Quickie S-1 1 Service Manual ©2005 Sunrise Medical Inc. 100740 Rev A.
Quickie S-1 1 T roubleshooting Guide INTRODUCTION ...................................................... 0.1 Specifications VSI Controller ........................................................... 0.2 Plugs/Connectors ...............................
QUICKIE S-1 1 SERVICE MANUAL P AGE 0.1 2005 SUNRISE MEDICAL INTRODUCTION Please read and follow instructions in this service manual before attempting to troubleshoot or repair this product for the first time.
QUICKIE S-1 1 SERVICE MANUAL P AGE 0.2 2005 SUNRISE MEDICAL VSI Controller VSI Controller SMASP0X01 SMASP0P01 VSI Controller Buttons Battery Condition Meter- A series of ten LED’s, which indi- cate charge level.
QUICKIE S-1 1 SERVICE MANUAL P AGE 0.3 2005 SUNRISE MEDICAL Plugs/Connectors 0 Vdc 9 0 Vdc 24 Vdc 24 Vdc LH Motor - LH Motor + 1 = 24 Vdc 2 = 0 Vdc 3 = Inhibit 1/ Programmer 1 = Brake Solenoid 2 = Bra.
QUICKIE S-1 1 SERVICE MANUAL P AGE 0.4 2005 SUNRISE MEDICAL RIGHT MOTOR CHARGER AC INPUT Indicator Light Circuit Breaker LEFT MOTOR SOLENOID 24v BRAKE SOLENOID BLACK FUSE FUSE BLACK BLACK ORANGE ORANG.
QUICKIE S-1 1 SERVICE MANUAL P AGE 1 2005 SUNRISE MEDICAL Section 1 T roubleshooting: No Power Check that the voltage is going to the controller , set the meter to dc volts and take a voltage reading from pin 1 (using the red lead) and pin 2 (using the black lead) to the charger port of the VSI controller (see figure A1.
QUICKIE S-1 1 SERVICE MANUAL P AGE 2 2005 SUNRISE MEDICAL 1 3 5 7 2 4 6 8 9 1.5 Battery Connection Check Check that the female Beau plug on the chair has voltage. Set the meter to dc volts and measure pins 5 (using the red lead of the meter) and 7 (using the black lead of the meter) as shown in (figure A1.
QUICKIE S-1 1 SERVICE MANUAL P AGE 3 2005 SUNRISE MEDICAL If the above steps did not correct the problem, change the main harness. 1.9 Main Harness 1.8 Circuit Breaker T est T o check the circuit breaker set the meter to ohms and measure the resistance across the circuit breaker as shown in (figure A1.
QUICKIE S-1 1 SERVICE MANUAL P AGE 4 2005 SUNRISE MEDICAL Section 2 Understanding Controller Display The wheelchair is locked. T o unlock the wheelchair , deflect the joystick forwards until the control system chirps. Then deflect the joystick in reverse until the control system chirps.
QUICKIE S-1 1 SERVICE MANUAL P AGE 5 2005 SUNRISE MEDICAL Section 3 Understanding VSI Controller Diagnostics Codes T est Left Motor Open Check that the batteries are fully charged and in good condition; and check all cables and connections. Check the connections to the left motor , look for a loose or dam- aged connector .
QUICKIE S-1 1 SERVICE MANUAL P AGE 6 2005 SUNRISE MEDICAL T est Left Motor Short Check that the batteries are fully charged and in good con- dition; and check all cables and connections. Check the connections to the left motor , look for a loose or damaged connector .
QUICKIE S-1 1 SERVICE MANUAL P AGE 7 2005 SUNRISE MEDICAL Understanding VSI Controller Diagnostics Codes (cont.) T est Right Motor Open Check that the batteries are fully charged and in good condition, and check all cables and connections. Check the connections to the right motor , look for a loose or dam- aged connector .
