Owner’s Operator and Maintenance Manual DEALE R: This manual M UST be given to the user of th e product. USER: BEFORE using this product, read this manual an d save for future reference. For mo re inform ation reg arding In va ca re p ro du cts , pa rts, and services, please v isit www.
SPECIAL NOTES MA50 Series 2 Part No 1148136 WARNING DO NOT use this product or any available optional equipment without first completely reading and understanding these instructions and any addition al instructional material such as owner’s manuals, service manuals or instruction sheets supplied with this product or optional equipment.
TABL E OF CON TENT S Part No 1148136 3 MA50 Series TABLE OF CONTENT S SPE CI AL NOT ES ..... ...... ...... ...... ....... ...... ...... ...... ....... ...... ...... ...... ...... . 2 REGISTER YO UR PRODUCT ..... ..... ...... ...... ...... ...... .....
TABLE OF CONTENTS MA50 Series 4 Part No 1148136 TABLE OF CONTENT S SECTION 3—MAINTENANCE AND TROUBLES HOOTING ....... ........ .... 2 1 Cleaning the System .............................. ........................... ........................... ......
TYPICAL PRODUCT PARAMETERS Part No 1148136 5 MA 50 Se rie s TYPICAL PRODUCT PARAMETERS Electrical Param eters Performance Parameters Mechanical Param eters MA50/MA51/MA55 INPUT VOLTAGE AC: 90 V INPUT .
TYPICAL PRODUCT PARAMETERS MA50 Series 6 Part No 1148136 Environmental Parameters MA50/MA51/MA55 OPERATING CONDITIONS AMBIENT TEMPERA TURE: RELATIVE HUMIDITY: ATMOSPHERIC PR ESSURE: 50° - 104° F 30%.
LABEL LOCATION Part No 1148136 7 MA 50 Se rie s LABEL LOCAT ION DANGER-EXPLOSION HAZARD: DO NOT us e in the presence of flammable anesthetics. CAUTION: Eq uipment should be connected to a properly grounded receptacle (3-prong). Risk of Electrical shock.
SECTION 1—GENERAL GUIDELINES MA50 Series 8 Part No 1148136 SECTION 1—GENERAL GUIDELINES WARNING SECTION 1 - GENERAL GUIDELINES contains important information for the s afe operation and use of this product. Contraindications ALW AYS consult the pati ent’ s physici an before using the M A50, MA51 and MA55 systems.
SECTION 1—GENERAL GUIDELINES Part No 1148136 9 MA 50 Se rie s Fire Hazard DANGER Smoking DO NOT SMOKE while using this device. This system uses r oom air for circulation through the mattress. A cigarett e can burn a hole i n the bed surface and cause damage to the m attress.
SECTION 2—OPERATION MA50 Series 10 Part No 1148136 SECTION 2—OPERATION CAUTION The control unit and mattress on the MA50 series are designed to be used as a system. DO NOT replace mattresses or control units with other models or other brands. Otherwise, damage to the system m ay occur.
SECTION 2—OPERATION Part No 1148136 11 MA50 Se rie s Mattress Replaceme nt System (MA51 and MA55) CAUTION DO NOT strap t he mattress to t he bed frame at the head and foo t ends. Secure ALL mattr ess straps. Secure the straps t o the bed deck at the head and foot ends and to the frame at t he center of the bed.
SECTION 2—OPERATION MA50 Series 12 Part No 1148136 Installing the Side Rails WARNING Patient entrapment with bed side rails may cause injury or death. Mattress MUST fit bed frame and side rails snugly to prevent patient entrapment. Follow the manufacturer’s instructions.
SECTION 2—OPERATION Part No 1148136 13 MA50 Se rie s FIGURE 2.4 Connecting the Hose Connecting the Power Cord WARNING DO NOT alter plug to fit a non-conforming outlet. Instead, have an electrician install a properly grounded 3-prong outlet. Fail ure to use t he correct plug and outlet can result in a potential safety hazard.
SECTION 2—OPERATION MA50 Series 14 Part No 1148136 Using the Front Panel NOTE: For this pr ocedure, r efer to FIGURE 2.5. FIGURE 2.5 Using the Front Panel Power Butt on 1. T o turn the contro l uni t on or off, press and release the P ower butt on ( ).
SECTION 2—OPERATION Part No 1148136 15 MA50 Se rie s Static/Alte rnating Pre ssure Button 1. Press the Alternating Pressure butto n ( ) to select betwe en Static Mode and one of four Alternating Pressure Times.
SECTION 2—OPERATION MA50 Series 16 Part No 1148136 Firm/Soft Buttons 1. Select comfort pressure sett in gs by pressing Firm ( ) or Soft ( ) button. • Soft Button - Pressing this button reduces the pressure setting in the mattress. • Firm Button - Pressing this bu tton increase the pressure setting in the mattress.
SECTION 2—OPERATION Part No 1148136 17 MA50 Se rie s Mode Button (Max Inflate/Lo w Air Loss) NOTE: The Mode button is on the MA55 system only . 1. Press the Mode button ( ) to select the Max Inflate mode or the Low Air Loss mode. • Max Inflate Mode - In this mode, the mattress inflates rapidly to maximum firmness (pressuri zed to 35 ± 6 mmHg).
SECTION 2—OPERATION MA50 Series 18 Part No 1148136 Low Pressure LED In the ev ent t hat the mattr ess hose disc onnects, an alar m sounds and Low Pressure LED flashes AMBER . Once the low pressure problem is fixed, the control unit resumes operation in the previously set mode.
SECTION 2—OPERATION Part No 1148136 19 MA50 Se rie s 9. V erify that the patient has not bottomed out by performing the following steps: A. Ensure that the patient is lying flat on his/h er back in the middle of the mattress. B. Place four fingers between the air cushions di rectly underneath the sacral reg ion of the patient’ s body .
SECTION 2—OPERATION MA50 Series 20 Part No 1148136 Transferring Patient From/To a Wheel chair WARNING ALWAYS engage the wheel locks of the bed and the wheel locks of the wheel chair before transferring the patient between the bed and the wheelchair.
SECTION 3—MAINTENANCE AND TROUBLESHOOTING Part No 1148136 21 MA50 Se rie s SECTION 3—MAIN TENANCE AND TROUBLESHOOTING Cleaning the System WARNING Before cleaning or disassembling the MA50 series, check the underside of the mattress folds for sharp objects such as scissors, needles, etc.
SECTION 3—MAINTENANCE AND TROUBLESHOOTING MA50 Series 22 Part No 1148136 CAUT ION Laundry workers should always wash their hands before working with clean bedding. DO NOT overload the machine. DO NOT use chlorine bleach because it may damage the fabric coating.
SECTION 3—MAINTENANCE AND TROUBLESHOOTING Part No 1148136 23 MA50 Se rie s Troubleshooting PRO BL EM CA USE SOL UTI ON Mattress not inflating Not alternating properly Mattress hose disconnected Conn.
LIMITED WARRANTY PLEASE NO TE: TH E WAR RANTY B ELOW H AS BEEN DRAFTE D TO COMPLY W ITH FED ERAL LAW APPLICAB LE TO PR ODUC TS MANUFAC TURED AFTER JULY 4, 1975. This wa rrant y is e xtended o nly t o the or igin al pu rchaser who pur chases t his prod uct when new and unused from I nvacar e or a dealer.
デバイスInvacare MA50の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Invacare MA50をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはInvacare MA50の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Invacare MA50の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Invacare MA50で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Invacare MA50を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はInvacare MA50の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Invacare MA50に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちInvacare MA50デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。