EmersonメーカーHASXEDE-IM-EXの使用説明書/サービス説明書
ページ先へ移動 of 140
www .EmersonProcess.com Instruction Manual HASXEDE-IM-EX 03/2012 Gas Analyzer X-STREAM Enhanced Series Flameproof V ariation For Use in Zone 1 and Division 2 Hazardous Areas Instruction Manual Addendu.
ESSENTIAL INSTRUCTIONS READ THIS P AGE BEFORE PROCEEDING! Emerson Process Management (Rosemount Analytical) designs, manufactures and tests its products to meet many national and international standards.
Emerson Process Management GmbH & Co. OHG T OC-1 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 TO C T able of contents T able of C onTenTs Introduction S-1 Denitions S-1 T erms used in this instruction manual ..........................
Emerson Process Management GmbH & Co. OHG TOC-2 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.5 F ront Panel Elements ................................................... 3-5 3.5.1 Display ................................................
Emerson Process Management GmbH & Co. OHG T OC-3 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 TO C T able of contents T able of Figures A.3 CSA Certicate of Compliance ........................................... A-12 A.4 Block Diagram .
Emerson Process Management GmbH & Co. OHG TOC-4 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 4-1: Leak T esting with U-tube Manometer (Flame arrestor) .......................................... 4-5 Fig. 4-2: Pressure Drop T est ....
Emerson Process Management GmbH & Co. OHG S-1 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 S Safety Instructions This instruction manual provides information about installing, operating .
Emerson Process Management GmbH & Co. OHG S-2 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 T erms Used In This Instruction Manual A TEX Directive 94/9/EC, commonly called the A TEX („Atmosphères Explosibles“) directive, dealing with equipment intended to be used in potentially explosive atmospheres.
Emerson Process Management GmbH & Co. OHG S-3 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 S Safety Instructions Symbols Used On A nd Inside The Unit This symbol at the instrument .
Emerson Process Management GmbH & Co. OHG S-4 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Symbols Used In This Manual Where one or more of the following symbols appear within this manua.
Emerson Process Management GmbH & Co. OHG S-5 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 S Safety Instructions SAFETY INSTRUCTIONS Intended Use Statement X-STREAM XE series gas analyzers are intended to be used as analyzers for industrial purposes.
Emerson Process Management GmbH & Co. OHG S-6 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Safety Instructions Authorized Personnel In-depth specialist knowledge is an absolutely necessary condition for working with and on the analyzer .
Emerson Process Management GmbH & Co. OHG S-7 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 S Safety Instructions Safety Instructions EXPLOSION HAZARD BY CONNECTIONS Consider the waiting .
Emerson Process Management GmbH & Co. OHG S-8 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Safety Instructions ELECTRICAL SHOCK HAZARD Installation, and connecting mains and signal cable.
Emerson Process Management GmbH & Co. OHG S-9 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 S Safety Instructions Safety Instructions EXPLOSION HAZARD BY DAMAGED FLAME P A THS T ake care .
Emerson Process Management GmbH & Co. OHG S-10 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Safety Instructions EXPLOSION HAZARD BY HIGH PRESSURE Risk of internal overpressure under leak.
Emerson Process Management GmbH & Co. OHG S-1 1 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 S Safety Instructions Safety Instructions HOW T O ST A Y IN COMPLIANCE WITH THE EUROPEAN DIRECTIVE 94/9/EC ("A TEX") WHEN PERFORMING GAS ANAL YSIS WITHIN A FLAMEPROOF ENCLOSURE.
Emerson Process Management GmbH & Co. OHG S-12 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Safety Instructions.
Emerson Process Management GmbH & Co. OHG S-13 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 S Safety Instructions Safety Instructions HIGH TEMPERA TURES Hot parts may be exposed when working on photometers and/or heated components in the unit.
Emerson Process Management GmbH & Co. OHG S-14 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 General Operating Notes GENERAL OPERA TING NOTES • The unit must be installed in a clean and dry area, protected from strong vibrations and frost .