QUICKIE S-1 1 SERVICE MANUAL P AGE 8 2005 SUNRISE MEDICAL Understanding VSI Controller Diagnostics Codes (cont.) T est Right Motor Short Check that the batteries are fully charged and in good con- dition, and check all cables and connections. Check the connections to the right motor , look for a loose or damaged connector .
QUICKIE S-1 1 SERVICE MANUAL P AGE 9 2005 SUNRISE MEDICAL 1 3 5 7 2 4 6 8 9 Understanding VSI Controller Diagnostics Codes (cont.) A joystick trip is indicated. Make sure that the joystick is in the center position before switching on the control system.
QUICKIE S-1 1 SERVICE MANUAL P AGE 10 2005 SUNRISE MEDICAL Understanding VSI Controller Diagnostics Codes (cont.) Battery Fault Check that the batteries are fully charged, the correct volt- age and in good condition. T ake a voltage reading from pin 1 and pin 2 of the charger port of the VSI controller , see (figure A3.
QUICKIE S-1 1 SERVICE MANUAL P AGE 1 1 2005 SUNRISE MEDICAL Section 4 Disassembly/Reassembly , and Adjustment Disassembly a. Squeeze down the locking legs of the 3 pin charger plug and unplug it from the rear of the chair . (figure A4.1). b. Rock the controller plug back and forth until it is loose and then unplug it.
QUICKIE S-1 1 SERVICE MANUAL P AGE 12 2005 SUNRISE MEDICAL Disassembly/Reassembly , and Adjustment (cont) Disassembly a. Pull the left spring loaded release pin to disengage post (figure A4.5). While holding the left spring loaded release pin, remove the seat post.
QUICKIE S-1 1 SERVICE MANUAL P AGE 13 2005 SUNRISE MEDICAL Disassembly/Reassembly , and Adjustment (cont) Disassembly a. Lift the center shroud from the main frame (figure A4.7). Note: Center shroud is attached using a hook and loop fas- tener system.
QUICKIE S-1 1 SERVICE MANUAL P AGE 14 2005 SUNRISE MEDICAL Disassembly/Reassembly , and Adjustment (cont) Disassembly a. Peel the velcro strip apart, and remove the on-board charger (figure A4.1 1). b. Slide the female charger plug out of the beau plug bracket (figure A4.
QUICKIE S-1 1 SERVICE MANUAL P AGE 15 2005 SUNRISE MEDICAL Disassembly/Reassembly , and Adjustment (cont) A4.14 A4.15 Step 6 - Battery Disassembly a. There are two ways to access the batteries. One is fold the back and fl ip the seat forward by release the rear seat post latches ( fi gure A4.
QUICKIE S-1 1 SERVICE MANUAL P AGE 16 2005 SUNRISE MEDICAL Disassembly Note: For ease of disassembly , set the base frame on a block where all four wheels are at least 1 inch above the ground. a. Unwrap the wire looms from the motor and pull the motor connector out (figure A4.
QUICKIE S-1 1 SERVICE MANUAL P AGE 17 2005 SUNRISE MEDICAL Disassembly/Reassembly , and Adjustment (cont) Disassembly Access to Motor Brushes a. For quick access to Motor Brushes, perform steps 6b of this section. b. Use a flat blade screwdriver to remove the motor brush cap and then remove the brushes for inspec- tion.
QUICKIE S-1 1 SERVICE MANUAL P AGE 18 2005 SUNRISE MEDICAL Disassembly/Reassembly , and Adjustment (cont) Disassembly a. Remove all the zip ties on the base frame. b. Unwrap all the wire loom ties securing the harness to the frame. c. For the main harness, remove the two screws that hold- ing the female beau plug (figure A4.
デバイスSunrise Medical S-11の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Sunrise Medical S-11をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはSunrise Medical S-11の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Sunrise Medical S-11の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Sunrise Medical S-11で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Sunrise Medical S-11を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はSunrise Medical S-11の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Sunrise Medical S-11に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちSunrise Medical S-11デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。