Manual Supplement © 2008-2009 Fluke Corporation . All rights reserv ed. Printed in the U.S .A. Manual Title: Impulse 6000D/7000DP Users Supplement Issue: 3 Print Date: August 2007 Issue Date: 3/09 Revision/Date: 1, 1/08 Page Count: 10 This supplement contai ns information nece ssary to en sure the accuracy of the above manual.
Impulse 6000D/7000DP Users Manual Supplement 4/08 1 Change #1, 46470, 46697, 47714, 48007, 49680 Replace pages 31 through 35, with the following: Defibrillator Analyzer Specifications Energy Output Measurement Compatible Defibrillator Wa veshapes ....
Manual Supplement Impulse 6000D/700 0DP Users 2 8/08 Accuracy ............................................................. ± 0.1 ms Voltage Range ................................................................. 20 to 5000 V Accuracy ...............
Impulse 6000D/7000DP Users Manual Supplement 4/08 3 Scope Output Autorange ........................................................... 2000:1, 400:1 and 80:1: depe ndant on the range Waveform Playback Output ...........................................
Manual Supplement Impulse 6000D/700 0DP Users 4 3/09 Ventricular Fibrillation ......................................... Coarse and fine Monomorphic Ventricular Tachycardia ............... 120 to 300 (by 5) BPM Polymorphic Ventricular Tachycardia .....
Impulse 6000D/7000DP Users Manual Supplement 8/08 5 Normal Sinus waves Lead # I II III V1 V2 V3 V4 V5 V6 Ref. Amp. 70 % 100 % 30 % 24 % 48 % 100 % 120 % 112 % 80 % Lead I reference Performance waves and R wave detection Lead # I II III V1 V2 V3 V4 V5 V6 Ref.
Manual Supplement Impulse 6000D/700 0DP Users 6 4/08 Sine waves ......................................................... 0.05, 0.5, 5, 10, 40, 50, 60, 10 0, 150, and 200 Hz Pulse ................................................................... 30 and 60 BPM, 60 ms pulse width R-Wave Detection Waveform .
Impulse 6000D/7000DP Users Manual Supplement 4/08 7 Accuracy ......................................................... ±(10 % of setting + 0.2 mV) Amplitude of Transvenous Pacer Pulse Simulation signals relative to amplitude setting (in percent) Lead II reference Lead # I II III V1 V2 V3 V4 V5 V6 Ref.
Manual Supplement Impulse 6000D/700 0DP Users 8 4/08 Change #2 On page 12, prior to Analyzing Pacemakers (7000DP only) , add the following section: SCOPE OUTPUT Each time a defibrillator is fired into.
Impulse 6000D/7000DP Users Manual Supplement 4/08 9 Pulse playback +4 V, 1 ms pulse Start of defibrillator pulse playback [1] [1] Pulse playback includes 0.4 ms before th e pulse was detected and theref ore displays any ear ly transitional signals. Use the synchronizing marker pulses to display specific information from the Scope Outpu t signal.
Manual Supplement Impulse 6000D/700 0DP Users 10 8/08 Change #3, 48038 On page 38, under Pacemaker Input, change: From: Accuracy …………..±1 % non-inductive (<2 μ H) To: Accuracy………….
デバイスFluke 6000Dの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Fluke 6000Dをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはFluke 6000Dの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Fluke 6000Dの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Fluke 6000Dで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Fluke 6000Dを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はFluke 6000Dの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Fluke 6000Dに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちFluke 6000Dデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。