USER GUIDE SCC-TC Series Thermocouple Input Modules The SCC-TC Series thermocouple input modules, SC C-TC01 and SCC-TC02, accept input signals from B-, E-, J-, K-, N-, R-, S-, and T-type thermocouples. Each module contains thermistor circ uitry po wer ed by 2.
SCC-TC Series Thermocouple Input Modules User Guide 2 ni.com bold Bold text denotes items that you must select in the software, such as menu items and dialog box options. Bold text also denotes parameter nam es. italic Italic text denotes variables, emphasis , a cross-reference, or an introduction to a key concept.
© National Instruments Corpor ation 3 SCC-TC Series Th ermocouple Input Mo dules User Guide ❑ Documentation – One of the following: • SC-2345/2350 User Manual • SCC-68 User Guide – SCC-TC S.
SCC-TC Series Thermocouple Input Modules User Guide 4 ni.com the inputs of the second-stage SCC. An example of dual-stage condit ioning in an SCC-TC0 X followed by an SCC-LP01 lowpass filter module. Sockets J9 to J16 also are av ailable for digital input/output (DIO) conditioning or control .
© National Instruments Corpor ation 5 SCC-TC Series Th ermocouple Input Mo dules User Guide Figure 1. SCC-TC0 X Signal Connections For more information about h ow to conf igure the SCC-TC0 X module usin g NI-D A Qmx, refer to the SCC Quick Start Guid e.
SCC-TC Series Thermocouple Input Modules User Guide 6 ni.com temperature measurements. T o make this con version, complete the following steps: 1. Measure the thermocouple v oltage. a. Read the thermocouple channe l on the E/M Series D A Q device V ESERIES [CH( X )].
© National Instruments Corpor ation 7 SCC-TC Series Th ermocouple Input Mo dules User Guide Detecting Open Thermocouples The SCC-TC0 X contains a 10 M Ω pul l-up resistor that connects to + 5 V to detect open thermocouples. To determine if you have an open thermocouple, check whether the co rresponding E/M Series DA Q device channel is saturated.
SCC-TC Series Thermocouple Input Modules User Guide 8 ni.com Cold-Junction Sensor The cold-junction sensor vol tage output varies from 1.91 to 0.58 V over a 0 to 55 ° C temperature range. Note NI-D A Qmx includes thermist or conv ersion utilities that implement the equations listed below .
© National Instruments Corpor ation 9 SCC-TC Series Th ermocouple Input Mo dules User Guide Calibrating System Offsets To calibrate the offset voltage of the SCC-TC0 X in the system using the E/M Series DAQ device, complete th e following steps. The E/M Series DAQ device must be powered for 30 minutes before you calibrate the offset voltage.
SCC-TC Series Thermocouple Input Modules User Guide 10 ni.com Specifications These specifications are typical at 25 °C unless otherwise stated. Analog Input Number of input channels.......................1 DIFF Input signals ..................... ...
© National Instruments Corpor ation 11 SCC-TC Series Thermocouple Input Modules U ser Guide Measurement Accuracy 1 1 T otal system measurement error fo r operating temper ature within ± 5 ° C of calibration temperature. Incl udes PCI/A T-MIO- 16XE-50 one-year accuracy specification of 0.
SCC-TC Series Thermocouple Input Modules User Guide 12 ni.com Cold-Junction Sensor Cold-junction sensor accuracy .......... ......± 0.4 ° C max from 15 to 35 ° C, ± 0.75 ° C max from 0 to 15 ° C and 35 to 55 ° C Output ............. ...........
© National Instruments Corpor ation 13 SCC-TC Series Thermocouple Input Modules U ser Guide Figure 3. SCC-TC02 Dimensions Weight TC-01 ................. ........... .............. .... 25 g (0.9 oz) TC-02 ................. ........... .............. .
SCC-TC Series Thermocouple Input Modules User Guide 14 ni.com Environmental Operating temperature ................ ............0 to 50 °C Storage temperature ............ .............. ......–20 to 70 °C Humidity ............ .............. .....
© National Instruments Corpor ation 15 SCC-TC Series Thermocouple Input Modules U ser Guide CE Compliance This product meets the essential re quirements of applicable European Directives, as amended .
National Instruments, NI, ni.com, and LabVIEW are tradem arks of National Instruments Corporation. Refer to t he Terms of Use section on ni.com/legal for mor e information abou t National Instruments trademarks. Other product and compan y names mentioned herein are trademarks or tr ade names of their respective companie s .
デバイスNational Instruments SCC-TC02の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
National Instruments SCC-TC02をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはNational Instruments SCC-TC02の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。National Instruments SCC-TC02の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。National Instruments SCC-TC02で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
National Instruments SCC-TC02を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はNational Instruments SCC-TC02の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、National Instruments SCC-TC02に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちNational Instruments SCC-TC02デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。