Emerson Process Management GmbH & Co. OHG 1-1 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Chapter 1 T echnical Description 1.1 Overview The new X-STREAM ameproof gas analy - zers are designed to be used in hazardous areas.
Emerson Process Management GmbH & Co. OHG 1-2 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.3 Protective Measures in Detail The cast aluminum enclosure consists of two parts: base and cover , connected by hinges.
Emerson Process Management GmbH & Co. OHG 1-3 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.4 High Pressure Option / Purge Option 1.
Emerson Process Management GmbH & Co. OHG 1-4 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.5 Compliances 1.5 Explosion Protection Compliances These products are available in two dif fe.
Emerson Process Management GmbH & Co. OHG 1-5 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.5 Compliances 1.5.1 Special Conditions for Safe Use • Only screws M16x45 ISO 4762 A2-70 as specied in the maintenance section of this manual shall be used ( spare part # 42716945).
Emerson Process Management GmbH & Co. OHG 1-6 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.6 Nameplate Label 1.6 Nameplate Label Fig. 1-3: Nameplate Label Details (exemplary) Area Desc.
Emerson Process Management GmbH & Co. OHG 1-7 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.7 T echnical Data 1.7 T echnical Data Cable inlets (enclosure threads; M20 x 1.5) Flame arre stors with gas tt ings (enclosure threads: M18 x 1.
Emerson Process Management GmbH & Co. OHG 1-8 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.7 T echnical Data Site of installation Humidity (non-consending) < 90 % r .
Emerson Process Management GmbH & Co. OHG 1-9 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.7 T echnical Data Standard interfaces up to 5 analog outputs (standard:.
Emerson Process Management GmbH & Co. OHG 1-10 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.7 T echnical Data Optional interfaces Digital I/O board 7 or 14 digital inputs electrical specication max. 30 V , internally limited to 2.3 mA HIGH: min.
Emerson Process Management GmbH & Co. OHG 1-1 1 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.7 T echnical Data Optional interfaces 2 Analog inputs electrical spec.
Emerson Process Management GmbH & Co. OHG 1-12 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.7 T echnical Data Increased sample and calibration gas pressure Increased gas pressure Above 1 100 hPa to max. 1500 hPa (take care of the measurement principles limits!) Maximum ow 1.
Emerson Process Management GmbH & Co. OHG 1-13 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 1-5: Signals T erminals 1.
Emerson Process Management GmbH & Co. OHG 1-14 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 1-6: Power T erminals / Fuse Holders 1.7 T echnical Data Power Connection 1 Power termina.
Emerson Process Management GmbH & Co. OHG 1-15 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.8 Measurements Specications Sample gas components and measuring ranges (standard congurations) In total, more than 60 gases are detectable, so the following table gives an overview only .
Emerson Process Management GmbH & Co. OHG 1-16 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.8 Measurements Specications Standard and Enhanced Performance Specications T ab. 1-7: IR, UV , VIS, TCD - Standard and Enhanced Measurement Performance Specications T ab.
Emerson Process Management GmbH & Co. OHG 1-17 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.8 Measurements Specications T ab. 1-9: Oxygen - Standard and Enhanced Measurement Performance Specications Note 1! Not all data listed are applicable to all analyzer versions (e.
Emerson Process Management GmbH & Co. OHG 1-18 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.8 Measurements Specications Special Performance Specications for Gas Purity Measurements (ULCO & ULCO 2 ) T ab.
Emerson Process Management GmbH & Co. OHG 1-19 1 T echn. Description X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.9 V apor Recovery Application 1.9 V apor Recovery Application (Simultaneous Measurement of CH 4 and Non-CH 4 ) This application is served by a special con - guration of the X-STREAM XEFD ameproof analyzer .
Emerson Process Management GmbH & Co. OHG 1-20 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1.9 V apor Recovery Application Converter efciency The measurement accuracy is highly depen- dent on the converter efciency: If this is too low , the converter material needs replace - ment.
