GEメーカー800の使用説明書/サービス説明書
ページ先へ移動 of 16
6060 Spine Road Boulder, CO 80301 USA T 303 444 2009 F 303 444 9543 GE Analytical Instruments, Inc. © 2005H Page 1 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI Sievers 800 TOC Analyzer Flow Switch Upgrade (Flow Switch Kit and Firmware 2.12 CBI) 1.
GE Analytical Instruments, Inc. © 2005H Page 2 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI FIGURE ONE: Flow Switch Kit 4. Mechanical installation Refer to Figures One or Two for this proced ure. Reference will depend on orientation of tubing from regulator.
GE Analytical Instruments, Inc. © 2005H Page 3 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 60 MICRON FILTER FLOW SWITCH REGULATOR SAMPLING BLOCK SWITCH OUTLET SWITCH BLOCK INLET INLET OUTLET FLOW SAMPLE REGULATOR FLOW REGULATOR INLET MODULE BINARY INPUT P2 FIGURE TWO: Installed Flow Switch on the Model 800 TOC Analyzer 5.
GE Analytical Instruments, Inc. © 2005H Page 4 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI TC ZERO CORRECTI ON -1.3250000 ENTER CLEAR ENT = SAVE CHAR/NUM CLR = RESTAR T/EXIT A RROWS = CHANGE CHA R FIGURE THREE: TOC ZERO OFFSET Screen 5.2.3.
GE Analytical Instruments, Inc. © 2005H Page 5 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 5.2.9. Select each consumable, pre ss Enter. Write the information on the Printout or the Constants Re cord Form. Record date and percent left for the OXIDIZER.
GE Analytical Instruments, Inc. © 2005H Page 6 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 5.3.4. Replace the EPROM from a Zero Insertion Force Socket. 5.3.4.1. There is a small brass lever on the socket to secure the EPROM chip. It should be in the locked position (end of the lever close to the PC board).
GE Analytical Instruments, Inc. © 2005H Page 7 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI carrier, gently press the pins of the chip to conform to the carrier. CAUTION: Be sure the NOTCH of the EPROM chip is toward the bottom of the analyzer and all the pins are engaged into th e socket.
GE Analytical Instruments, Inc. © 2005H Page 8 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CAUTION: IF THERE ARE ANY VARIATIONS, CONSTANTS MUST BE ENTERED THROUGH THE FACTORY ACCESS MENU. ACCESS TO THIS MENU IS LIMITED AND WILL REQUIRE A PASSWORD.
GE Analytical Instruments, Inc. © 2005H Page 9 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI FIGURE NINE : First Constant Screen 5.6.1.8. Use the Constants Record Form as a guide. If a number is on the Constants Record Form, the new constant value must be that value.
GE Analytical Instruments, Inc. © 2005H Page 10 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI to save digits. After resetting the cl ock, Warning 35 will occur until the power has been cycled two times. 5.7.2. Set Flowrates – The Firmware allows selection of different flow rates for on- line sampling and grab (container) samp ling.
GE Analytical Instruments, Inc. © 2005H Page 11 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI Constants Record CONSTANT VALUE NAME CONSTANT VALUE AFTER UPGRADE CONSTANT VALUE PRIOR TO UPGRADE C1 CONST: Same as prior value C2 CONST: Same as prior value C1 THERM A: Same as prior value C2 THERM.
GE Analytical Instruments, Inc. © 2005H Page 12 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CONSTANT VALUE NAME CONSTANT VALUE AFTER UPGRADE CONSTANT VALUE PRIOR TO UPGRADE ANALOG MAX: Sam.
GE Analytical Instruments, Inc. © 2005H Page 13 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI CONSTANT VALUE NAME CONSTANT VALUE AFTER UPGRADE CONSTANT VALUE PRIOR TO UPGRADE C1 XFER (50Hz): Same as prior value C2 XFER (50Hz): Same as prior value TC/IC OFFSET (50Hz): Same as prior value IC COND.
GE Analytical Instruments, Inc. © 2005H Page 14 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI 6. Firmware Configuration It is necessary to configure the binary inputs following the firmware upgrade. Follow these instructions to configure inputs and select proper polarity.
GE Analytical Instruments, Inc. © 2005H Page 15 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI ENABLE/DISABLE SET POLARITY ENTER CLEA R FIGURE THIRTEEN: Analyzer’s INPUT Menu 6.
GE Analytical Instruments, Inc. © 2005H Page 16 of 16 DIN 30001 Rev. E TOC,FLOW SWITCH UPGRADE 2.12CBI normally open or closed. Press Clear twice to return to the INPUT Menu. 6.2.8. Press Clear to return to the MAIN MENU. The analyzer will function as described in the Operation and Service Manual.
デバイスGE 800の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
GE 800をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはGE 800の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。GE 800の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。GE 800で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
GE 800を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はGE 800の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、GE 800に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちGE 800デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。