Muon Liquid Handling System User Guide C. Johnson, S.P. Cottrell et al Version 0.2 - 1 -.
Contents 1 Getting Started 3 1.1 Layout of the liquid handling system 3 1.2 Layout of the in situ sample stick 4 1.3 Layout of the pump 4 2 General procedures 6 2.1 Sample loading 6 2.1.1 Loading Vessel 1 6 2.1.2 Loading Vessel 2 6 2.2 Evacuation of the system 6 2.
1 Getting S t arted This manual describes the liquid handling system and the in situ sample cell as use d on the DEVA instrument with the “RF” spect rometer, for either norm al muon spins relaxation or RF resonance experiments.
1.2 Layout of the in situ sample stick The liquid sample stick is designed to fit into the DEVA f low cryostat, details of which can be found in the DEVA manual. It consists of a shapol target cell 30 mm x 30 mm with a mylar window upon which may be mounted an RF coil.
Roughing/ Turbo switch Off/Run switch Pressure reading Turbo control panel Turbo Pump Rotary Pump Figure 2 The lay out of the fr ont panel o n the vacuum pump use d with the muon liqui d handlin g system.
2 General procedures 2.1 Sample loading At this point, the system will be open to the atm osphere, ensure that valve V2 is closed to prevent air cond ensation in the cold trap. 2.1.1 Loading Vessel 1 • Open the tap on the right-hand side of Vessel 1.
2.2.1 Removing air from Vessel 1 Once the system has been pressuris ed, tap A on Vessel 1 can be opened. It is possible to bubbled helium gas through the sample liqui d against the non-retu rn valve, V13. If this is required, first ensure th at valves V 5, V7 and V8 are closed.
2.3 Freeze-Pump-Thaw cycle Using this system it is possible to o perat e freeze-pump-thaw cycles on sam ple liquids held in any of the three vessels. Procedur es appropriate to ea ch vessel are outlined below. 2.3.1 Sample in Vessel 1 • All taps and valves should be closed.
2.3.3 Sample in Vessel 3 • All taps and valves should be closed. • Open valves V2 and V3 along with tap C on Vessel 3 and freeze the sample liquid by slowly raising a dewa r of liquid nitrogen around it. • When it has been determined that the liquid ha s completely frozen switch the pump to roughing mode and open valve V1.
2.4.2 Vessel 3 to sample cell The transfer of liquid from Vessel 3 to the sample cell is brought about by the difference in pressure between that in the Vessel and a vacuum in the cell. However, in order to control the transfer of liquid th e pressure difference shou ld be as small as possible but not so low that the liquid starts to boil.
2.4.4 Recovery of liquid in sample cell to Vessel 1 The liquid in the sample cell cannot be recovered to Vessel 1 using the liquid handling rig as presently configured. 2.4.5 Recovery of liquid in sample cell to Vessel 2 The liquid in the sample cell can be r ecovered to Vessel 2 using the following procedure.
• All the liquid may not return in a single pass and repea ting the evacuation procedure may be required. However, some liquid m ay be present in the system and care should be taken during evacuation that th e trap does not become blocked. The liquid may also be encouraged to leave the sample cell by using a small amount of gas.
3 Example Experiment The paramagnetic signal in liquid n-hexane is difficult to observe because the muonium polarisation is found to decay on a microsecond time scale, which is further reduced by the presence of dissolved oxygen.
0.0 0.5 1.0 1.5 2.0 2.5 3.0 11.0 11.5 12.0 12.5 13.0 13.5 Initial asymmetry (%) Time ( µ s) 0.0 0.5 1.0 1.5 2.0 2.5 3.0 11.0 11.5 12.0 12.5 13.0 Initial asymmetry (%) Time ( µ s) 1.
デバイスBoca Research Noneの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Boca Research Noneをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはBoca Research Noneの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Boca Research Noneの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Boca Research Noneで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Boca Research Noneを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はBoca Research Noneの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Boca Research Noneに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちBoca Research Noneデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。