NI PXI/PCI-5124 Specifications 12-Bit 200 MS/s Digitizer This document lists the sp ecifications for the NI PX I/PCI-5124 (NI 5124) high-speed digitizer.
NI PXI/PCI-5124 Specifications 2 ni.com PFI 0 and PFI 1 (Programmabl e Function Interfac e, AUX Front Panel Connector) ................... .................... ................... .....20 TClk Specifications .............. .................... .......
© National Instruments Corporat ion 3 NI PXI/PCI-5 124 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable V ertical Offset 50 Ω 1M Ω — Range (V pk-pk ) V ertical Of fset Range (V) Range (V pk-pk ) V ertical Offs et Range (V) 0.
NI PXI/PCI-5124 Specifications 4 ni.com DC Drift Range (V pk-pk ) 50 Ω and 1 M Ω — 0.2, 0.4, 1, and 2 ±(0.057% of Input + 0.006% of FS +1 0 0 μ V) per °C 4, 10, and 20 (1 M Ω only) ±(0.057% of Input + 0.006% of FS + 900 µV) per °C A C Amplitude Accuracy 50 Ω 1M Ω W ithin ±5 °C of self-calibration temperature ±0.
© National Instruments Corporat ion 5 NI PXI/PCI-5 124 Specifications Figure 1. NI 5124 Frequency Response (T ypical) A C Coupling Cutof f (–3 dB) 12 Hz A C coupling av ailable on 1M Ω only Passband Flatness Filter Settings Range (V pk-pk ) 50 Ω and 1 M Ω Referenced to 50 kHz Filters Off All ranges except 0.
NI PXI/PCI-5124 Specifications 6 ni.com Specif ication Va l u e Comments Spectral Characteristics Spurious-Free Dynamic Range with Harmonics (SFDR) (Physical), T ypical Range (V pk-pk ) 50 Ω 1M Ω Fi.
© National Instruments Corporat ion 7 NI PXI/PCI-5 124 Specifications Intermodulation Distortion, T ypical 0.2 V pk-pk to 2.0 V pk-pk Ranges on 50 Ω Input Filters of f or antialias filter on T wo tones at 10.
NI PXI/PCI-5124 Specifications 8 ni.com Figure 2. NI 5124 Dynamic Performance, 50 Ω , 1 V pk-pk Range, 262,144 Point FF T (T ypical) –110 –100 –90 –80 –70 –60 –50 –40 –30 –20 –10 0 Amplitude (dBFS) 10 M 20 M 30 M 40 M 50 M 60 M 70 M 80 M 90 M 100 M 1 M F requency (Hz) 9.
© National Instruments Corporat ion 9 NI PXI/PCI-5 124 Specifications Specif ication Va l u e Comments RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Noise filter on 50 Ω terminator connected to input 0.2 NI PXI-5124: 94 μ V rms (0.047% FS) NI PCI-5124: 106 μ V rms (0.
NI PXI/PCI-5124 Specifications 10 ni.com RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Antiali as fil t e r on 50 Ω terminator connected to input 0.2 NI PXI-5124: 112 μ V rms (0.056% FS) NI PCI-5124: 126 μ V rms (0.063% FS) NI PXI-5124 : 130 μ V rms (0.065% FS) NI PCI-5124: 146 μ V rms (0.
© National Instruments Corporat ion 11 NI PXI/PCI-5 124 Specifications RMS Noise Range (V pk-pk ) 50 Ω 1M Ω Filters off 50 Ω terminator connected to input 0.2 NI PXI-5124: 114 μ V rms (0.057% FS) NI PCI-5124: 128 μ V rms (0.064% FS) NI PXI-5124 : 164 μ V rms (0.
NI PXI/PCI-5124 Specifications 12 ni.com Figure 3. Representation of NI 5124 Spectral Noise Density on 0.2 V Range, Noise Filter Enabled, 1 M Ω Input Impedance Figure 4.
© National Instruments Corporat ion 13 NI PXI/PCI-5 124 Specifications Horizontal Sample Clock Specif ication Va l u e Comments Sources NI PXI-5124 NI PCI-5124 * Internal Sample Clock is locked to th.
NI PXI/PCI-5124 Specifications 14 ni.com Sample Clock Jitter , T ypical ≤ 1 ps rms (10 0 Hz to 100 kHz) ≤ 2p s r m s ( 1 0 0 H z t o 1M H z ) Includes the ef fects of the con verter aperture uncer.
© National Instruments Corporat ion 15 NI PXI/PCI-5 124 Specifications External Sample Clock Sources NI PXI-5124 NI PCI-5124 — CLK IN (front panel SMB connector) PXI Star Trigger (backplane connect.
NI PXI/PCI-5124 Specifications 16 ni.com Phase-Locked Loo p (PLL) Reference Clock Sample Clock Exporting Exported Sample Clock Destinations Destination Maximum Frequency * Decimated Sample Clock only CLK OUT (front panel SMB connector) 210 MHz PXI_T rig<0.
