 , plus the array detector mounting fl an ge.
14 V : Ap pendices and Schemat ic Dra wings A ppendix V -A: Selecting Alternate Start-Up Para meters The following are the Start-U p parameters and their def ault values: GRATING #1 WAVELENGTH 0.0 nm SCAN SPEED 100.0 nm/min Each of the above may be changed through the RS - 232 port or USB port using the f ollowing commands.
15 A ppendix V-B: Cable Connection Diagrams - Monochromators & Spectrographs to Computer Figure 9 . RS -232 Cable Wir ing Figure 10 . Cable Connections (Back Panel ) POWER Power connection +5 and +24 Vol ts USB HUB USB Hub 1.1 compatible. USB USB connection to computer.
16 A ppendix V -C: Standard Slit Assembly Drawing.
17 A ppendix V -D: Accessories Listing SP -2300i M onochromators & Spectrographs SP - 2355 SP - 2300i, 300mm Monoch romator: Micrometer adjustabl e entrance/exit slits, 90° optical path, internal.
18 TDS- 429 Dual Light Source with 30 watt deuterium & 30 watt tungsten -halogen lamps. Inc ludes housing with light c ollection mirror, mounting flange, and power supply. Provides c ombined output from 190 to >2.0 µm. XS - 43 3 75 Watt Xenon Light Source includes power s upply, housing, mounting flange and light collection mirror.
19 Filter Wheels and Order Sorting Filter Ass emblies FA - 2448 Six Position Filter Wheel Assembl y. Manual sample index ing, holds up to six samples, each 1.
20 A ppendix V -E: Gratings for SP-2300i Grooves/mm Blaze Wavelength Standard Part Number Mechanical Limit Notes: † Bl az ed H ol og rap hi c, 19 0- 450 nm , op tim um ra ng e †† H ol ogr ap hi .
21 A ppendix V -F: Certification and Warranty Certification Princeton Instruments certifies that thi s instrument was thoroughly tested and f ound to meet the specif ications furnished by Princ eton Instruments when it was shipped f rom the factory.
22 A ppendix V -G: Drawings.
23.
24.
デバイスPrinceton Digital (USA) SP-2300iの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Princeton Digital (USA) SP-2300iをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはPrinceton Digital (USA) SP-2300iの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Princeton Digital (USA) SP-2300iの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Princeton Digital (USA) SP-2300iで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Princeton Digital (USA) SP-2300iを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はPrinceton Digital (USA) SP-2300iの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Princeton Digital (USA) SP-2300iに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちPrinceton Digital (USA) SP-2300iデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。