MODEL HPM 4/6 VACUUM GAUGE INSTRUCTION MANUAL TELEDYNE HASTINGS INSTRUMENTS ISO 9001 CER TIFIED.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 2 of 12 Manual Print History The print history shown below lists the p rinting dates of all revisions and a ddenda created for this manual. The revision level letter increases alphabetically as the manual undergoes subsequent u pdates.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 3 of 12 Table of Contents 1.0 QU IC K S TA RT I NS TR U CT ION S....................................................................................................... ............ 4 2.0 SPE CI FIC AT IO NS .
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 4 of 12 1. Unpack unit and insure battery is installed 2. Connect thermocouple tube. 3. Slide switch to DV4 or DV6 position according to tube type 4. Indicated pressure is displayed on LCD display. 5. Slide switch to center/off position when not in use.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 5 of 12 Operating Range .................................................................... DV-4 - 0.01 - 20 Torr (.01 to 20 mbar) ..........................................................................
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 6 of 12 3.1. Receiving Inspection Unpack unit and inspect all items for obvious si gns of damage. Immediately advise Tel edyne Hastings and the carrier of any suspe cted damage.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 7 of 12 Operation of the Hastings gauge sensor (DV-4 or DV-6) is based on a low voltage AC bridge that heats a noble metal thermopile. A change in pressure in the gauge tube alters the molecular collision rate and therefore the thermal conducti on of the gas or gas mixture surrounding the thermopile.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 8 of 12 c. Pressure under range “-0.00 to -0.05” (neg ative display values) indicating that the instrument requir es low-pressure adjustment. Ref er to Section 4. 3. Always place slide switch to the off position when not in use.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 9 of 12 5.1. Calibrat ion in a Vacuum Syste m 1 Install a DV-4 or DV-6 sensor in your vac uum system and ev acuate vacuum chamber to a pressure less than 1.00E-04 T orr. NOTE: Refer to Fig 3 for locations of calib ration pots.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 10 of 12 1 DV-4 low pressure adjust ments: a. Connect instrument gauge cable to a DB-16D Re ference Tub e and position slide switch to DV-4. b. Remove battery compartment cover, and then adj ust DV-4 calibration pot until display reads pressure value listed on reference tube label.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 11 of 12 6.1. Authorized Maintena nce The HPM 4/6 Power Supply/Display does not require maintenance other than routine battery replacement and calibration. Note: A low battery will be indicat ed by “LOBAT” appearing in lower left corner of LCD display.
Manual: 153-082010_HPM 4-6 Vacuum Gauge Page 12 of 12 7.1. Wa rranty Repair P olicy Hastings Instruments warrants this product for a period of one year from the date o f shipment to be free from defects in material and wor kmanshi p.
デバイスTeledyne HPM 4/6の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Teledyne HPM 4/6をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTeledyne HPM 4/6の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Teledyne HPM 4/6の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Teledyne HPM 4/6で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Teledyne HPM 4/6を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTeledyne HPM 4/6の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Teledyne HPM 4/6に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTeledyne HPM 4/6デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。