Emerson Process Management GmbH & Co. OHG 2-1 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Chapter 2 Installation On receipt, check the packaging and its contents thoroughly for damage. Inform the carrier immediately of any damage to packaging or contents, and keep dama- ged parts until clarication.
Emerson Process Management GmbH & Co. OHG 2-2 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 HEA VY INSTRUMENT HAZARD X-STREAM XEFD analyzers, to which this manual relates, intended to be wall mounted and/or outdoor installed, weigh up to approx.
Emerson Process Management GmbH & Co. OHG 2-3 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Install the analyzer to a stand or a wall by mea ns of 4 eyebolts, provided at the instru - ments rear side.
Emerson Process Management GmbH & Co. OHG 2-4 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 1 2 3 4 5 5 6 7 2.3 Installation - Gas Lines 2.3 Connecting Gas Lines Gas inlets and outlets are protected by ame arrestors, supporting stainless steel pipes of either 3.
Emerson Process Management GmbH & Co. OHG 2-5 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.3 Installation - Gas Lines Gas ttings are accessible at the instrument‘s outer bottom side.
Emerson Process Management GmbH & Co. OHG 2-6 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.3 Installation - Gas Lines The gas paths need additionally to be protected by suitable external inline ame arrestors, designed and approved for the in- stallation area and for the applied higher gas pressure (above 1 100 hPa to max.
Emerson Process Management GmbH & Co. OHG 2-7 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.3 Installation - Gas Lines 2.
Emerson Process Management GmbH & Co. OHG 2-8 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.4 Electrical Installation 2.4 Installation - Electrical EXPLOSION AND ELECTRICAL SHOCK HAZARD.
Emerson Process Management GmbH & Co. OHG 2-9 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical SELECT THE CORRECT TYPE OF CABLE ENTR Y Before sta.
Emerson Process Management GmbH & Co. OHG 2-10 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical EXPLOSION HAZARD WHEN INST ALLING CABLE GLANDS When using the AD.
Emerson Process Management GmbH & Co. OHG 2-1 1 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical T ype AD-U adapter installation instructions.
Emerson Process Management GmbH & Co. OHG 2-12 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 2-8: Allocation of terminals 1 T erminals for signal cables 2 Power EMI lter 3 4 cable entries for power and signal cables 4 Power terminals with integrated fuses 5 Ethernet and USB connectors 2.
Emerson Process Management GmbH & Co. OHG 2-13 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical All signal cables are to be connected via screw terminals, except the USB and ethernet connectors, located inside the analyzer .
Emerson Process Management GmbH & Co. OHG 2-14 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 **) Conguration of relay output terminals as per standard factory setting (NAMUR status signals) Fig.
Emerson Process Management GmbH & Co. OHG 2-15 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Pin Signal P2.1 Channel 1, (+) 4 (0) - 20 mA P2.2 Channel 1, GND P2.3 Channel 2, (+) 4 (0) - 20 mA P2.4 Channel 2, GND P2.5 Channel 3, (+) 4 (0) - 20 mA P2.
Emerson Process Management GmbH & Co. OHG 2-16 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 2-1 1: Modbus Interface - Ethernet connector 2.4 Installation - Electrical T o install this connection, a cable must be fed through the cable entry without a connector .
Emerson Process Management GmbH & Co. OHG 2-17 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Pin Signal P3.1 Input 1 P3.2 Input 2 P3.3 Input 3 P3.4 Input 4 P3.5 Input 5 P3.6 Input 6 P3.7 Input 7 P3.8 GND for Inputs 1-7 P3.9 not used P3.
Emerson Process Management GmbH & Co. OHG 2-18 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.4 Installation - Electrical 2 analog inputs 0 - 1 (10) V , congurable via software; R in = 100 k Ω optional (requires the use of wire-bridges, see.
Emerson Process Management GmbH & Co. OHG 2-19 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Connecting the power cord T o install the cable proceed according to the installation instructions for either conduits or cable glands, given on page 2-1 1.
