ADDENDUM TO M300E/EM MANUAL P/N 04288 MODEL 300EU CO ANALYZER WITH AUTO-REFERENCE © TELEDYNE INSTRUMENTS ADVANCED POLLUTI ON INSTRUMENTATION DIVISION (TAPI) 9480 CARROLL PARK DRIVE SAN DIEGO, CALIFORNIA 92121- 520 1 USA Toll-free Phone: 800-324-5190 Phone: 858-657-9800 Fax: 858-657-9816 Email: api-sales@teledyne.
.
M300 EU SAFETY MESSAGES Addendum to M300E/EM Manual P/N 04288 SAFETY MESSAGES Your safety and the safety of others are very important. We have provided many important safety message s in this manual. Please read these messag es carefully. A safety message alerts you to potential hazards that could hurt you or others.
SAFETY MESSAGES M300 EU Addendum to M300E/EM Manual P/N 04288 USER NOTES: ii 05904 Rev A.
M300 EU TABLE OF CONTENTS Addendum to M300E/EM Manual P/N 04288 TABLE OF CONTENTS 1. INTRODUCTION .........................................................................................................................................................
TABLE OF CONTENTS M300 EU Addendum to M300E/EM Manual P/N 04288 10.3. Electronic Operation ............................................................................................................................... ...................... 53 10.3.1.
M300 EU TABLE OF CONTENTS Addendum to M300E/EM Manual P/N 04288 Table 6-7: M300EU He ssen GAS ID List ......................................................................................................39 Table 6-8: Default Hessen Status Bit Assignments .
TABLE OF CONTENTS M300 EU Addendum to M300E/EM Manual P/N 04288 USER NOTES: iv 05904 Rev A.
M300 EU INTRODUCTION Addendum to M300EU Manual - P/N 04288 1. INTRODUCTION This addendum is based o n the M300E/EM Operators Manual (P/N 04288). It is intended as a supplemen t to the M300E/EM manual .
INTRODUCTION M300 EU Addendum to M300E/EM Manual P/N 04288 USER NOTES: 2 05904 Rev A.
M300 EU SPECIFICATIONS, A PPROVALS AND WARRANTY (Addendum to M300EU Manual - P/N 04145) 2. SPECIFICATIONS, APPROVALS AND WARRANTY 2.1. SPECIFICATIONS Table 2-1: M300EU Basic Unit Speci fications Range.
SPECIFICATIONS, APPR OVALS AND WARRANTY M300 EU Addendum to M300E/EM Manual P/N 04288 2.2. EPA EQUIVALENCY DESIGNATION The M300EU has not been certified by the EPA as an equivalent method at the ti me.
M300 EU GETTING STARTED (Addendum to M300EU Manual - P/N 04145) 3. GETTING STARTED 3.1. UNPACKING THE M300EU Unpack the M300EU as p er the directions in Section 3.1 of the M300E/EM manual (P/N 04288), with the exception that there are no shippi ng screws.
GETTING STARTED M300 EU Addendum to M300E/EM Manual P/N 04288 The pneumatic flow of the M300EU is different from the M300E/E M in two ways: • The addition of a high-flow multi-tube Nafion ® dryer. • An additional gas p ath that passes the sample gas though a CO scrubber u sed during the auto- reference cycle.
M300 EU GETTING STARTED (Addendum to M300EU Manual - P/N 04145) 3.3.2. ANALOG OUTPUT Connections The analog outputs for the M300E U are different from thos e of the M300E/EM.
GETTING STARTED M300 EU Addendum to M300E/EM Manual P/N 04288 Table 3-2: Analog Output Pin Outs Pin Analog Output Standard Voltage Output Current Loop Option 1 V Out I Out + 2 A1 Ground I Out - 3 V Out I Out + 4 A2 Ground I Out - 5 6 A3 NOT USED 7 V Out Not Available 8 A4 Ground Not Available 3.
M300 EU GETTING STARTED (Addendum to M300EU Manual - P/N 04145) 3.4. INITIAL OPERATION OF THE M300EU NOTE The analyzer’s cover must be installed and secur ely f astened to ensure that the convection ov en portion of the internal layout is capable of properly cr eating and controlling temperatur es of the analy zer’s optical bench.
