CanonメーカーCR-DA1の使用説明書/サービス説明書
ページ先へ移動 of 11
Canon Des Adapter CR-DA1 Operation Manual Before using the instrument, be sure to read this manual thoroughly. Also, read the Non-Mydriatic Retinal Camera's Operation Manual. Keep the manual where it is easily accessible. For sales, service, and support: Richard J Kinch, PhD http://www.
PLEASE NOTE 1. In no event will Canon be liable for direct or indirect consequential damage arising out of the use of this product. 2. The CR-DA 1 system has been approved by UL and GM with the EOS 030 digital camera powered by a battery.
CONTENTS 1. Outline............................................................................ 1 2. Notes for Using the Instrument.............................. 1 3. Description 2 4. Photography 3 5. Installation 5 6. Before Calling a Service Person 7 7.
1. Outline The Canon DCS (Digital Camera System) Adapter CR-DAI is for converting the size of the image taken with our Non-Mydriatic Retinal Camera CR6-45NMlCR6-45NMf into a size to allow the image to be taken with a Canon's EOS digital camera.
3. Description CD Retinal camera mount This mount is for attaching the adapter to the retinal camera. ® Lock release button This button releases the adapter from the retinal camera. ® Digital camera mount This mount is for attaching the digital camera to the adapter.
4. Photography Operation of the retinal camera in the system using this adapter is basically the same as in the general retinal photography. Please also refer to the manuals of the CR6-45NM1CR6-45NMf and the digital camera. NOTE: FCD (Flash Compensation by Diopter) does not operate in the photography using this adapter.
Flash intensity can be changed in 11 steps (increments: 1/3 steps). DA 1 DA2 DA3 DA4 DA5 DA6 DA7 DA8 DA9 DA10 DA 11 Equival- ent to Dark E08 ~ Bright - ~ D30 180400 (3) After changing the setting, press the switch under "RTN". Now the changed flash intensity is valid.
5. Installation & WARNING Be sure to turn OFF the power of respective equipments before connecting or disconnecting the cables. Also, do not handle them with wet hands. Step 1 Attaching the adapter to the retinal camera (I) Remove the cap from the retinal camera mount on the adapter.
Step 2 Attaching the digital camera to the adapter (1) Remove the caps from the lens mount on the ~ digital camera and from the digital camera mount on the adapter. (2) Align the guide mark on the adapter and the lens attachment mark on the digital camera.
Step 3 Connecting the signal cables (1) Connect the strobe timing signal cable to the PC terminal for flash syncronization on the side of the digital camera. Rotate the screw on the connector of the cable in clockwise direction to lock the cable. Rotate the screw in counterclockwise direction when disconnecting the cable.
7. Specifications Attachable digital cameras: Mounts: Magnification: Dimensions: Mass: Canon digital camera (such as EOS D30) Retinal camera side: Canon EF mount Digital camera side: Canon EF mount 22 (on 35mm film surface) : 17 (on imaging element) ~ 85 x 70mm Approx.
デバイスCanon CR-DA1の購入後に(又は購入する前であっても)重要なポイントは、説明書をよく読むことです。その単純な理由はいくつかあります:
Canon CR-DA1をまだ購入していないなら、この製品の基本情報を理解する良い機会です。まずは上にある説明書の最初のページをご覧ください。そこにはCanon CR-DA1の技術情報の概要が記載されているはずです。デバイスがあなたのニーズを満たすかどうかは、ここで確認しましょう。Canon CR-DA1の取扱説明書の次のページをよく読むことにより、製品の全機能やその取り扱いに関する情報を知ることができます。Canon CR-DA1で得られた情報は、きっとあなたの購入の決断を手助けしてくれることでしょう。
Canon CR-DA1を既にお持ちだが、まだ読んでいない場合は、上記の理由によりそれを行うべきです。そうすることにより機能を適切に使用しているか、又はCanon CR-DA1の不適切な取り扱いによりその寿命を短くする危険を犯していないかどうかを知ることができます。
ですが、ユーザガイドが果たす重要な役割の一つは、Canon CR-DA1に関する問題の解決を支援することです。そこにはほとんどの場合、トラブルシューティング、すなわちCanon CR-DA1デバイスで最もよく起こりうる故障・不良とそれらの対処法についてのアドバイスを見つけることができるはずです。たとえ問題を解決できなかった場合でも、説明書にはカスタマー・サービスセンター又は最寄りのサービスセンターへの問い合わせ先等、次の対処法についての指示があるはずです。