TeledyneメーカーT265の使用説明書/サービス説明書
ページ先へ移動 of 47
Manual Addendum Model T265 Chemiluminescence Ozone Analyzer (Addendum to Model T200 Operation Manual PN 06858) © TELEDYNE ADVANCED POLLUTION INSTRUMENTATION 9480 CARROLL PARK DRIVE SAN DIEGO, CA 92121-5201 USA Toll-free Phone: 800-324-5190 Phone: 858-657-9800 Fax: 858-657-9816 Email: api-sales@teledyne.
.
i IMPORTANT SAFETY INFORMATION Important safety messages are provided throughout this manual for the purpose of avoiding personal injury or instrument damage. Please read these messages carefully. Each safety message is associated with a saf ety alert sy mbol and placed throughout this manual and inside the instrument.
ii CONSIGNES DE SÉCURITÉ Des consignes de sécurité importantes sont fournies tout au long du présent manuel dans le but d’éviter des blessures corporelles ou d’endommager les instruments. 1tée par un pictogramme d’alerte de sécurité; ces pictogramm es se retrouvent dans ce m anuel et à l’intérieur des instruments.
iii ABOUT THIS MANUAL This addendum is to be used in conjunc tion with the T200 Operation Manual, PN 06858 . REVISION HISTORY This section provides information regarding changes to this manual.
iv This page intentionally left blank. 07337B DCN6324.
v TABLE OF CONTENTS IMPORTANT SAFETY INFORMATION ................................................................................................... .................... I ABOUT THIS MANUAL .............................................................
vi This page intentionally left blank. 07337B DCN6324.
7 1. INTRODUCTION This section presents a brief overview of the Model T265; supplemental information may be found in the Model T200 Operation Manual. 1.
Introduction Model T265 Chemiluminescence Ozone A nalyzer Addendum to Model T200 Manual 8 This page intentionally left blank. 07337B DCN6324.
9 2. SPECIFICATIONS, APPROVALS, & CERTIFICATIONS Table 2-1 presents the specifications for the Model T265 analyzer. This is followed by EPA and Safety and Emissions Compliance information.
Specifications, Approv als, & Certifications Model T265 Chemiluminescence Ozone A nalyzer Addendum to Model T200 Manual 10 Table 2-1. Model T265 Specifications PARAMETER SPECIFICATION Ranges 0-100 ppb to 0-2,000 ppb full scale, user selectabl e.
Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Specif ications, Approvals, & Certifications 11 2.1 EPA DESIGNATION Teledyne API’s Model T265 Chemiluminescence Ozone Analyzer received EPA approval for Designation EQOA-0611-199.
Specifications, Approv als, & Certifications Model T265 Chemiluminescence Ozone A nalyzer Addendum to Model T200 Manual 12 This page intentionally left blank.
13 3. GETTING STARTED This section addresses the procedures for unpack ing the instrument and inspecting for damage, and introduces the instrument layout, then presents the procedures for getting started: making electrical and pneum atic connections, and conducting an initial calibration check.
Getting Started Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual 14 CAUTION Remove dust plugs from pneumatic fittings on the rear panel of the analy zer, prior to applying power to the analyzer.
Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Getting Started 15 3.2 INSTRUMENT LAYOUT Instrument layout includes front panel and di splay, rear panel connectors, and internal chassis layout. 3.2.1 Front Panel Figure 3-1 illustrates the front panel of the Model T265 analyzer.
Getting Started Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual 16 3.2.2 Rear Panel Figure 3-2 illustrates the layout of the Model T265 analyzer’s rear panel.
Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Getting Started 17 3.2.3 Internal Chassis Layout Figure 3-3 illustrates the location of the internal chassis components.
Getting Started Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual 18 3.3 CONNECTIONS AND SETUP This section presents connection information for setting up the instrum ent to operate. 3.3.1 Electrical and Pneumatic Connections 1.
Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Getting Started 19 Figure 3-4. T265 Pneumatics, Basic Configuration 07337B DCN6324.
Getting Started Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual 20 STARTUP, FUNCTIONAL CHECKS, AND INITIAL CALIBRATION 3.4.1 Startup 1.
Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual Getting Started 21 Setting the range and inputting the expected ozone span gas concentration Performing a zero/span calibration Running a calibration quality check Part 1 - Set the range, then enter the expected ozone span gas concentration: Step Action Comment 1.
Getting Started Model T265 Chemiluminescence Ozone Analyzer Addendum to Model T200 Manual 22 Step Action Comment 6. Press EXIT, input Span gas The Model T265 returns to the CAL menu. Now allow span gas to flow passed the instrument. 7. Wait 30 min Wait for the O 3 reading to stabilize at the span value (watch O 3 STB).
