Flow Meter FC01-LQ User handbookp.1Circuit Protection & Controlp.11 Descriptionp.71.1 Calorimetric monitoring procedurep.81.2 System descriptionp.9DC 19 … 32 Vp.91.2.1 User interfacesp.102.1 Selection of materialp.122 Installationp.122.2 Installation of the monitoring headsp.13SWp.142.2.2 Monitoring head CSF-11 with adjustable immersion depthp.162.3.1 Mechanical installationp.202.3 Installation of electronic control unit FC01-LQp.202.3.2 Electrical connectionp.233 Operating systemp.30PARAMETER SELECTION (MENU)p.30CONFIGURATION LEVEL (MENU)p.30MAIN LEVEL (MENU)p.304.2 Measuring cyclep.324.1 Switch-on performancep.324 Operation and main menup.32TOTALISATORp.32LAST ERRORp.32PEAK VALUE MAXp.32PEAK VALUE MINp.324.2.1 Operating datap.33T F 37044.9 m 3 5.0 m/s -13.5 ° Cp.33T Fp.33T F Fp.335.0 m/s -13.5 ° Cp.3322206.9 l/minp.3337004567.9 lp.331332.4 m 3 /hp.33370.1 l/sp.33HEATING UP *p.355.2 Monitoring head data (menu option: SENSOR CODE )p.365.1 Selection of monitoring head (menu option: SENSOR SELECT )p.365 Configuration (menu option: CONFIGURATION )p.365.3 Limit switch combinations (menu option: LIMIT SWITCHES )p.375.6 Display (menu option: DISPLAY SELECT )p.385.5 Medium temperature unit (menu option: TEMP. UNIT )p.385.4 Flow rate unit (menu option: FLOW UNIT )p.385.7 Bar graph (menu option: BARGRAPH )p.39LS2 ONp.39T Tp.3923 ° Cp.395.9 Frequency output (menu option: FREQUENCY OUTPUT )p.405.8 Pipe diameter (menu option: PIPE SIZE )p.405.12 Quitting the configuration menup.415.11 Analogue output - medium temperature (menu option: ANA OUT TEMP. )p.415.10 Analogue output - flow rate (menu option: ANA OUT FLOW )p.415.13 Configuration menup.43CONFIGURATION LIMIT SWITCHESp.43CONFIGURATION SENSOR SELECTp.43CONFIGURATIONp.435.14 Configuration submenusp.44SENSOR SELECTp.44SENSOR CODEp.44Configuration submenus (cont’d)p.45LIMIT SWITCHES LS1 → F LS2 → Tp.45LIMIT SWITCHES LS1 → T LS2 → Fp.45FLOW UNIT PERCENT (%)p.45FLOW UNIT FEET/SEC. (FPS)p.45TEMP. UNIT CELSIUS ( ° C)p.45TEMP. UNIT FAHRENHEIT ( ° F)p.45TEMP. UNITp.45KELVIN (K)p.45PIPE SIZEp.46FREQUENCY OUTPUT?p.46FREQUENCY OUTPUTp.46° US-GALLONSp.466.1 Measuring time (menu option: MEAS. TIME )p.476 Parameter selection (menu option: PARAMETERS )p.47OFFp.47ONp.476.4 Scaling factor (menu option: FLOWSCALE )p.486.5 Quitting the parameter selection menup.496.6 Parameter selection menup.50PARAMETERSp.50
Page 1 of the ETA Systems FC01-LQ user manual — Circuit Protection & Control
1ページ · クリックして拡大
1ページ
ETA Systems FC01-LQ — Flow Meter FC01-LQ User handbook diagram (page 1)
図 1 · Flow Meter FC01-LQ User handbook
ETA Systems FC01-LQ — Flow Meter FC01-LQ User handbook diagram (page 1)
図 2 · Flow Meter FC01-LQ User handbook

Flow Meter FC01-LQ User handbook

Circuit Protection & Control

M_FC01_LQ_0408_e

Page 2 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
2ページ · クリックして拡大
2ページ

Please follow these installation and adjustment instructions carefully. Failure to comply with these instructions or misuse of this equipment will void your warranty coverage.

Equipment installation, connection and adjustment by qualified personnel only!

The instructions cover software version 1.00.

Page 3 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
3ページ · クリックして拡大
3ページ
Page 3 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
Page 4 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
4ページ · クリックして拡大
4ページ
Page 4 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
Page 5 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
5ページ · クリックして拡大
5ページ
Page 5 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
Page 6 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
6ページ · クリックして拡大
6ページ
Page 6 of the ETA Systems FC01-LQ user manual — Flow Meter FC01-LQ User handbook
Page 7 of the ETA Systems FC01-LQ user manual — 1 Description
7ページ · クリックして拡大
7ページ

The Flow Meter FC01-LQ may be used with calorimetric monitoring heads type CSx (CSP, CSF).

The transistor outputs enable the user to additionally process fault, status and volume pulse indications in the control system.

As relay outputs or transistor outputs the digital signals enable the user to integrate the FC01-LQ into a control and monitoring system.

These quantities are made available to the user as analogue electrical signals, physically isolated, as current or voltage output and may be monitored by means of a limit monitor.

The Flow Meter FC01-LQ is designed to detect flow speed, flow volume and medium temperature of water (other liquids upon request).

1 Description

Description 1

Page 8 of the ETA Systems FC01-LQ user manual — 1.1 Calorimetric monitoring procedure
8ページ · クリックして拡大
8ページ
ETA Systems FC01-LQ — 1.1 Calorimetric monitoring procedure diagram (page 8)
図 3 · 1.1 Calorimetric monitoring procedure

The flow rate is determined by mass flow.

Major benefits of this method are:

Two temperature-sensitive resistors (sensor elements RS and RM) are immersed in the medium. Sensor RM assumes the temperature of the medium ϑM whilst heater resistor RH heats element RS to temperature ϑS. As a function of the medium, the temperature differential Δϑ = ϑS - ϑM is preselected as a reference variable by the CTD control and is kept constant. The required heater current IH is a function of mass flow so that the control variable y of the control can be used for evaluation.

Fig. 1 is a schematic diagram of a CTD method based sensor.

The temperature difference Δϑ between the two sensors is kept constant and the mass flow is determined by measuring the calorific power.

The thermal measuring method CTD (Constant-Temperature-Difference) is based on following principle:

A body with a temperature higher than its surroundings supplies a medium flowing past that body with energy in the form of heat. The amount energy supplied is a function of temperature difference Δϑ and mass flow.

The calorimetric measuring procedure is based on the physics of heat dissipation.

1.1 Calorimetric monitoring procedure

  • Increased safety because the sensor cannot be overheated during flow standstill.
  • Medium temperature measurement, providing optimum temperature compensation.
  • Fast response, particularly to sudden flow standstill.

y m: mass flow w: reference variable ( ) x : actual value ( S- M)

xdsystem deviation y : control variable I H : heater current

2 Description

Page 9 of the ETA Systems FC01-LQ user manual — DC 19 … 32 V
9ページ · クリックして拡大
9ページ
5 Micro controller systemsignal processing and monitoring
4 Sensor interfacescalorimetric monitoring had type CSx_ (terminal XSK)
3 Keyboard and displaykeypads
2 User interfaces2.1 signal outputs 2-way or 4-way signal outputs (terminal XAH)

1 Input voltage DC supply (terminal XV)

The system comprises the following hardware functional modules:

1.2 System description

5 Controller system:

3 Keyboard/Display:

2.2 User interface 2

2.1 User interface 1

1 Input voltage:

Sensor- interface calorimetric monitoring head CSx_

2.2 analogue outputs 1 and 2 (terminal XAO)

signal processing I/O - controlling monitoring parameter memory fig. 2

calorimetric monitoring head type CSx_

keypads LC display 2 x 16 digits

relay outputs: 2 limit values transistor outputs: 2 limit values + 1 error indication + 1 busy signal or frequency output (software selected) analogue outputs current or voltage

DC 19 … 32 V

Show 8 more labels / short text on this page
Description 3 · 4 Sensor interface · liquid crystal display · Power supply DC/DC · Keyboard and Display · Micro controller system · User interface 2 · User interface 1
Page 10 of the ETA Systems FC01-LQ user manual — 1.2.1 User interfaces
10ページ · クリックして拡大
10ページ

(optional) Please see the ordering number to find out whether it is a current or voltage output.

Analogue outputsTwo-channel physical isolation , current or voltage output

(optional) Two-channel physical isolation, relay change over contact

Signal outputs1. R2 - Relays outputs (2 limit values)

1.2.1 User interfaces

This mainly applies to monitoring head selection, signal outputs (switch point setting) and analogue outputs (zero point setting and scaling).

System configuration and parameter setting are by means of the keyboard if default values need to be changed. (paras. 5 and 6)

Cables and user interface connections are shown in para. 2.3.2 and circuit diagram 2.3.2.1/ 2.3.2.2.

Connection of the monitoring heads is by means of precut cable links.

There is no electrical isolation between monitoring head and central electronic unit.

The power supply is physically isolated between power supply input and system power supply output. This also applies to the analogue and signal outputs which are physically isolated from each other as well as from the other electronics.

2. T4 - Transistor outputs (2 setpoints + 2 status outputs or 2 setpoints + 1 status output + frequency output)

The shields of the signal cables should be applied on one side only.

Shield connections are ungrounded.

20% zero elevation and FS value can be programmed. (See para. 5.11)

These FS (full scale) output quantities apply to both channels (flow and temperature) as standard.

