iERP: 126102 D5618/7 [English] © BW T echnologies 2008. All rights reserved. Quic k R eference Guide Automatic T est and Calibration Station.
Limited W arranty an d Limitati on Liability BW T echnologies LP ( BW) warrants the product to be free from defects i n material and workmanship under normal use and servi ce f or a period of two years, beginning o n the date of shipment to the buyer.
1 MicroDock II Introducti on a Wa r n i n g T o ensure personal safety , read Safety Information - Read First before using the MicroDock II base station.
MicroDock II Quick Reference Guide 2 Display Elements Pushbuttons Item Function AC power Batteries fully charged Batteries h alf-charged Batteries at low level MultiMediaCar d (MMC) MultiMediaCar d (M.
3 MicroDock II User Options Me nu User Options Menu Description Display Time /D at e modifies the time and date of the base station. Inlet Setup enters the gas type, the gas concentratio n levels, and the gas cylinder lot numbe rs. Pump Setup modifies the station pump speed.
MicroDock II Quick Reference Guide 4 Installation Item Description 1 Inlet filter assembly 2 C-cell batteries (4) 3 Battery cover 4 Philips pan head retaining screws (2) 5 MultiMediaCard (MMC) 6 Batte.
5 MicroDock II Installation a War n i n g The atmosphere must be free of background gas. Do not use the base station in a hazardous area. All required National Electrical Codes (NEC) and safety standards must be followed. Note The base station can operate from either an electrical power source or batteries.
MicroDock II Quick Reference Guide 6 Inserting the Detector a Caution Infrared or intense ambient light (sun or halogen) may interfere with the base station/detector communication. T o insert a detector into a docking module, complete the follow- ing: 1.
7 MicroDock II Using the Base Station • When the detector has been inserted correctly , the RUN LEDs light yellow , Unit Inserted displays on the base station LCD, and MicroDock displays on the detector LCD. The base station LCD displays the docking mod- ule number , and the type and serial num ber of the detector .
MicroDock II Quick Reference Guide 8 The base station pushbuttons are not labelled. The base station is operated by pressing the C pushbutton that is located directly below the option that displays on the LCD. Activating the S tation 1. T o activate the station, press and hold C (the left- most button) until the following screen displays.
9 MicroDock II Bump Tes t 5. Press C exit to return to normal operation. For more information on setting up inlets, gas types, and con- centration levels, refer to Inlet Setup in the MicroDock II Base St ation User Manual .
MicroDock II Quick Reference Guide 10 A calibration adjusts the sensor sensitivity to ensure an accu- rate response to gas. T o calibrate a detector , complete the follow ing: 1. From the docking module of the applicable detector, press C CALIBRA T ION.
11 MicroDock II Automatic Datal og Download ( GasAlertMic roClip and GasAl ertMax XT docking module only) Successful Data T ransfe r: The P ASS LED lights green on the docking module and Pass displays on the base station LCD. From the LCD, press C Pass to return to normal operation.
MicroDock II Quick Reference Guide 12 • GasAlertMicroClip • GasAlertMax XT The detector can be configured prior to performing a bump test or calibration to • change user options, • sensor options, and • program options. The detector can also be configured to disallow bump tests or calibrations unless performed with the MicroDock II.
13 MicroDock II Maintenanc e 4. Allow the battery to obtain a full charge (2 to 4 hours, depending upon how many docking modules are con- nected to this station). If a docking module is in q ueue waiting to charge, the CHARGE LED flashes red (all charging modules).
MicroDock II Quick Reference Guide 14 Specifications The MicroDock II base station and docking modules are for indoor use only . Instrument dimensions: (w x l x h) 21.2 x 26.3 x 8.2 cm (8.3 x 10.4 x 3.2 in.) base station and one docking module Weight: 0.
15 MicroDock II Charger Speci fications Charger Specifications Size: 8.6 x 8.2 x 7.8 cm (3.4 x 3.2 x 3.1 in.) Weight: 97 g (3.4 oz.) per model Charger system ingress protection: IP20 Operating temperature: 10°C to 35°C (50°F to 95°F) Humidity: 0 to 50% Altitude: 2000 m (6561.
MicroDock II Quick Reference Guide 16.
iERP: 126102 D5618/7 [English] © BW T echnologies 2008. All rights reserved. Quic k R eference Guide Automatic T est and Calibration Station.
デバイスHoneywell MicroDock IIの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Honeywell MicroDock IIをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはHoneywell MicroDock IIの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Honeywell MicroDock IIの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Honeywell MicroDock IIで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Honeywell MicroDock IIを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はHoneywell MicroDock IIの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Honeywell MicroDock IIに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちHoneywell MicroDock IIデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。