Dealer: This manual MUST be given to the end user. User: Before using this cylinder, read this manual and save for future reference. For more information regarding Invacare products, parts, and services, please visit www.
HomeFill ™ 2 Part No 114811 4 WARNING DO NOT OPERATE THIS EQUIPMENT WITHOUT FIRST READING AND UN DERSTANDING THIS MANUAL. IF YOU ARE UNABLE TO UNDERSTAND THE WARNINGS, CAUTIONS, AND INSTRUCTIONS.
TABLE OF CONTENTS Part No 1148114 3 HomeFill ™ TABLE OF CONTENTS SPECIAL NOTES ............................................... 4 RECEIVING INSPECTION .................................. 4 INTENDED USE ................................................ 4 SECTION 1—GENERAL GUIDELINES .
SPECIAL NOTES HomeFill ™ 4 Part No 114811 4 SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could re sult in personal injury or property damage. Refer to the following table for definitions of the signal words.
SECTION 1—GENERAL GUIDELINES Part No 1148114 5 HomeFill ™ SECTION 1—GENERAL GUIDELINES WARNING SECTION 1 - IMPORTANT SAF EGUARDS contains important information for the safe operation and use of this product.
SECTION 1—GENERAL GUIDELINES HomeFill ™ 6 Part No 114811 4 WARNING DO NOT attempt to repair product. DO NOT immerse product in any kind of liquid. Store product in a clean area when not in use. Only use an Inv acare Co rporation HF2POST6BAG or HF2POST9BAG carrying bag.
SECTION 1—GENERAL GUIDELINES Part No 1148114 7 HomeFill ™ WARNING NEVER oil or lubricate the cylinder fittings. NEVER use any penetrating oil or lubricant such as WD-40 or 3 in One. If connection difficulty is experien ced v erify that the cylinder fitting is being inserted straight down into the compressor coupler .
SECTION 2—FEATURES HomeFill ™ 8 Part No 114811 4 SECTION 2—FEATURES FIGURE 2.1 Product Diagram Cylinder Fill Port Toggle Valve.
SECTION 3—TYPICAL PRODUCT PARAMETERS Part No 1148114 9 HomeFill ™ SECTION 3—TYPICAL PRODUCT PARAMETERS DIMENSIONS : HF2POST6: Weight 4.0 lbs (1.81 kg) HF2POST6: Overall L ength 11.1 in. (2.8 cm) HF2POST6: Diameter 4.4 in. (11.1 cm) HF2POST9: Weight 5.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 10 Part No 1148114 SECTION 4—INSPECTION AND COMPRESSOR INTERFACE NOTE: Refer to the HomeFill ™ Compr essor Owner’s Manual for compressor operating instructions.
SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1148114 11 HomeFill ™ External Examination NOTE: For this procedur e, r efer to FIGURE 4.1 on page 12.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 12 Part No 1148114 FIGURE 4.1 External Examination Checking Cylinder Pressure WARNING NEVER use tools of any kind to conn ect/disconnect the cylinder and the compressor. Otherwise, severe injury and/or damage may occur.
SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1148114 13 HomeFill ™ 2. Ensure toggle v alve is open. Open the toggle v alve by rotating counterclockwise. 3. Examine the gauge on the connected conserver or regulator . P erform one of the following: • If the cylinder pressure is less than 1500 psig, fill the cylinder .
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 14 Part No 1148114 FIGURE 4.2 Component Identification 4. Remove the post v alve cov er (located on the cylinder) and the connector fillport covers (located on the compressor) (FIGURE 4.3). FIGURE 4.
SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1148114 15 HomeFill ™ NOTE: If the outer ring (sleeve) is in the UP position (GREEN dots not visible), the connector fillport will not be able to accept the cylinder fillport. Pushing DOWN momentarily will reset the connector fill port (GREEN dots visible) to accept the cylinder fillport.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 16 Part No 1148114 NOTE: Refer to Cylinder Fill T ime s on page 20 for the length of time it will take to fill the cylinder .
SECTION 4—INSPECTION AN D COMPRESSOR INTERFACE Part No 1148114 17 HomeFill ™ 11. After the cylinder is full , press the compressor powe r switch to the Off (O) position (FIGURE 4.6). 12. Grasp the cylinder assembly (FIGURE 4.6). 13. W ith the other hand, grasp the outer ring (sleeve) of the connector fillport and push DOWN (FIGURE 4.
SECTION 4—INSPECTION AND COMPRESSOR INTERFACE HomeFill ™ 18 Part No 1148114 WARNING The fillport cover on the co nnector fillport and the post valve fillport MUST be replaced after filling and whenever not in use. If either of th e fillport covers are missing contact your healthcare provider before using.
SECTION 5—MAINTENANCE Part No 1148114 19 HomeFill ™ SECTION 5—MAINTENANCE 1. After each use, clean exterior of the product with a dry , lint free cloth only . 2. Store product in a clean area free from grease, oil, and other sources of contamination.
SECTION 6—CYLINDER FILL TIMES HomeFill ™ 20 Part No 1148114 SECTION 6—CYLINDER FILL TIMES NOTE: All filling times ar e a ppr oximate and may vary on environmental conditions. * Troubleshooting If post v alv e cylinder fails to function, try the following solutions.
NOTES Part No 1148114 21 HomeFill ™ NOTES.
LIMITED WARRANTY HomeFill ™ 22 Part No 1148114 LIMITED WARRANTY PLEASE NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURED AFTER JULY 4, 1975. This warranty is extended only to the origina l purchaser/user of our products.
LIMITED WARRANTY Part No 1148114 23 HomeFill ™ LIMITATIONS AND EXCLUSIONS : THE FOREGOING WARRANTY SHALL NOT APPLY TO SERIAL NUMBERED PRODUCTS IF THE SERIAL NUMBER HAS BE EN REMOVED OR DEFACED; PROD.
Invacare Corporation www.invacare.co m USA One Invacare Way Elyria, Ohio USA 44036-2125 800-333-6900 Technical Services 800-832-4707 All rights reserved . Trademarks are identified by the symbols ™ and ®. All trademarks are owned by or licensed to Invacare Corporation unless otherwise noted.
デバイスInvacare HF2POST9の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Invacare HF2POST9をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはInvacare HF2POST9の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Invacare HF2POST9の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Invacare HF2POST9で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Invacare HF2POST9を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はInvacare HF2POST9の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Invacare HF2POST9に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちInvacare HF2POST9デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。