Output quantities0/1 - 5 V FS (option V1)

Please see para. 8.4.2 for electrical connection.

Channels 3 and 4: Both channels may be assigned individually or in pairs to the physical quantities of temperature or flow. The switch on or off values of each transistor output can be set as desired.

Channel 2busy signal or frequency output
Channel 1common error signal

Four-channel physical isolation, transistor output - collector/emitter (NPN) freely connectable

Please see para. 8.4.1 for electrical connection.

The channels may be assigned in menu “CONFIGURATION”, either individually or in pairs, to the physical quantities of temperature or flow. The switch on and off values can be set as desired in menu “PARAMETERS” (yet within the measuring range) for each contact.

0/4 - 20 mA FS (option C1)

0/2 - 10 V FS (option V2)

4 Description

Page 11 of the ETA Systems FC01-LQ user manual — 1.2.1 User interfaces
11ページ · クリックして拡大
11ページ
Power supplyDC 24 V supply

Please see para. 8.2.1 for technical characteristics.

A PTC resistor provides protection from overcurrent. The element automatically resets upon removal of the disturbance or after disconnection of the supply voltage of the FC01-LQ for approx. 1 s (e.g. remove terminal XV).

Noise emission on the connection cable is limited by circuit design and filter.

Internal switch mode power supply with physical isolation of the primary and secondary side.

Description 5

Page 12 of the ETA Systems FC01-LQ user manual — 2 Installation
12ページ · クリックして拡大
12ページ

Hastelloy 2.6410 is a material with a chemical resistance generally exceeding that of stainless steel and copper-based alloys. They are particularly suitable for alkaline media (pH > 7). They should however be examined for suitability for each specific application using resistance tables and pragmatical values.

Nickel-based alloy (Hastelloy 2.4610)

Stainless steel heads must not get in contact with steel parts other than stainless steel or with chemically dissimilar metals, as this would cause electrolytic corrosion.

Stainless steels owe their resistance to rust mainly to their alloy combination with chromium, the formation of chromic oxide on the steel surface resulting in a passive state. Contamination, deposits on the surface, or foreign rust may however neutralize the passivity. Therefore care should be taken to keep the surfaces clean.

The resistance of this stainless steel should however be verified by the user, particularly when it is used in medium mixtures that may from time to time be exchanged with cleansing agents. Its chemical resistance also depends on temperature, flow rate and concentration of the medium.

The standard monitoring head material is stainless steel 1.4571/AISI 316 Ti, an austenitic, acid- resisting stainless steel that is commonly used throughout industry. Manufacturers claim it also withstands oxidizing organic and anorganic acids, and partly even reductive media.

Stainless steel 1.4571 / AISI 316 Ti

2.1 Selection of material

These are general directions for the application of calorimetric measuring heads which from application to application should be reviewed by the user in accordance with individual requirements.

2 Installation

6 Installation

Page 13 of the ETA Systems FC01-LQ user manual — 2.2 Installation of the monitoring heads
13ページ · クリックして拡大
13ページ
ETA Systems FC01-LQ — 2.2 Installation of the monitoring heads diagram (page 13)
図 4 · 2.2 Installation of the monitoring heads

Material of the area exposed to medium: stainless steel 1.4571/AISI 316 Ti, electropolished O ring viton

Installationsensor adapter TP-.. (fig. 4) ball valve BV-.. (fig. 5)
Styleinsertion-type for sensor adapter and ball valve
Applicationgeneral industry and installation

2.2.1 Insertion head CSP for sensor adapter TP-.. and ball valve BV-..

2.2 Installation of the monitoring heads

Installation 7

Page 14 of the ETA Systems FC01-LQ user manual — SW
14ページ · クリックして拡大
14ページ

To ensure the safety label is clearly visible, it should be affixed on to or close to the measuring point.

2.2.1.2 Mounting instructions (monitoring head CSP-.. with sensor adapter TP-.. )

SW

Material of the area exposed to medium:

The sensor adapter TP-.. is available in 6 pipe diameters from 1/2" to 2". Sensor adapter is available as screw-in (fig. 4 top) or welding type (fig. 4 bottom).

2.2.1.1 Sensoradapter TP-..

  • Screw the sensor adapter into the pipe, using hemp or teflon tape for sealing.
  • Insert the monitoring head with the O ring into the sensor adapter and tighten the union nut (observe correct retention).
ETA Systems FC01-LQ — SW diagram (page 14)
図 5

Sealing of the monitoring head in the sensor adapter is by means of the O ring provided (fig. 3).

Correct immersion depth of the monitoring head is ensured by the stop provided.

The retention pin ensures correct alignment of the monitoring head after the union nut has been tightened.

The monitoring head should only be installed or removed when the pipes are unpressurised.

- stainless steel 1.4571/AISI 316 Ti

DN Type TP-01 … TP-02 … TP-03 … TP-04 … TP-05 … TP-06 …

8 Installation · - brass or · retention pin

Page 15 of the ETA Systems FC01-LQ user manual — A
15ページ · クリックして拡大
15ページ

The monitoring head may also be replaced in pressurised pipe systems on duty.

The ball valve of Postberg GmbH at D-63485 Bruchköbel is available in 4 nominal diameters from 1" to 2". The ball valve ensures the sensors are fully immersed in the medium.

❑Please see the applicable drawing of Postberg GmbH for mounting instructions and dimensions, if necessary.

ETA Systems FC01-LQ — SW diagram (page 15)
図 6

A

Ball valve BV-… · Installation 9

Page 16 of the ETA Systems FC01-LQ user manual — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth
16ページ · クリックして拡大
16ページ
ETA Systems FC01-LQ — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth diagram (page 16)
図 7 · 2.2.2 Monitoring head CSF-11 with adjustable immersi…

Material of the area exposed to medium:

Style: smooth shank, 18 mm dia., immersion depth adjustable within the PG16 cable gland (accessory) or mounting in the stainless steel cutting ring gland

Mediumliquids
Applicationpipelines with nominal pipe diameters > DN50

2.2.2 Monitoring head CSF-11 with adjustable immersion depth

- threaded installation bush stainless steel 316 (cutting ring) (see fig. 7)

- cable gland PG16 nickel-plated brass (see fig. 7)

Accessories:

- M7 sensor stainless steel 1.4571 /AISI 316 Ti, shank aluminium

- M1 sensor and shank stainless steel 1.4571 /AISI 316 Ti (sensor)

monitoring head should be aligned in direction of flow (see arrow)

10 Installation

Page 17 of the ETA Systems FC01-LQ user manual — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth
17ページ · クリックして拡大
17ページ
ETA Systems FC01-LQ — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth diagram (page 17)
図 8 · 2.2.2 Monitoring head CSF-11 with adjustable immersi…

2.2.2.1 Mounting instructions for monitoring head CSF-11

  • For horizontal pipelines the monitoring head should be mounted on the underside of the line (suspended).
  • In the case of vertical pipelines the monitoring head should be installed where the flow is rising, if possible.

The shoulder of the sensor (7 mm from the tip) must be at the position 1/8 from dia. i.

The two sensors (M) should be screwed into the pipeline far enough to ensure that they are aligned side by side directly across the direction of flow. The sensors are correctly positioned when the wrench flats (S) are aligned parallel with the pipeline.

monitoring head should be aligned in direction of flow (see arrow)

Installation 11

Page 18 of the ETA Systems FC01-LQ user manual — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth
18ページ · クリックして拡大
18ページ
ETA Systems FC01-LQ — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth diagram (page 18)
図 9 · 2.2.2 Monitoring head CSF-11 with adjustable immersi…
  • Fit monitoring head with locking set as follows (fig. 9):
  • When installing the monitoring head, please observe that the arrow corresponds to the flow direction.
  • The monitoring head should be installed only in a straight section of piping. There should be a distance of at least 20 pipe diameters before the monitoring head, and 5 pipe diameters after the monitoring head before or after bends and changes in pipe diameter, to avoid any effects of turbulence.

The locking chain must be mounted as tightly as possible.

- Put chain catch (2) into link and fasten with the tight chain.

- Fix first link of chain (1) into the clip (3) (tightening torque 10 Nm).

ETA Systems FC01-LQ — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth diagram (page 18)
図 10 · 2.2.2 Monitoring head CSF-11 with adjustable immersi…
ETA Systems FC01-LQ — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth diagram (page 18)
図 11 · 2.2.2 Monitoring head CSF-11 with adjustable immersi…

5 x D 20 x D

D = nominal pipe diameter

12 Installation

Page 19 of the ETA Systems FC01-LQ user manual — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth
19ページ · クリックして拡大
19ページ
ETA Systems FC01-LQ — 2.2.2 Monitoring head CSF-11 with adjustable immersion depth diagram (page 19)
図 12 · 2.2.2 Monitoring head CSF-11 with adjustable immersi…

1 chain 4 x 32 DIN 5685 (approx. 1 m) 2 catch for chain NG 5 3 clip with screws and nuts DN15 to DIN 11850 (tightening torque 10 Nm)

Cable Do + Ka type xx (depends on FC01-LQ version and sensor-type used - see figs. 15 and 16)

cable Kabel-Union LifYCY 4 x 2 x 0,2 mm2

Installation 13 · Locking set 01

Page 20 of the ETA Systems FC01-LQ user manual — 2.3.1 Mechanical installation
20ページ · クリックして拡大
20ページ
ETA Systems FC01-LQ — 2.3.1 Mechanical installation diagram (page 20)
図 13 · 2.3.1 Mechanical installation

2.3.1.1 Rail-mounted version FC01-LQ-U1...

2.3.1 Mechanical installation

2.3 Installation of electronic control unit FC01-LQ

  • Removal is by releasing the spring catch.
  • For thermal reasons, the modules should be spaced by at least 10 mm.
  • The electronic housing is mounted on a symmetric rail to EN 50022.

