Rev 24 .0 7. 09 FC - 446 - 02 1-U.DOC 15 Discovery Way, Acton, MA 0 1720 Phone: (978)263-3 584, Fax: (978)263- 5086 Web Site: www. princetoninstruments.
1 FC - 446 - 021 -U.DOC Description: The FC- 446 - 02 1-U is an adjustable f iber optic adapter designed fo r use with SpectraPr o series monochromators and spect rographs. It includes an integrated horizon tal and vertical adjust ment mechanism desig ned to position fib ers at the slit opening.
2 FC - 446 - 021 -U.DOC Fiber Cable Adaptability : The asse m bl y is compatible with 10 mm diameter fib er optic bundles, SMA connectors, and FC con nectors.
3 FC - 446 - 021 -U.DOC The following procedure is recommended fo r installing and aligning the FC - 44 6- 021 -U and fiber optic bu ndles. 1. Remove the entrance slit sh ipping cover to ex pose the s lit baff le as shown below. 2. Using a flat-bladed screwdrive r, remove the 4 slotted screws f rom the sli t assembly.
4 FC - 446 - 021 -U.DOC 4. Mount the fiber optic input assembly with the 4 X 8- 32 X 3/8 ” socket head cap screws provid ed. Fiber Cable and ada pters: The re are 4 diff erent fiber cable/adapter set up configu rations available f or the input of the FC - 446 - 021 -U adapter asse m b ly.
5 FC - 446 - 021 -U.DOC Set Up 1: FC Fibe r Cable If the fiber cable h as a single FC fibe r, attach the FC adap ter as shown below . S ingle fiber FC Cable FC Cable attached to FC Adapter Gently slide the ca ble/fiber asse m bly in to the fiber optic in put assembly so the end of the f iber adapter contacts the slits.
6 FC - 446 - 021 -U.DOC Set Up 2: SMA Fi ber Cable If the fiber cable h as a single SMA fi ber, attach the S MA adapter as shown below. Si ng le Fiber SMA Cable SMA Cable atta ched to SMA Adap ter Gently slide the ca ble/fiber assembly in to the fiber op tic input assembly so the end of the f iber adapter contacts the slits.
7 FC - 446 - 021 -U.DOC Set Up 3: 10 mm Dia m eter Sleeve with a single fiber Install the 10 mm sleeve to th e X- Y adapter and lock it into place b y tightening down the 3 set scre ws surrounding the adapter.
8 FC - 446 - 021 -U.DOC Set Up 4: 10 mm Dia m eter Sleeve with multiple fibers alig ned in a row Install the 10 mm sleeve to th e X- Y adapter and lock it into place b y tightening down the 3 set scre ws surrounding the adapter. Note the orientatio n of the fibers in the sleeve of the f iber cable.
9 FC - 446 - 021 -U.DOC 15 Discovery Way, Acton, MA 0 1720 Phone: (978)263-3 584, Fax: (978)263- 5086 Web Site: www. princetoninstruments.com RECOMMENDED MINIMUM BEND RADIUS FOR FUSED SILICA FIBERS: Fiber Diameters Momentary Min. Bend Radius Long Term Min.
デバイスPrinceton FC-446-021-Uの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Princeton FC-446-021-Uをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはPrinceton FC-446-021-Uの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Princeton FC-446-021-Uの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Princeton FC-446-021-Uで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Princeton FC-446-021-Uを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はPrinceton FC-446-021-Uの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Princeton FC-446-021-Uに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちPrinceton FC-446-021-Uデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。