Emerson Process Management GmbH & Co. OHG 2-20 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 2.4.1 Equipotential bonding conductor The X-STREAM XEFD enclosure provides an additional terminal for connecting an equi - potential conductor , located at the base part ange´s rear side (see g.
Emerson Process Management GmbH & Co. OHG 2-21 2 Installation X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 After all connections are established in the specied manner , and veried, • x the cables according to the installation instruction given in the cable glands/ con - duits manufacturer documentation.
Emerson Process Management GmbH & Co. OHG 2-22 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012.
Emerson Process Management GmbH & Co. OHG 3-1 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup Chapter 3 Start up OPERA TION A T LOW TEMPERA TURES When operating an instrument at t.
Emerson Process Management GmbH & Co. OHG 3-2 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.2 Leak T est 3.2 Performing a Leak T est Max.
Emerson Process Management GmbH & Co. OHG 3-3 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup Once the unit has been unpacked and in - stalled, we recommend to rst check the settings, and if necessary adjust them to the user ’s needs.
Emerson Process Management GmbH & Co. OHG 3-4 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.3 Symbols Used 3.4 Symbols and T ypographical Conventions In the following sections, the symbols and typographical conventions explained below are used to describe the software menus and navigation.
Emerson Process Management GmbH & Co. OHG 3-5 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.5 Front Panel Elements 3.5 Front Panel Elements Note! During an active webserver se.
Emerson Process Management GmbH & Co. OHG 3-6 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.5.3 Keys Six keys enable the use of the menu system. Depending on the operational mode (measu- ring, browsing menus, editing) they have the following functions: enter key: Mode Function Measuring Enter main menu Browsing Open submenu (.
Emerson Process Management GmbH & Co. OHG 3-7 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup left key: Mode Function Measuring Enter main menu or open 2 nd measurement display pa.
Emerson Process Management GmbH & Co. OHG 3-8 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.6 Software The analyzer software displays measurement results and status messages, allows parame - ters to be set and edited, and maintenance functions (e.
Emerson Process Management GmbH & Co. OHG 3-9 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.6 Software Some menus have more entries than can dis - played at once. In these menus, an indicator in the last ( ) and/or rst ( ) line indicates the direction the menu continues in.
Emerson Process Management GmbH & Co. OHG 3-10 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.6 Software Editing The editing mode enables changing parame - ters. It is initiated by pressing enter . If the selection is a parameter list , the current entry is highlighted and may be changed by up and down .
Emerson Process Management GmbH & Co. OHG 3-1 1 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.6.1 Access Levels & Codes 3.
Emerson Process Management GmbH & Co. OHG 3-12 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.7 Start Up Depending on the last action performed by the user , one of the following messages may be displayed to assist or inform the user . Conrmation of execution of function: Conrms that a function or procedure (e.
Emerson Process Management GmbH & Co. OHG 3-13 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup SETUP enables several additional congu - rations, e.
Emerson Process Management GmbH & Co. OHG 3-14 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 If the analyzer is operational and it becomes clear that the incorrect language has been set, which is unintelligible to the operator , the 3.8 Language Settings 3.
Emerson Process Management GmbH & Co. OHG 3-15 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.9 Checking the Settings Starting with the measurement display ( 3-12 ), pressing a.
Emerson Process Management GmbH & Co. OHG 3-16 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.9 Checking the Settings 3.9.1 Installed Options Press left to return to seTuP , highlight " In- stalled options" and press enter . This 2 pages menu indicates, which of the possible optional components are installed in the unit.
Emerson Process Management GmbH & Co. OHG 3-17 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.9 Checking the Settings 3.9.2 Conguring the Display Check the settings for the measurement display , temperature and pressure units, and for menu access: press enter to open DIS- PLA Y .