14 Installation

Page 21 of the ETA Systems FC01-LQ user manual — 2.3.1 Mechanical installation
21ページ · クリックして拡大
21ページ
ETA Systems FC01-LQ — 2.3.1 Mechanical installation diagram (page 21)
図 14 · 2.3.1 Mechanical installation

2.3.1.2 Surface mounted version FC01-LQ-FH-U1...

  • Replace the cover and tighten the retaining screws.
  • Install the housing in place using the 4 self-taping screws M4 (see fig. 12).
  • Remove the cover of the housing.

Installation 15

Page 22 of the ETA Systems FC01-LQ user manual — 2.3.1 Mechanical installation
22ページ · クリックして拡大
22ページ
ETA Systems FC01-LQ — 2.3.1 Mechanical installation diagram (page 22)
図 15 · 2.3.1 Mechanical installation

2.3.1.3 Front panel mounted housing FC01-LQ-ST-U1...

  • Insert housing in the front of mounting hole and fix with 4 screws (see fig. 13) from the rear.

16 Installation

Page 23 of the ETA Systems FC01-LQ user manual — 2.3.2 Electrical connection
23ページ · クリックして拡大
23ページ
ETA Systems FC01-LQ — 2.3.2 Electrical connection diagram (page 23)
図 16 · 2.3.2 Electrical connection

XTF - Keyboard release

Connection by 3 pole connector; max. 1.5 mm2; 3 x 0.75 mm2 cable recommended

XV - Power supply

Valid for all plug-in screw terminal strips:

2.3.2 Electrical connection

Bridge 2-3 inserted = keyboard blocked

Connection by 3 pole connector; factory-wired

Pin No. Signal name Function

Contact materialpre-tinned tin bronze
Clamping screwM2 (nickel-plated brass)
Stripping length6.5 mm
Cable size0.14 mm 2 to 1.5 mm 2 , single or stranded conductor

3 -UV negative pole of supply voltage

2 +UV positive pole of supply voltage

1 SGND general reference ground/shield ground

XSK - calorimetric monitoring head XTF - keyboard release

1 2 3 4 5 6 7 8 9 10 1 2 3

Installation 17

Page 24 of the ETA Systems FC01-LQ user manual — 2.3.2 Electrical connection
24ページ · クリックして拡大
24ページ

Connection by 8 pole connector; max. 1.5 mm 2; LifYCY 4 x 2 x 0.2 mm2 cable recommended

XAH - Limit switch signal outputs - transistor outputs NPN, freely connectable as emitter (-) and collector (+) have been brought out separately.

Connection by 8 pole connector; max. 1.5 mm2; LiYCY 3 x 0.38 mm2 cable recommended for each signal output

XAH - Limit switch signal outputs - relay outputs - change over contacts

* Apply shield on one side only.

Pin selection for analogue outputs (option: V1, V2, C1)

Connection by 8 pole connector; max. 1.5 mm2; LiYCY 2 x 0.25 mm2 cable recommended for each analogue output

XAO - - Analogue outputs

Pin No. Signal name Function

Pin No. Signal name Function

Pin No. Signal name Function

4 /BUSY/PULSE C availability signal or frequency output - collector terminal

3 /BUSY/PULSE E availability signal or frequency output - emitter terminal

2 /ERROR C summarized error indication - collector terminal

1 /ERROR E summarized error indication - emitter terminal

8 /LIM2 inverted signal output 2 (N/C)

7 LIM2COM common change over input 2

6 LIM2 non-inverted signal output 2 (N/O)

5 SGNDL2 shield ground 2

4 /LIM1 inverted signal output 1 (N/C)

3 LIM1COM common change over input 1

2 LIM1 non-inverted signal output 1 (N/O)

1 SGNDL1 shield ground 1

7 ANA2GND reference potential for analogue output 2

6 ANAO2 analogue output 2 - temperature

5 SGNDA2 shield ground for analogue output 2 (ungrounded) *

4 SGNDA1 shield ground for analogue output 1 (ungrounded) *

3 ANA1GND reference potential for analogue output 1

2 ANAO1 analogue output 1 - flow

18 Installation · 8 NC none · 1 NC none

Page 25 of the ETA Systems FC01-LQ user manual — 2.3.2 Electrical connection
25ページ · クリックして拡大
25ページ

Only for connection of cable shield (not released for user)

(XAS - secondary current supply)

Pre-sized connecting cable Do+Ka type 15 or Do+Ka type 18 with plug-in screw terminal strip (see 2.2.3).

XSK - Connection of calorimetric monitoring head type CS_x

Pin No. Signal name Function

3 SGND /shield ground

8 LIM1 C limit value 1 - collector terminal

7 LIM1 E limit value 1 - emitter terminal

6 LIM2 C limit value 2 - collector terminal

5 LIM2 E limit value 2 - emitter terminal

Installation 19

Page 26 of the ETA Systems FC01-LQ user manual — }
26ページ · クリックして拡大
26ページ

2.3.2.1 Circuit diagram FC01-LQ

Version24 V, relay outputs

* recommended ** SGNDA1 SGNDA2 Apply shield one side only.

ETA Systems FC01-LQ — 2.3.2 Electrical connection diagram (page 26)
図 17 · 2.3.2 Electrical connection

1 2 3 4 5 6 7 8 9 10 1 2 3

recommended3x0.75 mm 2

signal outputs analogue outputs ** C1, V1, V2

}

LifYCY 4x 2x0.2 mm2

}

20 Installation · power supply · calorimetric monitoring head

Page 27 of the ETA Systems FC01-LQ user manual — }
27ページ · クリックして拡大
27ページ

2.3.2.2 Circuit diagram FC01-LQ

Version24 V, transistor outputs (NPN)

* recommended ** SGNDA1 SGNDA2 Apply shield one side only.

E/ - emitter terminal C/+ collector terminal

}

ETA Systems FC01-LQ — } diagram (page 27)
図 18

1 2 3 4 5 6 7 8 9 10 1 2 3

0 V recommended3x0.75 mm 2

signal outputs analogue outputs ** C1, V1, V2

}

Installation 21 · power supply · calorimetric monitoring head

Page 28 of the ETA Systems FC01-LQ user manual — 2.3.2 Electrical connection
28ページ · クリックして拡大
28ページ
ETA Systems FC01-LQ — 2.3.2 Electrical connection diagram (page 28)
図 19 · 2.3.2 Electrical connection

Typical circuit (example 1)

If the frequency output of the FC01-LQ is connected to an electronic counter, computer or PLC, the load current should not exceed 10 mA so as to ensure low level is 0.8 V. The max. admissible voltage level of 48 V is irrelevant in this connection.

Electronic signal processing (fig. 17)

The shield cables can be connected to connector XAS, pin 3.

Select cable size ≤1.5 mm 2 to make the connections.

The pulse length is 50 ms (±1%) continuously.

Signal ground shall be connected to pin 3 (BUSY E/-) and the driving load to pin 4 (BUSY C/+).

The quantity-dependent pulse may be selected in the menu item “DISPLAY SELECT”. A square pulse signal is available for driving a counter or a primary control at the plug XAH /BUSY E/- and /BUSY C/+ (pins 3 and 4) (see fig. 16 - circuit diagram FC01-LQ - transistor outputs).

2.3.2.3 Electrical connection - frequency output (version FC01-LQ-U1T4)

22 Installation

Page 29 of the ETA Systems FC01-LQ user manual — 2.3.2 Electrical connection
29ページ · クリックして拡大
29ページ
ETA Systems FC01-LQ — 2.3.2 Electrical connection diagram (page 29)
図 20 · 2.3.2 Electrical connection
ETA Systems FC01-LQ — 2.3.2 Electrical connection diagram (page 29)
図 21 · 2.3.2 Electrical connection

Typical circuit (example 2)

If a separate relief network is preferred to the integral network, care should be taken when processing the max. frequency of 10 Hz to ensure the energy stored in the operating coil has dissipated by the time the counter output is reset. The time to do this should be below 40 ms, making due consideration to switching times and pulse variations.

It should therefore be ensured that the counter can be increased by one during the time available.

The counter should be able of processing a counting frequency of ≥10 Hz, as the puls e length is 50 ms (±0.1%) constinuously.

The FC01-LQ driver output comprises an integral safety circuit which when isolating the counter operating coil will limit overvoltages caused by inductance and convert the energy stored.

Electromechanical pulse counter (fig. 18)

❑ As there will be a reset pulse available at the output in the moment the supply voltage of the FC01-LQ is applied, make sure that the counter is switched on delayed or set to zero after it has been switched on

tON - switch-on time tL - time constent of switch-off current

Installation 23

Page 30 of the ETA Systems FC01-LQ user manual — PARAMETER SELECTION (MENU)
30ページ · クリックして拡大
30ページ
ETA Systems FC01-LQ — 3 Operating system diagram (page 30)
図 22 · 3 Operating system
ETA Systems FC01-LQ — 3 Operating system diagram (page 30)
図 23 · 3 Operating system

It is also required for setting the unit to simultaneously press UP and DOWN = .

Setting and configuration is by means of three front keypads MODE, UP and DOWN.

