Teledyneメーカー633の使用説明書/サービス説明書
ページ先へ移動 of 75
Model 633 Aethalometer ® Black Carbon Monitor User Manual © TELEDYNE ADVANCED POLL UTION INSTRUMENTATION 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 USA Toll-free Phone: 800-324-5190 Phone: 858-657-9800 Fax: 858-657-9816 Email: api-sales@teledyne.
.
07645A DCN6170 i NOTICE OF COPYRIGHT © 2013 Teledyne Advanced Pollution Inst rumentation, user m anual expanded and edited with permission from Aero sol d.o.o./Magee Scientific. All rights reserved. TRADEMARKS The Aethalometer ® is a registered trademark of Magee Scientific an d the instrument is manufactured by Ae rosol d.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation ii 07645A DCN6170 Thispageintentionallyleftblank.
07645A DCN6170 iii TABLE OF CONTENTS 1 INTRODUCTION 5 1.1 Description 5 1.1.1 ‘DualSpot™’ Technology 6 1.1.2 Automatic Zero and Span 10 1.1.3 User and Communications Interfaces 11 1.1.4 Modular Construction 12 1.2 Technical Specifications 13 1.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation iv 07645A DCN6170 5.4 Manual flow meter calibration 55 5.5 Manual stability test 61 5.5 Manual clean air test 61 5.6 Cleaning the optical head 62 5.
07645A DCN6170 5 1 INTRODUCTION 1.1 Description Figure 1. The Aethalometer® Model 633. The Model 633 represents the “Next Generation”.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 6 07645A DCN6170 The Model 633 Aethalometer has been developed with inp.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 7 The ‘spot loading effect’ is a v ariable phenomen.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 8 07645A DCN6170 Figure 2. Aethalometer da ta sorted and averag.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 9 It is clearthat the effect,when present, is linearwith loadi ng (‘ATN’).
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 10 07645A DCN6170 2000 3000 4000 0 20000 40000 60000 80000 BC concent rat i on ( ng/ m 3 ) time (sec) IR, 1 second raw BC1 raw BC2 compensated BC Figure 3.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 11 -500 0 500 1000 1500 2000 2500 14:30:00 14: 35:00 14:40:00 14:45: 00 14:50:00 14:55: 00 15:00:00 15:05: 00 15:10:00 1 secon d tim e-base 30 per .
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 12 07645A DCN6170 1.1.4 Modular Construction The Model 633 Aethalometer is constructed with a modular design , so that sub‐units may be e a s i l y s e r v i c e d .
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 13 1.2 Technical Specifications Operation: Supplyvolt.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 14 07645A DCN6170 1.3 Functional Description The Model 633 features mod.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 15 ENCLOSURE The Model 633 enclosure is v ery robust and made from sheet meta l .
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 16 07645A DCN6170 ETHERNET CONNECTOR The Ethernet connector allows the connection of the Model 633 m easurement system to Ethernetbasedcommunicationnetworks.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 17 FLOWMETERS Two flow meters measure the airflow in different points in the system. One of them is used alsointheairflowregulationloop.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 18 07645A DCN6170 LIGHT SOURCE The light source integrates groups of LEDs of different light w avelengths.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 19 ON/OFF SWITCH The secondary ON/OFF switch is located behind the front door.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 20 07645A DCN6170 Thispageintentionallyleftblank.
07645A DCN6170 21 2 SAFETY NOTES and LABELS CAUTION! READ THIS CHAPTER VERY CAREFULLY BEFORE OPERATING THE INSTRUMENT. Instrument operation R e a d t h i s U s e r ’ s m a n u a l B E F O R E o p e r a t i n g t h e i n s t r u m e n t .
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 22 07645A DCN6170 Fire and explosion NEVER install the instrument in explosion‐risk areas and never use the equipmentnearflammablesubstances.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 23 Instrument service and repair U N A U T H O R I S E D i n s t r u m e n t s e r v i c e a n d r e p a i r p r o c e d u r e s a r e N O T A LLOWED.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 24 07645A DCN6170 Moving parts During AUTOMATIC tape advance procedure, the measurement chambe r is lifted b y a m o t o r i z e d c h a m b e r l i f t m e c h a n i s m .
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 25 3. INSTRUMENT INSTALLATION 3.1 Unpacking the system Taketheinstrumentoutoftheb oxandplaceitonanevensurf ace.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 26 07645A DCN6170 e x h a u s t f r o m t h e H V A C u n i t – t h i s .
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 27 in v er ted so ra i nc a n no t en ter .
