USER MANUAL H ELI O 2 HELIUM-OxygEn BLEndEr (DISS and NIST Connections) Model No. PM5400 Series (shown) PM5500 Series SA VE THESE INSTRUCTIONS CAUTION Federal (USA) law restricts this device to sale by or on the order of a physician. 300 Held Drive T el: (+001) 610-262-6090 Northampton, P A 18067 USA Fax: (+001) 610-262-6080 www .
1 H ELI O 2 BLENDEr HELIUM-OXYGEN RECEIVING / INSPECTION Remove the Precision Medical, Inc. H ELI O 2 (Helium-Oxygen) Blender from the packaging and inspect for damage. If there is any damage, DO NOT USE and contact your Provider . INTENDED USE The Precision Medical, Inc.
2 H ELI O 2 BLENDEr HELIUM-OXYGEN READ ALL INSTRUCTIONS BEFORE USING This manual instructs a Professional to install and operate the H ELI O 2 Blender . This is provided for your safety and to prevent damage to the H ELI O 2 Blender . If you do not understand this manual, DO NOT USE the H ELI O 2 Blender and contact your Provider .
3 H ELI O 2 BLENDEr HELIUM-OXYGEN W ARNING • Only trained, qualied medical personnel under the direct supervision of a licensed physician should operate the H ELI O 2 Blender . • Use this H ELI O 2 Blender only for its Intended Use as described in this manual.
4 H ELI O 2 BLENDEr HELIUM-OXYGEN SPECIFICA TIONS High Flow Low Flow Model PM5580: 80/20 heliox PM5570: 70/30 heliox PM5480: 80/20 heliox PM5470: 70/30 heliox Primary Outlet Flow Range 15 - 120 lpm 3 - 30 lpm W i th b ot h s u p p l y p re s s u r e s a t 50 p s i ( 3.
5 H ELI O 2 BLENDEr HELIUM-OXYGEN SPECIFICA TIONS continued Dimensions: (without ttings) Depth: 4.9 in (12.5 cm) Width: 2.3 in (5.7 cm) Height: 4.1 in (10.
6 H ELI O 2 BLENDEr HELIUM-OXYGEN DIAGRAMS CAUTION Missing or illegible labels must be replaced, contact Precision Medical, Inc. PM5400 and PM5500 Models A B F E D C H G.
7 H ELI O 2 BLENDEr HELIUM-OXYGEN COMPONENT DESCRIPTION ITEM DESCRIPTION A Oxygen Concentration Dial A dia l used for sele cti ng oxyg en con cen tra tion s betw een 20% -100% or 30% -100%. The F I O 2 scale is used for reference only . The actual F I O 2 must be veried with an Alarmed Oxygen Monitor / Analyzer .
8 H ELI O 2 BLENDEr HELIUM-OXYGEN PRE-USE TESTING W ARNING • Read this User Manual before installing or operating the H ELI O 2 Blender . • Conrm the concentration of heliox with an alarmed Oxygen Monitor / Analyzer . • Conrm contents of heliox cylinder prior to use.
9 H ELI O 2 BLENDEr HELIUM-OXYGEN REVERSE GAS FLOW PROCEDURE ( CE Requirements ONL Y) 1. Assure bleed ow valve is not engaged. Disconnect the oxygen hose from the gas source. Remove all outlet connections from the H ELI O 2 Blender to ensure that there is no outlet ow .
10 H ELI O 2 BLENDEr HELIUM-OXYGEN 7. Conrm the concentration of heliox with an Oxygen Monitor / Analyzer . If necessary activate the bleed ow valve to maintain F I O 2 accuracy . 8. T o activate the bleed, turn and rotate the knurled collar back until it contacts the cover .
1 1 H ELI O 2 BLENDEr HELIUM-OXYGEN OXYGEN FLOWMETER CONVERSIONS Oxygen Flowmeter Setting Corrected Heliox Flow (LPM) at V arious F I O 2 Settings 20% 30% 40% 50% 60% 70% 80% 90% 100% 1 1.8 1.6 1.4 1.3 1.2 1.18 1.15 1.02 1.0 2 3.6 3.2 2.8 2.6 2.4 2.4 2.
12 H ELI O 2 BLENDEr HELIUM-OXYGEN TECHNICAL DESCRIPTION For a complete T echnical Description of the H ELI O 2 Blender and list of Replacement Parts, reference the H ELI O 2 Blender Service Manual (P/N 506124) available on the Internet; www .precisionmedical.
13 H ELI O 2 BLENDEr HELIUM-OXYGEN TROUBLESHOOTING If the H ELI O 2 Blender fails to function, consult the T roubleshooting Guide below . If problem cannot be solved by using T roubleshooting Guide, refer to the H ELI O 2 Blender Service Manual (P/N 506124) available on the Internet; www .
14 H ELI O 2 BLENDEr HELIUM-OXYGEN LIMITED W ARRANTY AND LIMIT A TION OF LIABILITY Precis ion Medical, Inc. warran ts that the HeliO 2 Blende r , (the Product), will be free of defects in workmanship .
DECLARA TION OF CONFORMITY Manufacturer: Precision Medical, Inc. 300 Held Drive, Northampton, P A 18067, USA CONT ACT : Quality Manager Phone: 610-262-6090 Authorized European Representative: Emergo Europe (European Ofce) Molenstraat 15 2513 BH, The Hague The Netherlands Phone: +31 (0) 70.
デバイスHelio PM5400の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Helio PM5400をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはHelio PM5400の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Helio PM5400の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Helio PM5400で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Helio PM5400を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はHelio PM5400の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Helio PM5400に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちHelio PM5400デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。