All functions are distributed on the three following menu levels:

Clear menu-driven control, via keyboard and display, enables easy definition of parameters and configuration. This provides high system flexibility, making the FC01-LQ the optimum solution for a wide variety of measuring, monitoring and display tasks.

3 Operating system

The FC01-LQ can only be set or operated when connector XTF (keyboard release) is removed!

PARAMETER SELECTION (MENU)

CONFIGURATION LEVEL (MENU)

MAIN LEVEL (MENU)

24 Operating system

Page 31 of the ETA Systems FC01-LQ user manual — 3 Operating system
31ページ · クリックして拡大
31ページ

reset by simultaneously pressing UP and DOWN = .

Selected data such as MIN, MAX values, totalized quantity or LAST ERROR can be deleted or

Afterwards the data are still available even after switching the FC01-LQ repeatedly on and off.

A permanent transfer of settings and values is only effected when quitting the menu, after a plausibility check of all entries.

Pressing MODE transfers the set value or the selected menu option to a volatile memory.

The longer UP or DOWN are pressed, the faster the increase or reduction.

Each time UP or DOWN are pressed, the value indicated will be increased and reduced respectively, by one numeral skip.

menu option, the value indicated can be changed by pressing UP or DOWN.

Some menu options require numerical values to be entered. After selecting the appropriate

Simultaneously pressing UP and DOWN = calls the selected menu option, or causes skipping to the selected submenu.

Calling a menu option

Pressing MODE after the last menu option will cause skipping to the first option of the menu.

The next menu option is selected by pressing MODE (forward paging).

After configuration and parameter selection re-connect plug XTF (keyboard release) to protect the system against unauthorised access!

▲ + ▼ ▼ ▲

▲ + ▼ ▼ ▲

Show 8 more labels / short text on this page
Deleting data · Transfer of entries · Entry of numerals · Menu paging · Operating system 25 · ▼ ▲ · ▼ ▲ · ▼ ▲
Page 32 of the ETA Systems FC01-LQ user manual — TOTALISATOR
32ページ · クリックして拡大
32ページ

The following operating data may be retrieved in the main menu during the measuring cycle: (see para. 4.2.1).

Analogue output, limit switches etc. can be set beyond the measuring range. When per cent display is selected, the defined measuring range will correspond to 0 ... 100% (>100% when the defined range is exceeded).

This feature will not affect the accuracy specified for the measuring ranges defined whilst no accuracy information can be given for conditions where the measuring ranges are exceeded.

Theoretically established measuring values will be used when the measuring range of calorimetric monitoring heads is exceeded (0.05 … 3 m/s for water). The FM-LQ 1 can thus be operated beyond the measuring values defined (up to 4 m/s for water).

Over limits of the measuring range

may be deleted simultaneously operating the UP and DOWN = switches without affecting the measuring operations.

All options of the main menu may be addressed without affecting the measuring and monitoring function.

Upon completion of the heating up period and availability of the first measured value, the display will change to measuring cycle, and the user interfaces such as analogue outputs or limit switches will be up-dated.

4.2 Measuring cycle

If during the test no error was found, the display will indicate HEATING UP. The FC01-LQ will then be in the heating up period required for the measuring procedure.

During this period, the integral controller will conduct test routines (see para. 7.1, Test and diagnosis).

Upon power application POWER ON TEST will be shown on the display for approx. 1 sec., with the software version number being indicated in the second line.

4.1 Switch-on performance

4 Operation and main menu

26 Operation and main menu

❒ It is not possible to configure or select parameters of the system during the measuring operations!

TOTALISATOR

LAST ERROR

PEAK VALUE MAX

PEAK VALUE MIN

▲ + ▼ ▼ ▲

Menu option values

Page 33 of the ETA Systems FC01-LQ user manual — 5.0 m/s -13.5 ° C
33ページ · クリックして拡大
33ページ

The analogue bar has different meanings, depending on its configuration (see para. 5.7 - menu option BARGRAPH).

The lower line of the display will optionally show the switching condition of the limit switches and an analogue bar with a 10-segment resolution, or the flow volume/time unit pertinent to the indicated flow rate or the totalized flow volume (totalizer function).

Flow rate and medium temperature are indicated by the units selected in the upper line of the LC display.

4.2.1.1 Measured value(s)

4.2.1 Operating data

Operation and main menu 27

T F 37044.9 m 3 5.0 m/s -13.5 ° C

T F

F

T F

T F

T F

T F F

5.0 m/s -13.5 ° C

5.0 m/s -13.5 ° C

5.0 m/s -13.5 ° C

5.0 m/s -13.5 ° C

5.0 m/s -13.5 ° C

5.0 m/s -13.5 ° C

22206.9 l/min

37004567.9 l

1332.4 m 3 /h

370.1 l/s

P

T

P

F

T

Frequency output selected (invers representation) „switch-on condition“

Temperature, active (invers representation) „switch-on condition“

Show 9 more labels / short text on this page
Totalizer · Totalizer · Totalizer · Flow volume indication · Flow volume indication · Flow volume indication · Bar graph indication · flow rate temperature · Flow rate, passive
Page 34 of the ETA Systems FC01-LQ user manual — 4.2.1 Operating data
34ページ · クリックして拡大
34ページ

The user may purposely delete the error memory in the condition selected by simultaneously pressing UP and DOWN = . ▲ ▼ ▲ + ▼

Other than the peak value memories described above, the contents of this memory will be retained even upon power failure.

This error memory comprises the number of the last error (see section 7). It may be very helpful when commissioning the FC01-LQ.

The last main menu option to be called is the error memory.

4.2.1.3 Last error (menu optionLAST ERROR )

▲ + ▼ ▼ ▲

After switch-on or NOT-BUSY indication, the minimum and maximum values are deleted and will be continuously updated (non-return pointer principle). The peak values may be retrived in the main menu and are deleted by simultaneously pressing

They store the lowest and highest value of flow rate and medium temperature.

The FC01-LQ comprises four specific measured-values memories.

4.2.1.2 Peak values (menu optionPEAK VALUE MIN / PEAK VALUE MAX )

The following figures show the display variants under menu option “Measured value(s)” (para. 5.6 - menu option DISPLAY SELECT and 5.9 - menu option FREQUENCY OUTPUT).

Limit switches lying within the analogue bar range are also represented at the appropriate place of the analogue bar (see para. 5.7).

If F and T are shown reversed, the limit switch is in the switch-on condition.

The limit switches are identified according to their physical assignment, i.e. by F for flow rate and T for medium temperature, at the first or last place of the second line on the display.

28 Operation and main menu

UP and DOWN = .

Power failure or disconnection of the power supply will delete the contents of the four measured-values memories.

MAX VALUE flow rate

MIN VALUE flow rate

MAX VALUE medium temperature fig. 21

MIN VALUE medium temperature

Page 35 of the ETA Systems FC01-LQ user manual — T F F
35ページ · クリックして拡大
35ページ

4.2.1.4 Main menu

Operation and main menu 29

ETA Systems FC01-LQ — HEATING UP * diagram (page 35)
図 24

T F F

HEATING UP *

* not available with turbine-type sensor

See section 6 · See section 5 · power-on

Page 36 of the ETA Systems FC01-LQ user manual — 5 Configuration (menu option: CONFIGURATION )
36ページ · クリックして拡大
36ページ

Available sizes are:

The nominal diameter of the ball valve is selected in menu option BV SIZE SELECT.

Enter the following characteristics when selecting a CSP monitoring head:

Enter the following characteristics when selecting a CSF monitoring head. The setting is menu driven:

These characteristics are specified by the sensor code which together with the monitoring head type number is marked on the monitoring head housing.

To operate the FC01-LQ with a calorimetric monitoring head, it is necessary to set sensor- specific characteristics.

5.2 Monitoring head data (menu option: SENSOR CODE )

The monitoring head programme is continuously extended - please enquire.

The appropriate sensor can be selected from the following menu (for type designation see type label or CPI catalogue).

This menu option comprises all monitoring heads that can be used with the FC01-LQ (issue 07.2003).

5.1 Selection of monitoring head (menu option: SENSOR SELECT )

The CONFIGURATION menu serves to adjust the FC01-LQ to the application within the entire system. During system configuration, measuring operations are not possible (see appendix 1). Configuration possibilities are:

5 Configuration (menu option: CONFIGURATION )

  • DN25 (1 in), DN32 (1 1/4 in), DN40 (1 1/2 in) und DN50 (2 in).
2. SENSOR CODE value characteristics range: 01 … 99
1. SENSOR CODE C value characteristics range: 700 … 1300
2. SENSOR CODE T value characteristics range: 01 … 99
1. SENSOR CODE C value characteristics range: 700 … 1300

Take care to repeat these settings after replacing the monitoring head or electronic control unit FC01-LQ, as the accuracy of measurements is determined by the sensor code.

Menu option SENSOR SELECT may influence data in the parameter selection menu (see para. 5.12, Quitting the configuration menu).

  • TYPE S-No. xxx (custom designed monitoring head for special liquids)
  • TYPE CSF11WM1 (monitoring head for water with adjustable immersion depth)
  • TYPE CSP11WM1BV (insertion head for water with ball valve BV-..)
  • TYPE CSP11WM1 (insertion head for water with sensor adapter TP-..)

30 Configuration

Page 37 of the ETA Systems FC01-LQ user manual — 5.3 Limit switch combinations (menu option: LIMIT SWITCHES )
37ページ · クリックして拡大
37ページ

Mode of operation, limit value and hysteresis of the limit switches are set in menu PARAMETER SELECTION.

