THE NEW FACE of Glucometer ® Diabetes Care System U SER G UIDE F or Use W ith Ascensia E LITE ™ XL Blood Glucose Meter Need Help with Reimbursement? Call Our Bayer Care Help Line: 1-877-787-1912.
2 The Meter ii T est end Meter end T est Strip Icon Button Display T est slot.
T esting Materials 2 iii • Ascensia E LITE ™ T est Strip – use for blood or control testing. Insert into Meter before drawing in blood or control sample . • Ascensia E LITE ™ Check Strip – use to check if the Meter is working properly before running a blood glucose test.
iv • T wo-step testing process: 1. Insert T est Strip in Meter 2. T ouch TEST END to blood • T est results are displayed in 30 seconds . • Meter turns on when inserting a strip and turns off when remov- ing the strip . • Up to 120 test results may be stored in the Meter and can be recalled when necessary .
1 What do I do first? If needed, install batteries and set the time — test the Meter electronics with a Check Strip — code the Meter to match Ascensia E LITE T est Strips — and then run a control test to test the whole system before running a blood glucose test.
2 Important Information 2 • T he Ascensia E LITE ™ XL Blood Glucose Meter has been designed for use with Ascensia E LITE ™ T est Strips . T he perfor- mance characteristics of the Meter may not be valid for any other brand of T est Strips . T he use of other T est Strips may cause erroneous results .
User Guide Contents 2 3 SECTION P AGE Set Up For T esting 1 Install Batteries 4 2 Set the Alarm, Time and Units 5 3 Check Strip T est 6 4 Code the Meter 7 5 Control T est 8 T esting 6 Blood Glucose T .
SECTION 1 Install Batteries 4 1. Press firmly on “open” arrow and slide forward in the direc- tion of the arrow . The lid releases . 2. Place new batteries into the bat- tery compartment with the “+” side up . Slide lid back into place , lining up with open slots , and close firmly .
SECTION Set the Alarm, Time and Units 2 5 Some Meter options may need to be reset when the Meter is new and when batteries are changed. The options may be changed any time to fit your needs. How do I set the Meter? Hold the button for five seconds to place Meter in “set up” mode .
SECTION 3 Check Strip T est 6 Check your Meter when new or any time after use, to make sure that electr onics are working pr operly . HOW : 1. Remove the Check Strip from the package (save the package). 2. Insert the Check Strip (“check” side up) fully into the Meter until you hear a beep .
SECTION Code the Meter 4 7 Code the Meter whenever a new box of T est Strips is opened. * This will match the Meter to the T est Strip reactivity . HOW : Y ou will need the Code Strip found in the new box of strips . 1. T ear open the packet found in the new carton of T est Strips and remove the Code Strip (save the pack et).
SECTION 5 Control T est 8 Prepare for the T est: 1. T o open the T est Strip foil pack et, carefully peel the foil back until the T est Strip is completely exposed (as shown). Save the foil pack et. Note: Numbers printed on each T est Strip are manufacturing numbers and are not to be confused with Code Numbers (F#).
SECTION Control T est 5 9 T ak e the T est: 5. Squeeze a small drop of Control Solution onto the inside of the foil pack et you saved. 6. T ouch and hold the TEST END of the T est Strip to the drop until the Meter beeps (the Control Solution has now been drawn into the T est Strip).
SECTION 5 Control T est 10 If you did not mark your control or want to delete the test result, do one of the following options before removing the strip . Remember , after marking or deleting, the test result will then not be a part of the blood glucose test average and will be stor ed in the memory and displayed with “CONT” or “DEL ”.
SECTION Blood Glucose T est 6 11 What does blood glucose monitoring do for me? Blood glucose levels change during the day due to medication, exercise , diet, insulins or hypoglycemics (oral agents) or possibly stress and illness . Blood glucose testing provides information that will help you understand the changes .
SECTION 6 Blood Glucose T est 12 Y ou need a Meter and T est Strip . T ak e the T est: 1. After washing your hands with warm soapy w ater and drying thoroughly , open a foil pack et and remove the T est Strip . 2. With the Meter off , hold the round end and insert the T est Strip fully into the Meter .
SECTION Blood Glucose T est 6 13 Note: Multiple Site T esting Consult with your healthcare professional prior to testing from a site other than your fingertips . Blood glucose testing results from sites other than your fingertip could be significantly different.
SECTION 6 Blood Glucose T est 14 T est Results: After 29 seconds , your test result is displayed. Do not remove strip until results are displayed. Record the test result/date in your Clinilog Diary . * Remove the T est Strip . T he Meter turns off and the test result is stored.
SECTION Important to Remember 7 15 DON’T FORGET— when using the Meter: • Insert the Check Strip , Code Strip or T est Strip with a firm motion until it comes to a full stop . • T he tab on the strip can no longer be seen if the T est Strip has been inserted fully into the Meter .
