Dealer: This manual MUST be given to the end user of this product. User: BEFORE using this product, read this manual and save for future reference. User Manual For more information regarding Invacare products, parts, and services, please visit www.invacare.
WARNING Perfecto 2 ™ Series 2 Part No 114348 2 WARNING DO NOT USE THIS PRODUCT OR ANY AVAILABLE OPTIONAL EQUIPMENT WITHOUT FIRST COMPLETELY READING AND UN DERSTANDING THESE INSTRUCTIONS AND ANY ADDITIONAL INSTRUCTIONAL MATERIAL SUCH AS OWNER’S MANUALS, SERVICE MANUALS OR INSTRUCTION SHEETS SUPPLIED WITH THIS PRODUCT OR OPTIONAL EQUIPMENT.
TABLE OF CONTENTS Part No 1143482 3 Perfecto 2 ™ Series Table of Contents ACCESSORIES .................................................. 2 SPECIAL NOTES ............................................... 4 LABEL LOCATION ...............................
SPECIAL NOTES Perfecto 2 ™ Series 4 Part No 114348 2 SPECIAL NOTES Signal words are used in this manual and apply to hazards or unsafe practices which could result in personal injury or property damage.
SPECIAL NOTES Part No 1143482 5 Perfecto 2 ™ Series CAUTION “Caution: Federal law restricts this device to sale or rental by or on order of .
LABEL LOCATION Perfecto 2 ™ Series 6 Part No 114348 2 LABEL LOCATION FOR OPERA TING INSTRUCTIONS or ALARMS (See Operator’s Manual) DANGER - ELECTRIC SHOCK DANGER - RISK OF FIRE NO SMOKING - Keep ALL sources of ignition out of the room in which this product is located and away from areas where o xygen is being delivered.
SECTION 1—GENERAL GUIDELINES Part No 1143482 7 Perfecto 2 ™ Series SECTION 1—GENERAL GUIDELINES In order to ensure the safe installation, assembly and operation of the Pe r f e c t o 2 concentrator these instructions MUST be followed.
SECTION 1—GENERAL GUIDELINES Perfecto 2 ™ Series 8 Part No 114348 2 Operating Information For optimum performance, Inv acare recommends that each concentrator be on and runni ng for a minimum of 30 minutes at a ti me.
SECTION 1—GENERAL GUIDELINES Part No 1143482 9 Perfecto 2 ™ Series Other devices may experience interference from ev en the low levels of electromagnetic emissions permitted by the above standards.
SECTION 2—FEATURES Perfecto 2 ™ Series 10 Part No 1143482 SECTION 2—FEATURES Oxygen Outlet Flowmeter Oxygen Purity Indicator Lights / Fault an d Power Indicator Lights Power Switch Circuit Break.
SECTION 3—HANDLING Part No 1143482 11 Perfecto 2 ™ Series SECTION 3—HANDLING The concentrator should AL W A YS be kept in the upright position to prevent cabinet damage while being transported.
SECTION 4—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™ Series 12 Part No 1143482 SECTION 4—TYPICAL PRODUCT PARAMETERS Electrical Requirements: 120 VAC ± 10% (132 VAC/10 8 VAC), 60 Hz Rated Current I.
SECTION 4—TYPICAL PRODUCT PARAMETERS Part No 1143482 13 Perfecto 2 ™ Series Change in maximum recommended flow when back pressure of 7kPa is applied: 0.7 L/min Filters: Cabinet, Outlet HEPA and Compressor Inlet Safety System: Current overload or line surge shutdown.
SECTION 4—TYPICAL PRODUCT PARAMETERS Perfecto 2 ™ Series 14 Part No 1143482 *NOTE: Stated concentration levels achieved after initial warm ‐ up period (appr oximately 30 minutes). Placement: No closer than three inche s from any wall , furniture, draperies, or similar surfaces.
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 15 Perfecto 2 ™ Series SECTION 5— OPERATING INSTRUCTIONS Introduction Yo u r oxygen concentrator is intended for individual use indoors. It is an electronically operated device that separates oxygen from room air .
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 16 Part No 1143482 Set Up Plug in Power Cord Plug in pow er cord to an electrical outlet. Connect Humidifier (If So Prescribed) WARNING DO NOT overfill humidifier .