The following combinations are available:

The FC01-LQ comprises two limit switches (LS1 and LS2) which are assigned to the physical quantity/quantities to be monitored in submenu LIMIT SWITCHES.

5.3 Limit switch combinations (menu option: LIMIT SWITCHES )

Available sizes are:

The internal diameter of the sensor adapter is selcted in menu option TYPE SELECT.

Enter the following characteristics when selecting a CSP monitoring head with sensor adapter TP-..:

  • LS1 T and LS2 F
  • LS1 F and LS2 T
  • LS1 T and LS2 T
  • LS1 F and LS2 F
  • TP03 (1 in)
  • TP06 (2 in)
  • TP02 (3/4 in)
  • TP05 (1 1/2 in)
  • TP01 (1/2 in)
  • TP04 (1 1/4 in)
2. SENSOR CODE T value characteristics range: 01 … 99
1. SENSOR CODE C value characteristics range: 700 … 1300

Menu option LIMIT SWITCHES may influence data in the parameter selection menu (see para. 5.12, Quitting the configuration menu).

limit switch 2 flow rate

limit switch 1 medium temperature

limit switch 2 medium temperature

limit switch 1 flow rate

limit switch 2 medium temperature

limit switch 1 medium temperature

limit switch 2 flow rate

limit switch 1 flow rate

Configuration 31

Page 38 of the ETA Systems FC01-LQ user manual — 5.4 Flow rate unit (menu option: FLOW UNIT )
38ページ · クリックして拡大
38ページ

In both cases the totalizer will restart from zero.

The content of the totalizer is deleted by simultaneously pressing UP and DOWN =

When the display changes from m3 to litre or gallons, or from litre or gallons to m3, the value already counted will be converted.

Where totalizer function has been selected, the totalizer will start at zero counting in the unit selected (litre, m3 or gallons).

When the first line of the LC display in the main menu indicates the flow rate in the unit selected as well as the medium temperature in °C, °F or K, it is possible to select the second line from the following menu options (see para. 5.14).

The FC01-LQ enables the user to define the 2nd line of the display 2 in certain points.

5.6 Display (menu option: DISPLAY SELECT )

All other entries relating to the medium temperature (limit value, analogue output, etc.) refer to the unit selected there. When the temperature unit is changed, all configuration and parameter data relating to medium temperature will automatically be converted.

Options are:

This submenu is used to select the medium temperature unit (1st line top left).

5.5 Medium temperature unit (menu option: TEMP. UNIT )

When the flow rate unit is changed, all configuration and parameter data relating flow rate will automatically be converted!

Standard percent is displayed when BLANK (no unit) is selected.

Any further entries relating to flow rate (e.g. limit value, analog ue output etc.), refer to that unit.

This menu option is used to set the desired flow rate unit:

At these point (1st line top left) the requested unit for the flow velocity will be set.

5.4 Flow rate unit (menu option: FLOW UNIT )

  • GALLONS°/MINUTE
  • METRE 3 / HOUR [m 3 /h]
  • GALLONS° [° = US-GALLONS]
  • LITRE/MINUTE [l/min]
  • METRE 3 [m 3 ]
  • LITRE/SECOND [l/s]
  • LITRE [l]
  • BARGRAPH Totalizer :
  • KELVIN [K]
  • GRAD FAHRENHEIT [°F]
  • GRAD CELSIUS [°C]
  • PERCENT [%]
  • BLANK [no unit]
  • METRE/SEC [m/s]
  • FEET/SEC [FPS]

, or when the max. display value (99999999.9 l, m3 or gallons) is reached.

▲ + ▼ · 32 Configuration · ▲ ▼

Page 39 of the ETA Systems FC01-LQ user manual — LS2 ON
39ページ · クリックして拡大
39ページ
ExampleLimit switch assignment: LS1 F and LS2 T

For representation details see para. 4.2.1 (Operating data).

The representation of the limit switches in the bar graph depends on the switch-on value of the limit switch.

The bar graph also comprises the representation of the limit switch(es) as far as they can be indicated in the bar range selected.

When entering the initial or final value, the user should observe reasonable resolution!

Independent of its assignment, the bar graph has a constant resolution of 10 segments.

This menu option allows the user to set the bar graph as desired. The following settings should be made:

5.7 Bar graph (menu option: BARGRAPH )

Skipping to the submenu BARGRAPH or PIPE SIZE is effected depending on the selected menu option.

  • FS = (final value of the bar graph)
  • ZERO = (initial value of the bar graph)
  • FLOW / TEMP = (bar graph assignment: flow rate/medium temperature)

resulting in the analogue bar graph display shown below:

Instantaneuous temperature value25 °C

Final value - analogue bar graph: 30 °C

Initial value - analogue bar graph: 20 °C

Analogue bar graph assignmentmedium temperature
Switch-off value LS229 °C
Switch-on value LS223 °C

The content of the totalizer is deleted in the event of power failure or disconnection of the power supply!

20 °C 21 °C 25 °C 29 °C 30 °C

LS2 ON

T T

23 ° C

Configuration 33

Page 40 of the ETA Systems FC01-LQ user manual — 5.8 Pipe diameter (menu option: PIPE SIZE )
40ページ · クリックして拡大
40ページ

This means that only the content of the totalizer is deleted.

As the frequency output refers to the content of the totalizer, although its operation is not dependent on the content of the totalizer, a totalized quantity that is smaller than that set per pulse will not be lost.

= in the main menu.

The content of the totalizer may be deleted by simultaneously pressing UP and DOWN

Behaviour of the frequency output when the content of the totalizer is deleted

When the measurement is stopped (as caused by priority II error and calling the configuration or parameter selection menus), the pulses for the quantity already counted will be available. Thereafter the output of pulses will be stopped, with the frequency output becoming high- resistive until the measurement is restarted.

Behaviour of the frequency output when the measurement is stopped

Error No. 20, 30, 60, 40, 41.

If a combination of priority III errors occurs simultaneously, they are indicated or stored in the error memory observing the following sequence:

The max. frequency being exceeded will not cause the measurement to stop but will rather cause the error output to signal error 60 on the display. This error is included in priority group III.

Behaviour of the frequency output when the max. frequency is exceeded

Potential setting range of the frequency output: 1 pulse per 0.1 … 999.9 [litre], [m3], [gallons]

When the quantity-proportional pulses are assigned, the frequency of the frequency output must no exceed 10 Hz. The limits that can be displayed are determined by the flow velocity range and the pipe diameter.

Example1 pulse / 10.0 [litre]

The proportional quantity pulses have been determined as follow:

The totalizer function of the FC01-LQ has been expanded by the output of proportional quantity pulses. The function can only be displayed by version FC01-LQ-U1T4 (transistor outputs).

5.9 Frequency output (menu option: FREQUENCY OUTPUT )

This is provided by selecting the pipe diameter in menu option PIPE SIZE comprising pipe diameters from 10.0 … 999.9 mm.

If flow volume/time unit display has been selected, it is necessary to indicate the pipe diameter to calculate mass flow.

5.8 Pipe diameter (menu option: PIPE SIZE )

The frequency output will supply 1 pulse per 10 litres (totalized quantity).

1 pulse / quantity (totalizer unit selected)

34 Configuration · ▲ + ▼ · ▼ ▲

Page 41 of the ETA Systems FC01-LQ user manual — 5.10 Analogue output - flow rate (menu option: ANA OUT FLOW )
41ページ · クリックして拡大
41ページ

Priority of entry errors in the CONFIGURATION menu:

“CONFIG. OK!” is indicated when the data are found correct. The menu may than be quitted by pressing MODE. Errors found during the plausibility check are indicated in the following sequence of priority.

To quit the configuration menu, the controller will check the data entered for plausibility.

Upon configuration of the analogue outputs, the menu may be quitted or re-set to the start (SENSOR SELECT).

5.12 Quitting the configuration menu

When entering the initial or final value, the user should observe a reasonable resolution!

Options are:

In conformance with the configuration “Flow rate analogue output” it is possible to adjust the medium temperature analogue output to the requirements of the entire system.

5.11 Analogue output - medium temperature (menu option: ANA OUT TEMP. )

With a flow volume/time unit selected in menu DISPLAY SELECT and when setting the initial and end values, the pertinent flow volumes will also be indicated.

When entering the initial or final value, the user should observe a reasonable resolution!

This menu option allows adjustment of the flow rate analogue output specifically to the requirement of the entire system.

5.10 Analogue output - flow rate (menu option: ANA OUT FLOW )

  • ERR. A-OUT TEMP. ZERO ≥ FS (initial value ≥ final value with temperature analogue output)
  • ERR. A-OUT TEMP. OUT OF RANGE (temperature analogue output outside measuring range)
  • ERR. A-OUT FLOW ZERO ≥ FS (initial value ≥ final value with flow analogue output)
  • ERR. A-OUT FLOW OUT OF RANGE (flow analogue output outside measuring range)
  • FS = ( final value 100% corresponds to a medium temperature of … [°C] [°F] [K])
  • ZERO = ( initial value 0(20) % corresponds to a medium temperature of … [°C] [°F] [K])
  • OFFSET = 0%/20% FS (0/4 … 20 mA, 0/1 … 5 V, 0/2 … 10 V)
  • FS = ( final value 100% corresponds to a flow rate of _ [m/s] [%] [FPS])
  • ZERO = ( initial value 0(20) % corresponds to a flow rate of _ [m/s] [%] [FPS])
  • OFFSET = 0%/20% FS (0/4 … 20 mA, 0/1 … 5 V, 0/2 … 10 V)

Options are:

Configuration 35

Page 42 of the ETA Systems FC01-LQ user manual — 5.12 Quitting the configuration menu
42ページ · クリックして拡大
42ページ

An overview of the configuration menu and a summary of the measuring ranges and menus available for the sensor type selected are shown on the following pages.