GETTING STARTED M300 EU Addendum to M300E/EM Manual P/N 04288 3.4.3. FUNCTIONAL CHECK The information found in Section 3.2.4 of the M300E/EM Operators Manu al (P/N 042 88) regarding performing an initial functional check of the analyzer is appli cable to the M300EU with the following exception(s).
M300 EU GETTING STARTED (Addendum to M300EU Manual - P/N 04145) 3.5. INITIAL CALIBRATION 3.5.1. PRE-CALIBRATION STABILIZATION Before initially calibrating the M300EU it must be allowed to op erate for a minimum of 12 hours.
GETTING STARTED M300 EU Addendum to M300E/EM Manual P/N 04288 3.5.3. BASIC CO CA LIBRATION PROCEDURE The initial calibration shoul d be carried out using the same reportin g range set up as used during the analy zer’s factory calibration.
M300 EU FREQENTLY ASKED QUESTIONS & GLOSSARY (Addendum to M300EU Manual - P/N 04145) 4. FREQENTLY ASKED QUES TIONS & GLOSSARY The information found in Chapter 4 of the M300E/EM O perators Manual (P/N 042 88) is applicable to the M300EU.
FREQENTLY ASKED QUESTIONS & GLOSSARY M300 EU Addendum to M300E/EM Manual P/N 04288 USER NOTES: 14 05831 Rev. 01.
M300 EU Optional Hardware and Software (Addendum to M300EU Manual - P/N 04145) 5. OPTIONAL HARDWARE AND SOFTWARE The following table lists the optional hard ware is availa ble for the M300Ex family of analyzers. There are no additional options available for the M300EU that are not also available for the M300E/EM.
Optional Hardware and Software M300 EU Addendum to M300E/EM Manual P/N 04288 For assistance with orderi ng these options, please co ntact the Sales department of Teledyne – Advanced Pollution Instruments at: TOLL-FREE: 800-324-5190 FAX: 858-657-9816 TEL: 858-657-9800 E-MAIL: api-sales@teledy ne.
M300 EU Optional Hardware and Software (Addendum to M300EU Manual - P/N 04145) 5.1.2. ZERO SPAN VALV ES WITH IZS (OPT 51B) Because of the addition of the auto-refe rence gas path and the mul ti-tube Nafion ® dryer, the internal pneumatics gas flow of the M300EU is somewh at different from the M300E/EM.
Optional Hardware and Software M300 EU Addendum to M300E/EM Manual P/N 04288 5.1.3. Zero Span Valves with IZS and Shutoff VALVE (OPT 51B) Because of the addition of the auto-refe rence gas path and the mul ti-tube Nafion ® dryer, the internal pneumatics gas flow of the M300EU is somewh at different from the M300E/EM.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 6. M300EU OPERATING INSTRUCTIONS 6.1. SUMMARY OF DIFFERENCES BETWEEN OPERATING THE M300EU AND THE M300E/EM ANALYZERS.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 6.2. OPERATING MODES The information found in Section 6.1 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exception(s). The following table supersedes Table 6 -2 of t he M300E/EM Operators Manu al (P/N 04288).
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 6.2.1. AUTO-REFERENCE MODE (AREF) One of the most significant differences between the M300E/EM and the M300EU analyze rs is the auto-reference measurement feature.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 6.2.2. TEST FUNCTIONS The information found in Section 6.2.1 of the M300E/EM Operators Manu al (P/N 04288) is applicable to the M300EU with the following exception(s): • The following table supersedes Table 6-2 of the M300E/EM Operat ors Manual (P/N 04288).