Emerson Process Management GmbH & Co. OHG 3-18 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Multi-channel unit: Select the component to be set in the gas component selection menu. Once the display settings have been chek - ked, press le f t to return to SETUP and open CALIBRA TION to check the calibration settings.
Emerson Process Management GmbH & Co. OHG 3-19 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup Signal damping (set in DAMPING) allows smoothing the output signal, but also affects the response time of outputs and display: • The factory setting is 0 seconds.
Emerson Process Management GmbH & Co. OHG 3-20 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.9 Checking the Settings Press left to return to SETUP , and then open IN/OUTPUTS. and from there enter ANALOG OUTPUTS. 3.9.4 Setting the A nalog Outputs Select the analog output you like to check.
Emerson Process Management GmbH & Co. OHG 3-21 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.9 Checking the Settings *) Numbers 1 to 5 refer to components [chan- nels] 1 to 5:.
Emerson Process Management GmbH & Co. OHG 3-22 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.9 Checking the Settings Next, select the output range: • 0-20 mA (dead zero) generates a 20 mA signal, if a concentration is measured at the upper limit of the signal range.
Emerson Process Management GmbH & Co. OHG 3-23 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.9 Checking the Settings T ab. 3-1: Analog Output Signals Settings and Operational Modes Operation Modes acc.
Emerson Process Management GmbH & Co. OHG 3-24 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Signal: Comp1 OutRange: 0-20 mA Low Scale: 0.
Emerson Process Management GmbH & Co. OHG 3-25 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.9 Checking the Settings Press l eft until SETUP is displayed, then open ALARMS - CONCENTRA TION 3.
Emerson Process Management GmbH & Co. OHG 3-26 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.9 Checking the Settings The most important settings have now been checked and the unit is congured to suite your needs. A backup copy of these conguration data can now be saved.
Emerson Process Management GmbH & Co. OHG 3-27 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.9 Checking the Settings Congratulations! Y ou´ve now successfully completed check- ing the unit’ s setup! • Press home to return to the MEASURE- MENT DISPLA Y .
Emerson Process Management GmbH & Co. OHG 3-28 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.10 We b Browser IP address for Ethernet1 connector Info Firmware 1.0 DSP version 1.0 Serial no 123456789 Components.. Installed options.. Ethernet1 IP 123.
Emerson Process Management GmbH & Co. OHG 3-29 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3 Startup 3.10 Web Browser IP address for Ethernet2 connector T o directly gain access to the instrument´s web browser interface, rst ensure the instrument is powered.
Emerson Process Management GmbH & Co. OHG 3-30 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 3.10 Web Browser Default password is: password Fig.
Emerson Process Management GmbH & Co. OHG 4-1 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Chapter 4 Service and Maintenance Note! This chapter deals with service and maint.
Emerson Process Management GmbH & Co. OHG 4-2 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4 Service and Maintenance EXPLOSION HAZARD BY FLAMMMABLE GASES Leaks may cause explosion when m.
Emerson Process Management GmbH & Co. OHG 4-3 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Modications made on the electrical appara - tus aff ecting the integrity of the type of protec - tion or the temperature of the apparatus shall be permitted only if the modied apparatus is resubmitted to a testing station.
Emerson Process Management GmbH & Co. OHG 4-4 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.1.1.3 T ests on Flame Arrestors The following procedure describes how to perform a leak test with the ame arrestor installed at the instrument.
Emerson Process Management GmbH & Co. OHG 4-5 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 4-1: Leak T esting with U-tube Manometer (Flame arrestor) Analyzer overpressure approx. 0.725 psi/50 mbar stop valve W ater ame arrestors 4.
Emerson Process Management GmbH & Co. OHG 4-6 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.2 Replacement of Parts EXPLOSION HAZARD BY REPLACEMENTS P ARTS Replacement of parts found defective is permitted only by using original parts! Violation voids the approvals and may cause explosions! 4.
