375 Marcus Boulevard • Hauppauge, NY 1 1 788 • USA 63 1 .273.0500 • Fax: 63 1 .273.077 1 e-mail: info@ultraviolet.com Extensive Product Information A vailable at: Document No. 98 - 1 123E1 • Revised February 2007 • © 1994-2007 Atlantic Ultraviolet Corporation www .
2 T ABLE OF CONTENTS S AFETY W ARNINGS .....................................................................................................................................................3 SAFETY INSTRUCTIONS .........................................
3 SAFETY W ARNINGS • All personnel should be alerted to the potential hazards indicated by the product safety labeling on this unit. • The following conventions are used to indicate and classify precautions in this manual and on product safety labeling.
4 PRODUCT APPLICA TION C ONSTRUCTION • The water purifier is designed to mount horizontally . • A drain port on the chamber aids in draining of the purifier . • The water purifier chamber is passivated and electropolished type 316 Stainless S teel.
5 INST ALLA TION L OCA TION 1. The water purifier is intended for indoor use only . The water purifier should be protected from the elements and from temperatures below freezing. The ambient temperature, in the area surrounding the water purifier , should be between 33 F and 100 F.
6 6 3. The water purifier should be mounted horizontally on a flat dry surface. Secure the water purifier using the mounting holes in the ballast housing or with the optional wall mounting kit. The purifier should not be solely supported by its plumbing connections.
7 R ECOMMENDED O PTIONS 1 . Guardian ™ Ultraviolet Monitor: V isually indicates the level of germicidal ultraviolet energy that penetrates the quartz sleeve and the water within the water purifier . The ultraviolet monitor is capable of operating an optional audio alarm and/or solenoid valve.
8 MAINTENANCE The water purifier is designed to operate with a minimal amount of maintenance, providing the water quality does not exceed maximum concentration levels, see “ W ater Quality ” in the “ Pr oduct Application ” section.
9 S tep 2 - Remove End-Cap S tep 3 - W ithdraw Lamp S tep 4 & 5 - Remove Sockets S tep 6 - Remove Lamp L AMP I NST ALLA TION OR R EPLACEMENT Figure 3 - Lamp Replacement 1. Disconnect power to water purifier . 2. Remove both easy-off end caps by pulling each cap of f static gland nut.
10 1. Disconnect power to water purifier . 2. Shut off water supply to water purifier via inlet and outlet shut of f valves. Drain chamber by removing drain plug.
11 R EPLACEMENT OF B ROKEN Q UARTZ S LEEVE CAUTION: Broken Quartz is SHARP . It is r ecommended that protective goggles and gloves are worn when handling br oken quartz sleeves. 1. Disconnect power to water purifier . 2. Shut off water supply to the water purifier via the inlet and outlet shut of f valves.
12 O PTIONAL G UARDIAN ™ ™ U L TRA VIOLET M ONIT OR S ENSOR P ROBE C LEANING Figure 5 - Ultraviolet Monitor and Sensor Probe If after cleaning the quartz sleeve, there is no significant improvement in the ultraviolet intensity , as shown on the intensity meter , it may be necessary to clean the Ultraviolet Monitor's sensor probe.
13 Problem Possible Cause Corrective Action No electrical power… V erify that the purifier is connected to a live power source. Ballast Disconnect the purifier from utility power . Replace lamp. Lamp-out Protection The purifier must remain disconnected from power for (MP16A, MP22A & MP36C) two (2) minutes to reset.
14 TECHNICAL SPECIFICA TIONS T able 4 - T echnical Specifications Model: MP13A MP16A MP22A MP36C MP49C Flow Rate (GPM): 2 3 6 12 20 InletOutlet Size: 1 / 2 " m NPT 3 / 4 " m NPT 3 / 4 " m NPT 1 " m NPT 1- 1 / 2 " m NPT Number of Lamps: 1 1 1 1 1 Lamp Model No.
15 OPTIONAL ACCESSORIES Optional accessories are available for operation at 120v 60Hz or 220v 50Hz, unless specified. Please specify . Consult Factory for 12v DC or other special input power requirements.
16 REPLACEMENT P AR TS Mighty Pure ™ ™ Models MP13A, MP16A & MP22A Figure 6 - Exploded V iew MP13A, MP16A & MP22A.
17 ITEM MP13A MP16A MP22A NO. DESCRIPTION QTY P AR T NO. QTY P AR T NO. QTY P AR T NO. 1 Easy-Off End Cap 2 25-1499A1 2 25-1499A1 2 25-1499A1 2 Rubber O-ring, Static Gland 2 00-1 108B 2 00-1108B 2 00-.
18 REPLACEMENT P AR TS Mighty Pure ™ ™ Models MP36C & MP49C Figure 7 - Exploded V iew MP36C & MP49C.
19 ITEM MP36C MP49C NO. DESCRIPTION QTY P AR T NO. QTY P AR T NO. 1 Easy-Off End Cap 2 25-1499A1 2 25-1499A1 2 Rubber O-ring, Static Gland 2 00-1108B 2 00-1 108B 3 Lamp Spacer - - 2 25-0209A 4 Lead W .
20 W ARRANTY W e warrant that this product will be free from defects in material and workmanship for a period of one year from the date of shipment thereof or the product’ s total rated life, whichever first occurs.
デバイスUltraViolet Devices MP36Cの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
UltraViolet Devices MP36Cをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはUltraViolet Devices MP36Cの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。UltraViolet Devices MP36Cの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。UltraViolet Devices MP36Cで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
UltraViolet Devices MP36Cを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はUltraViolet Devices MP36Cの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、UltraViolet Devices MP36Cに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちUltraViolet Devices MP36Cデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。