Model 109SS Temperature Probe 9/08 Copyright © 1983-2008 Campbell Scientific, Inc..
Warranty and Assistance The MODEL 109SS TEMPERATURE PROBE FOR HARSH ENVIRONMENTS is warranted by CAMPBELL SCIENTIFIC, INC. to be free from defects in m aterials and workm anship under normal use and servi ce for twelve (12) months from date of shipm ent unless specified otherwise.
109SS Table of Contents PDF viewers note: These page numbers refer to the printed version of this document. Use the Adobe Acrobat® bookmarks tab for links to specific sections. 1. General ..............................................................
.
Model 109SS Temperature Probe 1. General The –L portion of this probe’s m odel number indicat es the probe has a user- specified lead length. For readab ility purposes, the pr obe will be referred to as the 109SS throughout this docum ent. The 109SS consists of a thermistor encas ed in a stainless-steel sheath.
Model 109SS Temperature Probe 1.1 Specifications Temperature Range: -40° to +70°C Survival Range: -50° to +100°C (thermistor); -50°C to +70°C (overmolded joint and cable) Thermistor Interchangeability Tolerance: Temperature Tolerance -40°C ± 0.
Model 109SS Temperature Probe 2. Accuracy The overall probe accuracy is a combinati on of the thermi stor's interchangeability specification and the accuracy of the bri dge resistor. The Steinhart and Hart equation used to calculate temperature has a negligi ble error (Figure 2-1).
Model 109SS Temperature Probe 3. Installation and Wiring 3.1 Burial The 109SS is suitable for shallow burial onl y. It should be placed horizontally at the desired depth to avoid thermal conduction from the surface to the thermistor.
Model 109SS Temperature Probe The datalogger is programme d using eith er CRBasic or Edlog. Dataloggers that use CRBasic include our CR 200-series, CR800, CR850, CR1000, CR3000, CR5000, and CR9000(X); see Section 5.1. Dataloggers that use Edlog include our CR10, CR 10(X), CR23X, and CR7; refer to Secti on 5.
Model 109SS Temperature Probe 5.1.1.1 Sample Program for CR200 Series Datalogger 'CR200 Series Datalogger ‘This example program measures a single 109SS Thermist or Probe ‘once a second and stores the aver age temperature every 10 minutes.
Model 109SS Temperature Probe 5.1.1.3 Sample Program for CR5000 'CR5000 'This example program measures a single 109 Thermist or probe 'once a second and stores the aver age temperature every 10 minutes. 'Declare the variable for the temperature.
Model 109SS Temperature Probe Example Program for CR10X ;{CR10X} ; *Table 1 Program 01: 1 Execution Interval (seconds) 1: AC Half Bridge (P5) 1: 1 Reps 2: 25 2500 mV 60 Hz Reject ion Range 3: 1 SE Channel 4: 1 Excite all reps w/Exchan 1 5: 2500 mV Excitati on 6: 1 Loc [ V_Vx ] 7: 1.
Model 109SS Temperature Probe 9: Z=X+F (P34) 1: 8 X Loc [ Tk ] 2: -273.15 F 3: 9 Z Loc [ Air_Temp ] 10: If tim e is (P92) 1: 0 Minutes (Seconds --) into a 2: 10 Interval (same uni ts as above) 3: 10 S.
Model 109SS Temperature Probe 1: Excite-Delay (SE) (P4) 1: 1 Reps 2: 25 2500 mV 60 Hz Rejecti on Range (Delay must be zero) 3: 1 SE Channel 4: 1 Excite all reps w/Exchan 1 5: 2 Delay (0.01 sec units) 6: 2500 mV Excitati on 7: 3 Loc [V_Vx ] 8: .0004 Multiplier 9: 0.
Model 109SS Temperature Probe The resistance of the thermistor is: ⎟ ⎠ ⎞ ⎜ ⎝ ⎛ − = 1 900 , 24 V V R EX t The Steinhart and Hart equation i s used to calculate temperature from Resistance: 3 )) (ln( ) ln( 1 T T K R C R B A T + + = Where T K is the temperature in Kelvin .
Model 109SS Temperature Probe 12 Symptom: Unstable Tem perature Try using the 60 or 50 Hz integr ation option s, and/or increasing the settling time as descri bed in Sections 8 and 9. Make sure the clear shiel d wire is connected to datalogger ground, and the dat alogger is properly grounded.
This is a bl ank page..
Campbell Scientific Companies Campbell Scientific, Inc. (CSI) 815 West 1800 Nort h Logan, Utah 84 321 UNITED STATES www.campbellsci.com info@campbellsci.com Campbell Scientific Afri ca Pty. Lt d. (CSAf) PO Box 2450 Somerset West 7129 SOUTH AFRICA www.
デバイスCampbell Manufacturing 109SSの購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Campbell Manufacturing 109SSをまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはCampbell Manufacturing 109SSの技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Campbell Manufacturing 109SSの取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Campbell Manufacturing 109SSで得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Campbell Manufacturing 109SSを既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はCampbell Manufacturing 109SSの不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Campbell Manufacturing 109SSに関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちCampbell Manufacturing 109SSデバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。