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 28 07645A DCN6170 indicatedbygreencheckboxesforeachstep.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 29 3.4 Filter tape installation Inthe“Operation/General”menu (Section4providesinformati onabouttheuserinterface) press“Stop”tostopmeasurement s.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 30 07645A DCN6170 Thefirst“TapeChange Procedure”screenis displayedw.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 31 Slidetapefrombeneath theopticalchamberand removetheusedtape. Keeptheemptyrollfor usewithnew,replacement tape.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 32 07645A DCN6170 Tapedirectionisfrom bottom,lefttoright.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 33 Insertthenewtaperollon theleftsideofthe chamberandrunthetape undertheleftguide‐arm.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 34 07645A DCN6170 Ensurethetapeistaut betweenthetwosides. Replacethetransparent taperetentiondisks. Closethedoorandpress OKonthedisplayscreen.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 35 Thenext“TapeChange Procedure”screen promptsyoutowaitfor thechambertoreturnto homeposition.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 36 07645A DCN6170 3.4 Manually lifting/lowering chamber optical head In.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 37 4 USER INTERFACE, SETTINGS and OPERATION 4.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 38 07645A DCN6170 should you want the flow to be reported at 0 C, or 25 C for example.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 39 4.2 Instrument Status Theinstrument’scurrentstatus.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 40 07645A DCN6170 Single Status Condition IfthevaluedisplayedmatchesavalueintheStatusCondition column,itindicatesonlyone parameteranditsdescription.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 41 Gotothe“Data/Export”menuand press“ExporttoUSB”.The datawillbestoredinatextfile withaheader.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 42 07645A DCN6170 Data field description: Date (yyyy/MM/dd) :date. Time (hh:mm:ss) :time. ‐ Timebase :timebase–unitsinseco nds.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 43 K1; K2; K3; K4; K5; K6; K7 :K_iare th.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 44 07645A DCN6170 Establishcommunicationsusingthefollowingvalues: .BaudRate=115200 .
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 45 Example2:IfyouareusingHyper Terminaljustwritefollowing command$AE33:D1and pressenter(thisshouldaddCRc haracteratendofcommand).
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 46 07645A DCN6170 Thispageintentionallyleftblank.
07645A DCN6170 47 5 MAINTENANCE and SERVICE TheAethalometerneedstobemaintainedregularly.Werecommend thefollowingschedule forregularmaintainance.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 48 07645A DCN6170 5.1 Cleaning the cyclone Removethecycloneanddisassemb le(showninthefollowinginstructions),cleaning the variouspartsthoroughly.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 49 3.Loosenthethreesetscre wsthatholdtheinlet bodyinplace. 4.Removeinletbody. 5.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 50 07645A DCN6170 7.Removeallencapscrewsfromcyclonetop. 8.Cleanallpartsandthoroughlydry;then reassembleinre verseorder.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 51 3. Disconnectflowmeterfromtheinp ut.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 52 07645A DCN6170 Automatic flow calibration procedure: 1.Stopmeasurementsinthe OPERATION>GENERALmenu(press Stopbutton).
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 53 Automatic flow calibration procedure: 4.Waitforthemeasurementchamberto movetotheupposition.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 54 07645A DCN6170 Automatic flow calibration procedure: 7.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 55 5.4 Manual flow meter calibration First,ensurethataleaktestiscompleted(Section5.1)prior torunningthiscalibration procedure.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 56 07645A DCN6170 Manual flow calibration procedure: 4. Waitforthemeasurement chambertomovetotheup position.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 57 Manual flow calibration procedure: 7. Waitforthepumptorun forseveralsecondsto achieveastableflow.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 58 07645A DCN6170 Manual flow calibration procedure: 8.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 59 Manual flow calibration procedure: 10.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 60 07645A DCN6170 Manual flow calibration procedure: 13.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 61 5.5 Manual stability test Stabilitytestisconductedwith outflowtodeterminetheperfo rmanceofthelight source– detectorpair.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 62 07645A DCN6170 5.6 Cleaning the optical head Optical head cleaning procedure: 1. STOPmeasurementsinthe OPERATION>GENERALmenu.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 63 Optical head cleaning procedure: 4. Whilepressingtherelease buttonupward,graspoptical headandturntoloosenand remove.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 64 07645A DCN6170 Optical head cleaning procedure: 6. Alternatively,useacottonswab. Ensuretocleanbothtopand bottomofopticalhead.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 65 Optical head cleaning procedure: 5.7 Changing the bypass cartridge filter Bypassfilterchangingprocedure: 1.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 66 07645A DCN6170 Bypassfilterchangingprocedure: 3.Repeatonotherside. 4.Removecartridge. 5.Gentlywedgesmall,flattool betweenfittingandcartridge.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 67 Bypassfilterchangingprocedure: 8.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 68 07645A DCN6170 5.8 Calibrating the tape sensor Tapesensorcalib.
Teledyne Advanced Polluti on Instru mentation Model 633 Aethalometer ® Black Carbon Monitor User Manual 07645A DCN6170 69 Tapesensorcalibrationprocedure: 3.TheTapeSensors Adjustmentscreen promptstoplacethe smalltapeholder [cylinder]onleftand largeonright,then pressOK.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 70 07645A DCN6170 Thispageintentionallyleftblank.
07645A DCN6170 71 6 TECHNICAL SUPPORT a nd CONTACT INFORMATI ON Ifthismanualanditstroubleshooting&servicesectiondonot solveyourproblems,technical assis.
Model 633 Aethalometer ® Bla ck Carbon Monitor User Manual Teledyne Advanced Polluti on Instrumentation 72 07645A DCN6170 Thispageintentionallyleftblank.
07645A DCN6170 73 7 SPARE PARTS LIST Following is a list of spare parts availabl e for order through the Sales Department: DU0000146 Filter Tape for 633 DU0000196 Flow calibration pad TU0000034 Static dissipativ e inlet tubing (8-ft) DU0000148 PM2.
デバイスTeledyne 633の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Teledyne 633をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTeledyne 633の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Teledyne 633の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Teledyne 633で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Teledyne 633を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTeledyne 633の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Teledyne 633に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTeledyne 633デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。