OPERATING INSTRUCTIONS AND S PECIFICATIONS NI 9215 4-Channel, ±10 V, 16-Bit Simultan eous Analog Input Module ni.com/manuals Deutsch Français.
NI 9215 Operating I nstruction s and Specificati ons 2 ni.com This doc u ment describes ho w to u se the National Instr u ments 9 215 and incl u des specif ications and term inal assignments for the NI 9215. In this doc u ment, the NI 9215 with scre w terminal and NI 9215 with BNC are referred to incl u si v ely as the NI 9215.
© National Inst ruments Corp. 3 NI 9215 Operating Instructi ons and Specific ations Hot Surface This icon denotes that the component may be hot. T o u ching this component may res u lt in bodily inj u ry . Safety Guidelines for Hazardous V oltages Yo u can connect hazardo u s v oltages only to the NI 9215 with scre w terminal.
NI 9215 Operating I nstruction s and Specificati ons 4 ni.com and circ u its connected to the mod u le are properly ins u lated from h u man contact. Y o u m u st u se the NI 99 32 connector backshell kit to ens u re that the terminals are not accessible.
© National Inst ruments Corp. 5 NI 9215 Operating Instructi ons and Specific ations Safety Guidelines fo r Hazardous Loca tions The NI 9215 is s u itable for u se in Class I, Division 2, Gro u ps A, B, C, D, and T4 hazardo u s locations; Class I, Zone 2, AEx nC IIC T4 and Ex nC IIC T4 hazardo u s locations; and n onhazardo u s locations only .
NI 9215 Operating I nstruction s and Specificati ons 6 ni.com Caution For Zone 2 applications, connected signals m u st be within the following limit: Capacitance .
© National Inst ruments Corp. 7 NI 9215 Operating Instructi ons and Specific ations Caution T o meet radio freq u enc y emission req u irements for marine applications, u se shielded cables and install the system in a metal enclos u re.
NI 9215 Operating I nstruction s and Specificati ons 8 ni.com Connecting the NI 9215 The NI 9215 pro vides connections for fo u r dif ferential ana log inp u t channels.
© National Inst ruments Corp. 9 NI 9215 Operating Instructi ons and Specific ations Figure 3. Connector Assignments of the NI 9215 with BNC AI 3 + AI 3 – AI2+ AI2– AI0+ AI0– AI1+ AI1–.
NI 9215 Operating I nstruction s and Specificati ons 10 ni.com The NI 9215 with screw terminal has a 10-terminal detachable scre w-terminal connector .
© National Inst ruments Corp. 11 NI 9215 Operating Instructi ons and Specific ations Connecting Differen tial Voltage Signal s to the NI 9215 Yo u can connect gro u nded or floating dif feren tial signals to th e NI 9215. Connect the positi ve voltage signal to AI+ and the negati ve voltage signal to AI–.
NI 9215 Operating I nstruction s and Specificati ons 12 ni.com T o connect floating dif f erential signals to the NI 9215 with screw terminal, yo u m u st connect the nega tiv e lead of the signal to COM thro u gh a 1 M Ω resistor to keep the v oltage so u rce within the common-mode voltage ra nge, as sho wn in Fig u re 5.
© National Inst ruments Corp. 13 NI 9215 Operating Instructi ons and Specific ations Connecting Single-En ded V oltage Signals to the NI 9215 T o connect single-ende d vo ltage signals to the NI 9215.
NI 9215 Operating I nstruction s and Specificati ons 14 ni.com Wiring for High Vibrati on Applications If an application u sing the NI 9215 with scre w terminal is s u bject to high vibration, Nationa.
© National Inst ruments Corp. 15 NI 9215 Operating Instructi ons and Specific ations NI 9215 Circuitr y The NI 9215 channels share a common gro u nd that is isolated from other mod u les in the system. The NI 9215 protects each channel from ov erv oltages.
NI 9215 Operating I nstruction s and Specificati ons 16 ni.com Figure 8. Input Cir c uitry for One Channel on the NI 9215 with Screw T erminal COM AI + AI – NI 9215 with Screw T erminal Isolated ADC.
© National Inst ruments Corp. 17 NI 9215 Operating Instructi ons and Specific ations The NI 9215 with BNC has a resistor that ens u res the inp u t voltage does not drift o u tside of the common-mode range.
NI 9215 Operating I nstruction s and Specificati ons 18 ni.com Sleep Mode This mod u le s u pports a low-po wer sleep mode. S u pport for sleep mode at the system le vel depe nds on the chassis that the mod u le is pl u gged into. Refer to the chassis man u al for information abo u t s u pport for sleep mode.
