OPERATING INSTRUCTIONS NI 9219 4-Channel, 24-Bit, Universal Analo g Input Module.
NI 9219 Operating Instructions 2 ni.com These operating instructions desc ribe ho w to use the National Instruments 9219. For information about installing, conf iguring, and programming the syst em, refer to the system documentation.
© National Instruments Corp. 3 NI 9219 Operating Instructions Safety Guidelines for Hazardous V oltages If hazardous voltages are connect ed to the module, take the follow ing precautions. A hazardous voltage is a voltage greater than 42.4 V pk or 60 VDC to earth ground.
NI 9219 Operating Instructions 4 ni.com Figure 1 sho ws the NI 9972 connector backshell. Figure 1. NI 9972 Connector Backshell Safety Guidelines fo r Hazardous Loca tions The NI 9219 is suitable for u.
© National Instruments Corp. 5 NI 9219 Operating Instructions Caution Do not remove modules unle ss power has been switched off or the area is known to be nonhazardous. Caution Substitution of component s may impair suitability for Class I, Division 2.
NI 9219 Operating Instructions 6 ni.com Wiring the NI 9219 The NI 9219 has four 6-terminal connectors that pro vide connections for four analog input channels. Connect the positi ve signal of the signal source to the positive input signal terminal (HI) and the negati ve signal of the signa l source to the negati ve input signal terminal (LO).
© National Instruments Corp. 7 NI 9219 Operating Instructions Table 1. Signal N ames Module T erminal Signal Name Signal Description 1 T+ TEDS Data 2 T– TEDS COM 3 EX+/HI * Pos itive e xcitation or.
NI 9219 Operating Instructions 8 ni.com Table 2. Terminal Assignment s Mode T erminal 123456 V oltage T+ T – — HI LO — Current T+ T – HI — LO — 4-W ire Resistance T+ T – EX+ HI EX – LO.
© National Instruments Corp. 9 NI 9219 Operating Instructions Connecting Wires to the NI 9219 Connectors Use a flathead scre wdriv er with a blade smaller than 2.3 × 1.0 mm (0.09 × 0.04 in.) to connect wires to the detachable spring-terminal connectors.
NI 9219 Operating Instructions 10 ni.com Wiring TEDS Cha nnels The NI 9219 supports only Class II TEDS sensors. Connect th e two TEDS lin es to TEDS Data (T+) and TEDS COM (T–) and ensure that neither T+ nor T– is tied in common to an y of the signal inputs (terminals 3 through 6) on the NI 9219.
© National Instruments Corp. 11 NI 9219 Operating Instructions Figure 3. Connecting a G rounded Signal Sour ce to the NI 9219 Figure 4. Connecting a Floating Signal Source to the NI 9219 + – T wist.
NI 9219 Operating Instructions 12 ni.com NI 9219 Timing Options The NI 9219 supports four dif f erent timing options that are optimized for dif ferent types of applications by using dif ferent ADC con version times.
© National Instruments Corp. 13 NI 9219 Operating Instructions Figure 5. Input Circuitry for One Channel Volta ge and Current Modes In V oltage and Current modes, c onnect t he signal source to the NI 9219 across the HI and LO terminals. The current is computed from the v oltage that the ADC measures across an interna l shunt resistor .
NI 9219 Operating Instructions 14 ni.com 4-Wire Resistance and 4-Wire RTD Modes 4-W ire Resistance and 4-W ire R T D modes source a current, which va ries based on the resistance of the load, between the EX+ and EX– terminals. The me asured resistance is computed from the resulting v oltage readi ng.
© National Instruments Corp. 15 NI 9219 Operating Instructions lead wires have the same resistance . A gain of 2x is applied to the volta ge across the neg ativ e lead wire and the ADC uses this v oltage as the negati ve reference to canc el the r esistance error across the positiv e lead wire.
NI 9219 Operating Instructions 16 ni.com for lead wire resistance. Refer to Figure 9 for an illustration of the connections. Figure 9. Connections in 2-Wire Resistan ce and Quarter-Bridge Modes Half-B.
© National Instruments Corp. 17 NI 9219 Operating Instructions Figure 10. Connections in Half-Br idg e and Full-Bridge Modes Thermocouple Mode In Thermocouple mode, connect the positi ve end of the thermocouple to HI and the ne gati ve end of the thermocouple to LO.
NI 9219 Operating Instructions 18 ni.com en vironment and av oid placing heat sources near the module or its connectors. Refer to the Specifications section for more information about accuracy . Th e NI 9219 does not support open thermocouple detection.
© National Instruments Corp. 19 NI 9219 Operating Instructions circuit is open, make sure no more than ±60 V is sourced across the switch. Open Contact mode is supported only in CompactRIO systems. Refer to Figure 12 for an illustration of the connections.