SECTION 7 Important to Remember 16 DON’T FORGET— when using the Meter (cont.): • When the Meter memory is full and a new test result is entered, the oldest result is deleted (i.e., #1). • T each another person how to use the system, so you will have someone to help if you ever need assistance .
How to recall test r esults: 1. Press the button to turn Meter on. After a full display and year , the current date and time are displayed. T he 14-day average then appears with “MEM” (memory) and “ A ” (average) in the lower left corner .
SECTION 9 Data Management 18 • Use Win Glucofacts ™ XL Diabetes Management Software . • Y our Meter can transfer test results to a computer where results can be electronically summarized in a logbook report, graph, or chart. • T o make use of this feature , you need special softw are and a connecting cable .
19 SECTION Meter and T est Strip Care 10 • W ash hands and dry thoroughly before testing to k eep the Meter and T est Strips free of oils and other contaminants . • Handle the Meter carefully to avoid damaging the electronics or causing other malfunctions .
SECTION 10 W arning: P otential Biohazard 20 Healthcare professionals or persons using this system on multiple patients should be aware of the following and should follow the infection control procedure approved by their facility .
SECTION Meter and T est Strip Problems 11 MESSAGE CA USE SOLUTION See the possible solutions listed and if you need further assistance , contact Bayer Customer Service at: 1-800-348-8100 (24 hours a day) E-1 E-3 display E-4 E-2 display Meter electronics failure .
SECTION 11 Meter and T est Strip Problems 22 MESSAGE CA USE SOLUTION Meter fails to start countdown. 1. Blood applied wrong; see page 12. 2. T est Strip inserted incorrectly . 3. Meter turned on before T est Strip inserted. 4. Sample is applied while full display is showing.
SECTION Meter and T est Strip Problems 11 23 MESSAGE CA USE SOLUTION “Lo” display . CAUTION: Glucose levels below 50 mg/dL (2.8 mmol/L) may indicate a potentially serious medical condition. 1. T he blood glucose value may be below 20 mg/dL (1.1 mmol/L).
SECTION 11 Meter and T est Strip Problems 24 MESSAGE CA USE SOLUTION Blood glucose or Control T est results are inconsistent, or Control T est results are not within the specified range . 1. Not enough blood or Control Solution is drawn into the T est Strip .
SECTION Service 12 25 1-800-348-8100 IMPORT ANT: • Speak to a Customer Service Representative before returning your Meter for any reason. He/she will give you the informa- tion needed to get your problem handled correctly and efficiently .
SECTION 12 Service 26 Check List 1. T he Meter serial number is: ___________________ 2. T he date and time the problem occurred was: ___________________ 3.
SECTION Limitations of the Procedure 13 27 1. PRESERV A TIVES: • Fluoride – Do not use fluoride as a preservative for blood specimens . • Iodoacetic acid – Do not use iodoacetic acid as a preserv ative for blood specimens .
SECTION 14 Specifications 28 T est Sample: Whole blood T est Result: Plasma/serum glucose Reaction Chamber Sample V olume: 2 L Measuring Range: 20 – 600 mg/dL (1.
SECTION Supplies 15 29 When calling or writing for supplies be sure to include the number with the name of the replacement part or accessory item. REPLACEMENT P ARTS Part Number* Item 40030030 3-volt .
SECTION 15 Supplies 30 ACCESSOR Y ITEMS Product Code Number Item 3870A Ascensia E LITE ™ T est Strips – 50’ s (Medicare patient use only) 3871A Ascensia E LITE ™ T est Strips – 50’ s (mail.
SECTION W arranty 16 31 Manufacturer’ s Warr anty: Bayer Corporation (“Bayer”) warrants to the original purchaser that this instrument will be free from defects in materials and workmanship for 5 years from the date of original pur- chase (except as noted below).
SECTION 16 W arranty 32 BA YER MAKES NO OTHER EXPRESS WARRANTY FOR THIS PRODUCT . THE OPTION OF REPLACEMENT , DESCRIBED A BOV E, IS BA YER’S ONL Y OBLIGA TION UNDER THIS W ARRANTY . IN NO EVENT SHALL BA YER BE LIABLE FOR INDIRECT , SPECIAL OR CONSEQUENTIAL DAMA GES , EVEN IF BA YER HAS BEEN ADVISED OF THE POSSIBILITY OF SUCH DAMA GES .
デバイスBayer HealthCare Ascensia Elite XLの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Bayer HealthCare Ascensia Elite XLをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはBayer HealthCare Ascensia Elite XLの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Bayer HealthCare Ascensia Elite XLの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Bayer HealthCare Ascensia Elite XLで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Bayer HealthCare Ascensia Elite XLを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はBayer HealthCare Ascensia Elite XLの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Bayer HealthCare Ascensia Elite XLに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちBayer HealthCare Ascensia Elite XLデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。