National InstrumentsメーカーNI PCI-5114の使用説明書/サービス説明書
ページ先へ移動 of 23
NI PXI/PCI-5114 Specifications 8-Bit 250 MS/s Digitizer This document lists the sp ecifications for the NI PX I/PCI-5114 (NI 5114) high-speed digitizer.
NI PXI/PCI-5114 Specifications 2 ni.com Waveform Specifications ........... .................... ...................... ...................13 Calibration .............. .................... .................... .................... ...................
© National Instruments Corporat ion 3 NI PXI/PCI-5 114 Specifications Voltage Levels Full Scale (FS) Input Range and Programmable V ertical Of fset 50 Ω 1M Ω — Range (V pk-pk ) V ertical Of fset Range (V) Range (V pk-pk ) V ertical Offset Range (V) 0.
NI PXI/PCI-5114 Specifications 4 ni.com Figure 1. NI 5114 Frequency Response (T ypical) Bandwidth and Transient Response Bandwidth (–3 dB) Range (V pk-pk ) Bandwidth Rise/F all T ime, T ypical — All ranges except 0.04 125 MHz 2.8 ns 0.04 100 MHz 3.
© National Instruments Corporat ion 5 NI PXI/PCI-5 114 Specifications Specif ication Va l u e Commen ts Spectral Characteristics Spurious-Free Dynamic Range with Harmonics (SFDR), T ypical Range (V p.
NI PXI/PCI-5114 Specifications 6 ni.com Horizontal Sample Clock Specification Va l u e Comments Sources Internal, Onboard Clock (i nternal VCXO)* External, CLK IN (front panel SMB con nector) * Intern.
© National Instruments Corporat ion 7 NI PXI/PCI-5 114 Specifications Phase-Locked Loo p (PLL) Reference Clock External Sample Clock Sources CLK IN (front panel SMB connector) — Frequenc y Range 50.
NI PXI/PCI-5114 Specifications 8 ni.com CLK IN (Sample Clock and Reference Clock Input, Front Panel Connector) T r igger Reference (Stop) T rigger Specif ication Va l u e Input V oltage Range Sine wa ve: 0.65 V pk-pk to 2.8 V pk-pk (0 dBm to 13 dBm) Square wa ve: 0.
© National Instruments Corporat ion 9 NI PXI/PCI-5 114 Specifications Analog Trigger (Edge, Window, and Hysteresis Trigger Types) Sources CH 0 (front panel BNC connector) CH 1 (front panel BNC connec.
NI PXI/PCI-5114 Specifications 10 ni.com TRIG (External T rigger , Front Panel Connector) T ypes Specific Line Any Line Specific Field — Standards SDTV : M-NTSC, B/G-P AL, SECAM, M-P AL EDTV : 480i/59.94 fps, 480i /60 fps, 480p/59.94 Fps, 480p/60 Fps, 576i/50 f ps, 576p/50 Fps HDTV : 720p/ 50 Fps, 720p/59.
© National Instruments Corporat ion 11 NI PXI/PCI-5 114 Specifications PFI 0 and PFI 1 (Programmable Function Interface, AUX Front Panel Connectors) Specif ication Va l u e Connector 9-pin mini-circu.
NI PXI/PCI-5114 Specifications 12 ni.com TClk Specifications National Instruments TClk synchronizati on method and the NI-TClk driver are used to align the sample clocks on any numb er of SMC-based modules in a chassis.
© National Instruments Corporat ion 13 NI PXI/PCI-5 114 Specifications W aveform Specifications Specif ication Va l u e Comments Onboard Memory Size 8 MB per Channel Standard 8 megasamples per channe.
NI PXI/PCI-5114 Specifications 14 ni.com Calibration Power Specif ication Va l u e Self-Calibrati on Self-calibratio n is done on software command. The calibration corrects for gain, of fset, compensated 1 M Ω attenuator , triggerin g, and timing adjustment errors for all input ranges.
© National Instruments Corporat ion 15 NI PXI/PCI-5 114 Specifications Software Specification Va l u e Dri ver Software NI PXI-5114 : NI-SCOPE 2.9 or later NI PCI-5114 : NI-SCOPE 3.1 or later NI-SCOPE is a n IVI-compliant dri ver that allo ws you to configure, control, and calibrate the NI 5114.
NI PXI/PCI-5114 Specifications 16 ni.com Environment NI PXI-5114 Note T o ensure that the NI PXI-5114 cools ef fectiv ely , follow the guidelines in the Maintain Forced Air Cooling Note to Users included in the NI PXI-5114 kit. The NI PXI-5114 is intended for indoor use only .
© National Instruments Corporat ion 17 NI PXI/PCI-5 114 Specifications NI PCI-5114 Note T o ensure that the NI PCI-5114 cools effecti v ely , make sure that the chassis in which it is used has activ e cooling that provides at least some airflow acros s the PCI card cage.
NI PXI/PCI-5114 Specifications 18 ni.com Safety , Electromagnetic Compatibility , and CE Compliance Safety This product is designed to meet the requirements of the follow ing standards of safety for e.
© National Instruments Corporat ion 19 NI PXI/PCI-5 114 Specifications Environmental Management National Instruments i s committed to designing and manu facturing products in an environmental ly responsible mann er.
NI PXI/PCI-5114 Specifications 20 ni.com Dimensions and W eight NI PXI-5114 Front Panel Indicators Label Function Fo r mo re information, refer to the NI High-Speed Digitizers Help . A CCESS The A CCESS LED indicates the status of the PCI b us and the interface from the NI PXI-5114 to the controller .
© National Instruments Corporat ion 21 NI PXI/PCI-5 114 Specifications NI PCI-5114 Dimensions 35.5 × 2.0 × 11.3 cm (14.0 × 0.8 × 4.4 in.) We i g h t 421 g (14.
NI PXI/PCI-5114 Specifications 22 ni.com Where to Go for Support The National Instruments W eb site is your complete resource for technical support. At ni.com/support you ha ve access to e verything from troubleshooting and ap plication dev elopment self-help resources to email and phone assistance from NI Application Engin eers.
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for more inform ation about Natio nal Instruments trademarks. Other product and compan y names mentioned herein are trademarks or trade names of their respective companie s.
デバイスNational Instruments NI PCI-5114の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
National Instruments NI PCI-5114をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはNational Instruments NI PCI-5114の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。National Instruments NI PCI-5114の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。National Instruments NI PCI-5114で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
National Instruments NI PCI-5114を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はNational Instruments NI PCI-5114の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、National Instruments NI PCI-5114に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちNational Instruments NI PCI-5114デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。