Teledyneメーカー212Rの使用説明書/サービス説明書
ページ先へ移動 of 28
Teledyne Analytical Instruments 1 Thermal Conductivity Analyzer Model 212R I NSTR UCTION M ANU AL M ODEL 212R T HERMAL C ONDUCTIVITY A NAL YZER HIGHLY TOXIC AND OR FLAMMABLE LIQUIDS OR GASES MAY BE PRESENT IN THIS MONITORING SYSTEM. PERSONAL PROTECTIVE EQUIPMENT MAY BE REQUIRED WHEN SERVICING THIS SYSTEM.
Teledyne Analytical Instruments 2 Thermal Conductivity Analyzer Model 212R All Rights Reserved. No part of this manual may be reproduced, transmitted, transcribed, stored in a retrieval system, or tra.
Teledyne Analytical Instruments 3 Thermal Conductivity Analyzer Model 212R T ab le of Contents 1 Introduction 1.1 Method of Analysis ....................................................... 5 1. 2 Sensitivity ...........................................
Teledyne Analytical Instruments 4 Thermal Conductivity Analyzer Model 212R 4. 3 Standardization ............................................................ 16 5 P eriodic Maintenance 5. 1 Heater F an ..................................................
Teledyne Analytical Instruments 5 Thermal Conductivity Analyzer Model 212R 1 . INTRODUCTION 1.1 Method of Analysis. The Model 212R compares the thermal conductivity of a sample gas to that of a fixed composition reference gas and produces an electrical output signal that is calibrated to represent the difference between the two gases.
Teledyne Analytical Instruments 6 Thermal Conductivity Analyzer Model 212R 1.2 Sensitivity. Differences in thermal conductivity that produce a measuring bridge error signal of 0.
Teledyne Analytical Instruments 7 Thermal Conductivity Analyzer Model 212R panel, are located immediately behind a hinged front access door. The analyzer is suitable for installation in a sheltered non-hazardous area. A recording and/or indicating device will be required to transduce the electrical output signal into readable information.
Teledyne Analytical Instruments 8 Thermal Conductivity Analyzer Model 212R IMPORTANT: The foam lining the interior of the analyzer is an integral part of the environmental temperature control system. Removal or destruction of this lining will result in erratic instrument perfor- mance.
Teledyne Analytical Instruments 9 Thermal Conductivity Analyzer Model 212R Run the cable attached to the transformer unit in through the access hole that is equipped with the cable clamp and terminate it as specified on the interconnection diagram. 2.
Teledyne Analytical Instruments 10 Thermal Conductivity Analyzer Model 212R cause a large fluctuation in output signal. For best results, use electropolished SS tubing Note #4 : Because of the diffusi.
Teledyne Analytical Instruments 11 Thermal Conductivity Analyzer Model 212R 2.3.4 Installation of Cylinder Supplies. The reference and standard gas cylinders should be installed as close to the analyzer as possible. Each cylinder should be provided with a dual stage, metallic diaphragm, pressure reducing regulator.
Teledyne Analytical Instruments 12 Thermal Conductivity Analyzer Model 212R 2.3.7 Vent Lines. The sample path, reference path, and an integral bypass system vent from separate ports at the rear of the analyzer. Wherever possible, TAI recommends that the gases be permitted to vent directly to the atmosphere.
Teledyne Analytical Instruments 13 Thermal Conductivity Analyzer Model 212R 4. Determine what the sample pressure is and set the output pressure of the cylinder gas supplies to agree with the sample pressure. This procedure, although not absolutely necessary, will minimize flow adjustments when the sample path is switched to the various inputs.
Teledyne Analytical Instruments 14 Thermal Conductivity Analyzer Model 212R After the analyzer stabilizes adjust the "ZERO" control until the recorder indicates the impurity concentration of the zero gas composition. Be sure that the range switch is on position #1 and that the sample path flowrate is 0.