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 17 Perfecto 2 ™ Series FIGURE 5.2 Humidifer Bottle Adapter 5. Attach the humidifier bottle adapter to the humidifier .
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 18 Part No 1143482 FIGURE 5.3 Humidifier Compartmen t Power Switch NOTE: For this pr ocedure, refe r to FIGURE 5.4. 1. Press pow er sw itch to on position.
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 19 Perfecto 2 ™ Series Flowrate NOTE: For this procedure, refe r to FIGURE 5.5 on page 20. 1. T urn the flowrate knob to the setting prescribed by you r physician or therapist.
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 20 Part No 1143482 FIGURE 5.5 Flowrate Front Panel Indicators - Units with SensO 2 The SensO 2 feature monitors the purity of oxygen generated by the oxygen concentrator .
SECTION 5—OPERATING INSTRUCTIONS Part No 1143482 21 Perfecto 2 ™ Series Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial sta rt warm ‐ up time (approximately 30 min.
SECTION 5—OPERATING INSTRUCTIONS Perfecto 2 ™ Series 22 Part No 1143482 Initial Startup of the Concentrator NOTE: Concentrator may be used during the initial sta rt warm ‐ up time (approximately 30 min).
SECTION 6—MAINTENANCE Part No 1143482 23 Perfecto 2 ™ Series SECTION 6—MAINTENANCE WARNING The Invacare concentrators are specifically designed to minimize routine preventiv e mainte nance.
SECTION 6—MAINTENANCE Perfecto 2 ™ Series 24 Part No 1143482 2. Clean the cabinet filter with a va c u u m cleaner or wa s h in w arm soapy wa t e r and rinse thoroughly . 3. Dry the filter thoroughly before reinstallation.
SECTION 6—MAINTENANCE Part No 1143482 25 Perfecto 2 ™ Series FIGURE 6.2 Preventi ve Maintenance Record ON EACH INSPECTION MODEL NO. I R C SERIAL NO. PREVENTIVE MAINTENANCE RECORD Record Date of Service Record Elapsed Hours On Hour Meter Clean Cabinet Filter(s) Check Prescribed L/min.
SECTION 7—TROUBLESHOOTING GUIDE Perfecto 2 ™ Series 26 Part No 1143482 SECTION 7—TROUBLESHOOTING GUIDE SYMPTOM: PROBABLE CAUSE: SOLUTION: Alarm: Short beeps, long pause Concentrator not operating, power switch On. Beep......Beep. ..... Main Power Loss: 1.
SECTION 7—TROUBLESHOOTING GUIDE Part No 1143482 27 Perfecto 2 ™ Series YELLOW or RED light Illuminated. Alarm: Continuo us On RED light only. * Only applies to IRC5PO2. 1. Low oxygen purity.* 1. Clean or Replace filters. 2. Kinked or blocked tubing, cannula or humidifier.
NOTES Perfecto 2 ™ Series 28 Part No 1143482 NOTES.
NOTES Part No 1143482 29 Perfecto 2 ™ Series NOTES.
NOTES Perfecto 2 ™ Series 30 Part No 1143482 NOTES.
LIMITED WARRANTY Part No 1143482 31 Perfecto 2 ™ Series LIMITED WARRANTY NOTE: THE WARRANTY BELOW HAS BEEN DRAFTED TO COMPLY WITH FEDERAL LAW APPLICABLE TO PRODUCTS MANUFACTURE D AFTER JULY 4, 197 5. This warranty is extended only to th e original purchaser who purchases this product when new and unused from Invacare Corporation or a dealer.
LIMITATIONS AND EXCLUSIONS: THE FOREGOING WARRANTY SHALL NOT APPLY TO PRODUCTS SUBJECTED TO NEGLIGENCE, ACCIDENT, IMPROPER OPERATION, MAINTENANCE OR STORAGE, SOOT OR SMOKE-FILLED ENVI RONMENTS, OR OTH.
デバイスInvacare IRC5PO2Wの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Invacare IRC5PO2Wをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはInvacare IRC5PO2Wの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Invacare IRC5PO2Wの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Invacare IRC5PO2Wで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Invacare IRC5PO2Wを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はInvacare IRC5PO2Wの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Invacare IRC5PO2Wに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちInvacare IRC5PO2Wデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。