© National Inst ruments Corp. 19 NI 9215 Operating Instructi ons and Specific ations Specifications The follo wing specifications are typical for the range – 40 to 70 °C u nless otherwise noted. Input Characteristics N u mber of channels .........
NI 9215 Operating I nstruction s and Specificati ons 20 ni.com Inp u t voltage ranges 1 1 Refer to the Safety Guideli nes section for more information abo u t safe oper ating voltages. Measurement V oltage, AI+ to AI– Maximum V oltage (Signal + Common Mode) Minimum* (V) Ty p i c a l (V) Maximum (V) Screw Te r m i n a l BNC ±10.
© National Inst ruments Corp. 21 NI 9215 Operating Instructi ons and Specific ations Overv oltage protecti on ................ ..... ± 30 V Con version time Channel 0 only .......... ............ ..... 4 .4 μ s Channels 0 and 1 ............ .......
NI 9215 Operating I nstruction s and Specificati ons 22 ni.com CMRR ( f in = 60 Hz) ................ .......... –73 dB min Inp u t bandwidth (–3 dB) ..... ............. . 420 kHz min Inp u t impedance Resistance NI 9215 with screw terminal .......
© National Inst ruments Corp. 23 NI 9215 Operating Instructi ons and Specific ations NI 9215 with BNC 10 V step ....... ........... ......... ... 25 μ s 20 V step ....... ........... ......... ... 35 μ s No missing code s ....... ........... ......
NI 9215 Operating I nstruction s and Specificati ons 24 ni.com Thermal dissipation (at 70 °C) Active mode .............. ......... ........ 560 mW max Sleep mode . ......... ........... ......... ... 25 μ W ma x Physical Characteristics If yo u need to clean the mod u l e, wipe it with a dry to wel.
© National Inst ruments Corp. 25 NI 9215 Operating Instructi ons and Specific ations Safety NI 9215 with Screw T erminal Safety Voltages Connect only v oltages that are within the follo wing lim its. Channel-to-COM ........... ........... ........ ± 30 V max Isolation Channel-to-channel .
NI 9215 Operating I nstruction s and Specificati ons 26 ni.com Caution Do not connect the NI 9215 with scre w terminal to signals or u se for meas u rements within Meas u rement Categories III or IV . NI 9215 with BNC Safety Voltages Connect only v oltages that are within the follo wing lim its.
© National Inst ruments Corp. 27 NI 9215 Operating Instructi ons and Specific ations eq u ipment, limited-ene rg y parts of eq u ipment , circ u its po wered by reg u lated lo w-voltage so u rces, and electronics. Caution Do not connect the NI 9215 with BNC to signals or u se for meas u rements within Meas u rement Categories II, III, or IV .
NI 9215 Operating I nstruction s and Specificati ons 28 ni.com Hazardous Locations U.S. (UL) .............. ......... ........... ........ Class I, Division 2, Gro u ps A, B, C, D, T4; Class I, Zone 2, AEx nC IIC T4 Canada (C-UL) ... ............. ...
© National Inst ruments Corp. 29 NI 9215 Operating Instructi ons and Specific ations Operating h u midity (IEC 60068-2-56)....... ........... ............ 10 to 90% RH, noncondensing Storage h u midity (IEC 60068-2-56)....... ........... ............
NI 9215 Operating I nstruction s and Specificati ons 30 ni.com Electromagnetic Compatibility This prod u ct is designed to meet the req u ire ments of the follo wing standards of EMC for electrical eq.
© National Inst ruments Corp. 31 NI 9215 Operating Instructi ons and Specific ations ni.com/certification , search by mod u le n u mber or prod u ct line, and click the appropriate link in the Certification col u mn.
NI 9215 Operating I nstruction s and Specificati ons 32 ni.com Calibration Yo u can obtain the calibration cert ificate and information abo u t calibration services for th e NI 9215 at ni.com/calibration . Calibration interval ...... ........... .....
© National Inst ruments Corp. 33 NI 9215 Operating Instructi ons and Specific ations National Instr u ments also has off ices located aro u nd the world to help address yo u r s u pport needs. For telephone s u pport in the United States, c reate yo u r ser vice req u est at ni.
© 2004–2007 Nati onal Instruments Corp. All rights reserved. 373779E-01 Dec07 National Instruments, NI, n i.com, and LabVIEW are trademarks of National Instrument s Corporation. Refer to the Terms of Use section on ni.com/legal for more information about Natio nal Instruments trademarks.
デバイスNational Instruments NI 9215の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
National Instruments NI 9215をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはNational Instruments NI 9215の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。National Instruments NI 9215の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。National Instruments NI 9215で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
National Instruments NI 9215を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はNational Instruments NI 9215の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、National Instruments NI 9215に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちNational Instruments NI 9215デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。