Dealer: This manual MUST be given to the end user . User: Before using this cylinder, r ead this manual and save fo r future reference. For more informa tion re garding Invac are p roduc ts, par ts, and servi ces , please v isit www.
WARNING Oxygen Conservin g Cylinder 2 Part No 1143247 WARNIN G DO NOT OPERATE THIS EQ UIPMENT WITHOUT FIRST READING AND UNDERS TANDING THIS MANUAL.
TABLE OF CONTENTS Part No 1143247 3 Oxygen Conservin g Cyl inder TABLE OF CONTE NTS SPE CI AL NO TES ...... ....... .... ...... ...... ....... ...... .... 4 RECEIVING I NSPECTION . ...... ...... ...... ..... ...... .... 4 INTENDED USE ...... ..... ...
SPECIAL NOTES Oxygen Conservin g Cylinder 4 Part No 1143247 SPECIA L NO TES Signal wor ds are used in this manual and apply to hazards or unsafe practices whi ch could result i n personal injury or property damage. Refer to the following table for definitions of the signal words.
SECTION 1—GENERAL GUIDELINES Part No 1143247 5 Oxygen Conserving Cylinder SECT ION 1—GENE RAL G UIDELIN ES WARNING SECTION 1 - IMPORTANT SAFEGUARDS contains important information for the safe operation and use of this product.
SECTION 1—GENERAL GUIDELINES Oxygen Conservin g Cylinder 6 Part No 1143247 WARNING DO NOT attempt to r epair product. DO NOT immerse product in an y kind of liquid. Store pr oduct in a c lean area when not in use. Only use Inv acare Corpor ation HF2PCM2 carrying bag.
SECTION 1—GENERAL GUIDELINES Part No 1143247 7 Oxygen Conserving Cylinder WARNING DO NOT allow cylinders to tip o r fall. Secure gas cylinders so they cannot fall. For sto rage, ensure cylinder is AL W A Y S in upright position. DO NOT store cylinders near sources of heat or flame.
SECTION 2—FEATURES Oxygen Conservin g Cylinder 8 Part No 1143247 SECTIO N 2— FEATU RES DANGER This product is not intended as a life-sustaining or life-supporting device.
SECTION 3—TYPICAL PRODUCT PARAMETERS Part No 1143247 9 Oxygen Conserving Cylinder SECT ION 3—TYPICAL PRODUCT PARAM ETERS CONSERVER S/N R ANGE: 75,000+ INLET PRESSURE R ANGE: 300 - 2100 psig INTERNAL REGULATED P RESSUR E: 50 psig DIMENSIONS: HF2PCM2KIT: Weig ht 2.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conserving Cylinder 10 Part No 1 143247 SECT ION 4—INSPE CTION AND COMP RESSOR INT ERFACE NOTE: Refer to the HomeFill Compr essor Owner’s Manu al for compres sor operating instructions.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 11 Oxygen Conserving Cylinder External Exam inati on 1. Examine the outside of the cylinder for the following conditions, and replace th.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conservin g Cylinder 12 Part No 1143 247 FIGURE 4.1 External Examination DO NO T REMO VE THIS PRODUCT LABEL Not for Commer cial Resale Inv acare Cor poration 2101 E.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 13 Oxygen Conserving Cylinder Connectin g/Disconn ecting the C ylinder to/from th e Comp ressor Conn ect ing the Cylind er t o th e Comp res sor WARNING NEVER use tools of any kind to connect/disconn ect the cylinder and the compressor.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conservin g Cylinder 14 Part No 1143 247 4. Momentarily push DOWN on the outer rin g (sleev e) of the connector fillport to reset the connector as shown in FIGURE 4.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 15 Oxygen Conserving Cylinder FIGURE 4.3 Connecting the Cylinder to the Compressor Cylinder/Regulat or Assembly Compre ssor Cyli nde r G.