Model LXT-230 Transmitter Installation and Operating Manual for pH and ORP Teledyne Analytical Instruments 16830 Chestnut Street • City of I ndustry, CA 91748 888 -789-8168 • ask_tai@tele dyne.
Model LXT-230 Instruction Manu al / 2 TABLE OF CONTENTS UNPACKING THE INSTRUMENT ....... ...... ....... ...... ...... ....... ...... ....... ...... ...... ....... ...... ....... 4 1.0 GENER AL DESCRIPTION ....... ...... ....... ...... ...... ....... .
Model LXT-230 Instruction Manu al / 3 6.6 Calibrati on Error Dete ction & Electro de Operational Guidelines ....... ...... ....... ...... .... 15 6.7 Back-to-Factory Calibration ..... ...... ....... ...... ....... ...... ...... ....... ...... ....
Model LXT-230 Instruction Manu al / 4 UNPACKING THE INSTRUMENT Your Teledyne inst rument has been care fully packaged to protect it from damage during shipment an d dry storage.
Model LXT-230 Instruction Manu al / 5 Fahrenheit. A toggle function in the Temperature Calibration Menu allows this selection to be performed in the field. 1.7 PID Outpu t (Optional) An optional PID output is available and provides a 4-20 mA, three-mode control signal to a transducer or a final control element.
Model LXT-230 Instruction Manu al / 6 2.0 INSTALLATIO N 2.1 Mounting Three typical inst allation configurations ar e avail able for the LXT-230 transmitte r: universal mounting, han drail mounting, and panel mounting configurations. The universal mounting configuration allows the LXT-230 to be mounted on a wall or a 2” pipe stand.
Model LXT-230 Instruction Manu al / 7 3.0 FAMILIARIZATION This se ction will pro vide an overview of the front panel key functions and the disp lay menu s for a single channel pH and ORP transmitter. Multiple channel transmitters will have the same menus.
Model LXT-230 Instruction Manu al / 8 3.2.2 Main Menu This menu appears after the Copyright Display when power is first applied to the LXT-230. The Main Menu di splays the measured process variable, temperature, and the current output in pe rcent of f ull- scale.
Model LXT-230 Instruction Manu al / 9 The t op line displays the buffer, or calibr ation point , at which the last calibrat ion was performed. For pH transmitters, the bottom line displays the electro de slope, or efficiency, in millivolts per pH (mV/pH).
Model LXT-230 Instruction Manu al / 10 4.0 SENTINEL DIAGNOSTICS The basic menu structure of the LXT-230 transmitter is architecturally the same as the LXT- 230 pH , ORP and PION, with the only va riable being the diagnostic information on the main and special diagnostic menus.
Model LXT-230 Instruction Manu al / 11 To file the input, with the microprocessor, exit the calibrate mode by pressing either menu selection key. The u pper half of the split screen menu displays absolute reference potential (±2 millivolts) directly from the electrode vi a the diagnostic signal co conditione r located in the sensor.
Model LXT-230 Instruction Manu al / 12 5.0 MAINTENANCE If service of the internal components is required, always turn off the power to the instrument. 5.1 Clean ing Cleaning the front panel can be performed with a detergent and water. DO NOT USE ACETONE, ACID OR CAUSTIC SOLUTIO NS ON THE ENCLOSURE SURFACE.
Model LXT-230 Instruction Manu al / 13 6.0 CALIBRATION 6.1 Manual Output Mode To perform calibrations with out interferi ng with control or recorder functions, the LXT-230 incorporates a manual output mode. The current output is set to the desired level and saved until ch anged or released fro m the manual mode.
Model LXT-230 Instruction Manu al / 14 2 pH 4.00 -58.2 mV/pH 2 ORP 180 .0 mV .980 mV/ORP Procedure: 1. Press the appropriate MENU SELECTION key to reach the Electrode Standardization Menu and locate the cursor under the "1" as illustrated in section 3.
Model LXT-230 Instruction Manu al / 15 diagnostic tool in determining the degrada tion of the ele ctro de. Large differences between calibrations m ay indicate coating or damage to the measurement half-cell.
Model LXT-230 Instruction Manu al / 16 Procedure: 1. Press the appropriate MENU SELECTION key to reach the Output Calibration Menu. 2. Position the cu rsor on the top line to change the 4 mA point or the bottom line to change the 20 mA point. 3. Press both horizontal C ALIBRATE keys simultaneously to enter the calibrate mode.
Model LXT-230 Instruction Manu al / 17 4. To toggle back to the first unit of measurement, press the CALIBRATE keys again. Trim °C .0 Unit 25.0°C Trim °F .
Model LXT-230 Instruction Manu al / 18 7.0 LXT-230 MODEL NUMBER DESIGNATORS Channel 1 Input & Output LXT-230- PH/MA pH LXT-230- OR/MA ORP LXT-230- PION/MA Specific Ion (see n ote 1) Note 1: F or PION, the model number will reflect the ion being measured; for example, Ca f or calcium, Fl for fluoride, et c .
Model LXT-230 Instruction Manu al / 19 8.0 TROUBLESHOOTING GUIDE Symptom Possible Causes Suggested Acti ons No power to the instrument. Check power supply to be sure the correct voltage is being su pplied to the transmitter. (13.5 to 50 VDC) Check polarity of transmitter wiring.
Model LXT-230 Instruction Manu al / 20 Electrode Span Menu, then perform a buffer calibration. Check the sensor and electrode for proper operation. Wrong temperature correction has been calibrated in the temperature trim. Examine the temperature and the Temperature Calibration Menu.
Model LXT-230 Instruction Manu al / 21 APPENDIX A - SPECIFICATIONS MEASUREMENT RANGE pH: 0.00 to 14.00, fully expandable and reversible, standard. -2.00 to 14.00 pH , op tional. ORP: +1000 to -1000 mV, fully expandable and reversible, standard. OUTPUT 4-20 mA or 20-4 mA, linear and expan dable.
Model LXT-230 Instruction Manu al / 22 Temperature Trim Allows for compensation for any differen ces in RTDs b y progr amming the offset into the transmitter.
デバイスTeledyne LXT-230の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Teledyne LXT-230をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTeledyne LXT-230の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Teledyne LXT-230の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Teledyne LXT-230で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Teledyne LXT-230を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTeledyne LXT-230の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Teledyne LXT-230に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTeledyne LXT-230デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。