Flukeメーカー77の使用説明書/サービス説明書
ページ先へ移動 of 12
77/75/23/2 1 English Instru ction Sheet Page 1 ® 77/75/23/21 Series III Multimeter Instruction Sheet W Read First: Safe ty Information • Never use the m eter if t he meter or t est leads look dama ged. • Be sure the tes t leads an d switc h are in t he correct p osition for the desir ed measurem ent.
77/75/23/2 1 English Instru ction Sheet Page 2 Input Jacks V 600V 1000V FUSED CAT CAT Volts, Ohms, Diode Test Common Terminal Amp Milliamp hc1f.eps See Specific ations for overload protection. Autorange The meter defa u lts to autorange w hen you tur n on the meter.
77/75/23/2 1 English Instru ction Sheet Page 3 Automatic Touch Hold Mode W Warning To avoid electric shock, do not use the Touch Ho ld mode to determine if a circuit with high voltage i s dead. The Touch Hol d mode will not capture unstable or noisy re adings.
77/75/23/2 1 English Instru ction Sheet Page 4 Standby If the meter is on but is inactive for an ho ur (20 minutes in diode test), the dis play goes blank and displays four bar graph segments . To resume op eration, turn the rotary switc h or press a button.
77/75/23/2 1 English Instru ction Sheet Page 5 Diode Test ( G ) RANGE CAT + __ Forward Bias Single Beep RANGE CAT CAT + Reverse Bias _ RANGE CAT CAT + Shorted Open and Good Diode Good Diode Bad Diode _ RANGE CAT CAT + Bad Diode hc5f.
77/75/23/2 1 English Instru ction Sheet Page 6 Continuity Test ( R ) RANGE CAT CAT + _ RANGE CAT CAT + _ hc8f.eps If continuity exists (resistance < 210 Ω for Models 21/75 and <270 Ω for Models 23/77), the beeper s ounds c ontinuously. The meter beeps twice if it is in Touc h Hold mode.
77/75/23/2 1 English Instru ction Sheet Page 7 Probe Holder hc14f.eps Maintenanc e W Warning To avoid electr ic shock, r emove the test lead s before opening t he case, and close the case before using the met er. To prevent fire and possible arc- flash, use fuses w ith ratings shown on the back of the meter.
77/75/23/2 1 English Instru ction Sheet Page 8 Battery R epla cement Note Before opening the case, make sure t he test leads are removed and the rotary switch is turned to OFF.
77/75/23/2 1 English Instru ction Sheet Page 9 Service a nd Part s To contact Fl uke, call one of the follow ing telepho ne numbers: USA and Canada: 1-888-99-FLUKE (1-8 88-993-5853) Europe: +31 402- 678-200 Japan: +81-3- 3434-0181 Singapore: +65- * -276- 6196 Anywhere in the world: +1-425-356-5 500 Or, vis it Fluke’ s Web site at www .
77/75/23/2 1 English Instru ction Sheet Page 10 Electromagnet ic Compatibility In RF field of 3 V/m on all functi ons. Total a ccuracy = Specified a ccura cy + 0.
77/75/23/2 1 English Instru ction Sheet Page 11 Function Range Accuracy L 3.200 V, 32.00 V, 320.0 V 600 V (21/75), 1000 V (23/77) ± (0.3%+1) ± (0.4%+1) m L 320.0 mV ± (0. 3%+1) K (45 to 500 Hz, 3.2 V range. Other ranges 45 to 1 kHz) 3.200 V, 32.00 V, 320.
77/75/23/2 1 English Instru ction Sheet Page 12 Overload pro tection for all functions and r anges: Rated voltage . Function Input Impedance (Nominal) L , m L , K >10 M Ω , <50 pF Common Mo de.
デバイスFluke 77の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Fluke 77をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはFluke 77の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Fluke 77の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Fluke 77で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Fluke 77を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はFluke 77の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Fluke 77に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちFluke 77デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。