Reason: Changing the physical assignment of limit switch 2 will adjust its switch-on and switch-off values to the new assignment (flow rate).

Effects on parameter dataLS2 ON = 0.00

Example: Changing the limit switch assignment from LS1 →F / LS2 →T to LS1 →F / LS2 →F

of the configuration menu by pressing UP or DOWN and select the menu option with the incorrect entry for correction.

The menu can only be quitted after correction of the error(s). To do this, return to the beginning

  • ERR. BARGRAPH ZERO ≥ FS (bar initial value ≥ bar final value)
  • ERR. BARGRAPH OUT OF RANGE (bar value outside measuring range)

In this event it is imperative to branch into parameter selection menu to set the data in conformance with the desired application.

If during the configuration data are affected which are accessible in the parameter selection menu, (which may be the case for the options Sensor Selection, Medium Selection and Limit Switch Assignment), the option “PARAMETERS” in the main menu will be flashing.

LS2 OFF = end of measuring range (depending on the medium selected)

36 Configuration · ▼ ▲

Page 43 of the ETA Systems FC01-LQ user manual — CONFIGURATION LIMIT SWITCHES
43ページ · クリックして拡大
43ページ

5.13 Configuration menu

ETA Systems FC01-LQ — 5.13 Configuration menu diagram (page 43)
図 25 · 5.13 Configuration menu

END OF CONFIG.? M→yes ▲ or ▼ → no

CONFIGURATION LIMIT SWITCHES

CONFIGURATION SENSOR SELECT

CONFIGURATION

G

F

E

D

B

A

See:

Show 6 more labels / short text on this page
Configuration 37 · ▲ + ▼ · ▲ + ▼ · ▲ + ▼ · submenu LS combination · submenu sensor select
Page 44 of the ETA Systems FC01-LQ user manual — SENSOR CODE
44ページ · クリックして拡大
44ページ

5.14 Configuration submenus

ETA Systems FC01-LQ — 5.14 Configuration submenus diagram (page 44)
図 26 · 5.14 Configuration submenus

SENSOR SELECT

A

TYPE S-No. xxx

SENSOR CODE

SENSOR CODE

PIPE SIZE DIAM. = xxx.x mm

Conf.

Conf.→ return to configuration menu

Show 4 more labels / short text on this page
38 Configuration · or ▲ ▼ · or ▲ ▼ · T xxx
Page 45 of the ETA Systems FC01-LQ user manual — D
45ページ · クリックして拡大
45ページ

Configuration submenus (cont’d)

Conf.

Conf.

Conf.

Conf.

submenu limit switch combination

B

LIMIT SWITCHES LS1 → T LS2 → T LIMIT SWITCHES LS1 → F LS2 → F

LIMIT SWITCHES LS1 → F LS2 → T

LIMIT SWITCHES LS1 → T LS2 → F

or ▲ ▼ or ▲ ▼

Conf.

Conf.

Conf.

Conf.

FLOW UNIT METRE/SEC (m/s)

FLOW UNIT PERCENT (%)

FLOW UNIT FEET/SEC. (FPS)

FLOW UNIT BLANK (no unit)

or ▲ ▼ or ▲ ▼ or ▲ ▼ or ▲ ▼

Conf.

Conf.

Conf.

D

TEMP. UNIT CELSIUS ( ° C)

TEMP. UNIT FAHRENHEIT ( ° F)

TEMP. UNIT

KELVIN (K)

or ▲ ▼ or ▲ ▼ or ▲ ▼

Conf.→ return to configuration menu

Show 6 more labels / short text on this page
Configuration 39 · or ▲ ▼ · or ▲ ▼ · submenu · flow unit · submenu temperature unit
Page 46 of the ETA Systems FC01-LQ user manual — ° US-GALLONS
46ページ · クリックして拡大
46ページ

Configuration submenus (cont’d)

ETA Systems FC01-LQ — Configuration submenus (cont’d) diagram (page 46)
図 27 · Configuration submenus (cont’d)

E

F

BARGRAPH TYP=FLOW or TEMP

PIPE SIZE

BARGRAPH ZERO = 10.0 m/s

G

52.5 mm

ANA OUT FLOW OFFSET = (FS) ANA OUT TEMP. OFFSET = (FS)

BARGRAPH FS = 20.0 m/s

FREQUENCY OUTPUT?

Conf.

M → yes other → no

Conf.

FREQUENCY OUTPUT

PULSE/ (unit)

Conf.

Conf.

° US-GALLONS

Conf.→ return to configuration menu

Show 5 more labels / short text on this page
40 Configuration · yes · Totalizer function? · no · or ▲ ▼
Page 47 of the ETA Systems FC01-LQ user manual — OFF
47ページ · クリックして拡大
47ページ
Example 1Switch-on value lower than switch-off value

A specific definition of the operation (closed-current or open-circuit principle) may be dropped by separately entering the switch-on and switch-off value of the limit switch, because the definition is deducted from the switch-on and switch-off value..

The hysteresis is determined by entering different switch-on and switch-off values. Its magnitude should be reasonably adjusted to current operating conditions.

Limit switch up-date is by measuring rate, independent of the set measuring time.

Depending on the configuration (see configuration menu) limit value 1 may be set either for flow rate or medium temperature. The limit value may be set over the entire measuring range and is always related to the display value.

6.2 Limit switch 1 - switch-on value (menu option: LS1 ON = ) Limit switch 1 - switch-off value (menu option: LS1 OFF = )

The set measuring time does not influence the measuring rate and display up-date.

The effect of the measuring time may be compared to that of a filter; it is used to determine the average of the last measured values (measuring time 100 ms) after each measurement.

The measuring time may be set between 1 and 30 sec., referring both to flow rate and medium temperature.

6.1 Measuring time (menu option: MEAS. TIME )

The following parameters may be set in the Parameter selection menu:

During parameter setting, measuring operations are not possible (see Appendix 1).

After configuration of the FC01-LQ in conformance with its application (configuration menu), it is possible to set parameters (e.g. limit values).

6 Parameter selection (menu option: PARAMETERS )

OFF

ON

Show 8 more labels / short text on this page
Parameter selection 41 · switch-on value · hysteresis · switch-off value · ∞ · switching condition OFF · switching condition ON · measured value (flow/temperature)
Page 48 of the ETA Systems FC01-LQ user manual — OFF
48ページ · クリックして拡大
48ページ

For example, the scaling factor may be used to indicate the average flow rate in the pipeline rather than that available at the sensor.

The factor, which may be set between 0.01 and 9.99, allows flow rate indication changes (increasing or reducing the measured value in the display).

The scaling factor influences flow rate indication.

6.4 Scaling factor (menu option: FLOWSCALE )

See limit switch 1!

6.3 Limit switch 2 - switch-on value (menu option: LS2 ON = ) Limit switch 2 - switch-off value (menu option: LS2 OFF = )

With limit switch 1 set for flow rate and a flow volume/time unit selected in menu DISPLAY SELECT, and when setting the switch-on and switch-off value, the pertinent flow volumes will also be indicated.

Example for ONas described in example 1 (fig. 23)
Example 2Switch-on value higher than switch-off value

FC01-LQ with transistor outputs (option T4):

FC01-LQ with relay outputs (option R2):

Example for ON:

  • LIM1E - LIM1C = switched
  • LIM1 - LIM1COM = closed

/LIM1 - LIM1COM = open

OFF

ON

Show 8 more labels / short text on this page
42 Parameter selection · switch-off value · hysteresis · switch-on value · ∞ · switching condition OFF · switching condition ON · measured value (flow/temperature)
Page 49 of the ETA Systems FC01-LQ user manual — 6.5 Quitting the parameter selection menu
49ページ · クリックして拡大
49ページ

An overview of the parameter selection menu is shown on the following page.

The menu can only be quitted after correction of the error(s). To do this, return to the beginning of the parameter selection menu by pressing UP or DOWN and select the menu option with the incorrect entry for correction.

Priority of entry errors in the PARAMETER SELECTION menu:

Errors found during the plausibility check are indicated in the following sequence of priority.

quitted by pressing MODE.

“PARAMETERS OK!” is indicated when the data are found to be correct. The menu may then be

Before the parameter selection menu can be quitted, the controller will conduct a plausibility check of the data entered.

6.5 Quitting the parameter selection menu

  • ERROR LS2 ON = OFF
  • ERROR LS1 ON = OFF
  • ERROR LS2 OUT OF RANGE
  • ERROR LS1 OUT OF RANGE

switch-on value for limit switch 2 equals switch-off value for limit switch 2

switch-on value for limit switch 1 equals switch-off value for limit switch 1

switch-on and/or switch-off value for limit switch 2 outside measuring range

switch-on and/or switch-off value for limit switch 1 outside measuring range

Parameter selection 43 · ▲ ▼

Page 50 of the ETA Systems FC01-LQ user manual — PARAMETERS
50ページ · クリックして拡大
50ページ

6.6 Parameter selection menu

ETA Systems FC01-LQ — 6.6 Parameter selection menu diagram (page 50)
図 28 · 6.6 Parameter selection menu

END OF PARAM.? M→yes ▲ or ▼ → no

PARAMETERS LS1 ON = 1.24 m/s

PARAMETERS MEAS. TIME = 3 sec

PARAMETERS

Show 4 more labels / short text on this page
44 Parameter selection · ▲ ▼ or · ▲ + ▼ · ▲ ▼ or
Page 51 of the ETA Systems FC01-LQ user manual — 7.1 Test and diagnosis
51ページ · クリックして拡大
51ページ

Other than the above priority groups, errors No. 20, 30, 60, 40 and 41 will not cause measurements to stop; the error output will indicate and the number of the error will be shown on the display.

