PN 644226 July, 1997 Rev.1, 3/99 © 1997,1999 Fluke Corporation, Al l rights reserved. Printed in U .S.A. All p r oduct names are trademarks of their respective companies. 1 ® 16 Multimeter with Temperature Calibration Information W Warning To avoid electrical shock, remove test leads and any input signals from the Meter before opening the case.
16 Multimeter with Temperature 2 Accuracy specifications are given as follows: ±([% of reading] + [number of least significant digits]). Table 1. Specifications (cont) Function Range Resolution Accuracy f (50 to 400 Hz) 4000 mV 1 4.000 V 40.00 V 300.
Specifications 3 Table 1. Specifications (cont) Function Overload Protection 1 Input Impedance (Nominal) Common Mode Rejection Ratio (1 k Ω Unbalanced) 2 Normal Mode Rejection 2 f 600 V dc >5 M .
16 Multimeter with Temperature 4 Disassembly and Reassembly Referring to Figure 1, disassemble the Meter as follows. A Phillips-head screwdriver and small flat-blade screwdriver are required. 1. Remove the test leads and set the rotary knob to OFF. 2.
Recommended Test Equipment 5 Case, Top Decal Switch Assembly Actuator, Switch Main PCA Shield, Bottom Foot LCD Conn. Elastomeric Conn. Elastomeric LCD to PCA Support, Switch Assembly Battery Shock Absorber Case, Bottom Screw, THD Form, PH.P.STL, 5-14 x .
16 Multimeter with Temperature 6 1 2 3 Snap Snap Snaps ig2f.eps Figure 2. Removing and Reinserting the Printed Circuit Assembly.
Performance Test 7 Performance Test W Warning To avoid electric shock, do not execute the performance tests procedures unless the Meter is fully assembled. To ensure that the Meter is working properly and performs to specifications, use the following procedures.
16 Multimeter with Temperature 8 Table 3. Performance Tests Test No. Meter Range Input To Meter Switch Position Display Minimum Display Maximum 1 2 3 4 5 6 7 8 9 1 10 4.000 V ac 4.000 V ac, MIN MAX 4.000 V ac 4.000 V ac 4.000 V ac 4.000 V ac 40.00 V ac 40.
Calibration Adjustments 9 + + 5700A Ice Bath AK-80 Banana Plug Temperature Yellow Red Copper 0 ig6f.eps Figure 3. Performance Test for Temperature + _ 9V BATTERY + Clip to Base Contacts Adjust R4 for 3.997-4.003V DC Input Receptacles + ig1f.eps Figure 4.
16 Multimeter with Temperature 10 Temperature Note Adjust DC voltage (R4) before adjusting temperature (R38). The following describes three temperature calibration procedures: • Calibrating at room temperature. Use this procedure if it is not necessary to optimize the Meter’s performance.
Replacement Parts 11 ! CAL ENABLE 10A 100 mA 1000V 750V MAX ALL INPUTS 1kV MAX 500mA F 250V POWER AUTO 2ND MN MX RATE dB REL LOCAL THRESH ADDR BAUD HOLD LO HI COMP A A V FREQ DUAL DISPLAY MULTIMETER 45 V REF# REF V COM FUSED 80T-150U + _ 9V BATTERY + + R38 Clip to Base Contacts V m ig4f.
16 Multimeter with Temperature 12.
デバイスFluke Fluke 16の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Fluke Fluke 16をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはFluke Fluke 16の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Fluke Fluke 16の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Fluke Fluke 16で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Fluke Fluke 16を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はFluke Fluke 16の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Fluke Fluke 16に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちFluke Fluke 16デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。