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 6.3. SETUP Æ RNGE: ANALOG OUTPUT REPORTING RANGE CONFIGURATION 6.3.1. PHYSICAL RANGE VERSUS ANALOG OUTPUT REPORTING RANGE.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 The A1 and A2 channels output a signal that is proportional to the CO concentration of the sample gas. They can be configured: • With independent reporting rang es reporting a “single” output signal ( SNGL Mode, see Section 6.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 6.3.3. RNGE Æ MODE Æ SNGL: CONFIGURING TH E M300EU ANALYZER FOR SINGLE RANGE MODE The single range mode sets a sin gle maximum range for the both the A1 and A2 analog outputs.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 6.3.4. RNGE Æ MODE Æ DUAL: CONFIGURING THE M300EU ANALYZER FOR DUAL RANGE MODE DUAL range mode allows the A1 and A2 outputs to be config ured with se parate reporting range sp ans as well as separate electronic sign al levels.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 6.3.5. RNGE Æ MODE Æ AUTO: CONFIGURING THE M300EU ANALYZER FOR AUTO RANGE MODE AUTO range mode give s the analyzer to ab.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 6.3.6. SETUP Æ RNGE Æ UNIT: SETTING THE REPORTING RANGE UNIT TYPE The M300EU can display concentrations in ppb, ppm, ug/m 3 , mg/m 3 units. Changing units affects all of the COM port values, and all of the display values for all reporting rang es.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 6.4. SETUP Æ MORE Æ AREF: CONFIGURING AND PERFORMING AUTO-REFERENCE MEASUREMENTS 6.4.1. ADJUSTING THE A-REF MODE CYCLE TIME The auto-reference measurement feature is initiated by the M300EU at regul ar intervals (see Section 10.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 6.4.2. FORCING AN AUTO-R EFERENCE MEASUREMENT Sometimes it is advisable to perform an auto-referen c e measurement at other times such as just before calibrating the analyzer.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 05514 Rev A1 31.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 6.5. SETUP Æ MORE Æ VARS: INTERNAL VARIABLES (VARS) The information found in Section 6.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 6.6. SETUP Æ MORE Æ DIAG: USING THE DIAGNOSTICS FUN CTIONS The information found in Section 6.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 6.7. ANALOG OUTPUT CONFIGURATION 6.7.1. ANALOG I/O CONFIGURATION Because the manner in which the an alog outputs ar e implemented differs betwe en the M300E/EM and the M300EU, there are significant differences in the m anner in which the ANALOG I/O are ma nipulated.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) To configure the analyzer’s three anal og outputs, set the electronic signal type of each cha nnel and calibrate the outputs.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 ANALOG OUTPUT SIGNAL TYPE AND RANGE SELECTION The information found in Section 6.13.5.2 of the M300E/EM Operat ors Manual (P/N 04288) is applicable to the M300EU with the following exce ption(s): • There are minor differences in the displays ( e.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) 6.7.2. CALIBRATION OF THE ANALOG OUTPUTS This section replaces Section 6.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 6.7.3. SELECTING A TEST CHAN NEL FUNCTION FOR OUTPUT A4 This section replaces Section 6.
M300 EU M300EU OPERATING INSTRUCTIONS (Addendum to M300EU Manual - P/N 04145) To activate the TEST Channel and select a function, press: 6.7.4. HESSEN PROTOCOL The information found in Section 6.
M300EU OPERATING INSTRUCTIONS M300 EU Addendum to M300E/EM Manual P/N 04288 • The list of Hessen status flags for the M300EU is di fferent from that of the M300E/EM. The following table replaces Table 6-33 of the M 300E/EM Operator s Manual (P/N 04288).
M300 EU CALIBRATION PROCEDURES (Addendum to M300EU Manual - P/N 04145) 7. CALIBRATION PROCEDURES Calibration of the M300EU sho uld be performed according to the procedures described in Chapters 7 &.
CALIBRATION PROCEDURES M300 EU Addendum to M300E/EM Manual P/N 04288 MANUAL CALIBRATION WITH ZERO/SPAN VALV ES The information found in Section 7.4 of the M300E/EM Operators Manual (P/N 04288) is applicable to the M300EU with the following exceptio n(s).
M300 EU EPA Protocol Calibration (Addendum to M300EU Manual - P/N 04145) 8. EPA PROTOCOL CALIBRATION The information found in Chapter 8 of the M300E/EM O perators Manual (P/N 042 88) is applicable to the M300EU.
EPA Protocol Calibration M300 EU Addendum to M300E/EM Manual P/N 04288 USER NOTES: 44 05831 Rev. 01.