23 4. MAINTENANCE 4.1 MAINTENANCE SCHEDULE CAUTION The operations outlined in this section are to be performed by qualified maintenance personnel only. Please refer to the Instrument Maintenance section of the Model T200 manual for maintenance items and their interval schedules.
Maintenance Model T265 Chemiluminescence Ozone A nalyzer Addendum to Model T200 Manual 24 This page intentionally left blank. 07337B DCN6324.
25 5. PRINCIPLES OF OPERATION The Teledyne API Model T265 analyzer is designed to measure the concentration of ozone using the chemiluminescence reac tion below.
Principles of Operation Model T265 Chemiluminescence Ozone A nalyzer Addendum to Model T200 Manual 26 This page intentionally left blank. 07337B DCN6324.
A-1 APPENDIX A: Models T265 and 265E Menu Trees The following menu trees in this appendix show the structure of the Sam ple menu and the Setup submenu.
A-2 This page intentionally left blank. 07337B DCN6324.
A-3 APPENDIX A-1: Software Menu Trees: Sample Menu 1 Only appears when warning messages are active. 2 Appears only if AUTO RA NGE mode is activated. 3 Only appears if analyzer is equipp ed with Calibration Valves. Press to cycle through the active warning messages.
A-4 07337B DCN6324.
APPENDIX B - Spare Parts Note Use of replacement parts other than t hose supplied by Teledyne A dvanced Pollution Instrume ntation (TAPI) may result in non-compliance with Europe an standard EN 61010-1. Note Due to the dy namic nature of part num be rs, please refer to the TAPI Website at http://www.
This page intentionally left blank. B-2 07337B DCN6324.
T265 Spare Parts List (Reference: 07388 06/01/2011, 15:53) PARTNUMBER DESCRIPTION 000940600 CD, ORIFICE, .010 BROWN 000941800 CD, ORIFICE, .001, RED/BLK 001330000 SLEEVE, 303 SS (UNPLATED) (KB) 002160.
T265 Spare Parts List (Reference: 07388 06/01/2011, 15:53) PARTNUMBER DESCRIPTION 066220000 ASSY, VACUUM MANIFOLD, M265E 066970000 PCA, INTRF. LCD TOUCH SCRN, F/P 067240000 CPU, PC-104, VSX-6154E, ICO.
T265 Spare Parts List (Reference: 07388 06/01/2011, 15:53) PARTNUMBER DESCRIPTION PU0000011 REBUILD KIT, THOMAS 607(KB) PU0000052 PUMP, THOMAS 688, 220/240V 50HZ/60HZ PU0000054 PUMP, THOMAS 688, 100V,.
This page intentionally left blank. B-6 07337B DCN6324.
Appendix C Warranty/Repair Q u estionnaire M265E a n d T 2 6 5 (P/N 074990000 Rev A, DCN# 6167) TELEDYNE API CU STOMER SER VICE EMAIL: api-customerservice@teledyne.
Appendix C Warranty/Repair Q u estionnaire M 2 65E and T265 (P/N 074990000 Rev A, DCN# 6167) TELEDYNE API CU STOMER SER VICE EMAIL: api-customerservice@teledyne.
APPENDIX D – Wire List and Electronic Schematics 07337B DCN6324 D-1.
This page intentionally left blank. D-2 07337B DCN6324.
T265 INTERCONNECT LIST (Reference: 0739001A DCN6165) Cable Part # Signal Assembly PN J/P Pin Assembly PN J/P Pin 0364901 CBL, AC POWER AC Line Power Entry CN0000073 L Power Switch SW0000025 L AC Neutr.
T265 INTERCONNECT LIST (Reference: 0739001A DCN6165) Cable Part # Signal Assembly PN J/P Pin Assembly PN J/P Pin CONNECTION FROM CONNECTION TO 04671 CBL, MOTHERBOARD TO XMITTER BD (MULTIDROP OPTION) G.
T265 INTERCONNECT LIST (Reference: 0739001A DCN6165) Cable Part # Signal Assembly PN J/P Pin Assembly PN J/P Pin CONNECTION FROM CONNECTION TO 06651 CBL, RX CELL THERM Rcell thermistor A Reaction cell.
T265 INTERCONNECT LIST (Reference: 0739001A DCN6165) Cable Part # Signal Assembly PN J/P Pin Assembly PN J/P Pin CONNECTION FROM CONNECTION TO 06746 CBL, MB TO 06154 CPU GND Motherboard 058021100 P12 .
07337B DCN6324 D-7.
デバイスTeledyne T265の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Teledyne T265をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはTeledyne T265の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Teledyne T265の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Teledyne T265で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Teledyne T265を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はTeledyne T265の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Teledyne T265に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちTeledyne T265デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。