Teledyne Analytical Instruments QUICKS T AR T GUIDE M O D E L 3 0 0 0 T A T ra ce Oxygen Analyzer.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 3 GETTING S T ARTED This Quickstart Guide is designed to get you s et up and ope rating your Teledyne An a lytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 4 • Left & Right Arrows Select between functi ons currently displayed on the VF D screen. • Up & Down Arrows Increment or decrement values of functions currently displayed.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 5 • Analog Output C onnecti on: Terminals for current and volt age a n alog output signals. • Digital Input Terminals: Termi n als for connecti ng d igital input signals. • Alarm Connections: 2 concentration alarms and 1 system alarm.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 6 SETUP AND INS T ALL Mount the Ana lyzer The Model 3000T A is for indoor use in a general purpose are a. It is NOT for hazardous environme nts of any type. It is design ed fo r f lush p anel m ount ing.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 7 4. If using remote calibra tion feature: • Connect zero gas to t he Z ERO IN port. • Connect span ga s to the SP AN IN port. 5. Otherwise, c a li b ra t io n ga se s m u st b e T e e' d i nt o t h e s a m p l e i n le t wi t h a p p r o p ri a t e v a l ve s .
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 8 • Connect primary input power: Insert the power cord into the power cord rece ptacle. • Install the fuse: Fuses are not installed at the fac tory. Be sure to install the proper fuse as par t of installati on.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 9 • Connect Range ID Contacts: There a re four ded icated Range ID relay contacts on the rear panel. The first three ranges are assigned to relays in ascending order. Range 1 ID— Low range Range 2 ID— Medium ra nge Range 3 ID— High range Range 4 ID—Air Cal Ra nge (25%).
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 10 9. With t h e O-ring in place, align the guide pin with the hole on the cell holder.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 11 BASIC OPER A TION Once the a naly zer has been installed, it can be configured for your application. Prior to using the instrument you m ust : • Calibrate t he instrument • Define the analysis r anges and choose whether to use autoranging or a fixed range.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 12 A vailab le Functions The six function butt ons are used to control the analyzer. Some functi ons have submenus that allow additional choices as s hown below. Any function ca n be sel ected at any time by pressing the appropriate button (unless password restrictions appl y).
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 13 Password Protection If a password is assigned, then setting the following system parameters can be done only after t he passw ord is en.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 14 TRAK/HLD Auto — Cal PSWD Logou t M ore • Use the ◄► arrow keys to scroll the blin king over to PSWD , • Press Enter to select the password function.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 15 Auto Mode Zero Calibration • Press Zero to enter the zero function mode. • Use the ▲▼ arrow keys to toggle between AUT O and MAN zero settling. Stop when AUT O appears blinking. Zero : Settling: AUTO <ENT> To B e gin • Press Enter to begin zeroing.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 16 Span Calibration Make sure an appropriate span gas is connecte d to the sample port. Auto Mode Spanning: • Press Span to enter the span function. • Use the arrow keys to toggle between AUTO and MAN span settling.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 17 Calib. Holding time Cal hold: 5 min • Press Enter key to continue. Span Val : 000008.00 <ENT>Span <UP>Mod # • Press ▲ (<UP>) to permit modification ( Mod # ) of span value.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 18 • Latc hing? Ltch–Y/N (Yes/No). • To change any of th e alarm parameters, use the ◄► arrow keys to move the blinkin g over to the specific par ameter. • Use the ▲▼ arrow keys to change t he va lue of the parameter.
Model 3000TA Quickstart Guide Teledyne Analytical Instruments 19 • Use the ▲▼ arrow keys to switch from AUTO t o FX/L O , FX/MED , or F X /HI to set the instrument on the desired fixe d range (low, medium , or h igh).
デバイスTeledyne 3000TAの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Teledyne 3000TAをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTeledyne 3000TAの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Teledyne 3000TAの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Teledyne 3000TAで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Teledyne 3000TAを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTeledyne 3000TAの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Teledyne 3000TAに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTeledyne 3000TAデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。