NI 9219 Operating Instructions 20 ni.com Binary Data NI 9219 modules in the system retu rn calibrated binary data when used in a CompactRIO chassis. For these modules, you can con vert the data to engineering units in software. Refer to the software documentation for inform ation about con verti ng data.
© National Instruments Corp. 21 NI 9219 Operating Instructions Input Characteristics Number of channels ............ ............. . 4 analog i nput channels ADC resolution .............. .............. ..... 24 bits T ype of ADC ........... ........
NI 9219 Operating Instructions 22 ni.com Con version time, no ch annels in TC mode High speed .. ............. ........... ........ 10 m s for all channe ls Best 60 Hz rejection ......... .......... 110 ms for all channels Best 50 Hz rejection .......
© National Instruments Corp. 23 NI 9219 Operating Instructions Input impedance V oltage and Digital In modes ( ±60 V , ±15 V , ±4 V ) ........... .......... 1 M Ω Current mode .... .............. ........... . < 40 Ω All other modes ..... ...
NI 9219 Operating Instructions 24 ni.com Cold-junction compensa tion sensor accuracy ....... .............. ........... . ±1°C typ 4-W ire and 3-W ire R TD, Pt 1000 ±0.1, ±0.5 ±240, ±640 4-W ire and 3-W ire R TD, Pt 100 ±0.1, ±0.5 ±2400, ±6400 Quarter-Bridge, 350 Ω ±0.
© National Instruments Corp. 25 NI 9219 Operating Instructions Stability Mode, Range Gain Drift (ppm of Reading/°C) Offset Drift (ppm of Range/°C) V oltage, ± 60 V ±20 ±0.2 V oltage, ± 15 V ±20 ±0.8 V oltage, ± 4 V ±20 ±3.2 V oltage, ± 1 V ±10 ±0.
NI 9219 Operating Instructions 26 ni.com Input noise in ppm of Range rms Full-Bridge, ± 62.5 mV/V ±3 ±20 Full-Bridge, ± 7.8 mV/V ±3 ±20 Mode, Range Con version Time High speed Best 60 Hz rejection Best 50 Hz rejection High res o - lution V oltage, ±60 V 7.
© National Instruments Corp. 27 NI 9219 Operating Instructions 4-W ire and 2-W ire Resistance, 10 k Ω 4.1 1.3 0.8 0.3 4-W ire and 2-W ire Resistance, 1 k Ω 7.1 1.8 1.2 0.7 4-W ire and 3-W ire R TD, Pt 1000 7.6 1.7 1.1 0.4 4-W ire and 3-W ire R TD, Pt 100 10.
NI 9219 Operating Instructions 28 ni.com Input bias curre nt ................ ........... ... <1 nA INL ........... ........... ........... .............. ..... ±15 ppm CMRR ( f in = 60 Hz) ............ ........... ... >100 dB NMRR Best 60 Hz rej ection .
© National Instruments Corp. 29 NI 9219 Operating Instructions Excitation le vel for Resistance , R TD, and Quarter-Bridge modes MTBF .......... .....
NI 9219 Operating Instructions 30 ni.com Thermal dissipation (at 70 °C) Activ e mode .............. ............ ..... 625 mW max Sleep mode . ........... ........... .......... 25 μ W max Physical Characteristics If you need to clean the module, wipe it with a dry towel.
© National Instruments Corp. 31 NI 9219 Operating Instructions Channel-to-earth ground Continuous .... ........... ............ 250 V AC, Measurement Category II W ithstand .
NI 9219 Operating Instructions 32 ni.com by module number or product line, and clic k the appropriate link in the Certification column. Hazardous Locations U.
© National Instruments Corp. 33 NI 9219 Operating Instructions Maximum altitude.......... ......... .......... 2,000 m Pollution Degree (IEC 60664) .......... 2 Shock and Vibration T o meet these specif ications, you must panel mount the system. Operating vibration Random (IEC 60068-2-34).
NI 9219 Operating Instructions 34 ni.com Note For EMC compliance, operate this de vice with shielded cabling. CE Compliance This product meets the essentia l requirements of applicable European direct.
© National Instruments Corp. 35 NI 9219 Operating Instructions Calibration Y ou can obtain the calibration cert ificate and information about calibration services for th e NI 9219 at ni.com/calibration . Calibration interval .... ............. ......
© 2007 National I nstruments Corp. All rights reserved. 374473D-01 Jun07 National Instruments, NI, n i.com, and LabVIEW are trademarks of National Instruments Corporation. Refer to the Terms of Use section on ni.com/legal for more information about Nat ional Instruments trademarks.
デバイスNational Instruments NI 9219の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
National Instruments NI 9219をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはNational Instruments NI 9219の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。National Instruments NI 9219の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。National Instruments NI 9219で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
National Instruments NI 9219を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はNational Instruments NI 9219の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、National Instruments NI 9219に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちNational Instruments NI 9219デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。