Teledyne Analytical Instruments 15 Thermal Conductivity Analyzer Model 212R virtually instantaneous. Recovery is a function of how long it takes to purge the sampling system upstream from the cell. Accelerating the sample path flowrate will speed the response.
Teledyne Analytical Instruments 16 Thermal Conductivity Analyzer Model 212R The reference and sample flowrates should be checked daily. The instru- ment is somewhat flow sensitive. Operating on Range #1 a change in flowrate of 0.1 scfh will cause a corresponding change in sensitivity of from 1 to 20% of scale.
Teledyne Analytical Instruments 17 Thermal Conductivity Analyzer Model 212R Due to the delicate balance of the thermal conductivity cell employed in the analyzer, exposure to the atmosphere in an uncontrolled temperature environ- ment is undesirable.
Teledyne Analytical Instruments 18 Thermal Conductivity Analyzer Model 212R Check to see that both the analyzer and recorder are being furnished electrical power. Check to see that the analyzer and recorder fuses are intact. Check all electrical connections outside of the analyzer.
Teledyne Analytical Instruments 19 Thermal Conductivity Analyzer Model 212R 6.2.1 Dynamic Balance Procedure. The measuring bridge circuit (see schematic) incorporates two 10 turn potentiometers that a.
Teledyne Analytical Instruments 20 Thermal Conductivity Analyzer Model 212R 5) Using a screwdriver that has a shaft diameter of 1/8” and a shaft length of at least 6”, turn the “Reactive” balance adjustment all the way counterclockwise (CCW) until the potentiometer stop point is reached.
Teledyne Analytical Instruments 21 Thermal Conductivity Analyzer Model 212R or incorrect instrument installation. The following conditions can produce an erratic output from the analyzer: 1) Installation in an area that is subject to large changes in ambient temperature, direct sunlight, drafts from wind, blowers, or air condi- tioners.
Teledyne Analytical Instruments 22 Thermal Conductivity Analyzer Model 212R Check as previously stated, the analyzers entire sampling system has been checked under pressure with a leak detector.
Teledyne Analytical Instruments 23 Thermal Conductivity Analyzer Model 212R tom must be evident with zero gas flowing. If this symptom is displayed only when the sample is flowing, the problem is related to one (or more) of the five (5) points outlined in section 7.
Teledyne Analytical Instruments 24 Thermal Conductivity Analyzer Model 212R CAUTION: Do not touch the thermistor probe with the iron. The end of the probe is tied and taped to the sample tubing below the cell cover. If the meter reading suddenly drops to zero, remove the soldering iron immediately.
Teledyne Analytical Instruments 25 Thermal Conductivity Analyzer Model 212R 7. 2 Output Signal: Output coincides with 100% analysis. 4 m A = 0 ppm N 2 in H 2 20mA = 50,000 ppm N 2 in H 2 - Range #1 = 5000 ppm N 2 in H 2 - Range #2 = 500 ppm N 2 in H 2 - Range #3 7.
Teledyne Analytical Instruments 26 Thermal Conductivity Analyzer Model 212R 7.4.3 Span Gas. Containing an equivalent impurity. Again, the span gas must be procured from a supplier who will certify its composition. The range switch must be in range 3 when the span gas is introduced.
Teledyne Analytical Instruments 27 Thermal Conductivity Analyzer Model 212R Orders should be sent to: TELEDYNE Analytical Instruments 16830 Chestnut Street City of Industry, CA 91749-1580 Phone (626) 934-1500, Fax (626) 961-2538 TWX (910) 584-1887 TDYANYL COID Web: www.
Teledyne Analytical Instruments 28 Thermal Conductivity Analyzer Model 212R APPENDIX Specifications : Ranges: Three decade ranges: Minimum Full scale(ppm) 20 ppm Hydrogen balance Argon 25 ppm Hydrogen.
デバイスTeledyne 212Rの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Teledyne 212Rをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTeledyne 212Rの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Teledyne 212Rの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Teledyne 212Rで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Teledyne 212Rを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTeledyne 212Rの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Teledyne 212Rに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTeledyne 212Rデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。