These test routines are also continuously carried out during operation.

7.1.3 Priority group III

These test functions are continuously carried out during operation. The occurrence of errors No. 10 and 21 will cause measurements to stop, indicating the error and monitoring the source of the error. Upon rectification of the error, the system will automatically return to measuring operation.

7.1.2 Priority group II

Priority I errors cannot be rectified by the user!

If even after several trials the switch-on test cannot be conducted without error indication, the system should be returned to the supplier for rectification, indicating the error number.

The test routines may be repeated by pressing any of the switches.

Errors No. 1 to 5 do not allow system operation.

Their implementation is indicated.

Priority group I comprises the switch-on test routines (FC01-LQ self-test) which are carried out when the system is switched on.

7.1.1 Priority group I

The FC01-LQ is provided with extensive test and diagnosis functions which may be classified as follows:

7.1 Test and diagnosis

7 Errors

Errors 45

Page 52 of the ETA Systems FC01-LQ user manual — 7.2 Potential errors
52ページ · クリックして拡大
52ページ

If a combination of errors occurs simultaneously, they are indicated or stored in the error memory observing the following sequence.

In order to facilitate operation, the last error is stored in a non-volatile memory. The stored error may be retrieved and deleted in the main menu.

Independent of the priority group, all errors found are indicated with their relevant number.

7.2 Potential errors

No. 21 Medium temperature too high

No. 10 Sensor not connected; or cable Check cable or replace sensor. between FC01-LQ and sensor defective; or defective sensor

No. 5 Internal controller error Return to supplier.

No. 4 Incorrect test sum of data memory Return to supplier.

No. 3 Incorrect test sum of program Return to supplier. memory

No. 2 Incorrect test sum of parameter Return to supplier. memory

No. 1 No system parameter available Return to supplier.

Sensor selected (configuration) Correct sensor selection in differs from sensor connected configuration menu

Show 5 more labels / short text on this page
Priority group II · Priority group I · 46 Errors · Error Cause Rectification · Error Cause Rectification
Page 53 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
53ページ · クリックして拡大
53ページ

* Error No. 60 can only occur with version FC01-U1T4.

No. 41 Controller error (watchdog timer)

No. 40 Controller error (oscillator-watchdog)

No. 60 Assignment of quantity per pulse too low *

No. 30 Over limits of flow rate

No. 20 Medium temperature too low

Admissible EMC levels may have been exceeded

Admissible EMC levels may have been exceeded

Priority group III · Errors 47 · Error Cause Rectification

Page 54 of the ETA Systems FC01-LQ user manual — -U V 3
54ページ · クリックして拡大
54ページ

* DC 12 V possible if the FC01-LQ is used whithout option C1 (analogue outputs).

Power consumption may be up to 300 mA±10 % when analogue output C1 is fitted.

Admissible ripple: max. 20 % UV Rated current consumption: Ivnk = 170 mA with zero flow

Input voltage rangeU V = DC 19 V to DC 32 V (ripple incl.)
Supply voltageU VN = DC 24 V *

8.2.1.1 DC voltage supply

Pin selectionSignal name Pin XV

8.2.1 Power supply

8.2 Electrical characteristics

Degree of protectionIP20 IP65 IP65
Ambient temperature* +10 … +50 °C ** +10 … +50 °C +10 … +50 °C
Storage temperature-20 … +70 °C -20 … +70 °C -20 … +70 °C

8.1 Ambient conditions

** The max. ambient temperature of +40 °C applies to all systems fitted with current output C1.

* Only if the modules are spaced by at least 10 mm.

8 Technical data

rail-mounted surface mounted front panel mounted

Ivnk = 200 mA with max. flow (end of measuring range)

-U V 3

+U V 2

DC supply · 48 Technical data · shield 1

Page 55 of the ETA Systems FC01-LQ user manual — NC 1
55ページ · クリックして拡大
55ページ
Insulation voltageanalogue output - analogue output DC 500 V

The output is reverse polarity protected.

* Shield ungrounded - apply on one side only.

Analogue output 2 - ANA OUT TEMP. (temperature output)

Analogue output 1 - ANA OUT FLOW (flow output)

Pin selection for analogue outputs V1, V2 and C1

The analogue outputs are physically isolated from each other as well as from the electronic control unit FC01-LQ.

8.3 Analogue outputs

Insulation voltagesupply input - central electronic unit ≥ DC 500 V
Rated power consumptionP n = 4.1 W with zero flow, voltage outputs
Switch-off currentI kipp = typ. 0.75 A
Inrush currentI p = typ. 3 A (20 µs)

analogue output - central electronic unit DC 500 V

NC - not used

NC 8

reference ground 2 7

analogue output 2 - temperature 6

shield 2 * 5

shield 1 * 4

reference ground 1 3

analogue output 1 - flow rate 2

NC 1

Signal name Pin XAO

Pn = 4.8 W max. flow (end of measuring range), voltage outputs

Technical data 49

Page 56 of the ETA Systems FC01-LQ user manual — 8.3.1 Voltage output V1 - 5 V FS
56ページ · クリックして拡大
56ページ
Pin selectionSignal name Pin XAH

8.4.1 Relay outputs R2 (DC or AC)

The signal outputs are physically isolated from each other as well as from the electronic control unit FC01-LQ.

8.4 Signal outputs

Max. admissible load resistanceR l = 250 Ω
Min. admissible load resistanceR l = 0 Ω
Max. signal rippledI S = 5% FS
Signal current rangeI S = 0 mA (4 mA) to 20 mA ±2% FS

8.3.3 Current output C1 - 20 mA FS

Short circuit proofyes (XAO - between all terminals)
Max. admissible load inductanceL l = 100 nH
Max. admissible load capacityC l = 1 nF
Min. admissible load resistanceR l = 1 k Ω
Max. signal rippledU S = 5% FS
Signal voltage rangeU S = 0 V (2 V) to 10 V ±2% FS

8.3.2 Voltage output V2 - 10 V FS

Short circuit proofyes (XAO - between all terminals)
Max. admissible load inductanceL l = 100 nH
Max. admissible load capacityC l = 1 nF
Min. admissible load resistanceR l = 1 k Ω
Max. signal rippldU S = 5 % FS
Signal voltage rangeU S = 0 V (1 V) to 5 V ±2% FS

8.3.1 Voltage output V1 - 5 V FS

Limit Switch 2 / N/C 8

Limit Switch 2 / common 7

Limit Switch 2 / N/O 6

Limit Switch 2 / Schirm 5

Limit Switch 1 / N/C 4

Limit Switch 1 / common 3

Limit Switch 1 / N/O 2

Limit Switch 1 / shield 1

50 Technical data

Page 57 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
57ページ · クリックして拡大
57ページ
Short circuit protectionyes
Reverse polarity protectionyes
High level - passiveU ce < 48 V
Low level - activeU ce < 0.8 V at I C < 10 mA
Pin selectionSignal name Pin XAH Polarity

8.4.2 Transistor outputs (DC)

Insulation voltagesignal contact - central electronic unit DC 500 V
Contact life cos = 0.52.4 x 10 5 cycles
Max. admissible switching voltage100 V
Max. admissible continuous current1.25 A
Max. admissible switching current1.25 A
Max. admissible switching capacity125 VA

Inductive load - with safety circuit - AC voltage

Contact life at 1 A3 x 10 5 cycles
Max. admissible switching voltage50 V
Max. admissible continuous current1 A
Max. admissible switching current1 A
Max. admissible switching capacity50 W

max. leakage current ≤ 25 µA

Uce max = 60 V

Uce < 1 V at IC < 100 mA

Limit Switch 1 collector 8 +

Limit Switch 1 emitter 7 -

Limit Switch 2 collector 6 +

Limit Switch 2 emitter 5 -

/ BUSY / PULSE collector 4 +

/ BUSY / PULSE emitter 3 -

/ ERROR collector 2 +

/ ERROR emitter 1 -

signal contact - signal contact DC 500 V

Voltage level · Resistive load · Technical data 51

Page 58 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
58ページ · クリックして拡大
58ページ
Insulation voltagesignal contact - central electronic unit DC 500 V
Max. admissible switching voltage36 V
Max. admissible switching current1.5 A
Max. admissible switching capacity1.5 VA

Capacitive load - C < 20 µF

Max. admissible switching voltage36 V
Max. admissible switching current40 mA
Max. admissible switching capacity1.5 VA

Inductive load - L < 100 mH (DC voltage - without external safety circuit)

Max. admissible switching voltage36 V
Max. admissible switching current150 mA
Max. admissible switching capacity1.5 W

signal contact - signal contact DC 500 V

Resistive load · 52 Technical data

Page 59 of the ETA Systems FC01-LQ user manual — TP06 0.2 … 21.2 28.3
59ページ · クリックして拡大
59ページ

The flow measurement range is determined by the inner pipe diameter (see table). It can be calculated with the following equation:

Flow measurement ranges:

8.5.1.2 Monitoring head CSF

Flow rate measurement:

8.5.1.1 Monitoring head CSP- with sensor adapter type TP-..

Measuring is possible up to the flow rates indicated in the display range. However, the indi cated accuracy is no longer guaranteed. The repeatability value remains valid.