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conservin g Cylinder 16 Part No 1143 247 Disconnecti ng the Cyli nder from the Compre ssor WARNING NEVER use tools of any kind to connect/disconn ect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 17 Oxygen Conserving Cylinder 5. When the cylinder fillport is di sconnected from the connector fillport, release the outer ring (sleeve) of the connector fillport and use two hands to r emove the cylinder from the compressor cradle.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conservin g Cylinder 18 Part No 1143 247 Changi ng the Carryin g P osition of the Cyli nder NOTE: For this pr ocedure, r efer to FIGURE 4.6 on page 19. The cylinder MUST be kept in the cyl inder bag at all times except when cleaning.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Part No 1143247 19 Oxygen Conserving Cylinder ii. Thread the end of the belt that has no buckle and retaining loop through the tw o loops on the cylinder bag. See Detail “B”. N O T E : T h e w e b m a t e r i a l o n t h e b e l t s h o u l d b e f a c i n g a w a y f ro m t h e cylinder bag.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE Oxygen Conserving Cylinder 20 Part No 1 143247 Cleaning th e Cylind er Bag NOTE: Skip STEP one if the belt is to be cleaned with the cylinder bag. 1. Unhook and remov e the cylinder bag and belt from body .
SECTION 5—OPERATING INSTRUCTIONS Part No 1143247 21 Oxygen Conserving Cylinder SECTIO N 5— OPERA TING INSTRU CTIO NS WARNING NEVER smoke in an area where oxygen is being administered. NEVER u se near any type of flame or flammable/exp losive substances, vapors or atmosphere.
SECTION 5—OPERATING INSTRUCTIONS Oxygen Conservin g Cylinder 22 Part No 114324 7 5. Place the cannula over yo ur ears and positi on the prongs in yo ur nose as instructed by your Health Care prov ider or cannula manufacturer . Refer to FI GURE 5.1. 6.
SECTION 6—MAINTENANCE Part No 1143247 23 Oxygen Conserving Cylinder SECT ION 6—MAINTE NANCE 1. After each use, clean exterior of the product with a dry , li nt free cloth only . 2. Store pr oduct in a c lean area free fr om grease, oil, and ot her so ur ces of con t amina tio n.
SECTION 7—CYLINDER FILL TIMES Oxygen Conservin g Cylinder 24 Part No 114324 7 SECTIO N 7— CYLIN DER FIL L TIMES NOTE: All filling times ar e approximate and may vary on envir onmental conditions. * Troubl eshootin g If Conserving Cylinder fails to func tion, consult the T roubleshooting guide.
SECTION 7—CYLINDER FILL TIMES Part No 1143247 25 Oxygen Conserving Cylinder NOT ES.
LIMITED WARRANTY Oxygen Conservin g Cylinder 26 Part No 114324 7 LIMITE D WARRAN TY PLEASE NOTE: THE WAR RANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APP LICABLE TO PRODUCTS MAN UFACTUR ED AFTER JULY 4, 1975. This warra nty is extend ed only to the ori ginal p urchaser/u ser of our pro ducts.
LIMITED WARRANTY Part No 1143247 27 Oxygen Conserving Cylinder LIMITATIONS AND EXCLUS IONS: THE FOREGOING WARRANTY SHALL N OT APPLY TO S ERIAL NUMBERED PRODUCTS IF THE SER IAL NUMB ER HAS BEEN REMOVED.
Invacare Corporation www .invaca re.com USA One Invac are Way Elyria , Oh io USA 44036-2125 800-333- 6900 Techni cal Se rvices 800-832- 4707 Invacare, Yes, you can.
デバイスInvacare HF2PCM2KITの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Invacare HF2PCM2KITをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはInvacare HF2PCM2KITの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Invacare HF2PCM2KITの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Invacare HF2PCM2KITで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Invacare HF2PCM2KITを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はInvacare HF2PCM2KITの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Invacare HF2PCM2KITに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちInvacare HF2PCM2KITデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。