© National Instruments Corporat ion 17 NI PXI/PCI-5 124 Specifications CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) CLK OUT (Sample Clock and Reference Clock Output, Front Panel Connector) T rigger Reference (Stop) T rigger Specification Va l u e Comments Input V oltage Range Sine wav e: 0.
NI PXI/PCI-5124 Specifications 18 ni.com Ti m e Resolution TDC Onboard Clock External Clock TDC = T ime to Digita l Con version Circuit On 50 ps N/A Of f 5n s External Clock Period Rearm T ime TDC Rearm T ime Ho ldoff s et to 0 On 10 μ s Of f 2 μ s Holdoff TDC Onboard Clock External Clock — On 10 μ s to 85.
© National Instruments Corporat ion 19 NI PXI/PCI-5 124 Specifications TRIG (External T rigger , Front Panel Connector) T rigger Filters Lo w-Frequenc y (LF) Reject High-Frequency (HF) Reject — 50 kHz 50 kHz Digital Trigger (Digital Trig ger Type) Sources NI PXI-5124 NI PCI-5124 — PXI_T rig<0.
NI PXI/PCI-5124 Specifications 20 ni.com PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connector) Specif ication Va l u e Co mments Connector 9-pin mini-circu lar DIN — Direction.
© National Instruments Corporat ion 21 NI PXI/PCI-5 124 Specifications TClk Specifications National Instruments TClk synchronizati on method and the NI-TClk driver are used to align the sample clocks on any numb er of SMC-based modules in a chassis.
NI PXI/PCI-5124 Specifications 22 ni.com W aveform Specifications Specification Va l u e Comments Onboard Memory Size 8 MB per channel stan dard (4 megasamples per channel) * NI PXI-5124 only 32 MB pe.
© National Instruments Corporat ion 23 NI PXI/PCI-5 124 Specifications Calibration Power Specification Va l u e Comments Self-Calibrati on Self-calibration i s done on soft ware command. The calibration corrects for gain, offset, frequency response, triggering, and timing adjust ment errors for all input ranges.
NI PXI/PCI-5124 Specifications 24 ni.com Software Specif ication Va l u e Comments Dri ver Software NI-SCOPE 2.7 or later NI-SCOPE is an IVI-comp liant driv er that allows you to configure, control, and calib rate the NI 5124. NI-SCOPE provides application program ming interf aces for many dev elopment environments.
© National Instruments Corporat ion 25 NI PXI/PCI-5 124 Specifications Environment NI PXI-5124 Note T o ensure that the NI PXI-5124 cools ef fectiv ely , follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5124 kit.
NI PXI/PCI-5124 Specifications 26 ni.com NI PCI-5124 Note T o ensure that the NI PCI-5124 cools effecti v ely , make sure that the chassis in which it is used has activ e cooling that provides at least some airflow acros s the PCI card cage. T o maximize airflo w and extend the life of the de vice, leav e any adjacent PCI slots empty .
© National Instruments Corporat ion 27 NI PXI/PCI-5 124 Specifications Safety , Electromagnetic Compatibility , and CE Compliance Safety This product is designed to meet the requirements of the follo.
NI PXI/PCI-5124 Specifications 28 ni.com Environmental Management National Instruments i s committed to designing and manu facturing products in an environmental ly responsible mann er. NI recognizes that eliminating certa in hazardous substances from our products is beneficial not only to t he environment but also to NI customers.
© National Instruments Corporat ion 29 NI PXI/PCI-5 124 Specifications Dimensions and W eight NI PXI-5124 Front Panel I ndicators Label Function Fo r m or e information, refer to the NI High-Speed Digitizers Help . A CCESS The A CCESS LED indicates the status of the PCI b us and the interface from the NI PXI-5124 to the co ntroller .
NI PXI/PCI-5124 Specifications 30 ni.com NI PCI-5124 Dimensions 35.5 × 2.0 × 11.3 cm (14.0 × 0.8 × 4.4 in.) We i g h t 455 g (16 oz) 2.0 cm (0.8 in.
© National Instruments Corporat ion 31 NI PXI/PCI-5 124 Specifications Where to Go for Support The National Instrum ents W eb site is your complete resource for technical support.
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more inform ation about Natio nal Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s.
デバイスNational Instruments 12-Bit 200 MS/s Digitizerの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
National Instruments 12-Bit 200 MS/s Digitizerをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはNational Instruments 12-Bit 200 MS/s Digitizerの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。National Instruments 12-Bit 200 MS/s Digitizerの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。National Instruments 12-Bit 200 MS/s Digitizerで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
National Instruments 12-Bit 200 MS/s Digitizerを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はNational Instruments 12-Bit 200 MS/s Digitizerの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、National Instruments 12-Bit 200 MS/s Digitizerに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちNational Instruments 12-Bit 200 MS/s Digitizerデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。