M300 EU INSTRUMENT MAINTENANCE (Addendum to M300EU Manual - P/N 04145) 9. INSTRUMENT MAINTENANCE The information found in Chapter 9 of the M300E/EM Oper ators Manual (P/N 04288) is applicable to the M.
INSTRUMENT MAINTENAN 46 CE M300 EU Addendum to M300E/EM Manual P/N 04288 05831 Rev. 01 USER NOTES:.
M300 EU INSTRUMENT MAINTENANCE (Addendum to M300EU Manual - P/N 04145) Table 9-1: M300EU Test Function Record DATE RECORDED FUNCTION OPERATING MODE* STABILITY ZERO CA L CO MEAS ZERO CAL ZERO CAL MR RA.
CE M300 EU Addendum to M300E/EM Manual P/N 04288 INSTRUMENT MAINTENAN 48 05831 Rev. 01 USER NOTES:.
M300 EU THEORY OF OPERATION (Addendum to M300EU Manual - P/N 04145) 10. THEORY OF OPERATION 10.1. MEASUREMENT METHOD The information found in Section 10.1 of the M300E/E M Operators Manual (P/N 04288) is applicable to the M300EU with the following addition.
THEORY OF OPERATION M300 EU Addendum to M300E/EM Manual P/N 04288 The auto-reference measurement takes approximately 15 minutes. To ensure that the sample chambe r of the optical bench is properly p u.
M300 EU THEORY OF OPERATION (Addendum to M300EU Manual - P/N 04145) 10.2.2. THE A-REF VALVE ASSEMBLY The auto-reference valve assembly is located next to t he optical bench, at the re ar of the analyzer between the Nafion ® dryer and the outer wall of the instrument (see Figur e 3-1).
THEORY OF OPERATION M300 EU Addendum to M300E/EM Manual P/N 04288 The dryer consists of a bundle of 0.030" O.D. Nafion ® tubes housed within a single large tube shell. Sample gas flows within the Nafion tubes while water vapor absorbs into the tubular memb rane walls and is removed.
M300 EU THEORY OF OPERATION (Addendum to M300EU Manual - P/N 04145) 10.3. ELECTRONIC OPERATION 10.3.1. OVERVIEW The information found in Section 10.4.1 of the M300E /E M Operators Manual (P/N 04288) is applicable to the M300EU with the following exceptio n(s): • This figure replaces Figure 10-9of the M300E/EM Operators Manu al.
THEORY OF OPERATION M300 EU Addendum to M300E/EM Manual P/N 04288 10.3.2. THE RELAY PCA The information found in Section 10.4.5 of the M300E /E M Operators Manual (P/N 04288) is applicable to the M300EU with the following exceptio n(s): 10.3.2.1. Temperature Contro l of the Convection Oven The information found in Section 10.
M300 EU THEORY OF OPERATION (Addendum to M300EU Manual - P/N 04145) 10.3.2.2. Oven Heaters AC Power Configuration The two main heaters for the convection oven are a of t he M300EU are AC heaters.
THEORY OF OPERATION M300 EU Addendum to M300E/EM Manual P/N 04288 10.3.2.3. Status LED’s This information replaces t hat found in Sectio ns 10.4.5.5 of the M300E/EM Operators Manual.
M300 EU THEORY OF OPERATION (Addendum to M300EU Manual - P/N 04145) 10.3.4. POWER DISTRIBUTION The information found in Section 10.4.5 of the M300E /E M Operators Manual (P/N 04288) is applicable to the M300EU with the following exceptio n(s): • This Figure replaces Figure 10-15 of the M300E/EM Operators Manual (P/N 04288).
.
M300 EU TROUBLESHOOTING & REPAIR (Addendum to M300EU Manual - P/N 04145) 11. TROUBLESHOOTING & REPAIR 11.1. GENERAL NOTES CAUTION THE OPERATIONS OUTLINED IN THIS CHAP TER MUST BE PERFORMED BY QUALIFIED MAINTENANCE PERSO NNEL ONLY.
TROUBLESHOOTING & REPAIR M300 EU Addendum to M300E/EM Manual P/N 04288 11.2. FAULT DIAGNOSIS WITH WARNING MESSAGES The following Table replac es Table 11-1; of the M300E/EM Operators M anual.
