User Guide 1 Φ /3 Φ 1000 Amp True RMS Power Clamp-On Model 380976.
380976 V3.2 4/11 2 Introduction Congratulations on your purchase of the Ext ech 380976 Power Cla mp-On Meter. This device measures 1 Φ /3 Φ Power (True, Ap parent, and Reactive), Horsepo wer, Phase Angle, True RMS Current/Voltage, Resistance, Capacitance, Freque ncy, & Temperature.
380976 V3.2 4/11 3 Warranty EXTECH INSTRUMENTS CORPORATION warrants this instrument to be free of defects in parts and workmanship for on e year from date of shipment (a six month limited warranty applies to sensors and cables).
380976 V3.2 4/11 4 Meter Description 1. Transformer Jaws 2. Jaw opening trigger 3. Data Hold & MX/MN button 4. Function Selector 5. Range button 6.
380976 V3.2 4/11 5 Measurements AC + DC Voltage Measure ments WA R NI NG The maximum input is 600V. Do no attempt vo ltage measurements above this limit. Exceeding this limit could cause electrical sh ock and damage to the meter. 1. Set the rotary switch to the ‘ V’ position.
380976 V3.2 4/11 6 AC Power Measurements 1 Φ KW, KVA, KV A R, Pow er Factor & Phase Angle Measurement 1. Set the rotary switch to the ‘KW/KVA’ position. 2. Insert the test leads to the meter as follows: Black to ‘COM’ and Red to ‘V’. 3.
380976 V3.2 4/11 7 3 Φ 3-Wire KW, HP, KVA, KVAR, Pow er Factor & Phase Angle Measurement 1. First, measure W RS(L1 L2) (refer to the diagram be low). a. Set the rotary switch to the “ V ”. b. Press and hold the “ HOLD ” key while setting the rotary switch to “KW/KVA”, the 3 φ 3W and W L12 symbols will appear.
380976 V3.2 4/11 8 3. The power clamp will process these tw o sets of data (W L12 , W L23 ), and show the result on the LCD. The W L123 symbol will be shown to indicate 3 φ 3W power.
380976 V3.2 4/11 9 NOTE Once a phase is selected as COM, user s can not change this selection in the subsequent measurement. For example, if S (or L2) phase is selected, S (or L2) phase is always connected to t he COM during measurement of W RS (or W L1L2 ) and W TS (or W L3L2 ) in 3 φ 3W unbalanced power.
380976 V3.2 4/11 10 3 Φ 4-Wire KW, HP, KVA, KVAR, Pow er Factor & Phase Angle Measurement 1. First, measure W R(L1) (refer to t he diagram below).
380976 V3.2 4/11 11 3. Third, measure W T(L3) (refer t o the diagram that follows the steps below) a. Disconnect the test probe from the phase wher e th e jaws were clamped in the previous measurement. b. Connect the test probe of the V (red) terminal to th e third phase (e.
380976 V3.2 4/11 12 4. The power clamp will proce ss these t hree sets of data ( W L1, W L2 W L3 ) and s how the r esult o n the L CD. The WL 123 symbol will be s hown to indic ate the 3 φ 4W power (r efer to di agram) . The 3 φ 4W power value in w atts is now stored in the meter’s memory.
380976 V3.2 4/11 13 KW+KVA R, KVA+ θ ( Phase A ngle) or A+V di splays. 6. W 3 φ 4W = W R(L1) + W S(L2) W T(L3) KVA K W KVAR WW W 34 2 34 2 34 φφ φ =+ PF K W KVA W W W 34 34 34 φ φ φ = 7. Set the rotary switch to another position to exit this mode and clear the stored data.
380976 V3.2 4/11 14 Capacitance Me asurements 1. Fully discharge the capacitor under test before pro ceeding. 2. Insert the test leads into the input jacks. (Black to ‘COM’ and Red to ‘ ’). 3. Set the rotary switch to the ‘ ’ position. 4. Connect the red and black test leads to the capacitor.
380976 V3.2 4/11 15 Meter Control Keys HOLD - MX/MN Ke y Data Hold Function Press this key momentarily to put the meter into Data Hold mode ( HOLD will appear on the LCD). In this mode, the meter freezes the displayed reading. To e xit the Data Hold mode, press the key again (the HOLD icon will sw itch off).
380976 V3.2 4/11 16 Automatic Sleep Mode and Battery Replacement The Meter is powered by a 9V batt ery. An AUTO SLEEP feature is included to preserve battery life. Note: The Auto Power OFF feature is disabled when the meter is in the MIN/MAX m ode Note: In the sleep mode the met er continues to draw a small amount of battery curr ent.
380976 V3.2 4/11 17 Specifications General Specifi cations Display Dual Display; 4-digit, 10,000 count (0 to 9999) LC D Jaw Opening 1.6” (42mm) Max. Input limit Max. voltage between any terminal and ground : 600Vrms Sampling rate 2.5 times per second (Digital Disp lay); Once every 6 seconds (KW, KVA, and KVAR) Auto Sleep After approx.
380976 V3.2 4/11 18 Measurement Specific ations Accuracy: ± ( % of rdg + number of digits) from 18 o C to 28 o C (64 o F t o 82 o F) R.H. < 80% AC Current (50 Hz to 400Hz) Tr ue RMS Range Resolution Accuracy (% reading) Sensitivity Overload Protection 99.
380976 V3.2 4/11 19 M Ω (Resistance) Range Resolution Accuracy Overload Protection 9.999M Ω 1K Ω 99.99M Ω 10K Ω ±(5% + 10 d) 600V Capacitance Range Resolution Accuracy Overload Protection 10.000 μ F 1nF 100.00 μ F 10nF 1000.0 μ F 100nF ±(1.5% + 5 d) 7000 μ F 1 μ F ±(4.
380976 V3.2 4/11 20 Support line (781) 890-7440 Technical support: Ex tension 200; E-mail: supp ort@extech.com Repair & Returns: Ex tension 21 0; E-mail: repair@extech.
デバイスHP Extech 380976の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
HP Extech 380976をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはHP Extech 380976の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。HP Extech 380976の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。HP Extech 380976で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
HP Extech 380976を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はHP Extech 380976の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、HP Extech 380976に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちHP Extech 380976デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。