ww w .EmersonProcess.com Instruction Manual BINOS 100 F & ML T 2 Addendum for Pressurized Analyzers intended to be used in Hazardous Areas classified Zone 2 3 rd Edition 08/2004 Instruction Manual.
BINOS 100 F & ML T 2 Addendum for Zone 2 Instruction Manual HAS8XE-IM-ATEX 08/2004 Emerson Process Management GmbH & Co. OHG Industriestrasse 1 D-63594 Hasselroth Germany T +49 (0) 6055 884-0 F +49 (0) 6055 884-209 Internet: www .EmersonProcess.
A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG P - 1 PREF ACE The purpose of this manual is to provide addit.
ML T 2 / BINOS 100 F A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 Emerson Process Management Manufacturing GmbH & Co. OHG P - 2 NOTE Highlights an essential operating procedure, condition or statement. Highlights an operation or maintenance procedure, practice, condition, statement, etc.
A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG P - 3 IMPORT ANT SAFETY INSTRUCTIONS SAFETY SUMMAR Y If this equipment is used in a manner not specified in these instructions, protective systems may be impaired.
ML T 2 / BINOS 100 F A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 Emerson Process Management Manufacturing GmbH & Co. OHG P - 4 TERMS USED IN THIS MANUAL A TEX Directive 94/9/EC, commonly called the A TEX („Atmosphères Explosibles“) products directive.
A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG P - 5 T erms used in this manual Dilution The continuous supp.
ML T 2 / BINOS 100 F A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 Emerson Process Management Manufacturing GmbH & Co. OHG P - 6.
A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG i - 1 TABLE OF CONTENTS Preface P-1 Defi niti ons ..............................................................
ML T 2 / BINOS 100 F A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 Emerson Process Management Manufacturing GmbH & Co. OHG i - 2 SECTION 3 S t artup 3 -1 3-1 Final Check .................................................................
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 1 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 SECTION 1 T echnical Description 1-1 Application and Principle of Operation The gas analyzers of type BINOS 100 F and ML T 2 are intended to measure gas components within gas mixtures.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 2 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 1-2 Instrument Layout 1- 2 Unused.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 3 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 1-3 Specification 1-3 Specification 1-3-2 Explosion Pro.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 4 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 1-3-4 Purge Gas Conditions 1-3-4 Purge Gas Conditions Purge gas: • Inert gas (e.g. Nitrogen) • Air (from an ex-free zone) T emperature: As ambient, but min.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 5 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 1-3-5 Sample Gas Conditions Applicable gases: Nonfla mm.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 1 - 6 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 1-4 Additional Safety Measures • Exhaust gas lines must end outside the hazardous area at a safe point. Sample gas lines may be returned to the sampling point.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 1 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 SECTION 2 Inst allation Before starting to install this.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 2 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 Fig. 2-1a Single enclosure analyzer (dimensions in mm) .
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 3 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 with Category 3 pressurization system Fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 4 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 Fig. 2-2a Dual enclosure analyzer (dimensions in mm) S .
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 5 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-3-2 Dual Enclosure Analyzer Fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 6 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-4 Connection of Sample Gas and Protective Gas (Purge Gas) When the analyzer with assembled EEx p safety device has been fixed, connect the gas lines according the diagram (fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 7 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-4-1 Hints on Flammable Gases • When applying flammable gases (either as sample gas or span gas) the corresponding gas inlets and outlets must be equipped with flame arrestors.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 8 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-4-2 ML T 2 Gas Flow Diagrams 2-4-2 ML T 2 Gas Flow Diagrams Fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 9 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-4-2 ML T 2 Gas Flow Diagrams Fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 10 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-4-3 BINOS ® 100 F Gas Flow Diagrams 2-4-3 BINOS ® 100 F Gas Flow Diagrams Fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 11 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-4-3 BINOS ® 100 F Gas Flow Diagrams 2-4-3 BINOS ® 100 F Gas Flow Diagrams (cont’d) Fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 12 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-4-3 BINOS ® 100 F Gas Flow Diagrams 2-4-3 BINOS ® 100 F Gas Flow Diagrams Fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 13 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-5 Safety Device Settings The EEx safety device is delivered with fixed basic parameters. The purge gas pressure is preset to 14 hPa (rel.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 14 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-7 Electrical Connections • All connections have to be done according to the schematics delivered together with the equipment.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 15 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-7-2 Mains Connection 2-7-2 Mains Connection 2-7-2-1 Purge System Remove the pressurization system control unit’ s cover .
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 16 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-7-2 Mains Connection The mains power line cord has to be connected to the terminals provided at the analyzer enclosure’s inner left side, near the EMC line filter (fig.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 17 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-7-3 Pressurization System Alarm Cont acts 2-7-3 Pressurization System Alarm Contact s The pressurization control unit provides alarm contacts which are widely programmable.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 2 - 18 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 2-7-4 Differential Pressure Switch(es) The contact of the differential pressure switch is connected to terminals X30 installed inside the analyzer .
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 3 - 1 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 SECTION 3 S t artup Startup can only be done properly b.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 3 - 2 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 4 - 1 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 SECTION 4 Maintenance After maintenance or replacement .
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 4 - 2 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 4 - 4 V erifications and T ests on Modified or Repaired.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 4 - 3 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 4-4-3 Leakage T est The leakage of the enclosure has to be measured with an overpressure of 25 mbar applied to the pressurized enclosure.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG 4 - 4 A TEX Instruction Manual HAS8xE-IM-A TEX January 2004 4-4-4 Removal of Modifications Remove all the modifications described in section 4-4-1. T ake special care of the gas connections to be tightend.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 1 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 APPENDIX A - 1 Declarations.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 2 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-1 Declarations.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 3 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-1 Declarations.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 4 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A - 2 Circuit Diagrams A-2-1 M L T 2, single A-2 Circuit .
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 5 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-1 ML T 2 Circuit Diagrams.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 6 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-1 ML T 2 Circuit Diagrams.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 7 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-1 ML T 2 Circuit Diagrams.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 8 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-2 M L T 2, dual A-2-2 ML T 2 Circuit Diagrams.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 9 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-2 ML T 2 Circuit Diagrams.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 10 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-2 ML T 2 Circuit Diagrams.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 11 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-2 ML T 2 Circuit Diagrams.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 12 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-2 BINOS ® 100 F Circuit Diagrams A-2-3 BINOS ® 100.
ML T 2 / BINOS 100 F Emerson Process Management Manufacturing GmbH & Co. OHG A- 13 A TEX Instruction Manual HAS8xE-IM-A TEX August 2004 A-2-2 BINOS ® 100 F Circuit Diagrams.
BINOS 100 F & ML T 2 Addendum Zone 2 Instruction Manual HAS8xE-IM-A TEX 08/2004 © Emerson Process Management GmbH & Co. OHG 2007 WORLD HEADQUARTERS ROSEMOUNT ANAL YTICAL EUROPE Emerson Process Management GmbH & Co.
デバイスEmerson HAS8xE-IM-ATEXの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Emerson HAS8xE-IM-ATEXをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはEmerson HAS8xE-IM-ATEXの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Emerson HAS8xE-IM-ATEXの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Emerson HAS8xE-IM-ATEXで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Emerson HAS8xE-IM-ATEXを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はEmerson HAS8xE-IM-ATEXの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Emerson HAS8xE-IM-ATEXに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちEmerson HAS8xE-IM-ATEXデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。