M300 EU TROUBLESHOOTING & REPAIR (Addendum to M300EU Manual - P/N 04145) Table 11-1: Warning Messages – Indi cated F ailures (cont.) WARNING MESSAGE FA ULT CONDITION POSSIBLE CAUSES SAMPLE FLOW WA RN Sample flow rate is <1400 cm 3 /min or > 2200 cm 3 /min.
TROUBLESHOOTING & REPAIR M300 EU Addendum to M300E/EM Manual P/N 04288 11.2.1. FAULT DIAGNOSIS WITH TEST FUNCTIONS The following table supersedes Table 1 1.
M300 EU TROUBLESHOOTING & REPAIR (Addendum to M300EU Manual - P/N 04145) Table 11-2: Test Functions - Indicated Failures TEST FUNCTIONS ( As Disp layed ) INDICATED FAILURE(S) PHT DRIVE If this dri.
TROUBLESHOOTING & REPAIR M300 EU Addendum to M300E/EM Manual P/N 04288 11.2.2. RELAY BOA RD STATUS LED’S The information found in Section 11.1.4.3 of the M300 E/EM Operat ors Manual (P/N 04288) is applicable to the M300EU with the following exception: • The following table replaces Table 11-5 of the M300E/EM Operators Manu al.
M300 EU TROUBLESHOOTING & REPAIR (Addendum to M300EU Manual - P/N 04145) 11.3. GAS FLOW PROBLEMS Along with the information found in Section 11.2 of the M300E/EM Operators Manual (P/N 0 4288), also be aware that problems with the flow of gas through the Nafion ® dryer could also be a cause of pressure or flow problems through the M30 0EU.
TROUBLESHOOTING & REPAIR M300 EU Addendum to M300E/EM Manual P/N 04288 11.5. SUBSYSTEM CHECHOUT 11.5.1. RELAY BOARD The information found in Section 11.5.5 of the M300E /E M Operators Manual (P/N 04288) is applicable to the M300EU with the following exception: • The following table replaces Table 11-8 of the M300E/EM Operators Manu al.
M300 EU TROUBLESHOOTING & REPAIR (Addendum to M300EU Manual - P/N 04145) 11.6. TECHNICAL ASSISTANCE If this addendum and its trouble-shooting / repair secti ons do not solve your problems, technical assistance may be obtained from Teledyne Instrument s, Custome r Service, 9480 Carroll Park Drive, San Diego, CA 92121.
TROUBLESHOOTING & REPAIR M300 EU Addendum to M300E/EM Manual P/N 04288 USER NOTES: 68 05514 Rev A1.
M300EU Addendum to M300E/EM Manual APPENDIX A - Version Specific Software Documentation APPENDIX A - Version Specific Software Documentation APPENDIX A-1: M300EU Soft ware Menu Trees APPENDIX A-2: M30.
APPENDIX A-1: M300EU Software M enu Trees, Re vi sion K.6 M300EU Addendum to M300E/EM Manual APPENDIX A-1: M300EU Soft ware Menu Trees, Revision K.6 RANGE= [Value] PPM RANGE1= [Value] PPM 2 RANGE2= [V.
M300EU Addendum to M300E/EM Manual APPENDIX A- 1: M300EU Software Menu Trees, Revision K.6 Figure A-2: Primary Setup Menu (Except iDAS) 05924 Rev K.6 A-3.
APPENDIX A-1: M300EU Software M enu Trees, Re vi sion K.6 M300EU Addendum to M300E/EM Manual Figure A-3: Primary Setup Menu Æ iDAS Submenu A-4 05924 Rev K.
M300EU Addendum to M300E/EM Manual APPENDIX A- 1: M300EU Software Menu Trees, Revision K.6 Figure A-4: Secondary Setup Menu Æ COMM and VARS Sub menus 05924 Rev K.
APPENDIX A-1: M300EU Software M enu Trees, Re vi sion K.6 M300EU Addendum to M300E/EM Manual Figure A-5: Secondary Setup Menu Æ Hessen P rotoc ol Submenu A-6 05924 Rev K.