Emerson Process Management GmbH & Co. OHG 4-7 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.2 Replacement of Parts Specically for CSA-C/US approved analyzers Metric-to-NPT adapters Any CSA-C/US certied ameproof type Male thread: M20 x 1,5 Female thread: 1 ⁄ 2 “ or 3 ⁄ 4 “ per request e.
Emerson Process Management GmbH & Co. OHG 4-8 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.3 V apor Recovery Application - Special Information 4.
Emerson Process Management GmbH & Co. OHG 4-9 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.3 V apor Recovery Application - Special Information • Ensure „Pump 1“ within the control menu is set to ON .
Emerson Process Management GmbH & Co. OHG 4-10 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.3 V apor Recovery Application - Special Information Fig.
Emerson Process Management GmbH & Co. OHG 4-1 1 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 4-6: Heated jacket 1: Jacket 2: Union nuts 1 1. Loosen the lace cord that keeps the jacket closed. 2. Open the jacket to get access to the con - verter .
Emerson Process Management GmbH & Co. OHG 4-12 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 4-8: Converter lling sequence 4.3 V apor Recovery Application - Special Information 5. Remove the end cap. 6. Remove the washer . Behind the washer , the vessel is lled with converter material.
Emerson Process Management GmbH & Co. OHG 4-13 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Fig. 4-9: Jacket in converter assembly 1: Jacket 2: Boot hooks (8 pcs) 1 2 13. Lay the lace cord around the boot hooks and close the jacket.
Emerson Process Management GmbH & Co. OHG 4-14 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.3 V apor Recovery Application - Special Information 4.3.3 Failure Situation Separated from the converter assembly an independent over temperature protection de- vice is installed.
Emerson Process Management GmbH & Co. OHG 4-15 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.4 Perform a Calibration 4.4 Perform a Calibration We rec ommend to perform at least a zero calibration, after startup of the instrument, to ensure proper measuring results.
Emerson Process Management GmbH & Co. OHG 4-16 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 4.4 Manual Calibration 4.4.1 Manual Zero Calibration Starting from the MEASUREMENT SCREEN press d own to open the MAIN MENU and enter CONTROL .
Emerson Process Management GmbH & Co. OHG 4-17 4 Maintenance X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Zero C alibration.. Span C alibration.. Advanced C alibration.. Apply Gas.. Lock M enus! Acknowledgements Pump 1: O ff Pump 2: O ff Ranges.
Emerson Process Management GmbH & Co. OHG 4-18 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Zero C alibration.. Span C alibration.. Advanced C alibration.. Apply Gas.. Lock M enus! Acknowledgements Pump 1: On Pump 2: O ff Ranges.. 4.
Emerson Process Management GmbH & Co. OHG 5-1 5 Dismounting X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Chapter 5 Dismounting and Disposal 5.
Emerson Process Management GmbH & Co. OHG 5-2 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 5.1 Dismounting & Disposal When the instrument has reached the end of its useful life, When.
Emerson Process Management GmbH & Co. OHG A-1 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Appendix This chapter contains EC Declaration of Conformity A.1, page A-2 A TEX EC T ype Examination Certicate A.2, page A-3 CSA Certicate of Compliance A.
Emerson Process Management GmbH & Co. OHG A-2 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.1 EC Declaration of Conformity A.1 EC Declaration of Conformity.
Emerson Process Management GmbH & Co. OHG A-3 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate A.
Emerson Process Management GmbH & Co. OHG A-4 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate.
Emerson Process Management GmbH & Co. OHG A-5 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate.
Emerson Process Management GmbH & Co. OHG A-6 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate.
Emerson Process Management GmbH & Co. OHG A-7 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate.
Emerson Process Management GmbH & Co. OHG A-8 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate.
Emerson Process Management GmbH & Co. OHG A-9 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate.
Emerson Process Management GmbH & Co. OHG A-10 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate.
Emerson Process Management GmbH & Co. OHG A-1 1 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.2 A TEX EC T ype Examination Certicate.
Emerson Process Management GmbH & Co. OHG A-12 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.3 CSA Certicate of Compliance A.3 CSA Certicate of Compliance Certificate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 DQD 507 Rev .
