4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 70-82-25-19 Rev. 1 7/99.
ii 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 1999 by Honeywell Inc.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual iii A bout This Document Abstract The purpose of this document is to support the installation operation and m aintenance of the 4909 CPVC Conductivity Cell Remov al/Insertion Assem bly.
iv 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Organization Telephone Address Honeywell TAC 1-800-243-9883 Voice 1100 Virginia Drive Fort Washington, PA 19034.
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual v Contents 1. INTRODUCTION ....................................................................................................1 1.1 Overview ............................
vi 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4. MA INTENA NCE A ND REPLAC EMENT PA RTS .................................................17 4.1 To Clean The Cell ...................................................
7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual vii Tables Table 4-1 Replacem ent Parts __________________________________________________________ 18 Table 4-2 Voltage.
.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 1 1. Introduction 1.1 Overview The 4909 Conduct ivit y C ell Ins ertion/R emoval A ssembly is design ed fo.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 2 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 1.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 3 1/ 4" NP T Plug Cabl e 1 - 1/ 2" NPT Nipple Conducti vity Cell P urg e Po rt C ompr es s ion H.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 4 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Cable C onductivity Cell Ad ap t er * * Au t oma t ic Tempera t ur e Com pensato r , w hen specifie d by cat a log nu mber, is m ounted insid e ad apter.
Introduction 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 5 A D C B Ter mina l Boar d As s e m b l y #6- 32 Bi ndi ng He ad Sc r e w & Wa s h e r Vi e w of B a .
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 6 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99.
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7 2. Specifications and Model Selection Guide 2.1 Specifications Cell Constants 0.01, 0.1,1.0, 5, 10, 20, 25 and 50 cm -1 . Electrode Material Nickel or Platinum , as specified.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 8 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Integral Automatic Temperature Compensation Refer to Section 2.2 – Model Selection Guid e Leadwire Tefzel covered, 18-g age cable (7 feet, 20 f eet, or Universal Head, as specified).
Specifications and Model Selection Guide 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 9 049 TABLE IV Selection 09 08 Leadwire Length ’ 7 ft.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 10 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 049 TABLE VII - OPTIONS Tagging None 0_ Linen L_ Stainless Steel S_ Certificate of Calibration No _ 0 Yes _ 1 Notes: 1.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 11 3. Installation 3.1 Requirements To insur e that a re presenta tive sample is being measured at all times, the solutio n must m o ve thr ough and completely purge out the cell channels or guard tube.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 12 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 Allow at least 1/2 inch clearance bey ond the end of th e cell and 1/8 to 3/16 inch radius clearance surrounding the cell to p e rmit circulation of the solutio n.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 13 3. Using a strap wrench on the m ounting nipple, tighten the assembly an additional 1-1/2 to 2 turns. Do not use stillson or chain type wrenches as they m ay damage and weaken the CPVC plastic.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 14 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 3.5 Removal 1. REDUCE PROCESS PRESSURE TO 50 PSIG OR LESS. Sh ut off purge line, if used. Dis connect wi ring conn ections if X1 is selected for Table IV in the MSG.
Installation 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 15 2.5" (64mm) App r ox 4.1" (105mm) App r ox 12.5" (317mm) X " See Bel ow App r ox 19.7" max . (502mm) Allow a minim um clearance o f 36" (914 m m) beyond this poin t for removal of electrode.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 16 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 2. 5" (64mm) Ap prox 4 .1" (105mm) Approx 12. 5" (317mm) X " See Below A pprox 19.
Maintenance and Replacement Parts 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 17 4. Maintenance and Replacement Parts If a series of below normal condu ctivity readings or abov e normal resistiv ity readings occu r, this m ay indicate that the cell is not filled with process solution.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 18 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 4.3 Replacement Parts Table 4-1 Replacement Parts Description Part Number Complete Ass embly See Figs.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 19 5. Platinization and Platinum Black Only the electrodes havi ng cons tants from 5 t.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 20 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 A D C B View o f J unct ion Box Head w ith Cap Re moved WH I T E BLA CK RED GREEN Cell Temp. Com p.
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 21 A A D C B View o f J unct i on Box He ad with Cap Removed WH I T E BLA CK RED GREEN Cell Temp.
4909 CPVC Conductivity Cell Insertion/Removal Assembly Operations Manual 22 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 7/99 WH I T E BLACK RED GREE N Cell Temp. Comp. WK S H G R Five Point Termin al Board. Each Terminal W i ll Accept #16 G age Max .
Platinization and Platinum Black 7/99 4909 CPVC Conductivity Cell Insertion/Removal Assembly– Operations Manual 23 WH I T E BLA CK RED GREEN Cell Temp.
.
.
Industrial Auto mation and Co ntrol Honey well, Inc. 1100 Virginia D rive Fort Wa shington, Pennsy lvania 19034.
デバイスHoneywell 4909 CPVCの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Honeywell 4909 CPVCをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはHoneywell 4909 CPVCの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Honeywell 4909 CPVCの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Honeywell 4909 CPVCで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Honeywell 4909 CPVCを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はHoneywell 4909 CPVCの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Honeywell 4909 CPVCに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちHoneywell 4909 CPVCデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。