Honeywell Process Solutions 4905 Style Conductivity Cells Installation and Maintenance Manual 70-82-25-18 Revision 6.
ii 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 Copyright, Notices, and Trademarks Printed in U.S.A. – © Copyright 2007 by Honeywell Inc.
12/07 4905 Series Conductivity Cells – Installation and Maintenance iii About This Document Abstract This document is intended to support the installation, operation and m aintenance of the 4905 Series of Conductivity Cells. Revision Notes The following list provides notes concerning all revisions of this document.
iv 4905 Series Conductivity Cells – Installation and Maintenance 12/07 Contents 1. INTRO DUCTION ................................................................................................... 1 1.1 Overview .....................................
12/07 4905 Series Conductivity Cells – Installation and Maintenance v Figures Figure 1-1 4905 Type Conductivity Cell Mounted in a 1-1/4” Schedule 80 Tee Using an Adapter Bushing __________________.
vi 4905 Series Conductivity Cells – Installation and Maintenance 12/07.
Introduction 12/07 4905 Series Conductivity Cells – Installation and Maintenance 1 1. Introduction 1.1 Overview These cells form the sensing network for industrial an alyzers and transm itters de signed to make continuous measurem ents of electrolyti c conductivity.
Introduction 2 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 The cell constant is selected according to the range of the measuring instrum ent used and the solution measured.
Specifications 12/07 4905 Series Conductivity Cells – Installation and Maintenance 3 2. Specifications 2.1 Specifications for 04905 Series Parameter Description Cell Constant 04905 Series: 0.
Specifications 4 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 2.2 Specifications for 276127 Flow Chamber Parameter Description Maximum Flow 2 gpm @ 40psig and atmospheric discharge Maximum Pressure 200 psig @ 25°C Maximum Temperature 140°C (284°F) at atmospheric pressure Dimensions 1-1/2” (3.
Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 5 3. Installation 3.1 Overview The conductivity cell is secured permanentl y to the 1” N.
Installation 6 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 3.4 Flow-Type Mounting The cross-channel or guard-tube hole in the cell m ust always be covered by t he solution and the solution level must be 1-1/2 inches above these holes.
Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 7 3.6 Insertion-Type Mounting The cell can be inserted into a 1” N.P.T. threaded op ening, but it is imperative that the tan k or chamber be full under all process conditions.
Installation 8 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 001 0.1 1 10 50 X" Ap pr ox. 5. 4" 5. 4" 5. 4" 5. 8" 6. 9 mm 138 138 138 146 175 0. 940" ( 24mm) Di a 1. 5" (38 mm ) Octagon 1" NPT 1.
Installation 12/07 4905 Series Conductivity Cells – Installation and Maintenance 9 Figure 3-5 Mounting Dimensions for 04905 Series w ith Quick Disconnect Option CABLE CONNECTOR WILL PASS THRU A 0.625(15,9) MIN. DIA PIPE OR CONDUIT OPENING QUICK DISCONNECT COAX ELECTRODE CABLE APPROX.
Electrical Connections 10 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 4. Electrical Connections 4.1 Overview The terminal board connections for the various Honeywell m easuring instrum ents are given in the appropriate Fig ures in this section.
Electrical Connections 12/07 4905 Series Conductivity Cells – Installation and Maintenance 11 Direct Ce ll to Analyz er Insta llation Cell to Analyz er thr ough Junction Box NOTE S: 1. FOR PURE W ATER S A M PLES I N NO N-C ONDUCTI V E (PL ASTIC, G LASS, ET C) PIPING, G ROUND THE B LACK CELL E L ECTRODE L EAD NEAR THE CEL L.
Electrical Connections 12 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 4.2.2 Model 4905 Series with Integral Cable to APT Series Analyzer/Transmitter 04905 series cells with .
Electrical Connections 12/07 4905 Series Conductivity Cells – Installation and Maintenance 13 4.3 Instrument Wiring for 4905 Cells with Quick Disconnect Cable 4.3.1 Wiring Model 4905 with Qu ick Disconnect Cable to UDA2182 Figure 4-4 Wiring Diagram for 4905 Cells w ith Quick Disconnect Cable Connected to UDA2182 Analyzer 4.
Electrical Connections 14 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 COAX YELLOW Note: Ignore black and blue wires. COAX YELLOW Note: Ignore black and blue wires.
Maintenance 12/07 4905 Series Conductivity Cells – Installation and Maintenance 15 5. Maintenance 5.1 Introduction If abnormal readings occur, this m ay indicate poor re sponse because the cell is not filled with process solution. Check the cell installation.
Maintenance 16 4905 Series Conductivity Cells – In stallation and Maintenance 12/07 5.4 Platinizing the Cell Electrodes Only the electrodes having const ants 10 and 50 must be replat inized if the v.
Replacement Parts and Accessories 12/07 4905 Series Conductivity Cells – Installation and Maintenance 17 6. Replacement Parts and Accessories Description Part Number Flow Cell Housing, PES 276127 Junction Box 31316260 Legacy updates cap 50028816-001 Extension Cables for Se nsors with Quick Disconnect Option 2m (6.
.
.
Honeywell Process Solutions Honeywell, Inc. 512 Virginia Drive Fort Washington, Pennsylvania 19034 70-82-25-18 1207 Printed in USA http://hpsweb.honey well.
デバイスHoneywell 70-82-25-18の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Honeywell 70-82-25-18をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはHoneywell 70-82-25-18の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Honeywell 70-82-25-18の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Honeywell 70-82-25-18で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Honeywell 70-82-25-18を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はHoneywell 70-82-25-18の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Honeywell 70-82-25-18に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちHoneywell 70-82-25-18デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。