Teledyne Analytical Instruments QUICKS T AR T GUIDE M O D E L 3 0 0 0 T A - X L - E U T ra ce Oxygen Analyzer.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 1 GETTING S T ARTED This Quickstart Guide is designed to get you s et up and ope rating your Teledyne An a lytical Instruments Analyzer quickly. It shortcuts the details so you can install and use your new analyzer with a minimum of fuss.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 2 Function Keys : Six touch-sensiti v e membrane swit ches are used to cha ng e the specific function performed by the analyzer: • Analyze Perform anal ysis for oxygen cont ent of a sample gas.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 3 Rear Panel (Equipment Int erfa ce) At the rea r pane l you make the gas and electrical connections to the instrument. • Power Connection: Universal AC power source. • Gas Inlet and Outlet: One inlet and one exhaust out.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 4 SETUP AND INS T ALL Mount the Ana lyzer The Model 3000T A -EU is for indoor use in a general purpose area. It is NOT for ha zardous environme nts of any type. It is design ed fo r f lush p anel m ount ing.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 5 Electrical Connections CAUT ION: RE AD ALL SAFETY INFORMATION REGARDING ELECTRICAL HAZARDS ASSOC IATIED WITH THIS I NSTRUMENT BEF ORE MAKING ELECTRICAL CONNECTIONS.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 6 • Install 50-Pin Equipment Interface Connector Connect t h e 50 pin c onn ector to the mating connector on the rear panel. • Connect the 9-pin D conn ector to the RS-232 serial port.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 7 1. Place t he c ell on t h e holder with the screen side faci ng down. Note: There is a small location hole drilled in the holder. This hole mates with a guide pin on th e bottom rear of the cell blo ck.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 8 BASIC OPER A TION Once the a naly zer has been installed, it can be configured for your application. Prior to using the instrument you m ust : • Calibrate t he instrument • Define the analysis r anges and choose whether to use autoranging or a fixed range.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 9 A vailab le Functions The six function butt ons are used to control the analyzer.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 10 Any function ca n be sel ected at any time by pressing the appropriate button (unless password restrictions appl y).
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 11 ◄► arrow keys to scroll back a nd forth b e tween letters, and the ▲▼ arrow keys to change t he l etters to the proper password. • Press Enter to enter the pa ssword. Change Password? <ENT>=Yes <ESC>=No • Press Escape .
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 12 The ze roing proc ess will automatically conclude when t he output is w ithin t h e acceptable range for a good zero. Then the analyzer automatically returns to the Analyze mode.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 13 #### pp m Sp a n Slop e=#### ppm/s The instrument a utomati cally returns to the Analyze mode at the conclusion of the auto span cycle .. Manual Mode Spanning: • Press Span to start the Span function.
Quickstart Guide Trace Oxygen Analyzer Teledyne Analytical Instruments 14 • Press the Alarm button on the front panel to ente r the Alarm func tion. Make sure that AL–1 is blinking. AL—1 AL —2 Choose Al arm • Using the ◄► arrow keys to move the b linking over to AL–1.
Model 3000 TA-XL-EU Quickstart Guide Teledyne Analytical Instruments 15 • Use the ◄► arrow keys to blink the r ange to be set: low (L), medium (M), or high (H). • Use the ▲▼ arrow keys to enter t he uppe r value of t he rang e (all ranges begin at 0 ppm).
デバイスTeledyne 3000TA-XL-EUの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Teledyne 3000TA-XL-EUをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTeledyne 3000TA-XL-EUの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Teledyne 3000TA-XL-EUの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Teledyne 3000TA-XL-EUで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Teledyne 3000TA-XL-EUを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTeledyne 3000TA-XL-EUの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Teledyne 3000TA-XL-EUに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTeledyne 3000TA-XL-EUデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。