No. EJL- 106A Density (Consistency) Meter LQ500 Introduction The LQ500 density (consistency) meter uses microwave phase shift technology to determine concentrations of solids in the material to be measured flowing through pipes.
LQ500 2 range, which is the upper limit setting range. Resolution: Meter size 50 mm (2”) 80 mm (3”)to 300 mm (12”) Resolution 0.002%TS 0.001%TS Note 1: Above values are the results of comm uting ability in the phase measurements of the converter.
LQ500 3 200 mm (8”) 8 mS/cm maximum 250 mm (10”) 8 mS/cm maximum 300 mm (12”) 6 mS/cm maximum Note 1: The LQ500 can not have an accurate density (consistency) me asurement when it is over the specification according to reduce the microwave s ignal.
LQ500 4 hours when programming every 5 minutes. • Moving average function: In order to keep the average density (consistency) output, or in order to suppress the deflection width of the output. It helps for densit y (consistency) control. <Specification> Enable to determine a number from 1 to 99.
LQ500 5 11 0 (4.3”) (1 19) (4.7”) Converter commu nicatio n port Steel pipe (50A) (2”) 360 (14.2”) 390 (15.4”) 234 (9.2”) Contact sign al I/O port (1) Contact sign al I/O port (2) Densi ty.
LQ500 6 (7) Install an adjustable piping mechanism if there is a possibility that the detector pipe m ay not fit between mating flanges. (8) Install the LQ500 in a place where enough water pressure is applied. Therefore, it is recommended that the LQ500 be installed as far away as possible from the pipe outlet opened to atmosphere.
LQ500 7 L1 L2 Tx+ TX- Rx+ Rx- SG FG +24V 0V FG Ec+ Ec- FG Tx+ Tx- Rx+ Rx- SG FG +24V 0V FG Communication cable Power su pply cable Converter Power su pp l y 100 to 240 V ac , 50/60Hz Ground terminal (PE) (Ground re sistan ce 100 ohm or less ) Densi ty ( cons ist en cy) mea surem ent outp ut (4-20mAdc, 750 ohm max .
LQ500 No. EJL-106A Ordering Information When ordering the LQ500, refer to Table 4. Type Specification Code. An entry must be made for each of the columns. The following items must also be specified: 1. Fluid characteristics: • Type of material to be measured • Density (consistency) (max.
LQ500 9 Table 4. Type Specification Code (LQ500 Density (consistency) Meter) TYPE CAT Code SPECIF ICATION 1 2 3 4 5 6 7 8 9 10 11 12 13 14 L Q 5 0 0 Microwave Density (Consistency) Meter A S t a n d a.
デバイスToshiba LQ500の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Toshiba LQ500をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはToshiba LQ500の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Toshiba LQ500の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Toshiba LQ500で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Toshiba LQ500を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はToshiba LQ500の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Toshiba LQ500に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちToshiba LQ500デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。