Leica Lamp housing and mercury burner lamp User Manual.
2 Leica lamp housing and mercury burner T able of contents Page 1. Safety concept . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 1.1 Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
1.1 Description 1.2 Safety precautions For the Leica MZ16 FA 3 Leica lamp housing and mercury burner 1. Safety concept The Leica 106 z lamp housing is the light source for the fluorescence stereomicroscope Leica MZ16 F A as well as the fluorescence module for Leica stereomicroscopes.
Lamp housing The Leica MZ16 F A fluorescence stereomicroscope and the fluorescence module for Leica stereomicroscopes may only be used in combination with the Leica lamp housing for 100W or 50W mercury burners, their corresponding supply units and stray-light protection.
Supply units The 100W mercury burners are powered by the ebq 100 supply unit. Read the separate instruction manual provided with this supply unit. The supply unit for 50W mercury burners is described in this manual. For the ebq supply unit: only fuses of the specified type with the listed rated current may be used.
• Use the stray-light protection. Danger to eyes from UV radiation. • During assembly work on xenon burners, always wear the protective gloves and face protection provided. Risk of explosion • Electrical components must be assembled at least 10 cm away from the wall and from flammable substances.
The Leica Microsystems 106 z lamp housing is used. If the 105 z lamp housing model is available, it can also be used. Follow the safety instructions! The following gas discharge lamps can be used. They require different types of lamp mounts. Order No.
Unscrew the 2 hexagon-head screws. Carefully pull out the base with lamp mount and connecting cable. Unscrew the top Phillips screw . Pull up the cooling element and pull it to the side. Unscrew the bottom clamping system. Remove the transport anchor and store it for later transport.
Adjust the collector on a trial basis using the focusing knob. • During this process, the collector may not come into contact with the flexible power supply . If necessary , bend the power supply away from the collector . Carefully close the lamp housing and make sure that the contact plug is set firmly in the socket.
Unscrew the hexagon-head screw on the adapter part using an Allen key . Insert the lamp housing into the adapter part and tighten the hexagon-head screw .
Pull the focusing knob outward. Carefully replace the collector and push it to the left until the focusing knob comes to a stop in the guide groove. Adjust the collector on a trial basis. Reinstall the lamp base and close the lamp housing.
4.1 Supply unit for lamp housing 106 z with 50W mercury lamp 12 Leica lamp housing and mercury burner T ype 301-185.003/10372 Input: 220V/240V ±10% 220V ~ 50/60Hz Hg 50W L1/L2 Osram No. 0-958 P . Max. 320V A 1. On/off switch 2. Operating time meter 3.
T o minimize the risk of explosion, replace a discolored high-pressure mercury burner promptly when it has reached the end of its rated life. Follow the manufacturer’ s specifications and the minute meter on the supply unit! Switch on the supply unit.
14 Leica lamp housing and mercury burner After installing the mercury burner , adjust it exactly as de- scribed. Never let the bright tips of the electric arcs, the focal spots, cross when projected. Danger of explosion from overheating Adjustment is also important to obtain a homogenous light surface and good fluorescence.
15 Leica lamp housing and mercury burner 3 2 1 5 4 6 Hg 50W Hg 100W Fig. a: Focus the arc. Fig. b: Position the arc at the crosshairs. Fig. c: Focus and position the second arc.
Leica Microsystems – the brand for outstanding products Leica Microsystems – an international company with a strong network of customer services Australia: Gladesville, NSW T e l. +1 800 625 286 Fax +61 2 9817 8358 Austria: Vienna T e l. +43 1 486 80 50 0 Fax +43 1 486 80 50 30 Canada: Richmond Hill/Ontario Tel.
デバイスLeica 106 Zの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Leica 106 Zをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはLeica 106 Zの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Leica 106 Zの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Leica 106 Zで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Leica 106 Zを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はLeica 106 Zの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Leica 106 Zに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちLeica 106 Zデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。