8.5.1 Flow rate measurement

8.5 Metrological data

Sensor adapter Measuring range Display range type in m3/h in m3/h

TP06 0.2 … 21.2 28.3

TP05 0.14 … 13.6 18.1

TP04 0.1 … 8.7 11.6

TP03 0.05 … 5.3 7.1

TP02 0.04 … 3.4 4.5

TP01 0.02 … 2.2 2.9

AR [m2] - inner pipe cross-section

VN [m/h] - average standard velocity

Q = VN x AR Q [m3/h] - flow quantity

Technical data 53

Page 60 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
60ページ · クリックして拡大
60ページ
ETA Systems FC01-LQ — 6.6 Parameter selection menu diagram (page 60)
図 29 · 6.6 Parameter selection menu

Failure diagram for water

Accuracy±1% MBE * at 2 m/s see failure ddiagram)
Repeatability1% MW ** (5% MBE to 100% MBE)
Response delay2.5 s
Display range0 … 4 m/s
Measuring range0.05 … 3 m/s

Setting range for internal pipe diameter: 50.0 mm … 999.9 mm

Inner pipe diamater Measuring range Display range D in mm in m3/h in m3/h

0.20 0.40 0.60 0.80 1.00 1.20 1.40 1.60 1.80 2.00 2.20 2.40 2.60 2.80 3.00 0

calibration point at 2 m/s± 1% of upper range value
Show 9 more labels / short text on this page
54 Technical data · 500 2120 2830 · 350 1040 1380 · 250 530 700 · 200 340 450 · 150 190 250 · 100 85 110 · 80 55 70 · 50 21 28
Page 61 of the ETA Systems FC01-LQ user manual — 8.5.2 Temperature measurement:
61ページ · クリックして拡大
61ページ

*** MB - measuring range

** MW - measured value

* MBE - of final value

Heating up period until full accuracy is reached: 15 min.

Temperature drift0.1%/K/MBE *

8.5.3 Electronic control unit FC01-LQ

Accuracy±1% MB ***
Measuring range-40 °C … +130 °C

8.5.2 Temperature measurement:

Technical data 55

Page 62 of the ETA Systems FC01-LQ user manual — 8.6.1 Electrical data of the terminal for calorimetric monitoring heads
62ページ · クリックして拡大
62ページ

* RTD = Resistive Temperature Device

8.6.1 Electrical data of the terminal for calorimetric monitoring heads

8.6 Sensor interface

XSK10 R(Tdiff)-HI Function: terminal for positive pole of the heated RTD * Input resistance: > 1 GΩ Dielectric strength: -17 V … +30 V DC

XSK9 R(Tdiff)-LO Function: terminal for negative pole of the heated RTD * Input resistance: > 1 GΩ Dielectric strength: -17 V … +30 V DC

XSK7 SGND Functionshield ground XSK8 Terminals for sensor cable shielding

XSK6 IS Function: output of exitation current source for RTD * operation Exitation current: 1 mA ± 1% Admissible load range: Rload = 0 … 2 k⏐ Dielectric strength: ±15 V DC

XSK5 AGND Functionanalogue ground Reference potential of exitation current source for RTD * operation

XSK4 R(Tref)-LO Function: terminal for negative RTD * pole for medium temperature measurement Input resistance: > 1 GΩ Dielectric strength: -17 V … +30 V DC

XSK3 R(Tref)-HI Function: terminal for positive RTD * pole for medium temperature measurement Input resistance: > 1 GΩ Dielectric strength: -17 V … +30 V DC

XSK2 R(HEIZ)-HI Function: terminal for positive pole of heater element Hi-potential of heater source Output voltage range (load dependent) Ua = 21 V … 24 V DC Max. current output: Imax = 100 mA Not short-circuit proof

XSK1 R(HEIZ)-LO Function: terminal for negative pole of heater element Drain output of heating current control Max. sink current: Isink = 88 mA Dielectric strength: -0.5 V … +20 V DC

56 Technical data · Terminal Mnemonics Data

Page 63 of the ETA Systems FC01-LQ user manual — 9 Accessories
63ページ · クリックして拡大
63ページ

9 Accessories

7 locking set 01 (for monitoring head CSF-11) 0Z122Z00024

6 ball valve BV

5 sensor adapter (screw-in or welding type) TP

4 calorimetric monitoring heads CSP / CSF

3 connecting cable for calorimetric monitoring head

2 front mounted housing FC01-LQ-ST

1 surface mounted housing FC01-LQ-FH

No. Accessory Ordering configuration

- Typ 18/ -60 °C … +200 °C non-halogenuous/highly flexible/paired

- Typ 15/ -10 °C … +80 °C highly flexible/paired

cable type LifYCY 4 x 2 x 0.2 mm 2 Do+Ka

Accessories 57

Page 64 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
64ページ · クリックして拡大
64ページ
Page 64 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
Page 65 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
65ページ · クリックして拡大
65ページ
Page 65 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
Page 66 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
66ページ · クリックして拡大
66ページ
Page 66 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
Page 67 of the ETA Systems FC01-LQ user manual — Y
67ページ · クリックして拡大
67ページ

Appendix 1 - Performance of the digital and analogue outputs during the operating and error modes

NO ERROR NOT BUSY and

LIMIT SWITCH 1

LIMIT SWITCH 2

FREQUENCY

ANA OUT

ANA OUT

OUTPUT

FLOW

TEMP.

ON

ON

ON

ON

MAX

MAX

OFF

OFF

OFF

OFF

MIN

MIN

Error No. 1

OFF

OFF

OFF

OFF

MIN

MIN

Error No. 2

OFF

OFF

OFF

OFF

MIN

MIN

Error No. 3

OFF

OFF

OFF

OFF

MIN

MIN

Error No. 4

OFF

OFF

OFF

OFF

MIN

MIN

Error No. 5

OFF

OFF

OFF

OFF

MIN

MIN

OFF

OFF

ON

OFF

MIN

MIN

Normal duty

ON

ON

OFF

OFF

ON

OFF

MIN

MIN

OFF

OFF

ON

OFF

MIN

MIN

Error No. 10

OFF

OFF

OFF

OFF

MIN

MIN

Error No. 20

OFF

ON

Error No. 21

OFF

OFF

OFF

OFF

MIN

MIN

Error No. 30

OFF

ON

Error No. 60 *

OFF

FA

Error No. 40

Y

ON

Error No. 41

Y

ON

X = standard performance Y = OFF pulse FA = frequeny output 10 Hz

* When frequency output has been selected.

Show 7 more labels / short text on this page
Appendix 1 · Duty-/ Error status · Start-up (reset) · Start-up test active · Heating period active · Configuration active · Parameter selection active
Page 68 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
68ページ · クリックして拡大
68ページ
Page 68 of the ETA Systems FC01-LQ user manual — 6.6 Parameter selection menu
Page 69 of the ETA Systems FC01-LQ user manual — FS
69ページ · クリックして拡大
69ページ
ETA Systems FC01-LQ — 6.6 Parameter selection menu diagram (page 69)
図 30 · 6.6 Parameter selection menu

Appendix 2 - Menu structure of the FC01-LQ (operator dialog)

PERCENT (%) METRE/SEC (m/s) FEET/SEC (FPS) BLANK (no unit)

LS1 → F / LS2 → F LS1 → T / LS2 → F LS1 → F / LS2 → T LS1 → T / LS2 → T

CELSIUS (°C) FAHRENHEIT (°F) KELVIN (K)

TYPE CSP-11 WM1 TYPE CSP-11 WM1BV TYPE CSF-11 WM1 TYPE S-No. xxx

GALLONSϒ/MINUTE (GPM) LITRE (l) METRE3 (m3) GALLONS° (G)

BARGRAPH LITRE/SEC (l/s) LITRE/MIN (l/min) m3/HOUR (m3/h)

FREQUENCY OUTPUT? M →yes other → no

FREQUENCY OUTPUT PULSE/ (unit)

ZERO

→ TYPE

return to configuration menu

nur bei gewählter Totalisatoreinheit

not available with sensor type CSP

FS

call menu or delete function

select and return to configuration menu

DN25 (1 in) DN32 (1.1/4 in) DN40 (1.1/2 in) DN50 (2 in)

TP01 (1/2 in) TP02 (3/4 in) TP03 (1 in) TP04 (1 1/4 in) TP05 (1 1/2 in) TP06 (2 in)

Show 20 more labels / short text on this page
Appendix 2 · yes Totalizer function? · CODE T xxx · CODE C xxx · ▲ ▼ or · ** · * · ▼ · ** · * · ▲ + ▼ · ▲ · Menus · action · page down · page up · CODE C xxx · CODE T xxx · CODE T xxx · CODE C xxx

ETA Systems FC01-LQについて質問する

ユーザーまたはスタッフが通常1日以内に回答します。あなたの質問が次の人の助けにもなります。

Add your email to get notified when it’s answered — no spam, ever.

0 questions
ETA Systems FC01-LQに関する質問はまだありません

最初の質問を投稿しましょう。他のユーザーや当チームが回答します。

原本

ETA Systems FC01-LQ 取扱説明書(PDF)

69 ページ·4.6 MB·酸素機器・酸素吸入器·PDF
上のページをクリックするとここに表示されます
ページ 1 / …
100%
ページを読み込み中…