300EU Addendum to M300E/EM Manual APPENDIX A- 1: M300EU Software Menu Trees, Revision K.6 4 Rev K.6 A-7 ANALOG OUTPUT SETUP PASS DAS RNGE CLK MORE ACAL CFG SAMPLE COMM VARS DIAG NEXT PREV SIGNAL I/O A.
APPENDIX A-2: Setup Variables For Serial I/O, Revision K.6 M300EU Addendum to M300E/EM Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision K.
M300EU Addendum to M300E/EM Manual APPENDIX A-2: Setup Variables For Serial I/O, Revision K.6 SETUP VARIABLE NUMERIC UNITS DEFAULT VA LUE VAL UE RA NG E DESCRIPTION OVEN_PROP 1/ºC 0.5 0–100 Internal box temperature #2/oven PID proportional coefficien t.
APPENDIX A-2: Setup Variables For Serial I/O, Revision K.6 M300EU Addendum to M300E/EM Manual SETUP VARIABLE NUMERIC UNITS DEFAULT VA LUE VA LUE RA NGE DESCRIPTION O2_FILT_PCT % 2 0.1–100 Relative change in O 2 concentration to shorten filter. O2_FILT_DELAY Seconds 20 0–300 Delay before leaving O 2 adaptive filter mode.
M300EU Addendum to M300E/EM Manual APPENDIX A-3: Warnings and Tes t Functions, Revis ion K.6 APPENDIX A-3: Warnings and Test Fu nctions, Revision K.6 Table A-3: Additional or Changed Relevant to the M300E U Warning Messages, Revision K.6 NAME MESSAGE TEXT DESCRIPTION WAUTOZERO AZERO WARN 1.
APPENDIX A-3: Warnings and Test Function s, Revision K.6 M300EU Addendum to M300E/EM Manual Table A-5: Additional or Changed Test Function Relevant to the M300EU , Revi sion K.6 TEST FUNCTION NA ME MESSAGE TEXT DESCRIPTION STABILITY STABIL=0.0 PPM These gases not applicable to the M300EU AUTOZERO AZERO RATIO=1.
M300EU Addendum to M300E/EM Manual APPENDIX A- 4: Signal I/O Defi nitions for 300E/EM Series Analyzers, Rev. K.6 APPENDIX A-4: Signal I/O Definitions for 300E/EM Series Analy zers, Rev. K.6 Table A-7: Additional or Changed Test Signal I/O Definition s for M30 0EU, Revision K.
APPENDIX A-5: M300EU iDAS Pa rameter s Revisio n K.6 M300EU Addendum to M300E/EM Manual APPENDIX A-5: M300EU iDAS Parame ters Revision K.6 Table A-9: Additional or Changed iDAS Trigger Ev ents for M 300EU, Revision K.
M300EU Addendum to M300E/EM Manual APPENDIX A-5: M300EU iDAS Para meters Revision K.6 Table A-9: Additional or Changed iDAS Functions for M3 00 EU, Revision K.6 NA ME DESCRIPTION UNITS AZE RO Auto-zero reading M/R OVNTMP Internal box temperature #2/oven ° C OVNDTY Internal box temperature #2/o ven control duty cycle Fraction (0.
APPENDIX A-6: Terminal Command D esig nators, Revision K.6 M300EU Addendum to M300E/EM Manual APPENDIX A-6: Terminal Command Designators, Rev ision K.6 The information found in Appendix A-6 of the M300E /EM Operators Manual (P/N 04906) is applicable to the M300EU .
M300EU Addendum to M300E/EM Manual APPENDIX B - M300EU SPARE PARTS & EXPENDABLES 05925 Rev A B-1 APPENDIX B - M300EU SPARE PARTS & EXPENDABLES NOTE Use of replacement parts other than those supplied by API may result in non-compliance with European standar d EN 61010-1.
APPENDIX B - M300EU SPARE PARTS & EXPENDABLES M300EU Addendum to M300E/EM Manual B-2 05925 Rev A USER NOTES.
デバイスTeledyne Model 300EUの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Teledyne Model 300EUをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTeledyne Model 300EUの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Teledyne Model 300EUの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Teledyne Model 300EUで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Teledyne Model 300EUを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTeledyne Model 300EUの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Teledyne Model 300EUに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTeledyne Model 300EUデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。