Emerson Process Management GmbH & Co. OHG A-13 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.3 CSA Certicate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 DQD 507 Rev .
Emerson Process Management GmbH & Co. OHG A-14 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.3 CSA Certicate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 DQD 507 Rev .
Emerson Process Management GmbH & Co. OHG A-15 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.3 CSA Certicate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 DQD 507 Rev .
Emerson Process Management GmbH & Co. OHG A-16 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.3 CSA Certicate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 DQD 507 Rev .
Emerson Process Management GmbH & Co. OHG A-17 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.3 CSA Certicate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 DQD 507 Rev .
Emerson Process Management GmbH & Co. OHG A-18 X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.3 CSA Certicate of Compliance Certificate: 1714037 (LR 105173) Master Contract: 185562 Project: 2507282 Date Issued: March 14, 2012 DQD 507 Rev .
Emerson Process Management GmbH & Co. OHG A-19 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Netzteil 100V -240V ~ Power Supply AC 100V -240V ~ wide range 24V - 24V DC Netzteil 100V -240V ~ Power Supply AC 100V -240V ~ wide range *1 24V - 24V DC 15.
X-STREAM XEFD Emerson Process Management GmbH & Co. OHG A-20 Instruction Manual HASXEDE-IM-EX 03/2012 Intern Internal Fieldhousings (all) XSP 01 / XIN 01 siehe Blatt 9, 1 1 see sheet 9, 1 1 P2 UFPSW 01 T astatur Keyboard Stromlaufplan / s chematics: 4.
Emerson Process Management GmbH & Co. OHG A-21 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Druck Ch1 P31 P30 Pressure Ch2 T emperatur Ch1 P18 P17 P16 P1 1 T emperature Ch4 Ch.
X-STREAM XEFD Emerson Process Management GmbH & Co. OHG A-22 Instruction Manual HASXEDE-IM-EX 03/2012 Druck Ch1 P30 Pressure T emperatur Ch1 P16 P1 1 T emperature Ch2 Durchfluss Ch1 P4 P5 Flow Ch2.
Emerson Process Management GmbH & Co. OHG A-23 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 Spurensauerstoffmessung Trace Oxygen 4 P1 XOX 01 Stromlaufplan / s chematics: 4.350-0440 XTEC 02 Stromlaufplan/schematic 4.350-1280/3 4.
X-STREAM XEFD Emerson Process Management GmbH & Co. OHG A-24 Instruction Manual HASXEDE-IM-EX 03/2012 2-adrige Leitung / cable with 2 lines Symbolerklärung: Description of used icons: Leitungsfarbe / cable colour bl Steckverbindung / plug connection /2.
Emerson Process Management GmbH & Co. OHG A-25 A Appendix X-STREAM XEFD Instruction Manual HASXEDE-IM-EX 03/2012 A.5 Assignment of T erminals and Socket A.
X-STREAM XEFD Emerson Process Management GmbH & Co. OHG A-26 Instruction Manual HASXEDE-IM-EX 03/2012.
Instruction Manual HASXEDE-IM-EX 03/2012 X-STREAM XEFD.
Instruction Manual HASXEDE-IM-EX 03/2012 X-STREAM XEFD ©2012 Emerson Proce ss Manage ment. All rights re served. The Emerson logo is a tradem ar k and service mark of Eme r son Electric Co. Ro semoun t Analytical is a mark of one of the Emerson Proce ss Manage ment f amily of companie s.
デバイスEmerson HASXEDE-IM-EXの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Emerson HASXEDE-IM-EXをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはEmerson HASXEDE-IM-EXの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Emerson HASXEDE-IM-EXの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Emerson HASXEDE-IM-EXで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Emerson HASXEDE-IM-EXを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はEmerson HASXEDE-IM-EXの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Emerson HASXEDE-IM-EXに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちEmerson HASXEDE-IM-EXデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。